VERI-Q SELF-TESTING AND VERI-Q PLUS MGD-2001 BLOOD GLUCOSE MONITORING SYSTEMS

K123016 · Mico Biomed Co., Ltd. · NBW · Jun 18, 2013 · Clinical Chemistry

Device Facts

Record IDK123016
Device NameVERI-Q SELF-TESTING AND VERI-Q PLUS MGD-2001 BLOOD GLUCOSE MONITORING SYSTEMS
ApplicantMico Biomed Co., Ltd.
Product CodeNBW · Clinical Chemistry
Decision DateJun 18, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The Veri-Q self-testing MGD-2001 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from fingertips, palm, or forearm. The Veri-Q self-testing MGD-2001 Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared. The Veri-Q self-testing MGD-2001 Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Veri-Q self-testing MGD-2001 Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly). The Veri-Q self-testing MGS-01 test strips are for use with the Veri-Q self-testing MGD-2001 meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from fingertips, palm, or forearm. The Veri-Q self-testing Glucose Control Solutions are for use with the Veri-Q self-testing MGD-2001 meter and Veri-Q self-testing MGS-01 test strips to check that the meter and test strips are working together properly and providing accurate results. The Veri-Q plus MGD-2001 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose (sugar) in fresh capillary whole blood drawn from fingertips, palm, or forearm. The Veri-Q plus MGD-2001 Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use). The Veri-Q plus MGD-2001 Blood Glucose Monitoring System is intended for multi-patient use in a professional healthcare setting, as an aid to monitor the effectiveness of diabetes control. This system is only used with single-use, auto-disabling lancing devices. The Veri-Q plus MGD-2001 Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly). The Veri-Q plus MGS-01 test strips are for use with the Veri-Q plus MGD-2001 meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from fingertips, palm, or forearm. The Veri-Q plus Glucose Control Solutions are for use with the Veri-Q plus MGD-2001 meter and Veri-Q plus MGS-01 test strips to check that the meter and test strips are working together properly and providing accurate results.

Device Story

System uses electrochemical biosensor technology with glucose oxidase enzyme to measure capillary blood glucose. User applies 0.5 µL blood sample to test strip via capillary action; meter detects electrons generated by reaction; magnitude of current proportional to glucose concentration. Results displayed in 5 seconds. Self-testing version for home use by patients; 'plus' version for multi-patient professional settings using auto-disabling lancing devices. System includes meter, test strips, and control solutions. Provides quantitative glucose readout to aid diabetes management. Not for diagnosis or screening. Cleaning/disinfection validated for multi-patient use.

Clinical Evidence

Clinical performance evaluated via user performance study (n=152 lay users) and alternative site testing (n=100 lay users). Accuracy compared to YSI 2300 reference method. Results showed 100% of results within ±15 mg/dL (for <75 mg/dL) and 100% within ±20% (for ≥75 mg/dL) for fingerstick testing. AST accuracy also met criteria. Analytical performance included precision, linearity, interference, hematocrit (20-60%), altitude, and temperature/humidity studies.

Technological Characteristics

Glucose oxidase-based electrochemical biosensor. No-code meter technology with automatic strip calibration. Dimensions/form factor not specified. Standalone device. Sterilization method not specified.

Indications for Use

Indicated for quantitative measurement of glucose in fresh capillary whole blood (fingertip, palm, forearm) as an aid to monitor diabetes control. Intended for single-patient home use (self-testing) or multi-patient professional healthcare use. Contraindicated for neonatal use, diabetes diagnosis/screening, critically ill patients, dehydrated/hypotensive/shock patients, or individuals in hyperglycemic-hyperosmolar state. Alternative site testing restricted to steady-state glucose conditions.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary K123016 This summary of 510(k)-safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92. Date: -_ 06/01/2013 # JUN 1 8 2013 #### 1. Submitter: MiCoBioMed Co., Ltd. 55-13, Singeonji-Dong, Anseong-Si, Gyenggi-Do, 456-781, Republic of Korea Tel : +82-31-8056-5940 Fax : +82-31-8056-5518 #### 2. Submission Correspondent: Priscilla Chung LK Consulting Group USA, Inc. 1515 E Katella Ave Unit 2115, Anaheim, CA 92805 Phone: 714-202-8637 Fax: 714-409-3357 Email: info@lkconsultinggroup.com #### 3. Device: | Proprietary Name: | Veri-Q self-testing MGD-2001 Blood Glucose. Monitoring System Veri-Q plus MGD-2001 Blood Glucose Monitoring System | |------------------------------|--------------------------------------------------------------------------------------------------------------------| | Common Name: | Blood Glucose Test System | | Classification Name: | System, Test, Blood Glucose, Over The Counter Glucose Oxidase Single(specified) analyte controls | | Classification: | Class II, 21 CFR 862.1345 | | Classification Product Code: | NBW | | Subsequent Product Codes | CGA, JJX | #### 4. Predicate Device: CareSens N(K083468) by i-SENS, Inc. {1}------------------------------------------------ #### 5. Description: The proposed Veri-Q self testing/ plus MGD-2001 Blood Glucose Monitoring System consists of a meter, test strips, control solutions (3 levels), a lancing device and sterile lancets. The blood glucose test system is an in vitro diagnostic device designed for measuring the concentration of glucose in whole blood sample by means of an electrical current produced in the test strip and sent to the meter for measurement. The Veri-Q testing/plus MGD-2001 meter is a no-code meter, so the responsibility of coding has been removed from the user. Each individual test strip contains technology that allows it to automatically calibrate and code the meter once it is inserted. #### 6. Indications for use: 6.1 Veri-Q self-testing MGD-2001 Blood Glucose Monitoring System The Veri-Q self-testing MGD-2001 Blood Glucose Monitoring System is intended for use in The quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from fingertips, palm, or forearm. The Veri-Q self-testing MGD-2001 Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared. The Veri-Q self-testing MGD-2001 Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Veri-Q sclf-testing MGD-2001 Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly). The Veri-Q self-testing MGS-01 test strips are for use with the Veri-Q self-testing MGD-2001 meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from fingertips, palm, or forearm. The Veri-O self-testing Glucose Control Solutions are for use with the Veri-Q self-testing MGD-2001 meter and the Veri-Q self-testing MGS-01 test strips to check that the meter and test strips are working together properly and providing accurate results. 6.2 Veri-Q plus MGD-2001 Blood Glucose Monitoring System The Veri-Q plus MGD-2001 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose (sugar) in fresh capillary whole blood drawn from fingertips, palm, or forearm. The Veri-Q plus MGD-2001 Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use). The Veri-Q plus MGD-2001 Blood Glucose Monitoring System is intended for multi-patient use in a professional healthcare setting, as an aid to monitor the effectiveness of diabetes control. This system is only used with single-use, auto-disabling lancing devices. The Veri-O plus MGD-2001 Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly). {2}------------------------------------------------ The Veri-O plus MGS-01 test strips are for use with the Veri-Q plus MGD-2001 meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from fingertips, palm, or forearm. The Veri-Q plus Glucose Control Solutions are for use with the Veri-Q plus MGD-20(01 meter and the Veri-Q plus MGS-01 test strips to check that the meter and test strips are working together properly and providing accurate results. #### 7. Technological Characteristics: The Veri-Q self-testing/plus MGD-2001 Blood Glucose Monitoring System has the same fundamental scientific technology as the predicate device. The indications for use, technology, design of glucose meter, and strips are almost the same. The major difference is that for the Veri-Q self-testing/plus meter, no coding is required on user side but the predicate device employs auto-coding method. There are also minor difference such as shelf life and memory capacity. #### 8. Performance Data: The testing for the Veri-Q self-testing/plus MGD-2001 Blood Glucose Monitoring System was performed in accordance with ISO 15197:2003. Clinical evaluation included method comparison, user performance and alternative-site blood glucose measurement. Non-clinical performance evaluations were conducted to establish the performance, functionality and reliability characteristics of the Veri-Q self-testing/plus MGD-2001 Blood Glucose Monitoring System. The device passed all of the tests based on pre-determined Pass/Fail criteria. #### 9. Conclusions: MiCoBioMed Co., Ltd. concludes that the Veri-O sclf-testing/plus MGD-2001 Blood Glucose Monitoring System is safe and effective also, substantially equivalent to the predicate device. {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wavy lines representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. ### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 18, 2013 MiCoBioMed Co., Ltd. C/O Priscilla Chung C/O LK Consulting Group 1515 E. Katella Ave, Unit 2115 ANAHEIM CA 92805 Re: K123016 Trade/Device Name: Veri-Q self-testing MGD-2001 Blood Glucose Monitoring System Veri-Q plus MGD-2001 Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW, CGA, JJX Dated: May 08, 2013 Received: May 09 2013 Dear Ms. Chung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices-good-manufacturing-practice-labeling-and-prohibitions-against-misbranding-andadulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2-Ms. Chung If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. # Carol C. Benson -S for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use ### 510(k) Number: K123016 Device Name: Veri-Q self-testing MGD-2001 Blood Glucose Monitoring System Indication for use: The Veri-Q self-testing MGD-2001 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from fingertips, palm, or forearm. The Veri-Q self-testing MGD-2001 Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared. The Veri-Q self-testing MGD-2001 Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Veri-Q self-testing MGD-2001 Blood Glucose Monitoring System should not be used for the diagnosis of on screening for diabetes or for neonatal use. Alternative site testing should be done only during steadystate times (when glucose is not changing rapidly), The Veri-Q self-testing MGS-01 test strips are for use with the Veri-Q self-testing MGD-2001 meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from fingertips, palm. or forcarm. The Veri-Q self-tcsting Glucose Control Solutions are for use with the Veri-Q self-testing MGD-2001 meter and the Veri-Q self-testing MGS-01 test strips to check that the meter and test strips are working together properly and providing accurate results. Prescription Use AND/OR Over-The-Counter_Use_ (Part 21CFR801 Subpart D) (Part 21CFR807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH. Office of In Vitro and Radiological Health (OIR) # Katherine Serrano -S Division Sign-Off Office of In Vitro and Radiological Health Evaluation and Safety 510(k) k123016 Page I of 2 {6}------------------------------------------------ ## Indications for Use ### 510(k) Number: K123016 Device Name: Veri-Q plus MGD-2001 Blood Glucose Monitoring System Indication for use: The Veri-Q plus MGD-2001 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose (sugar) in fresh capillary whole blood drawn from fingertips, palm, or forearm. The Veri-Q plus MGD-2001 Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use). The Veri-Q plus MGD-2001 Blood Glucose Monitoring System is intended for multi-patient use in a professional healtheare setting, as an aid to monitor the effectiveness of diabetes control. This system is only used with single-use, auto-disabling lancing devices. The Veri-Q plus MGD-2001 Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly). The Veri-Q plus MGS-01 test strips are for use with the Veri-Q plus MGD-2001 meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from fingertips, palm, or forearm. The Veri-Q plus Glucose Control Solutions are for use with the Veri-Q plus MGD-2001 meter and the Veri-Q plus MGS-01 test strips to check that the meter and test strips are working together properly and providing accurate results. Prescription Use AND/OR Over-The-Counter Use (Part-21CFR801-Subpart-D) (Part 21CFR807 Sübpart C) (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In Vitro and Radiological Health (OIR) Katherine Serrano -S Division Sign-Off Office of In Vitro and Radiological Health Evaluation and Safety 510(k) k123016 Page 2 of 2
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...