The MDN-CGS™ Insulin Dosing Calculator is intended to evaluate the current patient blood glucose values. Based on the measurements, the software will calculate either a dose of glucose or insulin to drive the blood glucose level up or down toward a predetermined target range. Once that target blood glucose range has been reached, the software program calculates dosing of insulin or glucose for the purpose of maintaining the patient's blood glucose level in that target range. The system is programmed to provide intravenous dosing calculations of glucose or insulin. The device is intended for use with patients with no known insulin allergies and for patients over the age of 18. The MDN-CGSTM Insulin Dosing Calculator's programmed logic is a tool and not a substitute for, but rather an assist to, clinical reasoning. The measurements and calculations generated by the MDN-CGSTM Insulin Dosing Calculator are intended to be used by qualified and trained medical personnel directed by physician order. No medical decisions are to be made solely on the guidance provided by this software program, but are to include consideration of clinical history, symptoms, other diagnostic measurements, and the professional's clinical judgment.
Device Story
Web-based software tool for clinical decision support; automates intravenous insulin and glucose dosing calculations. Inputs: current patient blood glucose values. Processing: programmed logic calculates required insulin/glucose doses to reach or maintain target blood glucose ranges. Outputs: dosing recommendations and alerts for blood glucose monitoring. Used in clinical settings by trained medical personnel. Acts as an assist to clinical reasoning; does not replace professional judgment. Benefits: facilitates glucose management in patients requiring intravenous therapy.
Technological Characteristics
Web-based software application; standalone architecture. Provides automated intravenous insulin/glucose dosing calculations and monitoring alerts. Operates as a clinical decision support tool.
Indications for Use
Indicated for patients over age 18 with no known insulin allergies requiring intravenous insulin or glucose dosing to manage or maintain blood glucose levels within a target range. Intended for use by qualified, trained medical personnel under physician order.
Regulatory Classification
Identification
A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for MDN, which stands for Medical Device Network. Next to the logo is the text "510(k) Summary". The text is in a simple, sans-serif font and is aligned to the right of the logo. The image is likely a document header or title page for a medical device summary.
K071713
Medical Decisions Network
## 5. 510(k) Summary
| Date: | June 21, 2007<br>OCT 12 2007 |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Medical Decisions Network, Inc.<br>2000 Holiday Drive, Suite 200<br>Charlottesville, VA 22901 |
| Contact: | Medical Decisions Network, Inc.<br>c/o Mr. Jonathan S. Kahan<br>Hogan & Hartson LLP<br>555 Thirteenth Street, NW<br>Washington, DC 20004 |
| | Phone: (202) 637-5600<br>Fax: (202) 637-5910<br>Email: jskahan@hhlaw.com |
| Trade Name: | Clarian Glucose Stabilizer Insulin<br>Dosing Calculator |
| Common Name: | MDN-CGSTM Insulin Dosing Calculator |
| Classification Name: | Drug Dose Calculator |
| Classification Number: | Class II (per 21 CFR 868.1890) |
| Product Code: | NDC |
| Predicate Device: | EndoTool <sup>TM</sup> Drug Dose Calculator<br>(K053137)<br>Glucommander Plus <sup>TM</sup> (K061110) |
| Description of Device: | The MDN-CGS <sup>TM</sup> Insulin Dosing<br>Calculator is a web-based software<br>solution that automates calculations used<br>by healthcare professionals to determine<br>the appropriate intravenous insulin drip<br>rate necessary to manage blood glucose<br>levels across a variety of patient<br>populations. The MDN-CGSTM Insulin<br>Dosing Calculator also provides alerts<br>for subsequent blood glucose testing and<br>monitoring. |
| Intended Use: | The MDN-CGS <sup>TM</sup> Insulin Dosing<br>Calculator is intended to evaluate the |
September 20, 2007
-- - ---
-----------
Page of this document of MDN CONFIDENTIAL AND PROPRIETARY
The information in this document is restricted for use by MDN personnel and its Partners and Customers under non-disclosure.
{1}------------------------------------------------
current patient blood glucose values. Based on the measurements, the software will calculate either a dose of glucose or insulin to drive the blood glucose level up or down toward a predetermined target range. Once that target blood glucose range has been reached, the software program calculates dosing of insulin or glucose for the purpose of maintaining the patient's blood glucose level in that target range. The system is programmed to provide intravenous dosing calculations of glucose or insulin. The device is intended for use with patients with no known insulin allergies and for patients over the age of 18.
The MDN-CGSTM Insulin Dosing Calculator's programmed logic is a tool and not a substitute for, but rather an assist to, clinical reasoning. The measurements and calculations generated by the MDN-CGSTM Insulin Dosing Calculator are intended to be used by qualified and trained medical personnel directed by physician order. No medical decisions are to be made solely on the guidance provided by this software program, but are to include consideration of clinical history, symptoms, other diagnostic measurements, and the professional's clinical judgment.
The submission device and the predicate devices have the same general intended use and similar indications, technological characteristics, and principles of operation. Each of the devices is intended to be used by trained clinicians. The only technological difference between the MDN-CGSTM Insulin Dosing Calculator and its predicates is that the MDN-CGSTM
Image /page/1/Picture/5 description: The image shows the text "September 20, 2007". The text appears to be a date, indicating the 20th day of September in the year 2007. The text is black and the background is white. The text is slightly blurry.
Image /page/1/Picture/6 description: The image shows a long, narrow object against a white background. The object appears to be a thin rod or stick, possibly made of metal or wood. There is a small, round object attached to the rod near the bottom. The rod is slightly curved and has a rough texture.
Comparison with Predicate Device:
ﺮ ﺍﻟﻤﺮﺍﺟﻊ
Image /page/1/Picture/8 description: The image shows a black and white graphic. The graphic appears to be a partial view of a head or object, with a large, dark, solid area dominating the upper portion. The lower part of the image features a curved line, possibly indicating a chin or the base of the object. The overall impression is that of a silhouette or a close-up of a dark, rounded form against a light background.
MDN CONFIDENTIAL AND PROPRIETARY The information in this document is restricted for use by MDN personnel and its Partners and Customers under non-disclosure.
Page 2 of 3
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a long, narrow object that appears to be a metal rod or pole. The object is dark in color and has a smooth, reflective surface. There are some small, round objects attached to the rod at various points. The background is white, which helps to highlight the shape and details of the object.
Conclusion:
Insulin Dosing Calculator is a standalone software device, while the predicate software programs are accessories to automated insulin pumps.
The MDN-CGS™ Insulin Dosing Calculator and the Endotool and Glucommander Plus have the same intended use and similar indications, technological characteristics and principles of operation. The minor technological differences between the MDN-CGS™ and its predicates do not present any new issues of safety or effectiveness. Thus, the MDN-CGSTM Insulin Dosing Calculator is substantially equivalent to the EndoTool™ Drug Dose Calculator (K053137) and the Glucommander Plus™ (K061110).

September 20, 2007
Page 3 of 3 MDN CONFIDENTIAL AND PROPRIETARY The information in this document is restricted for use by MDN personnel and its Partners and Customers under non-disclosure.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the edge. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
007 1 % 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Medical Decisions Network, Incorporated C/O Mr. Jonathan S. Kahan Partner Hogan & Hartson LLP 555 Thirteenth Street, North West Washington, DC 20004
Re: K071713
Trade/Device Name: Clarian Glucose Stabilizer Insulin Dosing Calculator Regulation Number: 21 CFR 868.1890 Regulation Name: Predictive Pulmonary-Function Value Calculator Regulatory Class: II Product Code: NDC Dated: September 21, 2007 Received: September 21, 2007
Dear Mr. Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2 - Mr. Kahan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Clues
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## 4. Indications for Use
510(k) Number (if known): K071713
Device Name: Clarian Glucose Stabilizer Insulin Dosing Calculator
Indications for Use:
The MDN-CGS™ Insulin Dosing Calculator is intended to evaluate the current patient blood glucose values. Based on the measurements, the software will calculate either a dose of glucose or insulin to drive the blood glucose level up or down toward a predetermined target range. Once that target blood glucose range has been reached, the software program calculates dosing of insulin or glucose for the purpose of maintaining the patient's blood glucose level in that target range. The system is programmed to provide intravenous dosing calculations of glucose or insulin. The device is intended for use with patients with no known insulin allergies and for patients over the age of 18.
The MDN-CGS™ Insulin Dosing Calculator programmed logic is a tool and not a substitute for, but rather an assist to, clinical reasoning. The measurements and calculations generated by the MDN-CGS™ Insulin Dosing Calculator are intended to be used by qualified and trained medical personnel directed by physician order. No medical decisions are to be made solely on the guidance provided by this software program, but are to include consideration of clinical history, symptoms, other diagnostic measurements, and the professional's clinical judgment.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | |
|-------------------------------------------------|--------|
| | AND/OR |
Over-The-Counter Use _ (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER
| (Division Sigh-Off) | PAGE OF NEEDED) |
|--------------------------------------------------------|--------------------|
| Division of Anesthesiology, General Hospital | |
| Infection Control, Dental Devices | |
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
| 510(k) Number: | k071713 |
| Page 1 of 1 | September 20, 2007 |
MDN CONFIDENTIAL AND PROPRIETARY The information in this document is restricted for use by MDN personnel and its Partners and Customers under non-disclosure.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.