WAVESENSE KEYNOTE CODELESS BLOOD GLUCOSE MONITORING SYSTEM
Applicant
Agamatrix
Product Code
NBW · Clinical Chemistry
Decision Date
Jan 30, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The WaveSense™ KeyNote Codeless Blood Glucose Monitoring System is intended to quantitatively measure blood glucose levels, also known as blood sugar, from fresh capillary whole blood samples taken from the fingertips, palm, or forearm. The WaveSense™ KeyNote Codeless Blood Glucose Monitoring System test strips are for in vitro diagnostic (outside of the body) use only. The WaveSense™ KeyNote Codeless Blood Glucose Monitoring System is not intended for use with neonates.
Device Story
Portable, battery-operated blood glucose meter; uses test strips to analyze fresh capillary whole blood samples from fingertips, palm, or forearm. Device measures glucose levels quantitatively; provides results to patients or healthcare professionals for diabetes management. Modification of WaveSense Jazz system; excludes meal tagging and positive feedback features; includes shell redesign. Used in home or clinical settings; provides immediate feedback for clinical decision-making regarding glycemic control.
Clinical Evidence
Bench testing only; design control verification and validation activities performed to assess impact of modifications.
Technological Characteristics
Portable, battery-operated meter. In vitro diagnostic system. Complies with ISO 15197:2003, ISO 14971:2000, IEC 61010-1, IEC 61010-2-101, and IEC 61000-4-3. Codeless technology.
Indications for Use
Indicated for quantitative blood glucose measurement in fresh capillary whole blood from finger, palm, or forearm. For OTC home use by persons with diabetes or clinical use by healthcare professionals to monitor diabetes control effectiveness. Contraindicated for neonates.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Related Devices
K052762 — LIBERTY BLOOD GLUCOSE MONITORING SYSTEM, MODEL 8000-0067 · Agamatrix, Inc. · Jan 23, 2006
K072413 — MODIFICATION TO: JAZZ BLOOD GLUCOSE MONITORING SYSTEM · Agamatrix · Sep 27, 2007
K241304 — WaveSense Jazz Blood Glucose Monitoring System · Agamatrix · Jun 6, 2024
K053529 — ONETOUCH ULTRA 2 BLOOD GLUCOSE MONITORING SYSTEM · Lifescan, Inc. · Jan 17, 2006
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification
ODE Review Memorandum (Decision Making Document is Attached)
To: THE FILE
RE: DOCUMENT NUMBER k073573
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary):
1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) k072413
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was for:
A. An engineering change in the mechanical appearance.
B. Removal of two user interface applications: meal tagging and positive feedback.
C. Re-naming the device.
4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and analytes.
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
c) A declaration of conformity with design controls. The declaration of conformity should include:
i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and
ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review.
6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices).
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.