MODIFICATION TO: JAZZ BLOOD GLUCOSE MONITORING SYSTEM

K072413 · Agamatrix · NBW · Sep 27, 2007 · Clinical Chemistry

Device Facts

Record IDK072413
Device NameMODIFICATION TO: JAZZ BLOOD GLUCOSE MONITORING SYSTEM
ApplicantAgamatrix
Product CodeNBW · Clinical Chemistry
Decision DateSep 27, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Indications for Use

AgaMatrix JAZZ™ Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the finger stick, palm, and/or forearm. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter (OTC) ) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. AgaMatrix JAZZ™ Blood Glucose Meter is intended for use with AgaMatrix JAZZ™ Blood Glucose Test Strips for the quantitative measurement of glucose in fresh capillary whole blood from the finger stick, palm, and/or forearm. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter (OTC) ) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. AgaMatrix JAZZ™ Blood Glucose Test Strips are intended for use with AgaMatrix JAZZ™ Blood Glucose Meter for the quantitative measurement of glucose in fresh capillary whole blood from the finger stick, palm and/or forearm. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter (OTC) ) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. AgaMatrix JAZZ™ Control Solutions are intended for use with the AgaMatrix JAZZ™ Meter and AgaMatrix JAZZ™ Test Strips as a quality control check to verify the accuracy of blood glucose test results.

Device Story

WaveSense JAZZ is a portable, battery-operated blood glucose monitoring system; includes meter, lancing device, lancets, and control solution. System measures glucose in fresh capillary whole blood samples (fingertip, palm, forearm). User applies blood to test strip; meter performs quantitative measurement. Key feature: no coding required for test strips. Used by diabetics for home monitoring or by clinicians in healthcare settings to assess diabetes control. Provides immediate glucose readings to aid clinical decision-making and patient self-management.

Clinical Evidence

Bench testing only; design control activities and verification/validation testing performed to assess impact of modifications on device performance and safety.

Technological Characteristics

Portable, battery-operated glucose meter. Uses electrochemical test strips for in vitro diagnostic measurement. Complies with ISO 15197:2003, ISO 14971:2000, IEC 61010-1, IEC 61010-2, and IEC 61000-4-3. No coding required.

Indications for Use

Indicated for use as a glucose test system for the quantitative measurement of glucose in capillary whole blood.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification ODE Review Memorandum (Decision Making Document is Attached) To: THE FILE RE: DOCUMENT NUMBER k072413 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary): 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) k071393 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for the modification to the user interface, adding a notch to the test strips and the use of one calibration code at the time of manufacturing so that the user does not manually enter a calibration code prior to using the glucose test system. 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics. Labeling changes were made in association with the above modification in the calibration section, expected value section and performance characteristics. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied c) A declaration of conformity with design controls. The declaration of conformity should include: i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review. 6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices). The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device. {1} 2
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...