CA

K981578 · Abbott Laboratories · CJY · Jun 16, 1998 · Clinical Chemistry

Device Facts

Record IDK981578
Device NameCA
ApplicantAbbott Laboratories
Product CodeCJY · Clinical Chemistry
Decision DateJun 16, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1145
Device ClassClass 2

Intended Use

The Calcium assay is used for the quantitation of calcium in human serum, plasma, or urine. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease, and tetany (intermittent muscular contractions or spasms).

Device Story

In vitro diagnostic assay for quantitative calcium measurement in human serum, plasma, or urine. Principle: Arsenazo-III dye reacts with calcium in acid solution to form blue-purple complex; color intensity measured at 600 nm is proportional to calcium concentration. Used in clinical laboratory settings. Provides quantitative results to clinicians for diagnosing and monitoring parathyroid, bone, and renal diseases, and tetany.

Clinical Evidence

Bench testing only. Method comparison against predicate (Hitachi 717) showed correlation coefficients of 0.9771 (serum) and 0.9987 (urine). Precision studies (within-run, between-run, between-day) demonstrated total %CVs between 1.0% and 1.7%. Linearity established up to 27.24 mg/dL (serum) and 30.8 mg/dL (urine). Limit of quantitation is 0.03 mg/dL.

Technological Characteristics

In vitro diagnostic clinical chemistry assay. Uses arsenazo-III dye-based colorimetric detection. Measures absorbance at 600 nm. Compatible with AEROSET System. No specific materials of construction or software algorithm details provided.

Indications for Use

Indicated for the quantitative determination of calcium in human serum, plasma, or urine to aid in the diagnosis and treatment of parathyroid disease, bone diseases, chronic renal disease, and tetany.

Regulatory Classification

Identification

A calcium test system is a device intended to measure the total calcium level in serum. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 1981578 Fax (972) 753-3367 JUN 1 6 1998 # 510(k) Summary Contact Person Submitter's Name/Address Mark Littlefield Abbott Laboratories Section Manager MS 1-8 1920 Hurd Drive Regulatory Affairs Irving, Texas 75038 (972) 518-6062 May 1, 1998 Date of Preparation of this Summary: Ca Device Trade or Proprietary Name: Device Common/Usual Name or Classification Name: Calcium 75CLC/Class II Classification Number/Class: This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. 81578 The assigned 510(k) number is: / #### Test Description: Calcium is an in vitro diagnostic assay for the quantitative determination of calcium in human serum, plasma, or urine. The Calcium assay is a clinical chemistry assay in which arsenazo-III dye reacts with calcium in an acid solution to form a blue-purple complex. The color developed is measured at 600 nm and is proportional to the calcium concentration in the sample. #### Substantial Equivalence: The Calcium assay is substantially equivalent to the Boehringer Mannheim® Calcium assay on the Hitachi® 717 Analyzer assay (K850281) for both the serum and urine applications. Similarities to Boehringer Mannheim: - . Both assays are in vitro clinical chemistry methods. Calcium 510(k) May 1, 1998 Calciuf.lwp Section II Page 1 Image /page/0/Picture/15 description: The image shows the number 0000012 in a bold, sans-serif font. The numbers are black against a white background. The number is evenly spaced and clearly legible. {1}------------------------------------------------ - Both assays can be used for the quantitative determination of calcium. . - . Both assays vield similar clinical results. Differences to Boehringer Mannheim: - There is a minor difference between the assay range. . #### Intended Use: The Calcium assay is used for the quantitation of calcium in human serum, plasma, or urine. #### Performance Characteristics: Comparative performance studies were conducted using the AEROSET™ System. The Calcium assay method comparison yielded acceptable correlation with the Boehringer Mannheim Calcium assay on the Hitachi 717 Analyzer. For the serum application, the correlation coefficient = 0.9771, slope = 1.011, and Y-intercept = 0.456 mg/dL. For the urine application, the correlation coefficient = 0.9987, slope = 1.039, and Y-intercept = 0.048 mg/dL. Precision studies were conducted using the Calcium assay. Within-run, between-run, and between-day studies were performed using two levels of control material. For the serum application, the total %CV is for Level 1/Panel 101 is 1.7% and Level 2/Panel 102 is 1.1%. For the urine application, the total % CV for Level 1/Panel 201 is 1.4% and Level 2/Panel 202 is 1.0%. The Calcium assay is linear up to 27.24 mg/dL for the serum application, and 30.8 mg/dL for the urine application. The limit of quantitation (sensitivity) of the Calcium assay is 0.03 mg/dL. These data demonstrate that the performance of the Calcium assay is substantially equivalent to the performance of the Boehringer Mannheim Calcium assay on the Hitachi 717 Analyzer for both serum and urine applications. Calcium 510(k) May 1, 1998 Calciuf two Section II Page 2 ## 0000013 {2}------------------------------------------------ ### Conclusion: 動物用品の収益をついている。 このコートレイト - ﯽ The Calcium assay is substantially equivalent to the Boehringer Mannheim Calcium assay on the Hitachi 717 Analyzer for the serum and urine applications as demonstrated by results obtained in the studies. Calcium 510(k) May 1, 1998 Calciuf.lwp Section II Page 3 A .. . . . . . . . . . . . . . . . 0000014 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is facing to the right. 14 Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 The 1 6 1998 Mark Littlefield Section Manager, Regulatory Affairs Abbott Laboratories 1920 Hurd Drive 75038 Irving, Texas K981578 Re : Calcium Requlatory Class: II Product Code: CJY Dated: May 1, 1998 Received: May 4, 1998 Dear Mr. Littlefield: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical General requlation (21 CFR Part 820) and that, Devices: through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Fail Failure to In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ Page 2 ছে তার Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Steven Litman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (if known): K981578 Calcium Device Name: Indications For Use: The Calcium assay is used for the quantitation of calcium in human serum, plasma, or urine. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease, and tetany (intermittent muscular contractions or spasms). | Division Sign-Off) | | |-----------------------------------------|---------| | Division of Clinical Laboratory Devices | | | 510(k) Number | K981575 | (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrey se of CDRH, Office of Device Evaluation (ODE) 机 Prescription Use OR Over-The-Counter Use (Per 21 CFR 801.109) (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...