AUDIT MICROCV T-UPTAKE LINEARITY

K112143 · Aalto Scientific, Ltd. · JJX · Jan 31, 2012 · Clinical Chemistry

Device Facts

Record IDK112143
Device NameAUDIT MICROCV T-UPTAKE LINEARITY
ApplicantAalto Scientific, Ltd.
Product CodeJJX · Clinical Chemistry
Decision DateJan 31, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Indications for Use

Audit® MicroCV™ T-Uptake Calibration/Verification Set consists of five levels of human and bovine albumin based serum containing T-Uptake. The Audit® MicroCV™ T-Uptake Calibration/Verification Set is a quality control material intended for use in the quantitative verification of calibration and reportable range of the Roche T-uptake Assay when performed on the P-Modular Analyzer. When used for quality control purposes, it is recommended that each laboratory establish its own means and acceptable ranges and use the values provided only as guides. The Audit® MicroCV™ T-Uptake Calibration/Verification is for In Vitro Diagnostic use only.

Device Story

Audit® MicroCV™ T-Uptake Calibration/Verification Set is a liquid, human and bovine albumin-based serum containing T-Uptake analyte. It consists of five levels (A-E) with concentrations spanning the reportable range of the Roche T-uptake Assay on the P-Modular Analyzer. Used in clinical laboratories to verify instrument calibration and reportable range. Healthcare providers use the output to confirm assay performance and establish laboratory-specific quality control ranges. Benefits include ensuring accuracy and reliability of T-Uptake diagnostic testing.

Clinical Evidence

No clinical data. Bench testing only. Stability was validated via accelerated studies (24-month closed vial shelf life) and real-time open vial studies (10 days at 2-8°C). Value assignment was performed using multiple replicates on the P-Modular analyzer to establish target means and ±10% ranges.

Technological Characteristics

Liquid, human and bovine albumin-based serum matrix. Contains T-Uptake analyte. Five-level set. Storage at 2-8°C. In vitro diagnostic quality control material.

Indications for Use

Indicated for use as a quality control material for the quantitative verification of calibration and reportable range of the Roche T-uptake Assay on the P-Modular Analyzer. For in vitro diagnostic use by clinical laboratories.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k112143 B. Purpose for Submission: New Device C. Measurand: Quality control materials for T- Uptake D. Type of Test: Not Applicable. E. Applicant: Aalto Scientific, Ltd. F. Proprietary and Established Names: Audit® MicroCV™ T-Uptake Calibration/Verification Set G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | JJX | Class I, reserved | 21 CFR § 862.1660, Quality control material | Clinical Chemistry (75) | H. Intended Use: 1. Intended use(s): Please see intended use below. 2. Indication(s) for use: Audit® MicroCV™ T-Uptake Calibration/Verification Set consists of five levels of human and bovine albumin based serum containing T-Uptake. The AuditR MicroCV™ {1} T-Uptake Calibration/Verification Set is a quality control material intended for use in the quantitative verification of calibration and reportable range of the Roche T-uptake Assay when performed on the P-Modular Analyzer. When used for quality control purposes, it is recommended that each laboratory establish its own means and acceptable ranges and use the values provided only as guides. The Audit® MicroCV™ T-Uptake Calibration/Verification is for In Vitro Diagnostic use only. 3. Special conditions for use statement(s): For prescription Use. 4. Special instrument requirements: Performance was established on P-Modular analyzer. I. Device Description: The Audit® MicroCV™ T-Uptake Calibration/Verification is a human and bovine albumin based liquid set of QC Material. Each of the five levels contains T-Uptake analyte. It is used to confirm the proper calibration, linear operating range, and reportable range of T-Uptake. Level A is near the lower limit level and Level E has concentrations near the upper limit of the P-Modular analyzer. Levels B, C, and D are prepared in a manner such that an equal distance exists between each consecutive level. J. Substantial Equivalence Information: 1. Predicate device name(s): Audit® MicroCV™ M Immunoassay Linearity Set. 2. Predicate 510(k) number(s): k062668 3. Comparison with predicate: | Attribute | Audit® MicroCV™ T-Uptake Calibration/Verification Set (Candidate Device - k112143) | Audit® MicroCV™ Immunoassay Linearity Set (Predicate - k062668) | | --- | --- | --- | | Indication for use / Intended for Use | The Audit® MicroCV™ T-Uptake Calibration/Verification Set is a quality control material intended for use in the quantitative verification of calibration and reportable range. | Same | | Test System (Instrumentation / technology) | P-Modular analyzer / Enzymatic | Immunoassay | {2} | Analyte | T-Uptake | Cortisol, Digoxin, Estradiol, Ferritin, Folate, Free T4, FSH, hCG, LH, Progesterone, Prolactine, Testosterone, Total PSA, Total T3, Total T4, TSH and Vitamin B12. | | --- | --- | --- | | Open Bottle Stability | 10 days at 2-8 °C | 5 days at 2-8 °C | | Shelf life | 24 months at 2-8 °C | 12 months at 2-8 °C | # K. Standard/Guidance Document Referenced (if applicable): CLSI Guideline, EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline # L. Test Principle: Not applicable. # M. Performance Characteristics (if/when applicable): # 1. Analytical performance: a. Precision/Reproducibility: Not applicable. b. Linearity/assay reportable range: Not applicable. c. Traceability, Stability, Expected values (controls, calibrators, or methods): # Traceability All components of the Audit® MicroCV™ T-Uptake Calibration/Verification Set are obtained from a commercial vendor and inspected in-house. # Stability Stability testing protocols and acceptance criteria were reviewed and found to be acceptable. Stability characteristics of the Audit® MicroCV™ T-Uptake Calibration/Verification Set were determined using an accelerated study to predict closed vial shelf life for each level. A closed vial shelf life of 24 months is expected at the recommended storage temperature (2 to $8^{\circ}\mathrm{C}$ ). Real-time closed vial stability is ongoing. Real-time opened vial stability studies were also performed. The sponsor states the open vial stability is 10 days if stored tightly capped at 2 to $8^{\circ}\mathrm{C}$ . Storage recommendations are provided in the labeling. # Value Assignment {3} Multiple replicates of the Audit® MicroCV™ T-Uptake Calibration/Verification Set are analyzed on the P- Modular analyzer during several days. Several replicates are tested and the target value of each level is the mean of the observed values. The target ranges are set at ± 10% of the target value. d. Detection limit: Not applicable. e. Analytical specificity: Not applicable. f. Assay cut-off: Not applicable. 2. Comparison studies: a. Method comparison with predicate device: Not applicable. b. Matrix comparison: Not applicable. 3. Clinical studies: a. Clinical Sensitivity: Not applicable. b. Clinical specificity: Not applicable. c. Other clinical supportive data (when a. and b. are not applicable): Not applicable. 4. Clinical cut-off: Not applicable. 5. Expected values/Reference range: 4 {4} | Level | Target | Expected Values / Range | | --- | --- | --- | | Calibrator Level A | 17.3 | 15.6-19.0 | | Calibrator Level B | 20.5 | 18.5-22.6 | | Calibrator Level C | 25.1 | 22.6-27.6 | | Calibrator Level D | 33.2 | 29.9-36.5 | | Calibrator Level E | 44.9 | 40.4-49.4 | # N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. # O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...