Software as a Medical Device, Therapeutic, Pediatric
Indications for Use
The GlucoStabilizer Insulin Dosing Calculator 3.0 is designed for use by order of a physician for hospitalized patients. It is intended to evaluate the current as well as cumulative patient blood glucose values, and based on those measurements, calculate and recommend a dose of insulin or dextrose to drive the either up or down towards a predetermined target range. Once that target blood glucose range has been reached, the system's function is to recommend dosing of insulin or dextrose for the purpose of maintaining the patient's blood glucose level in that target range. The system is programmed to provide intravenous dosing calculations of insulin and dextrose for both adult and pediatric (ages 2 to <18 years) patients with no known insulin allergies. The GlucoStabilizer Insulin Dosing Calculator's programmed logic is not a substitute for, but rather an assist to clinical reasoning. The measurements and calculations generated by the GlucoStabilizer Insulin Dosing Calculator are intended to be used by qualified and trained medical personnel in evaluating patient conditions in conjunction with clinical history, symptoms, and other diagnostic measurements, as well as the medical professional's clinical judgment. No medical decisions should be based solely on the recommended guidance provided by this software program.
Device Story
Web-based software; automates intravenous insulin/dextrose dosing calculations for hospitalized patients. Inputs: current and cumulative blood glucose values; patient weight (for pediatric patients). Operation: programmed logic evaluates glucose trends; calculates recommended insulin/dextrose drip rates to reach/maintain target ranges. Output: dosing recommendations displayed to clinicians. Used in hospital settings by trained medical personnel. Features: multi-view screen for managing multiple patient treatment programs; alerts for glucose testing/monitoring. Clinical role: decision support tool; assists clinical reasoning; does not replace professional judgment. Benefits: standardized dosing; improved glucose management efficiency.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and intended use comparisons.
Technological Characteristics
Web-based software solution; programmed logic for intravenous insulin/dextrose dosing calculations. Includes pediatric-specific protocol using patient weight and 3-digit multiplier. Features multi-view interface for multi-patient monitoring. Connectivity: networked/web-based.
Indications for Use
Indicated for hospitalized patients, including adults and pediatric patients (ages 2 to <18 years), requiring intravenous insulin or dextrose dosing to achieve or maintain blood glucose target ranges. Contraindicated for patients with known insulin allergies or patients under the age of 2 years.
Regulatory Classification
Identification
A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 15, 2014
Medical Decision Network, LLC C/O Jane Keathley, MS Regulatory Project Manager 2220 Ivy Road, Suite 403 Charlottesville, Virginia 22903
Re: K141321
Trade/Device Name: GlucoStabilizer® Insulin Dosing Calculator Regulation Number: 21 CFR 868.1890 Regulation Name: Predictive pulmonary-function value calculator Regulatory Class: II Product Code: NDC Dated: August 5, 2014 Received: August 6, 2014
Dear Ms. Keathley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/0/Picture/10 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular design with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle, there is a stylized graphic of three human profiles facing to the right, stacked on top of each other.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Mary S. Runner -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141321
Device Name GlucoStabilizer Insulin Dosing Calculator
#### Indications for Use (Describe)
The GlucoStabilizer Insulin Dosing Calculator 3.0 is designed for use by order of a physician for hospitalized patients. It is intended to evaluate the current as well as cumulative patient blood glucose values, and based on those measurements, calculate and recommend a dose of insulin or dextrose to drive the either up or down towards a predetermined target range.
Once that target blood glucose range has been reached, the system's function is to recommend dosing of insulin or dextrose for the purpose of maintaining the patient's blood glucose level in that target range. The system is programmed to provide intravenous dosing calculations of insulin and dextrose for both adult and pediatric (ages 2 to <18 years) patients with no known insulin allergies.
The GlucoStabilizer Insulin Dosing Calculator's programmed logic is not a substitute for, but rather an assist to clinical reasoning. The measurements and calculations generated by the GlucoStabilizer Insulin Dosing Calculator are intended to be used by qualified and trained medical personnel in evaluating patient conditions in conjunction with clinical history, symptoms, and other diagnostic measurements, as well as the medical professional's clinical judgment. No medical decisions should be based solely on the recommended guidance provided by this software program.
#### Contraindications:
The device is not intended for use with patients with known insulin allergies or patients under the age of 2 years.
### Warnings and Precautions:
The GlucoStabilizer Insulin Dosing Calculator's programmed logic is not a substitute for, but rather an assist to clinical reasoning. The measurements and calculations generated by the GlucoStabilizer Insulin Dosing Calculator are intended to be used by qualified and trained medical persomel in evaluating patient conditions in conjunction with clinical history, symptoms, and other diagnostic measurements, as well as the medical professional's clinical judgment. No medical decisions should be based solely on the recommended guidance provided by this software program.
Type of Use (Select one or both, as applicable)
ال Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for MDN, which stands for "Sullivan Street O'Sullivan Network, LLC". The logo is in bold, dark blue letters. The text "Sullivan Street O'Sullivan Network, LLC" is in a smaller font size and is located below the MDN letters.
September 10, 2014
#### 5. 510(k) Summary
| Date: | Sep-10-2014 | |
|--------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| Owner/Submitter: | Greg Menke<br>2220 Ivy Road, Suite 403<br>Charlottesville, VA 22903 | Medical Decision Network, LLC |
| | Phone: 866-791-6108<br>Fax: 434-202-7282 | |
| Contact: | Jane Keathley, MS<br>Regulatory Project Manager<br>2220 Ivy Road, Suite 403<br>Charlottesville, VA 22903 | Medical Decision Network, LLC |
| | Phone: (434) 409-3535 | Email: jane.keathley@mdnllc.net |
| Type of 510(k) Submission: | | Traditional 510(k) |
| Trade Name: | | GlucoStabilizer® Insulin Dosing<br>Calculator<br>(Formerly Clarian Glucose Stabilizer<br>Insulin Dosing Calculator) |
| Common Name: | | GlucoStabilizer<br>(Formerly MDN-CGSTM Insulin Dosing<br>Calculator) |
| Establishment Registration Number: | | 3010817588 |
| Review Panel/Medical Specialty:<br>Product Code:<br>Classification Name:<br>Classification Number:<br>Class: | | Anesthesiology<br>NDC<br>Drug Dose Calculator<br>21 CFR 868.1890<br>2 |
| Predicate Device: | | MDN-CGS™ Insulin Dosing Calculator (K071713)<br>Glytec LLC, Glucommander™ System (K113853) |
{5}------------------------------------------------
Description of Device: The GlucoStabilizer Insulin Dosing Calculator 3.0 is a Web-based software solution that automates calculations used by healthcare professionals to determine the appropriate intravenous insulin drip rate necessary to manage blood glucose levels across a variety of patient populations. GlucoStabilizer also provides alerts for subsequent blood glucose testing and monitoring. The GlucoStabilizer software release described in this 510(k) submission contains a modification that allows its use for pediatric patients. The pediatric protocol considers patient weight in insulin calculations for patients under the age of 18 years and provides a 3-digit multiplier for more granular calculation results. In addition, this release of GlucoStabilizer provides an optional 'multi-view'screen that allows healthcare personnel to more easily manage multiple, on-going patient treatment programs. Other features and technological characteristics of GlucoStabilizer are unchanged. Intended Use: The GlucoStabilizer Insulin Dosing Calculator 3.0 is designed for use by order of a physician for hospitalized patients. It is intended to evaluate the current as well as cumulative patient blood glucose values, and based on those measurements, calculate and recommend a dose of insulin or dextrose to drive the blood glucose level either up or down towards a predetermined target range. Once that target blood glucose range has been reached, the system's function is to recommend dosing of insulin or dextrose for the purpose of maintaining the patient's blood glucose level in that target range. The system is programmed to provide intravenous dosing calculations of insulin and dextrose for both adult and pediatric (ages 2-<18) patients with no known insulin allergies. The GlucoStabilizer Insulin Dosing Calculator's programmed logic is not a substitute for, but rather an assist to clinical reasoning. The measurements and calculations generated by the GlucoStabilizer Insulin Dosing Calculator are intended to be used by qualified and trained medical personnel in evaluating patient
{6}------------------------------------------------
| | conditions in conjunction with clinical history, symptoms, and other diagnostic measurements, as well as the medical professional's clinical judgment. No medical decisions should be based solely on the recommended guidance provided by this software program. |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison with<br>Predicate Device: | The submission device and the predicate devices have the same general intended use and similar indications, technological characteristics, and principles of operation. Each of the devices is intended to be used by trained clinicians. |
| | This release of GlucoStabilizer includes support for use with pediatric patients, similar to Glytec LLC, Glucommander™ System. |
| Conclusion: | GlucoStabilizer Insulin Dosing Calculator 3.0 has the same technological characteristics and principles of operation as the MDN-CGSTM Insulin Dosing Calculator and Glytec LLC, Glucommander™ System. GlucoStabilizer 3.0 has the same intended use and similar indications as the Glucommander System. The minor technological differences between the GlucoStabilizer 3.0 and its predicates do not present any new issues of safety or effectiveness. Thus, the GlucoStabilizer Insulin Dosing Calculator 3.0 is substantially equivalent to the MDN-CGS™ Insulin Dosing Calculator (K0717130) and the Glytec LLC, GlucommanderTM System (K113853). |
September 10, 2014
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.