K101371 · ACON Laboratories, Inc. · NBW · Jan 7, 2011 · Clinical Chemistry
Device Facts
Record ID
K101371
Device Name
ON CALL PLUS DIABETES MONITORING SOFTWARE
Applicant
ACON Laboratories, Inc.
Product Code
NBW · Clinical Chemistry
Decision Date
Jan 7, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The On Call® Plus Diabetes Software is an optional software accessory to be used with the On Call® Plus Glucose Meters for transferring data to a computer and organizing it in tables and graphs to be used at home and by health care professionals. The software does not recommend any medical treatment or medication dosage level.
Device Story
Software accessory for On Call® Plus Glucose Meters; transfers glucose data to PC via USB or RS232 cable. Used in home or clinical settings by patients or healthcare professionals. Features 'Home' version for single-user tracking and 'Professional' version for multi-provider/multi-patient database management. Transforms raw meter data into tables, line charts, bar charts, and pie charts (e.g., average day/week reports). Does not provide medical advice or dosage recommendations. Enables password-protected database management; allows configuration of glucose target levels and time periods. Benefits patients and providers by facilitating glucose trending and data visualization for diabetes management.
Clinical Evidence
Clinical usability study conducted with lay persons. Participants performed software installation, data transfer, setting configuration, database management, and report generation. Results demonstrated that inexperienced users could successfully operate the software and obtain glucose trending reports. Users reported satisfaction with ease of operation and performance.
Technological Characteristics
System requirements: Pentium processor, USB/RS-232 port, 128MB RAM, 70MB disk space. Compatible with Windows 2000, XP, Vista, 7. Data transfer via USB/RS232 TTL cable. Features password-protected databases, multi-user/multi-provider support, and configurable glucose target levels. Complies with EN 62366:2008 (usability) and EN 62304:2006 (software life cycle).
Indications for Use
Indicated for use by patients with diabetes or healthcare professionals to transfer, organize, and visualize blood glucose data from On Call® Plus Glucose Meters. Suitable for home and clinical settings.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Glucofacts® Deluxe Diabetes Management Systems (K091820)
Related Devices
K131469 — ACON ON CALL DIABETES MANAGEMENT SOFTWARE · ACON Laboratories, Inc. · Jan 6, 2014
{0}
1 of 6
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k101371
B. Purpose for Submission:
New 510(k) for diabetes data management software accessory for use with the On Call® Plus Glucose Meter (k090057)
C. Measurand:
Whole blood glucose
D. Type of Test:
Data management system
E. Applicant:
Acon Laboratories, Inc.
F. Proprietary and Established Names:
On Call Plus Diabetes Monitoring Software
G. Regulatory Information:
1. Regulation section:
21CFR Sec.- 862.1345-Glucose test system.
21CFR Sec.- 862.2100-Calculator/data processing module for clinical use.
2. Classification:
II, I
3. Product code:
NBW - System, Test, Blood Glucose, Over The Counter
JQP - Calculator/Data Processing Module, For Clinical Use
4. Panel:
Chemistry – 75
H. Intended Use:
1. Intended use(s):
See Indication(s) for use below
2. Indication(s) for use:
The On Call® Plus Diabetes Software is an optional software accessory to be used with the On Call® Plus Glucose Meter for transferring data to a computer and organizing it in tables and graphs to be used at home and by health care
{1}
professionals. The software does not recommend any medical treatment or medication dosage level.
3. Special conditions for use statement(s):
Over-the-counter use
4. Special instrument requirements:
On Call® Plus Glucose Meter
I. Device Description:
The On Call® Plus Diabetes Monitoring Software is for downloading glucose data from the On Call Plus Glucose meter to a PC through an USB to RS232 TTL level cable or RS232 to RS232 TTL level cable, with tracking and trending capabilities of glucose measurements.
The Diabetes Data Management Software is intended to be used in a clinical setting with multiple patients by a healthcare professional, or directly by the patient.
The user has the choice to install either a personal "Home" or "Professional" version. In the "Home" version, glucose results are assembled for a single user. In the "Professional" version, the software user can configure a database with multiple providers and patients. The number of providers, patients and data points is limited only by the computer's capabilities.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Bayer Healthcare, Glucofacts® Delux Diabetes Management Software
2. Predicate 510(k) number(s):
K091820
3. Comparison with predicate:
| Features | On Call® Plus Diabetes Monitoring Software | Glucofacts® Delux Diabetes Management Software (K091820) |
| --- | --- | --- |
| Similarities | | |
| Intended use /Indication(s) for use | The On Call® Plus Diabetes Software is an optional software accessory for transferring data from glucose meter to a computer and organizing it in tables and graphs to be used at home and by health care professionals. The software does not recommend any medical treatment or medication dosage level. | Same |
| COM Ports scan | Automatic scan for connected ports | Same |
| Multiple patients | Can list data for multiple patients | Same |
| Data list/ Summary report | List of all readings inside selected time frame | Same |
{2}
| Features | On Call® Plus Diabetes Monitoring Software | Glucofacts® Deluxe Diabetes Management Software (K091820) |
| --- | --- | --- |
| Reports and charts | Data List, Log Book report, Standard Day report, Trend report, Pie charts | Same |
| Target levels | High, low, hyper and hypo glucose target levels can be changed | Same |
| Time periods | Daily time periods can be changed | Same |
| Units of measure | Can choose either mg/dL or mmol/L mg/dL is default | Same |
| Data Base | Can Set up multiple patient data bases | Same |
| Differences | | |
| Change Meter settings | Does not allow change to meter settings through the Diabetes Monitoring Software | Allows target range, alarm sounds and optional level changes to supported meters through the Management Software |
| Operating System | Microsoft Windows 2000, XP Home/Professional, Vista, windows 7 | Microsoft Windows XP, Vista or Mac OS 10.5.7 or later (Mac is only an option with the Contour USB meter) |
| Associate a meter to a person | No association | Software can be set to recognize meters and associate them to specific patients |
| Standard week report | No Standard Week report | A graph that displays all readings by day overlapping all the days on a one week graph |
| Average Week | A bar graph that averages all readings for each day by day of the week, Sunday through Saturday and before and after meals. | No average week report |
| Average Day | A bar graph that averages all readings for each day by time slot and before and after meals. | No average day report |
| Password | Allows providers accounts to be password protected | Does not have password protection on the software |
| Providers | Allows for multiple providers each with their own data base of patients | Does not have providers but does allow for multiple databases |
# K. Standard/Guidance Document Referenced (if applicable):
- ISO 14971:2007, Medical devices - Application of risk management to medical devices.
- IEC 62366, Medical devices - Application of usability engineering to medical devices.
- IEC 62304 Ed. 1.0, Medical device software - Software life cycle processes.
# L. Test Principle:
Data Transmission of whole blood glucose data from glucose meter
{3}
M. Performance Characteristics (if/when applicable):
The below performance characteristics as applicable, were presented in specific glucose meter clearance under k090057
1. Analytical performance:
a. Precision/Reproducibility:
See above statement under section M.
b. Linearity/assay reportable range:
See above statement under section M.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
See above statement under section M.
d. Detection limit:
See above statement under section M.
e. Analytical specificity:
See above statement under section M.
f. Assay cut-off:
Not Applicable
2. Comparison studies:
a. Method comparison with predicate device:
See above statement under section M.
b. Matrix comparison:
Not Applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
See above statement under section M.
4 of 6
{4}
5 of 6
N. Instrument Name:
On Call Plus Diabetes Monitoring Software
O. System Descriptions:
1. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?:
Yes ☐ X or No ☐
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?:
Yes ☐ or No ☐ X
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
Pentium or equivalent processor, USB or RS-232 9-pin serial port, 128 MB memory, 70 MB hard drive space, minimum 800 x 600 display, CD-ROM drive
Microsoft Windows 2000, XP Home/Professional, Vista, windows 7
3. Specimen Identification:
Specimen identification is based on data transmitted from the meter.
4. Specimen Sampling and Handling:
Data transmission from glucose meters using capillary whole blood samples
5. Calibration:
Glucose meter specific See above statement under section M.
6. Quality Control:
Glucose meter specific See above statement under section M.
{5}
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
A study to verify the accuracy of data transmission from the On Call Plus Blood Glucose Monitoring Meter to the On Call Plus Diabetes Monitoring Software was conducted. The data management system was challenged under the conditions in which it is intended to be used in a professional health care setting.
20 meters were used for the study each with a full memory of 300 readings. Data stored on the meters incorporated a range of dates and values and included out of range readings for both high and low and in range data including readings marked as controls.
This study verified that the data transmitted to the PC from the meters was accurate and that the data was accurately manipulated by the software 100% of the time.
In addition, 50 participants with varying demographics (age, sex, and education level) were included for individual consumer use and multiple patient use with health care provider using both the Home Edition and Professional Edition. The user guide was used in the study (Flesch Kincaid Grade Level 8.3). Participants were required to install the software on a computer, download meter data and use the various features of the program. 96% of the participants evaluated the overall easy of use for the program as OK or above, on a scale for 1 – 5 where 1 (very hard) 3 (OK) and 5 (very easy).
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
6 of 6
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.