RANDOX CALCIUM

K974607 · Randox Laboratories, Ltd. · CJY · Feb 2, 1998 · Clinical Chemistry

Device Facts

Record IDK974607
Device NameRANDOX CALCIUM
ApplicantRandox Laboratories, Ltd.
Product CodeCJY · Clinical Chemistry
Decision DateFeb 2, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1145
Device ClassClass 2

Indications for Use

The Randox Laboratories Limited Calcium Test Kit is an in vitro diagnostic reagent for the quantitative determination of calcium in human serum. This method is based on the interaction of calcium with Arsenazo III to form a coloured complex which can be measured at 650nm. The intensity of the final colour is directly proportional to the calcium concentration. Measurements of calcium are used in the diagnosis and treatment of parathyroid disease, malignant diseases with and without bone involvement, chronic renal failure and a variety of bone diseases. This test kit must be used by suitably qualified laboratory personnel under appropriate laboratory conditions.

Device Story

In vitro diagnostic reagent kit for quantitative calcium measurement in human serum; utilizes Arsenazo III dye interaction to form colored complex; absorbance measured at 650nm; color intensity proportional to calcium concentration. Used in clinical laboratory settings by qualified personnel; results assist physicians in diagnosing/treating parathyroid, malignant, renal, and bone diseases.

Technological Characteristics

In vitro diagnostic reagent; Arsenazo III colorimetric assay; spectrophotometric measurement at 650nm; manual or automated clinical chemistry analyzer compatible.

Indications for Use

Indicated for quantitative determination of calcium in human serum for diagnosis and treatment of parathyroid disease, malignant diseases with/without bone involvement, chronic renal failure, and bone diseases. For use by qualified laboratory personnel.

Regulatory Classification

Identification

A calcium test system is a device intended to measure the total calcium level in serum. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract human figures, possibly representing people or families, with their arms raised. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 FEB - 2 1000 Dr. P. Armstrong Requlatory Affairs Randox Laboratories, Ltd. Ardmore, Diamond Road Crumlin, Co., Antrim, United Kingdom, BT29 4QY Re : K974607 Randox Calcium Requlatory Class: II Product Code: CJY Dated: December 5, 1997 Received: December 10, 1997 Dear Dr. Armstrong: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination-assumes compliance with " the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. {1}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation-entitled, - "Misbranding by -reference to - -------------------------------------------------------------------------------------------------------------------Other general premarket notification" (21 CFR 807.97). information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Steven Litman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ | Page of | 1 of 1 | |---------|--------| |---------|--------| K974607 510(k) Number (if known)_ NOT KNQWN Device Name ___CALCIUM - ARSENAZO III ## Indications For Use : The Randox Laboratories Limited Calcium Test Kit is an in vitro diagnostic reagent for the quantitative determination of calcium in human serum. This method is based on the interaction of calcium with Arsenazo III to form a coloured complex which can be measured at 650nm. The intensity of the final colour is directly proportional to the calcium concentration. Measurements of calcium are used in the diagnosis and treatment of parathyroid disease, malignant diseases with and without bone involvement, chronic renal failure and a variety of bone diseases. This test kit must be used by suitably qualified laboratory personnel under appropriate laboratory conditions. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number K974607 Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...