WAVESENSE KEYNOTE CODELESS BLOOD GLUCOSE MONITORING SYSTEM
Applicant
Agamatrix
Product Code
NBW · Clinical Chemistry
Decision Date
Jan 30, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Intended Use
The WaveSense™ KeyNote Codeless Blood Glucose Monitoring System is intended to quantitatively measure blood glucose levels, also known as blood sugar, from fresh capillary whole blood samples taken from the fingertips, palm, or forearm. The WaveSense™ KeyNote Codeless Blood Glucose Monitoring System test strips are for in vitro diagnostic (outside of the body) use only. The WaveSense™ KeyNote Codeless Blood Glucose Monitoring System is not intended for use with neonates.
Device Story
Portable, battery-operated blood glucose meter; uses test strips to analyze fresh capillary whole blood samples from fingertips, palm, or forearm. Device measures glucose levels quantitatively; provides results to patients or healthcare professionals for diabetes management. Modification of WaveSense Jazz system; excludes meal tagging and positive feedback features; includes shell redesign. Used in home or clinical settings; provides immediate feedback for clinical decision-making regarding glycemic control.
Clinical Evidence
Bench testing only; design control verification and validation activities performed to assess impact of modifications.
Technological Characteristics
Portable, battery-operated meter. In vitro diagnostic system. Complies with ISO 15197:2003, ISO 14971:2000, IEC 61010-1, IEC 61010-2-101, and IEC 61000-4-3. Codeless technology.
Indications for Use
Indicated for quantitative blood glucose measurement in fresh capillary whole blood from finger, palm, or forearm. For OTC home use by persons with diabetes or clinical use by healthcare professionals to monitor diabetes control effectiveness. Contraindicated for neonates.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Related Devices
K052762 — LIBERTY BLOOD GLUCOSE MONITORING SYSTEM, MODEL 8000-0067 · Agamatrix, Inc. · Jan 23, 2006
K072413 — MODIFICATION TO: JAZZ BLOOD GLUCOSE MONITORING SYSTEM · Agamatrix · Sep 27, 2007
K241304 — WaveSense Jazz Blood Glucose Monitoring System · Agamatrix · Jun 6, 2024
K053529 — ONETOUCH ULTRA 2 BLOOD GLUCOSE MONITORING SYSTEM · Lifescan, Inc. · Jan 17, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
K073573
10 Manor Parkway Salem, NH 03079 Office: +1 (603) 328-6000 Fax: +1 (603) 893-4191
JAN 30 2008
# 510 (k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: unknown
- 1. Submitter's Name, Address, Telephone Number, Contact Person, and date the summary was prepared.
| Submitter's Name: | AgaMatrix, Inc.<br>10 Manor Parkway<br>Salem, NH 03079 |
|-------------------|--------------------------------------------------------|
| Contact Person: | Connie Hertel<br>Director Quality & Regulatory Affairs |
| Telephone: | (603) 328-6051 |
| Fax: | (603) 893-4191 |
Date the summary prepared: December 6, 2007
#### 2. Device Name
| Trade/Propritary Name: | WaveSense™ KeyNote Codeless Blood Glucose |
|------------------------|-------------------------------------------|
| | Monitoring System |
| Common/Usual Name: | Blood Glucose Monitoring System |
| Classification Name: | Glucose Test System (per 21 CFR 862.1345) |
| Class: | II |
| Panel: | Chemistry |
#### 3. Modification Device
The modification of the WaveSense Jazz TM Blood Glucose Monitoring System (Codeless) excludes two User Interface features: meal time tagging and positive feedback (smiley face). In addition, the meter's shell has been modified.
#### 4. Description of the Device:
The AgaMatrix WaveSense™ KeyNote Codeless Blood Glucose Monitoring System includes a meter with batteries, compact carrying case, lancing device, lancets, control solution and owner's booklet. Test Strips are sold separately.
AgaMatrix
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for AgaMatrix. The logo consists of the word "AgaMatrix" in a simple, sans-serif font. A curved line underlines the text. To the right of the word, there are four small squares arranged in a curved line, suggesting a matrix or a connection of elements.
10 Manor Parkway Salem, NH 03079 Office: +1 (603) 328-6000 Fax: +1 (603) 893-4191
It is intended for over-the counter home use by diabetics to monitor their blood glucose levels, or for use in a clinical setting by healthcare professionals. The system tests fresh capillary whole blood. The meter is a portable, battery-operated instrument.
### 5. Intended use of the device:
The WaveSense™ KeyNote Codeless Blood Glucose Monitoring System is intended to quantitatively measure blood glucose levels, also known as blood sugar, from fresh capillary whole blood samples taken from the fingertips, palm, or forearm. The WaveSense™ KeyNote Codeless Blood Glucose Monitoring System test strips are for in vitro diagnostic (outside of the body) use only. The WaveSense™ KeyNote Codeless Blood Glucose Monitoring System is not intended for use with neonates.
#### Testing:
The manufacturer of the WaveSense™ KeyNote Codeless Blood Glucose Monitoring System certifies that the device complies with the following:
ISO 15197:2003 In vitro diagnostic test systems - Requirements for bloodglucose monitoring systems for self-testing in managing diabetes mellitus
ISO 14971:2000 Medical devices – Application of risk management to Medical devices
IEC 61010-1 Medical electrical equipment – General·requirements for safety
IEC 61010-2-101 Safety requirements for electrical equipment for measurement, control and laboratory use - particular requirements for in vitro diagnostic (IVD) medical equipment
IEC 61000-4-3 Electromagnetic compatibility (EMC)
#### 8. Conclusions
Based upon the testing and comparison to the predicate device, the WaveSense™ KeyNote Codeless Blood Glucose Monitoring System Blood Glucose Monitoring System has the same intended uses, with similar technological characteristics as the predicate device. The system performs as intended and raises no new safety or Effectiveness issues.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and three stripes extending from its body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
JAN 30 2008
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
AgaMatrix, Inc. c/o Ms. Connie Hertel Director Quality & Regulatory Affairs 10 Manor Parkway Salem, NH 03079
Re: k073573
> Trade Name: WaveSense™ Keynote Codeless Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Monitoring System Regulatory Class: Class II Product Codes: NBW, CGA Dated: January 16, 2008 Received: January 18, 2008
Dear Ms. Hertel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Jean M. Cooper, M.S., D.V.M.
Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indication for Use
510(k) Number (if known): k073573
## Device Name: AgaMatrix WaveSense™ KeyNote Codeless Blood Glucose Monitoring System
Indication For Use:
AgaMatrix WaveSense™ KeyNote Codeless Blood Glucose Monitoring System: AgaMatrix WaveSense KeyNote Codeless Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the finger stick, palm, and/or forearm. Testing is done outside the body (in vitro diagnostic use). It is indicated for use at home (over the counter (OTC) ) by persons with diabetes, or in a clinical setting by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.
Prescription Use X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use X. (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Carl C. Benson
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K073573
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.