MEG-2 BLOOD GLUCOSE MONITORING SYSTEM, MEG-2 50'S TEST STRIP, MEG-2 CONTROL SOLUTION
Applicant
Apex BioTechnology Corp.
Product Code
CGA · Clinical Chemistry
Decision Date
Aug 31, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Intended Use
MEG-2 Blood Glucose Monitoring System: The MEG-2 Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. Testing is done outside the body (In Vitro diagnostic use). It is indicated for lay use by people with diabetes, as an aid to monitoring levels in Diabetes Mellitus and should only be used by a single patient. This system should not be shared. It is not indicated for the diagnosis or screening of diabetes or for neonatal use. Alternative site testing can be used only during steady-state blood glucose conditions. MEG-2 Blood Glucose Test Strips: The MEG-2 Blood Glucose Test Strips are to be used with the MEG-2 Blood Glucose Meter to quantitatively measure glucose in capillary whole blood taken from fingertips, palm, or forearm. It is for use outside of the body (in vitro diagnostic use). The MEG-2 Blood Glucose Monitoring Systems are plasma-calibrated for easy comparison to lab results. They are intended for lay use by people with diabetes and should only be used by a single patient. This system should not be shared. They are not indicated for the diagnosis or screening of diabetes or for neonatal use. MEG-2 Glucose Control Solution: The MEG-2 Glucose Control Solutions are used with the MEG-2 and MEG-2 Multi Blood Glucose Monitoring Systems to indicate appropriate user technique and to indicate that the test strip and meter are functioning properly.. MEG-2 Multi Blood Glucose Monitoring System: The MEG-2 Multi Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. Testing is done outside the body (In Vitro diagnostic use). It is indicated to be use for multiple patients in a clinical setting by healthcare professionals, as an aid to monitoring levels in Diabetes Mellitus. This system is only used with single-use, auto-disabling lancing device. It is not indicated for the diagnosis or screening of diabetes or for neonatal use. Alternative site testing can be used only during steady-state blood glucose conditions. MEG-2 Multi Blood Glucose Test Strips: The MEG-2 Multi Blood Glucose Test Strips are to be used with the MEG-2 Multi Blood Glucose Meter; it measures glucose in capillary whole blood taken from a fingertip, palm, or forearm. It is for use outside of the body (in vitro diagnostic use). It is indicated in a clinical setting to be used for multiple patients by healthcare professionals. This system is only used with single-use auto-disabling lancing device
Device Story
System uses electrochemical biosensor technology with glucose oxidase to measure glucose in capillary whole blood (fingertip, palm, forearm). Capillary action draws blood into test strip chamber; meter displays result in 6 seconds. MEG-2 intended for single-patient home use; MEG-2 Multi intended for multi-patient clinical use with auto-disabling lancing devices. Healthcare providers/patients use results to monitor diabetes mellitus. System includes meter, test strips, and control solutions for performance verification. Disinfection protocols validated for multi-patient use. Provides 7-, 14-, and 30-day averages; stores 300 results.
Clinical Evidence
Bench testing only. Precision studies (within-run and between-run) performed over 20 days. Linearity evaluated across 15-610 mg/dL range (R^2 > 0.998). Accuracy compared to YSI reference method (N=111 samples); 100% of results within ±15 mg/dL for <75 mg/dL and within ±15% for ≥75 mg/dL. Lay-user study confirmed accuracy across finger, palm, and forearm sites. Interference testing performed for various substances. Robustness testing confirmed performance after 1,825 (home) and 10,950 (clinical) disinfection cycles.
Technological Characteristics
Glucose test system; electrochemical sensing. Features 3-button interface, battery-powered. Test strips include automatic calibration code detection. Control solutions contain red dye. Software performs quantitative glucose measurement.
Indications for Use
Indicated for people with diabetes (lay use) or multiple patients in clinical settings (professional use) to monitor blood glucose levels. Not for diagnosis/screening of diabetes or neonatal use. Alternative site testing restricted to steady-state conditions.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
GlucoSure STAR meter and test strips
Contrex Plus glucose control solutions
Related Devices
K120448 — MEG-2B BLOOD GLUCOSE MONITORING SYSTEM, PRO BLOOD GLUCOSE MONITORING SYSTEM, BLOOD GLUCOSE TEST STRIPS, GLUCOSE CONTROL · Apex BioTechnology Corp. · Apr 6, 2012
K123016 — VERI-Q SELF-TESTING AND VERI-Q PLUS MGD-2001 BLOOD GLUCOSE MONITORING SYSTEMS · Mico Biomed Co., Ltd. · Jun 18, 2013
K140210 — RIGHTEST BLOOD GLUCOSE MONITORING SYSTEM ( GM720, RIGHTEST PROFESSIONAL BGMS GM720 · Bionime Corporation · Aug 22, 2014
K103116 — AP-2000, AP-2010, AND AP-2020 BLOOD GLUCOSE MONITORING SYSTEM · Bestgen Biotech Corp. · Mar 13, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
AUG 3 1 2011
# 510(k) Summary
な
i
| Submitter: | Hsue-mei Lee<br>Manager of Quality Assurance Department<br>Apex BioTechnology Corp.<br>No. 7, Li-Hsin Road V, Hsinchu Science Park<br>Hsinchu, 30078<br>CHINA (TAIWAN)<br><br>email: hsue-mei@apexbio.com<br>Phone: 011-886-3-5641952<br>FAX: 011-886-3-5678302 |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Hsue-mei Lee<br>Manager of Quality Assurance Department<br>Apex BioTechnology Corp.<br>No. 7, Li-Hsin Road V, Hsinchu Science Park<br>Hsinchu, 30078<br>CHINA (TAIWAN)<br><br>email: hsue-mei@apexbio.com<br>Phone: 011-886-3-5641952<br>FAX: 011-886-3-5678302 |
| Date Prepared: | August 18, 2011 |
| Trade Names: | MEG-2 Blood Glucose Monitoring System, MEG-2 Blood Glucose Test<br>Strips, MEG-2 Glucose Control Solutions, MEG-2 Multi Blood Glucose<br>Monitoring System, MEG-2 Multi Blood Glucose Test Strips |
| Classification: | Glucose test system, 21 CFR 862.1345, Class II<br><br>Single (Specified) Analyte Controls (Assayed And Unassayed), 21 CFR<br>862.1660, Class I, reserved |
| Product Codes: | CGA, NBW, JJX |
| Predicate Devices: | GlucoSure STAR meter and test strips<br><br>Contrex Plus glucose control solutions |
| Device Description: | The MEG-2 blood glucose monitoring system consists of a meter, test<br>strips and three levels of control solution. It is used for testing of blood<br>glucose by self-testers at home.<br><br>The MEG-2 Multi blood glucose monitoring system consists of a meter,<br>test strips and three levels of control solution. It is used for testing of<br>blood glucose by professional testers in healthcare facilities. |
| 510(k) Summary (Continued) | |
| <b>Intended Use:</b><br>MEG-2 Blood Glucose Monitoring System: The MEG-2 Blood Glucose<br>Monitoring System is intended for the quantitative measurement of glucose in fresh<br>capillary whole blood samples drawn from the fingertips, forearm, or palm. Testing<br>is done outside the body (In Vitro diagnostic use). It is indicated for lay use by<br>people with diabetes, as an aid to monitoring levels in Diabetes Mellitus and should<br>only be used by a single patient. This system should not be shared. It is not<br>indicated for the diagnosis or screening of diabetes or for neonatal use. Alternative<br>site testing can be used only during steady-state blood glucose conditions.<br><br>MEG-2 Blood Glucose Test Strips: The MEG-2 Blood Glucose Test Strips are to<br>be used with the MEG-2 Blood Glucose Meter to quantitatively measure glucose in<br>capillary whole blood taken from fingertips, palm, or forearm. It is for use outside<br>of the body (in vitro diagnostic use). The MEG-2 Blood Glucose Monitoring<br>Systems are plasma-calibrated for easy comparison to lab results. They are intended<br>for lay use by people with diabetes and should only be used by a single patient.<br>This system should not be shared. They are not indicated for the diagnosis or<br>screening of diabetes or for neonatal use.<br><br>MEG-2 Glucose Control Solution: The MEG-2 Glucose Control Solutions are used<br>with the MEG-2 and MEG-2 Multi Blood Glucose Monitoring Systems to indicate<br>appropriate user technique and to indicate that the test strip and meter are<br>functioning properly..<br><br>MEG-2 Multi Blood Glucose Monitoring System: The MEG-2 Multi Blood<br>Glucose Monitoring System is intended for the quantitative measurement of glucose<br>in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm.<br>Testing is done outside the body (In Vitro diagnostic use). It is indicated to be use<br>for multiple patients in a clinical setting by healthcare professionals, as an aid to<br>monitoring levels in Diabetes Mellitus. This system is only used with single-use,<br>auto-disabling lancing device. It is not indicated for the diagnosis or screening of<br>diabetes or for neonatal use. Alternative site testing can be used only during steady-<br>state blood glucose conditions.<br><br>MEG-2 Multi Blood Glucose Test Strips: The MEG-2 Multi Blood Glucose Test<br>Strips are to be used with the MEG-2 Multi Blood Glucose Meter; it measures<br>glucose in capillary whole blood taken from a fingertip, palm, or forearm. It is for<br>use outside of the body (in vitro diagnostic use). It is indicated in a clinical setting to<br>be used for multiple patients by healthcare professionals. This system is only used<br>with single-use auto-disabling lancing device | |
·
.
{1}------------------------------------------------
## 510(k) Summary (Continued)
」な
ん
.
.
.
{2}------------------------------------------------
### 510(k) Summary (Continued)
ア
.
| Comparison of<br>Technological<br>Characteristics: | The MEG-2 meter uses the same test algorithm as the predicate. The meter<br>has been modified from a 4-button system to a 3-button one with a new<br>battery type. The test strip and test strip holder have been modified to allow<br>automatic detection of the calibration code upon insertion of the test strip.<br>The core functionality of the predicate meter has been retained in the MEG-2<br>meter. |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The MEG-2 Multi meter and MEG-2 Multi test strips are identical to their<br>MEG-2 counterparts except for labeling to distinguish that they are for<br>professional use (MEG-2 Multi) and home use (MEG-2), respectively. |
| | The MEG-2 control solutions have the same formula as the predicate with<br>the addition of a red dye and slight adjustment of the glucose values for each<br>level of control solution. |
| Non-Clinical<br>Testing: | Testing was conducted as follows: Software verification and validation plus<br>software integration testing, Battery Life verification, Test Strip Holder Life<br>verification, Control Solution Qualification, Control solution real-time and<br>use-life stability testing, Drop testing, EMC and Electrical Safety testing,<br>Linearity testing, Precision testing, Hematocrit range testing, Interference<br>testing, test strip stability testing, test strip open bottle use life testing, and<br>testing of the test strip autocode manufacturing and quality control processes.<br>Results demonstrate substantial equivalence to the predicate devices. |
| Clinical Testing | An Accuracy and User Performance Evaluation study was conducted with<br>141 test subjects. Blood testing was done by healthcare professionals and<br>non-professional self-testers against the YSI reference instrument. Control<br>solution testing was also conducted. Results demonstrate substantial<br>equivalence to the predicate device. |
| Conclusion: | Clinical and non-clinical testing demonstrated that the MEG-2 and MEG-2<br>Multi systems (meters, test strips and control solutions) perform in a<br>substantially equivalent manner to that of the predicate. We conclude that<br>the MEG-2 and MEG-2 Multi meters, test strips and control solutions are<br>substantially equivalent to their predicate. |
・
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white and appears to be a scanned image.
Food & Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993
AUG 3 1 2011
Apex BioTechnology Corporation c/o Hsue-mei Lee Manager of Quality Assurance Department No. 7 Li-Hsin Road V, Hsinchu Science Park Hsinchu, China (Taiwan) 30078
Re: k101204
> Trade Name: MEG-2 Blood Glucose Monitoring System, MEG-2 Multi Blood Glucose Monitoring System Regulation Number: 21 CFR §862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Codes: NBW, CGA, JJX Dated: August 19, 2011 Received: August 22, 2011
Dear Hsue-mei Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to Mav 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, thereforc, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2 -
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Viro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm.
Sincerely yours,
signature
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
{5}------------------------------------------------
#### Indications for Use Statement
510(k) Number (if known): k101204
Device Name: MEG-2 Blood Glucose Monitoring System
Indications for Use:
MEG-2 Blood Glucose Monitoring System:
The MEG-2 Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, Testing is done outside the body (In Vitro diagnostic use). It is indicated for lay use by people with diabetes, as an aid to monitoring levels in Diabetes Mellitus and should only be used by a single patient. This system should not be shared. It is not indicated for the diagnosis or screening of diabetes or for neonatal use. Atternative site testing can be used only during steady-state blood glucose conditions.
MEG-2 Blood Glucose Test Strips:
The MEG-2 Blood Glucose Test Strips are to be used with the MEG-2 Blood Glucose Meter to quantitatively measure glucose in capillary whole blood taken from fingertips, palm, or forearm. It is for use outside of the body (in vitro diagnostic use). They are intended for lay use by people with diabetes and should only be used by a single patient. This system should not be shared. They are not indicated for the diagnosis or screenng of diabetes or for neonatal use.
MEG-2 Glucose Control Solution:
The MEG-2 Glucose Control Solutions are used with the MEG-2 and MEG-2 Multi Blood Glucose Monitoring Systems to indicate appropriate user technique and to indicate that the test strip and meter are functioning properly.
Prescription Use Over-The-Counter Use X = AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign Off
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k): k101204
Page 1 of 1
1 of 2
{6}------------------------------------------------
### Indications for Use Statement
510(k) Number (if known): k101204
Device Name: MEG-2 Multi Blood Glucose Monitoring System
MEG-2 Multi Blood Glucose Monitoring System:
The MEG-2 Multi Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. Testing is done outside the body (In Vitro diagnostic use). It is indicated to be use for multiple patients in a clinical setting by healthcare professionals, as an aid to monitoring levels in Diabetes Mellitus. This system is only used with single-use, auto-disabling lancing device. It is not indicated for the diagnosis or screening of diabetes or for neonatal use. Alternative site testing can be used only during steady-state blood glucose conditions.
MEG-2 Multi Blood Glucose Test Strips:
The MEG-2 Multi Blood Glucose Test Strips are to be used with the MEG-2 Multi Blood Glucose Meter; it measures glucose in capillary whole blood taken from a fingertip, palm, or forearm. It is for use outside of the body (in vitro diagnostic use). It is indicated in a clinical setting to be used for multiple patients by healthcare professionals. This system is only used with single-use, auto-disabling lancing device.
Prescription Use Over-The-Counter Use × AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Richie Glenn Coffey
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k): k101204
Page 1 of 1
2 of 2
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.