AP-1000 BLOOD GLUCOSE MONITORING SYSTEM

K090389 · Bestgen Biotech Corp. · NBW · Jul 14, 2009 · Clinical Chemistry

Device Facts

Record IDK090389
Device NameAP-1000 BLOOD GLUCOSE MONITORING SYSTEM
ApplicantBestgen Biotech Corp.
Product CodeNBW · Clinical Chemistry
Decision DateJul 14, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Indications for Use

The AP-1000 Blood Glucose Monitoring System (BGMS) is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. The AP-1000 Meter is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. AP-1000 Blood Glucose Test Strips must be used with the AP-1000 Meter. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. The AP-1000 Blood Glucose Test Strips are intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. AP-1000 Blood Glucose Test Strips must be used with the AP-1000 Blood Glucose Meter. Testing is done outside the body (In Vitro diagnostic use). They are indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. MAJOR Level I/Level II Control Solutions are for use with the AP-1000 meter and AP-1000 Blood Glucose Test Strips as a quality control check to verify the accuracy of blood glucose test results.

Device Story

AP-1000 Blood Glucose Monitoring System uses electrochemical biosensor technology; capillary action draws small blood sample into test strip chamber. Meter processes signal to provide glucose result in 6 seconds. Used at home by patients or in clinical settings by professionals to monitor diabetes control. Output displayed on meter screen; assists healthcare providers in assessing diabetes management effectiveness. Benefits patient through convenient, rapid glucose monitoring.

Clinical Evidence

Pre-clinical and clinical data submitted per FDA Guidance for In Vitro Diagnostic Test Systems. No specific performance metrics (e.g., sensitivity, specificity) provided in the summary document.

Technological Characteristics

Electrochemical biosensor; capillary action sample intake; 6-second measurement time. System includes meter, test strips, and control solutions. Class II device.

Indications for Use

Indicated for persons with diabetes for quantitative glucose measurement in fresh capillary whole blood from fingertips. For home (OTC) or clinical use as an aid to monitor diabetes control. Contraindicated for neonates; not for diabetes diagnosis or screening.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k090389 B. Purpose for Submission: New device C. Measurand: Whole blood glucose D. Type of Test: Whole blood glucose concentration through a quantitative amperometric assay (Glucose Oxidase) E. Applicant: Bestgen Biotech Corporation F. Proprietary and Established Names: AP-1000 Blood Glucose Monitoring System Major Glucose Control Solution G. Regulatory Information: 1. Regulation section: 21 CFR § 862.1345 Glucose Test System 21 CFR 862.1660, Quality Control Material (assayed and unassayed) 2. Classification: Class II (assay) and Class I, reserved (controls) 3. Product code: NBW, Blood Glucose Test System, Over-the-Counter CGA, Glucose Oxidase, Glucose JJX, Single (specified) analyte controls (assayed and unassayed) 4. Panel: Clinical Chemistry (75) H. Intended Use: 1. Intended use(s): See indication for use below. 2. Indication(s) for use: The AP-1000 Blood Glucose Monitoring System (BGMS) is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an {1} aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. The AP-1000 Meter is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. AP-1000 Blood Glucose Test Strips must be used with the AP-1000 Meter. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. The AP-1000 Blood Glucose Test Strips are intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. AP-1000 Blood Glucose Test Strips must be used with the AP-1000 Blood Glucose Meter. Testing is done outside the body (In Vitro diagnostic use). They are indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. MAJOR Level I/Level II Control Solutions are for use with the AP-1000 meter and AP-1000 Blood Glucose Test Strips as a quality control check to verify the accuracy of blood glucose test results. 3. Special conditions for use statement(s): - Not intended for diagnosis of diabetes mellitus - Not intended for use on neonates - For in vitro diagnostic use only - Not for use on critically ill patients, patients in shock, dehydrated patients or hyperosmolar patients 4. Special instrument requirements: AP-1000 Blood Glucose Meter I. Device Description: The AP-1000 Blood Glucose Monitoring System is comprised of the AP-1000 Blood Glucose Meter, AP-1000 Blood Glucose Test Strips, Major Control Solutions (2 levels) and a lancing device. J. Substantial Equivalence Information: 1. Predicate device name(s): Diacare Easy Check Blood Glucose Monitoring System 2. Predicate K number(s): k062538 {2} 3. Comparison with predicate: | Item | Proposed Device | Easy Check (k062538) | | --- | --- | --- | | Similarities | | | | Detection method | Amperometry | same | | Enzyme | Glucose oxidase (A. Niger) | same | | Mediator | Potassium hexacyanoferrate | same | | Electrode | Carbon | same | | Sample type | Capillary whole blood | same | | Humidity range (operational) | 20 to 80 % | same | | Temperature range (operational) | 10 to 40 °C | same | | Power supply | 3V lithium battery (CR2032) | same | | Battery life | Over 1000 tests | same | | Differences | | | | Test range | 20 to 600 mg/dL | 30 to 600 mg/dL | | Hematocrit | 30 to 55 % | 30 to 50 % | | Altitude | Up to 10,183 feet above sea level | Only at sea level | | Test time | 6 seconds | 9 seconds | | Sample volume | 0.6 μl | 1.5 μl | | Test sample | Fingertip | Fingertip, palm and forearm | | Coding | Internal code | Code card | | Memory capacity | 960 measurements | 180 measurements | | Size L x W x H (mm) | 54 x 93 x 16 | 58 x 80 x 19 | | Weight (g) | 53 with battery | 50 without battery | | Item | Proposed Control Device | Easy Check Control (k062538) | | --- | --- | --- | | Similarities | | | | Number of Levels | 2 | same | | Analyte | Glucose | same | | Container | Plastic bottle with dropper tip | same | | Color | Red | same | | Temperature range | 2 to 30 °C | same | | Fill Volume | 2.5 mL | same | | Matrix | Buffer aqueous with glucose, sodium benzoate, viscosity modifier and non-reactive ingredient. | same | | Target Population | Professional and home use | same | K. Standard/Guidance Document Referenced (if applicable): - ISO 15197:2003, In Vitro Diagnostic Test Systems—Requirements for Blood Glucose Test Systems for Self Managing Diabetes Mellitus. - IEC/EN 61010-1, Medical electrical equipment Part 1. General requirements for safety, 2001. {3} - IEC/EN 61601-1-2, Medical electrical equipment, Part 2. Electromagnetic compatibility—Requirements and tests, 2001. - IEC/EN 61010-2-101, Safety particular requirements for IVD medical equipment, 2002. - FCC 47 CFR, Part 18, 2004. - CLSI EP5-A, Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline. - CLSI EP6-A, Evaluation of the Linearity of Quantitative Analytical Methods; Approved Guideline. - CLSI EP7-A, Interference Testing in Clinical Chemistry; Approved Guideline. ## L. Test Principle: The AP-1000 Blood Glucose Monitoring System uses electrochemical methodologies. The system quantitatively measures blood glucose levels using an amperometric method, which involves detecting the current produced from glucose oxidation. The electrons generated during this reaction are transferred from the blood to the electrodes. The magnitude of the resultant current is proportional to the concentration of glucose in the specimen and the signal is converted into a readout displayed on the meter. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: #### a. Precision/Reproducibility: The sponsor evaluated within run and between run precision of the device using a protocol based on CLSI EP5-A. Whole blood samples spiked with six different glucose concentrations were used. Ten replicate measurements were tested for each concentration each day, using two lots of test strips and ten meters for a total of twenty days. Results for each test strip lot are summarized in the tables below: Lot 1 | Mean (mg/dL) | Within run SD (mg/dL) | Between run SD (mg/dL) | Coefficient of variation (CV %) | | --- | --- | --- | --- | | 31.1 | 3.9 | 3.4 | 16.2 | | 49.9 | 1.5 | 3.8 | 11.9 | | 74.4 | 3.0 | 3.2 | 6.0 | | 121 | 3.0 | 3.1 | 4.5 | | 218 | 5.6 | 4.2 | 3.2 | | 332 | 11.4 | 5.5 | 3.7 | {4} Lot 2 | Mean (mg/dL) | Within run SD (mg/dL) | Between run SD (mg/dL) | Coefficient of variation (CV %) | | --- | --- | --- | --- | | 30.4 | 3.9 | 2.8 | 16.2 | | 49.9 | 1.5 | 3.9 | 12.3 | | 74.4 | 2.9 | 3.2 | 5.9 | | 120 | 3.0 | 2.9 | 4.6 | | 219 | 5.8 | 3.0 | 3.0 | | 330 | 10.4 | 6.2 | 3.6 | In addition, the sponsor evaluated five replicates of two levels of control solutions with two test strip lots and twenty meters. Results are summarized in the following tables: Lot 1 | Level | Mean (mg/dL) | SD (mg/dL) | Coefficient of variation (CV %) | | --- | --- | --- | --- | | 1 | 101.2 | 4.0 | 3.9 | | 2 | 221.3 | 6.6 | 3.0 | Lot 2 | Level | Mean (mg/dL) | SD (mg/dL) | Coefficient of variation (CV %) | | --- | --- | --- | --- | | 1 | 106.2 | 3.6 | 3.4 | | 2 | 231.8 | 6.9 | 3.0 | b. Linearity/assay reportable range: The sponsor evaluated linearity of the device using a protocol based on CLSI EP6-A. Testing was performed using oxygenated venous blood samples at eleven different blood glucose levels, ranging from 20 to 726 mg/dL, tested on two lots of test strips, on ten meters in replicates of ten. The YSI 2300D STAT plus glucose analyzer was used as the reference method. The measuring range of the device is 20 to 600 mg/dL. Linear Regression Analysis: Lot 1: $y = 1.0375x + 3.8754, r^2 = 0.9991.$ Lot 2: $y = 1.0563x + 0.8588, r^2 = 0.9986.$ c. Traceability, Stability, Expected values (controls, calibrators, or methods): The values for the MAJOR Control Solutions are assigned by repeat analysis using three different lots of AP-1000 test strips and five AP-1000 meters. The mean and standard deviation are used to establish the acceptable range for the strips. {5} Stability characteristics of the Major Glucose Control Solutions were determined using real-time studies. The unopened shelf-life is 24 months and the open vial stability is 90 days at the recommended storage of 36 °F to 86 °F. Stability characteristics of the AP-1000 blood glucose test strips were determined using real-time studies. The unopened shelf-life is 18 months and the open vial stability is 90 days at the recommended storage of 36 °F to 86 °F. d. Detection limit: The measuring range of the system is 20 - 600 mg/dL. This range was verified by the linearity study (see section M.1.b.). e. Analytical specificity: Interference testing was performed using a protocol based on CLSI EP7-A. A blood sample was collected from a volunteer. Hematocrit was adjusted to 40 ± 2 %, and the sample was separated into two aliquots. The glucose levels were adjusted to a low (100 to 150 mg/dL) and a high (200 to 250 mg/dL) level by supplementing each sample with a 50% aqueous glucose solution. The YSI 2300D STAT plus glucose analyzer was used as a reference instrument. The following substances were evaluated, Ascorbic acid (1.2, 1.5, 2 and 3 mg/dL), Acetaminophen (2, 6, 15 and 20 mg/dL), Dopamine (3.2, 6.5, 9.7 and 13 mg/dL), Ibuprofen (4.2, 15, 30 and 40 mg/dL), L-Dopa (5 and 10 mg/dL), Tetracycline (0.4, 1.5, 3 and 4 mg/dL), Tolbutamide (10, 35, 70 and 100 mg/dL), Cholesterol (300, 350, 400 and 500 mg/dL), Creatinine (1.5, 15, 22.5 and 30 mg/dL) and Uric acid (7, 12, 16 and 20 mg/dL). Each potential interfering substance was added to an aliquot of the low glucose blood sample and to an aliquot of the high glucose to the final concentrations listed above and compared to the same sample without the potential interfering substance (control). The results are as follows: Lot 1 | Exogenous Substance | | | Concentration at which drug interference was observed | | | --- | --- | --- | --- | --- | | Therapeutic levels (mg/dL) | | Maximum Test Concentration (mg/dL) | At low glucose level | At high glucose level | | Acetaminophen | 1 - 2 | 20 | 2.0 | 15 | | Ascorbic acid | 0.8 - 1.2 | 3 | 2.0 | 30 | | Dopamine | ……… | 13 | 3.2 | 9.7 | | Ibuprofen | 0.5 - 4.2 | 40 | none | | | Tetracycline | 0.4 | 4 | none | | | Tolbutamide | 5.3 - 10 | 100 | none | | | L-dopa | ……… | 10 | 5 | 5 | {6} 7 | Endogenous substance (mg/dL) | | | Concentration at which drug interference was observed | | | --- | --- | --- | --- | --- | | Physiological levels (mg/dL) | | Maximum Test Concentration (mg/dL) | At low glucose level | At high glucose level | | Cholesterol | 300 | 500 | none | | | Creatinine | 1.5 | 30 | none | | | Uric acid | 7 | 20 | 7 | 7 | Lot 2 | Exogenous Substance | | | Concentration at which drug interference was observed | | | --- | --- | --- | --- | --- | | Therapeutic levels (mg/dL) | | Maximum Test Concentration (mg/dL) | At low glucose level | At high glucose level | | Acetaminophen | 1 - 2 | 20 | 2.0 | 2.0 | | Ascorbic acid | 0.8 - 1.2 | 3 | 2.0 | 30 | | Dopamine | ……… | 13 | 3.2 | 9.7 | | Ibuprofen | 0.5 - 4.2 | 40 | none | | | Tetracycline | 0.4 | 4 | none | | | Tolbutamide | 5.3 - 10 | 100 | none | | | L-dopa | ……… | 10 | 5 | 5 | | Endogenous substance (mg/dL) | | | Concentration at which drug interference was observed | | | Physiological levels (mg/dL) | | Maximum Test Concentration (mg/dL) | At low glucose level | At high glucose level | | Cholesterol | 300 | 500 | none | | | Creatinine | 1.5 | 30 | none | | | Uric acid | 7 | 20 | 7 | 7 | Ibuprofen, Tetracycline, Tolbutamide, Cholesterol and Creatinine did not significantly interfere with the measurements (bias less than $\pm 10\%$). Acetaminophen (2 mg/dL), Ascorbic Acid (2 mg/dL at low glucose levels and 30 mg/dL at high glucose levels), Dopamine (3.2 mg/dL at low glucose levels and 9.7 mg/dL at high glucose levels), L-dopa (5 mg/dL) and Uric acid (7 mg/dL) significantly interfered with the measurements (bias more than $10\%$). The sponsor included the following warnings in the label: - Therapeutic levels of acetaminophen, normal to high levels of uric acid and high levels of ascorbic acid/ vitamin C in blood may result in inaccurate glucose readings. Blood glucose readings from these cases should be interpreted with caution. - Therapeutic levels of L-dopa or dopamine may result in inaccurate glucose readings with the system. {7} Hematocrit Study: the effect of hematocrit was evaluated in a study using a venous blood sample with five glucose concentrations (20, 70, 170, 285 and 600 mg/dL) and hematocrit levels adjusted to 25, 30, 40, 50, 55, 60 and 70 %. The YSI 2300D STAT plus glucose analyzer was used to check the glucose concentrations of the blood samples. Each glucose level/hematocrit combination was tested on ten meters per lot of strips (two lots of strips). The results of samples at each of the varying hematocrit levels were compared to the YSI value of the samples. The average of the measurements had a deviation less than 15% at ≥ 75 mg/dL (individual bias ranging from 0 to 17.6 %) and less than 10 mg/dL at < 75 mg/dL (individual bias ranging from 0 to 6 mg/dL) at 30 to 55% hematocrit. The sponsor included the following warning in the label: - Abnormal blood cell counts (hematocrit levels below 30 % and above 55 %) may cause inaccurate test results. Altitude Study: the effect of altitude was evaluated using five venous blood samples with five glucose concentrations (56, 72, 123, 253 and 462 mg/dL) with hematocrit levels between 30 and 55 %. The YSI 2300D STAT plus glucose analyzer was used to check the glucose concentrations of the blood samples. The results of this study are acceptable: all measurements must have a deviation less than 20% at ≥ 75 mg/dL and less than 15 mg/dL at < 75 mg/dL. The sponsor included the following statement in the label: - Clinical testing demonstrates that altitudes up to 10,183 feet (3104 meters) above sea level do not affect results with the AP-1000 Blood Glucose Monitoring System. Temperature and humidity studies: studies were performed to demonstrate that the meter can be used at temperatures from 10 to 40 °C and at relative humidity ranging from 20 to 80 %. A venous blood sample with three glucose concentrations (approximately 50, 100 and 250 mg/dL) was used for the studies. The YSI 2300D STAT plus glucose analyzer was used to check the glucose concentrations of the blood samples. Five meters and one lot of strips were used for each study. The results were reviewed and found to be acceptable. f. Assay cut-off: Not applicable {8} 9 2. Comparison studies: a. Method comparison with predicate device: Reference Method Comparison: This study was performed in accordance with ISO 15197 to demonstrate the accuracy of the proposed device when compared to measurements obtained with the YSI 2300D STAT plus glucose analyzer (reference method). One hundred blood samples with glucose concentrations ranging from 21.8 to 600 mg/dL were prepared by spiking a blood sample with glucose. The hematocrit was adjusted to 40 ± 2 %. The YSI 2300D STAT plus glucose analyzer was used to check the glucose concentrations of the blood samples. One meter and two lots of strips were used for the study. The study met the ISO 15197 standard where ninety-five percent (95%) of the individual glucose results shall fall within ±15mg/dL of the results at glucose concentrations <75mg/dL and within ±20% at glucose concentrations ≥75mg/dL. The results are as follows: Linear Regression Analysis: Lot 1: y = 1.0106x - 1.001, r² = 0.9961 Lot 2: y = 1.0029x - 0.4772, r² = 0.9970 For samples < 75 mg/dL | Lot | Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL | | --- | --- | --- | --- | | 1 | 16/17 (94 %) | 17/17 (100 %) | 17/17 (100 %) | | 2 | 16/17 (94 %) | 17/17 (100 %) | 17/17 (100 %) | For samples ≥ 75 mg/dL | Lot | Within ± 5 % | Within ± 10 % | Within ± 15 % | Within ± 20 % | | --- | --- | --- | --- | --- | | 1 | 64/83 (77 %) | 82/83 (99 %) | 83/83 (100 %) | 83/83 (100 %) | | 2 | 71/83 (86 %) | 82/83 (99 %) | 83/83 (100 %) | 83/83 (100 %) | Lay-User Study: This study was performed in accordance with ISO 15197 with 150 total patients. The 150 study participants obtained and tested their own capillary whole blood sample using the AP-1000 Blood Glucose Monitoring System (the proposed device) and only the directions in the labeling. Health care professionals then tested capillary whole blood using the proposed device and the YSI 2300D STAT plus glucose analyzer (the reference method) on the same individuals. These samples ranged in concentration from 58 to 350 mg/dL. The results met the ISO 15197 standard criteria where ninety-five percent (95%) of the individual glucose results shall fall within ±15mg/dL of the reference results at glucose concentrations <75mg/dL and within ±20% at reference glucose concentrations ≥75mg/dL. The results are as follows: {9} 10 # Linear Regression Analysis: Patient vs YSI $y = 1.0102x - 1.4672, r^2 = 0.9877$ Healthcare professional vs YSI $y = 1.0146x - 1.2854, r^2 = 0.9904$ For samples < 75 mg/dL | Sample | Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL | | --- | --- | --- | --- | | Patient | 21/21 (100 %) | 21/21 (100 %) | 21/21 (100 %) | | Professional | 19/22 (86.4 %) | 22/22 (100%) | 22/22 (100%) | For samples ≥ 75 mg/dL | Sample | Within ± 5 % | Within ± 10 % | Within ± 20 % | | --- | --- | --- | --- | | Patient | 89/129 (69.0 %) | 129/129 (100 %) | 129/129 (100 %) | | Professional | 94/128 (73.4 %) | 128/128 (100 %) | 128/128 (100 %) | b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Expected blood glucose levels for people without diabetes (referenced from the American Diabetes Association Clinical Practice Recommendations 2004, Diabetes Care, Vol. 27, Supplement 1, p. S9.) | Time | Range (mg/dL) | Range (mmol/L) | | --- | --- | --- | | Fasting | 70 to 100 | 3.9 to 6.1 | | Two hours after meals | less than 140 | less than 7.8 | N. Instrument Name: AP-1000 Blood Glucose Meter {10} 11 O. Systems Descriptions: 1. Modes of Operation: Each test strip is single use and must be replaced with a new strip for additional readings. Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?: Yes ☐ or No ☑ Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?: Yes ☐ or No ☑ 2. Software: FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types: Yes ☑ or No ☐ 3. Specimen Identification: There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected. 4. Specimen Sampling and Handling: This device is intended to be used with capillary whole blood from the finger, which can be applied directly to the test strip. 5. Calibration: The device must be coded with the code found on the current test strip label. No further calibration is required. 6. Quality Control: The sponsor has two levels of controls supplied with this meter. When a test strip is inserted into the meter, each control can be measured by following the instructions for “Testing Control Samples” provided in the User’s Guide for the meter. An acceptable range for each control level is printed on the test strip vial label. The user is instructed to contact the Customer Assistance line during the operational times or a healthcare provider outside the operational times if the control results fall outside these ranges. P. Other Supportive Instrument Performance Characteristics Data Not Covered In The “Performance Characteristics” Section above: Not applicable Q. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. {11} R. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 12
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...