K964062 · Behring Diagnostics, Inc. · JIX · Nov 21, 1996 · Clinical Chemistry
Device Facts
Record ID
K964062
Device Name
N PROTEIN STRANDARD SL
Applicant
Behring Diagnostics, Inc.
Product Code
JIX · Clinical Chemistry
Decision Date
Nov 21, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Intended Use
N Protein Standard SL is intended to be used for the establishment of reference curves.
Device Story
N Protein Standard SL is a liquid human serum-based calibrator containing stabilizers and preservatives. Used with Behring Nephelometer systems (K860894, K892223, K943997) to establish reference curves for various protein assays. Operates by providing known concentrations to calibrate nephelometric immunology assays. Used in clinical laboratory settings by trained personnel. Output allows healthcare providers to quantify specific protein levels in patient samples, aiding in diagnosis and monitoring of protein-related conditions.
Clinical Evidence
Bench testing only. Precision studies on Behring Nephelometer yielded CVs ranging from 0.9% to 3.9%. Stability testing confirmed shelf life of at least 12 months at +2 to +8°C (unopened) and 15 days at +2 to +8°C (opened).
Technological Characteristics
Liquid human serum-based calibrator; contains stabilizers and preservatives. Designed for use with nephelometric systems. Stable for 12 months at 2-8°C.
Indications for Use
Indicated for use as a multi-constituent calibrator for in vitro diagnostic immunology assays on Behring Nephelometer systems to establish reference curves for proteins including IgG, IgA, IgM, C3c, C4, Transferrin, Albumin, Ceruloplasmin, RbP, Kappa/Lambda light chains, alpha-1-antitrypsin, alpha-2-macroglobulin, Haptoglobin, alpha-1-acid glycoprotein, Pre-albumin, beta-2-microglobulin, Ferritin, and IgE.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K991181 — MODIFICATION OF N PROTEIN STANDARD SL · Dade Behring, Inc. · Jun 1, 1999
K012470 — MODIFICATION TO N PROTEIN STANDARD SL · Dade Behring, Inc. · Sep 28, 2001
K962410 — N PROTEIN STANDARD PY · Behring Diagnostics, Inc. · Jul 16, 1996
K031608 — RANDOX LIQUID PROTEIN CALIBRATOR · Randox Laboratories, Ltd. · Aug 8, 2003
K991182 — MODIFICATION OF N/T PROTEIN CONTROL SL · Dade Behring, Inc. · May 19, 1999
Submission Summary (Full Text)
{0}
Behring Diagnostics Inc.
N Protein Standard SL
510(k) Notification
K964062
NOV 21 1996
# ATTACHMENT 2
# 510(k) SUMMARY FOR N PROTEIN STANDARD SL
1. Manufactures Name, Address, Telephone, and contact person, date of preparation:
Manufacture
Behringwerke AG
Postfach 1140
35001 Marburg
Germany
Distributor
Behring Diagnostics Inc.
151 University Avenue
Westwood, MA 02090
617-320-3000
Contact Person: Laura LeBarron
Date of Preparation: October 4, 1996
2. Device Name/ Classification:
N Protein Standard SL: calibrator
Classification Number: class II (862.1150)
3. Identification of the legally marketed device:
N Protein Standard SY
4. Proposed Device Description:
The proposed standard, N Protein Standard SL is a standard prepared from human serum (liquid) with stabilizers and preservative. It is intended to be used together with the Behring Nephelometer Systems (Behring Nephelometer K860894, Behring Nephelometer 100 K892223 and the Behring Nephelometer II K943997) to establish reference curves for the following tests:
IgG Ceruloplasmin
IgA RbP
IgM Ig/L-chain, Kappa
C3c Ig/L-chain, Lambda
C4 IgG 1
Transferin IgG 2
Albumin IgG 3
CONFIDENTIAL
000008
{1}
Behring Diagnostics Inc.
N Protein Standard SL
510(k) Notification
Albumin
α1-antitrypsin (α1-proteinase inhibitor)
α2-macroglobulin
Haptoglobin
α1-acid glycoprotein
Pre-albumin (transthyretin)
IgG 3
IgG 4
β2-microglobulin
Ferritin
IgE
5. Proposed Device Intended Use:
N Protein Standard SL is intended to be used for the establishment of reference curves.
6. Medical device to which equivalence is claimed and comparison information.
N Protein Standard SL is substantially equivalent in intended use to the N Protein Standard SY. Both standards are in vitro diagnostic reagents intended for use as a multi-constituent standard for the establishment of reference curves. The N Protein Standard SL, like the N Protein Standard SY, is a multi-constituent standard in a blood based matrix. Also, both standards are used for the calibration of immunology assays.
The N Protein Standard SL differs from the N Protein Standard SY in that the N Protein Standard SY is a lyophilized standard while the N Protein Standard SL is a liquid standard.
7. Proposed Device Performance Characteristics:
Precision and reproducibility:
Precision studies using one lot of N Protein Standard SL were run on the Behring Nephelometer. The percent CVs ranged from 0.9 to 3.9 %.
Stability
Stability was run according to in-house protocols and the standard was found to be stable for at least 12 months at +2 to +8°C, as originally packaged and for at least 15 days at +2 to +8°C, once opened.
CONFIDENTIAL
300079
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.