G5 INFINITY BLODO GLUCOSE MONITORING SYSTEM WITH MODEL IGM, 0024A
K082201 · U. S. Diagnostics, Inc. · NBW · Feb 20, 2009 · Clinical Chemistry
Device Facts
Record ID
K082201
Device Name
G5 INFINITY BLODO GLUCOSE MONITORING SYSTEM WITH MODEL IGM, 0024A
Applicant
U. S. Diagnostics, Inc.
Product Code
NBW · Clinical Chemistry
Decision Date
Feb 20, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Infinity ™ Meter device is used along with the Infinity ™ Test Strips is used for the quantitative measurement of glucose level in whole blood as an aid in monitoring the effectiveness of diabetes management in the home and in clinical settings. Infinity TM System is for testing outside the body (in vitro diagnostic use only). Testing sites include the traditional fingertip testing along with alternate site testing on the dorsal palm, ventral palm, forearm, upper arm, calf and thigh.
Device Story
System measures capillary whole blood glucose via amperometric assay using glucose oxidase enzyme. User applies blood sample to single-use test strip inserted into G5 Infinity Meter; meter converts enzymatic reaction current into glucose concentration. Used in home by patients or clinical settings by professionals. Supports fingertip and alternate site testing (palm, forearm, upper arm, calf, thigh). Requires manual entry of batch-specific code number for calibration. Output displayed on meter screen; aids in diabetes management monitoring. Does not support wireless data transmission.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
In vitro diagnostic blood glucose monitor. Electrochemical sensing principle using dehydrogenase glucose reaction. Handheld meter form factor. System includes meter and test strips. No specific materials, software architecture, or connectivity standards disclosed.
Indications for Use
Indicated for quantitative blood glucose measurement in patients with diabetes to monitor management effectiveness. For home and clinical use. Testing sites: fingertip, dorsal/ventral palm, forearm, upper arm, calf, thigh. Contraindicated for neonates and not for diabetes screening or diagnosis.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k082201
B. Purpose for Submission:
Clearance of a new device
C. Measurand:
Whole blood glucose
D. Type of Test:
Whole blood glucose concentration through a quantitative amperometric assay (Glucose Oxidase)
E. Applicant:
U.S. Diagnostics, Inc.
F. Proprietary and Established Names:
G5 Infinity Blood Glucose Monitoring System
G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1345, Blood Glucose Test System, Glucose Oxidase
21 CFR § 862.1660, Quality control material (assayed and unassayed)
2. Classification:
Class II
Class I (reserved)
3. Product code:
NBW, CGA
JJX
4. Panel:
75 (Clinical Chemistry)
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
The G5 Infinity Meter device is used along with the G5 Infinity Test Strips for the quantitative measurement of glucose level in whole blood as an aid in monitoring the
{1}
effectiveness of diabetes management in the home by patients with diabetes and in clinical settings by healthcare professionals. G5 Infinity System is for testing outside the body (in vitro diagnostic use only). Testing sites include the traditional fingertip testing along with alternate site testing on the dorsal palm, ventral palm, forearm, upper arm, calf and thigh. The G5 Infinity Blood Glucose Monitoring System is not intended for screening or diagnosis of diabetes and also not for use with neonates.
3. Special conditions for use statement(s):
- Not for use with neonates
- Not intended for the screening or diagnosis of diabetes
- For In Vitro diagnostic use only
4. Special instrument requirements: G5 Infinity Blood Glucose Meter
I. Device Description:
The device consists of the G5 Infinity Blood Glucose Meter along with G5 Infinity Test Strips, a lancing device, lancets, owner's manual, log book, quick reference guide, carrying case, and control solution. The G5 Infinity Blood Glucose Meter, when used with the G5 Infinity Test Strips, quantitatively measures glucose in capillary whole blood. The control solution is used to verify the performance of the G5 Infinity Test Strips.
J. Substantial Equivalence Information:
1. Predicate device name(s): LifeScan, Inc., OneTouch Ultra Blood Glucose Monitoring System
2. Predicate 510(k) number(s): k024194
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Detection Method | Amperometry | Amperometry |
| Enzyme | Glucose Oxidase | Glucose Oxidase |
| Test Range | 20 – 600 mg/dL | 20 – 600 mg/dL |
| Equivalency | Plasma equivalent | Plasma equivalent |
| Test Time | 5 seconds | 5 seconds |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Volume Required | 0.5 μL | 1.0 μL |
| Hematocrit Range | 20-60% | 30-55% |
| Alternate Site Testing | Yes | No |
{2}
3
K. Standard/Guidance Document Referenced (if applicable):
None referenced
L. Test Principle:
Glucose measurement is based on electrical potential caused by the reaction of glucose with the reagents contained on the strip's electrodes. The glucose in the sample is oxidized by the enzyme glucose oxidase, and the current resulting from this enzymatic reaction is measured and converted to glucose concentration by the meter. The magnitude of the current is proportional to the concentration of glucose in the sample.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The sponsor evaluated within-run precision of the device using whole blood collected in an EDTA collection tube. The blood was then spiked with glucose to create five different levels for testing. Each level was tested 50 times using a single lot of test strips and a single device lot. Results are summarized below:
| Average glucose concentration (mg/dL) | 40.0 | 87.5 | 129.2 | 217.6 | 367.6 |
| --- | --- | --- | --- | --- | --- |
| SD (mg/dL) | 1.6 | 2.9 | 3.6 | 6.5 | 6.5 |
| CV (%) | 4.0 | 3.3 | 2.8 | 3.0 | 1.8 |
The sponsor evaluated day-to-day precision of the device using 3 levels of control solution. Each of the controls was measured twice a day, once in the morning (Run 1) and once in the afternoon (Run 2) during a month's time. For each Run, two measurement readings were taken in which the mean, standard deviation (SD), and coefficient variation percent (CV%) were calculated. The same test strip lot and device lot were used for the evaluation. In addition, the same lot of the 3 control solutions (low, normal, high) was used. Results are summarized below.
| Average glucose concentration (mg/dL) | 50.4 | 109.7 | 302.6 |
| --- | --- | --- | --- |
| SD (mg/dL) | 1.8 | 3.0 | 7.3 |
| CV (%) | 3.6 | 2.7 | 2.4 |
b. Linearity/assay reportable range:
To establish the linearity of the system through the range of 20 to 600 mg/dL, 14 glucose adjusted whole blood samples (actual range tested was 20 to 602 mg/dL according to YSI and 20.8 to 602.8 mg/dL when tested with the meters) were compared to YSI 2300 using 5 replicate measurements. Linear regression yields the following statistics:
| | Slope | y-intercept | r² |
| --- | --- | --- | --- |
| G5 Infinity BGMS | 1.0006 | 0.3031 | 0.9999 |
{3}
The sponsor claims 20 mg/dL as the lowest detectable limit in the labeling.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The reference instrument used is the YSI 2300 Glucose Analyzer and is calibrated by YSI 2747 Glucose Standard, a NIST traceable glucose standard.
The three control solutions (low, normal, and high) consist of buffered aqueous solutions of D (+)-glucose containing stabilizers, preservatives, and other non-reactive ingredients.
Closed Vial Stability of Control Solution: The measured glucose levels of 3 lots of control solutions during the shelf lifetime period (26 months) are all within the acceptance criteria of ±15%, and the average bias with respect to the mean value throughout the lifetime is less than ±4%.
Opened Vial Stability of Control Solution: The measured glucose levels of 3 lots of control solutions during the use lifetime period (3 months) are all within the specified ranges (50 ± 15 mg/dL for low level, 110 ± 16.5 mg/dL for normal level, 300 ± 45 mg/dL for high level solutions), and the average bias with respect to the mean value throughout the lifetime is less than ±0.7%.
The expected values of the glucose control solutions were established by testing the solutions 5 times with YSI 2300. If the solutions are within the acceptance criteria, the solutions are then tested 5 times on the Nova Biomedical Stat Profile M.
Solutions that pass these two QC checks are then tested 100 times on the candidate device, and the reported range of the control solutions are based on the average of these 100 tests.
d. Detection limit:
The measuring range of the system is 20 - 600 mg/dL. This range was verified by the linearity study (above section M.1.b.).
e. Analytical specificity:
Spiked whole blood samples containing 3 levels of glucose (low glucose at <60 mg/dL, normal glucose at 150 mg/dL, and high glucose at >300 mg/dL), with and without interfering substances, were prepared to test common endogenous and exogenous substances for interference. Each interferent was tested at five different levels, in accordance with CLSI EP7-P. The highest levels tested for which no interference was observed are summarized below:
| Interferent | (mg/dL) |
| --- | --- |
| Acetaminophen | 20 |
| Bilirubin | 40 |
| Gentistic Acid | 50 |
| Uric Acid | 20 |
{4}
| Interferent | (mg/dL) |
| --- | --- |
| Levo-Dopa | 4 |
| Creatinine | 30 |
| Methyl-Dopa | 2.5 |
| Tolazamide | 5 |
| Dopamine | 13 |
| Ascorbate | 3 |
| EDTA | 640 |
| Glutathione | 1 |
| Heparin | 1000 |
| Ibuprofen | 40 |
| Salicylic Acid | 50 |
| Tetracycline | 0.4 |
| Tolbutamide | 100 |
| Urea | 500 |
| Cholesterol | 500 |
| Triglycerides | 3000 |
| Galactose | 50 |
| Xylose | 10 |
| Maltose | 300 |
The acceptance criteria were as follows:
At <75 mg/dL samples, interferences over ± 15 mg/dL
At ≥75 mg/dL samples, interferences over ± 15 %
Interference testing showed that the candidate device results were within the acceptance criteria at the above interferent concentrations. The sponsor states that acetaminophen, uric acid, ascorbic acid, and other reducing substances do not significantly affect results when occurring in normal blood or normal therapeutic concentrations. However, an abnormally high concentration may cause inaccurately high results. The sponsor also states that lipemic samples, with cholesterol up to 500 mg/dL or triglycerides up to 3000 mg/dL do not significantly affect results.
An altitude study was performed in an altitude simulation chamber with capillary whole blood from volunteers which was then spiked to 3 levels (52, 153, and 352 mg/dL). The simulation chamber was set to 10,000 feet. The test data using capillary blood samples using both meters showed that the bias versus YSI at 10,000 feet is the same bias versus YSI as that observed at sea level. These data indicate no additional effect due to altitude up to 10,000 feet.
To test the accuracy of the hematocrit effect, blood samples adjusted to hematocrit levels of 20%, 30%, 40%, 50% and 60% at glucose concentrations of 40, 70, 110, 160, 260, 350, 450, and 550 mg/dL were tested a total of 15 times at each concentration. The acceptance criteria were as follows:
{5}
At < 75 mg/dL sample, all data should be within ± 15 mg/dL difference and that over 90% of data within ± 10 mg/dL difference (Test hematocrit range 20~60%)
At ≥ 75 mg/dL sample, all data should be within ± 15% error and that over 90% of data within ± 10 % error (Test hematocrit range 20~60%)
*Difference (mg/dL): assay value-average value of 40% Hct sample (at < 75 mg/dL)
* Error (%): (assay value-average value of 40% Hct sample)/average value of 40% Hct sample x 100 (at ≥ 75 mg/dL)
The results are as follows:
| Glucose Concentration | number of samples within ± 10 mg/dL | percentage of sample | number of samples within± 15 mg/dL | percentage of sample |
| --- | --- | --- | --- | --- |
| < 75 mg/dL | 119 | 99% | 120 | 100% |
| At < 75 mg/dL sample, total test number is 120 | | | | |
| Glucose Concentration | number of samples within ± 10 % error | percentage of sample | number of samples within± 15 % error | percentage of sample |
| ≥ 75 mg/dL | 356 | 99% | 360 | 100% |
| At ≥ 75 mg/dL sample, total test number is 360 | | | | |
f. Assay cut-off:
Not Applicable.
2. Comparison studies:
a. Method comparison with predicate device:
To demonstrate the accuracy performance of the candidate device, trained specialists tested the capillary blood samples from the finger of 160 patients (with samples ranging from 38 to 496 mg/dL) and compared the results to the results from a laboratory reference method (Hitachi 747) using samples taken at the same time.
*Samples with blood glucose levels below 50 mg/dL were obtained by glycolysis of patient samples, and samples above 400 mg/dL were supplemented with glucose.
Results are summarized below:
| System accuracy results for glucose concentration <75 mg/dL (4.2 mmol/L) | | |
| --- | --- | --- |
| Within ± 5mg/dL (within ± 0.28 mmol/L) | Within ± 10 mg/dL (within ± 0.56 mmol/L) | Within ± 15mg/dL (within ± 0.83 mmol/L) |
| 43/56 (77 %) | 56/56 (100%) | 56/56 (100%) |
{6}
7
| System accuracy results for glucose concentration ≥75 mg/dL (4.2 mmol/L) | | | |
| --- | --- | --- | --- |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| 149/264 (56%) | 224/264 (85%) | 251/264 (95%) | 262/264 (99%) |
The linear regression was $\mathrm{Y} = 0.9977\mathrm{x} + 0.4729$ , $\mathrm{R}^2 = 0.9832$
Lay-user studies were also performed involving 150 individuals fluent in English at 3 clinical sites (50 users per site). Each participant was given a readability questionnaire and draft labeling and instructions for the US market. No other instruction was given. The participants performed their own fingersticks using the G5 Infinity system. These results were then compared to fingerstick results obtained by health care professionals within 5 minutes of the lay user test, also using the G5 Infinity. The acceptance criteria were as follows:
95% of the individual glucose results shall fall within $\pm 15\mathrm{mg / dL}$ of the results of the manufacturer's measurement procedure at glucose concentration $< 75\mathrm{mg / dL}$ and within $20\%$ at glucose concentration $(\geq 75\mathrm{mg / dL})$ .
The results are as follows:
| System accuracy results for glucose concentration <75 mg/dL (4.2 mmol/L) | | | |
| --- | --- | --- | --- |
| Within ± 5mg/dL (within ± 0.28 mmol/L) | Within ± 10 mg/dL (within ± 0.56 mmol/L) | Within ± 15mg/dL (within ± 0.83 mmol/L) | |
| 8/25 (32 %) | 24/25(96 %) | 25/25(100%) | |
| System accuracy results for glucose concentration ≥75 mg/dL (4.2 mmol/L) | | | |
| --- | --- | --- | --- |
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| 74/125(59%) | 108/125(86%) | 121/125(97%) | 123/125(98%) |
The linear regression was $\mathrm{Y} = 0.9934\mathrm{x} + 0.955$ , $\mathrm{R}^2 = 0.9849$
To demonstrate the accuracy performance of the candidate device using alternate sites, a study was performed with lay users who read the user manual on their own, without additional instruction on use of the device by technicians. Samples were taken by the lay users from the fingertip, palm, upper arm, forearm, thigh, and calf and were compared to the users fingerstick result. Samples ranged from 64 - 482 mg/dL. Results are summarized below for samples $< 75\mathrm{mg / dL}$ being within $\pm 15$ mg/dL and $\geq 75\mathrm{mg / dL}$ within $\pm 20\%$ of the G5 infinity fingerstick:
| G5 Infinity AST vs. G5 Infinity Fingerstick | Palm vs. Fingerstick | Forearm vs. Fingerstick | Upper Arm vs. Fingerstick | Calf vs. Fingerstick | Thigh vs. Fingerstick |
| --- | --- | --- | --- | --- | --- |
| Total | 201/202 (99%) | 99/100 (99%) | 99/100 (99%) | 98/100 (98%) | 99/100 (99%) |
{7}
Within 5 minutes of the lay-users testing their own sample, a healthcare professional collected blood from the fingerstick site and tested it using the Hitachi 747 analyzer. The results are summarized below for the G5 Infinity alternate site samples obtained by the lay user, with $< 75\mathrm{mg / dL}$ being within $\pm 15\mathrm{mg / dL}$ and with $\geq 75\mathrm{mg / dL}$ within $\pm 20\%$ of samples taken by healthcare professionals and analyzed using the Hitachi 747.
| G5 Infinity BGM AST vs. Hitachi 747 Fingerstick | Palm vs. Fingerstick | Forearm vs. Fingerstick | Upper Arm vs. Fingerstick | Calf vs. Fingerstick | Thigh vs. Fingerstick |
| --- | --- | --- | --- | --- | --- |
| Total | 199/202 (98%) | 99/100 (99%) | 100/100 (100%) | 96/100 (96%) | 100/100 (100%) |
b. Matrix comparison:
Not applicable. Capillary whole blood is the only indicated matrix.
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable.
b. Clinical specificity:
Not Applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
The sponsor provided a readability study that indicated that the user manual, test strip labeling, and control solution labeling is at an $8^{\text{th}}$ grade reading level or below.
4. Clinical cut-off:
Not Applicable.
5. Expected values/Reference range:
The sponsor included the following expected glucose values for people without diabetes in their strip labeling:
Fasting: $70\mathrm{mg / dL} - 110\mathrm{mg / dL}$ $(3.9\mathrm{mmol / L} - 6.1\mathrm{mmol / L})$
Two hours after meals: less than $140\mathrm{mg / dL}$ $(7.8\mathrm{mmol / L})^2$
They also include the following statement: "For people with diabetes: your diabetes team will determine the appropriate blood glucose target range individually and jointly with you."
1) Stedmans Medical Dictionary, $27^{\text{th}}$ edition, 1999, p755.
{8}
2) American Diabetes Association Clinical Practice Recommendations 2004, Diabetes Care Supplement 1, pS9
## N. Instrument Name:
G5 Infinity Blood Glucose Monitoring Meter
## O. System Descriptions:
1. Modes of Operation:
Each test strip is single use and must be replaced with a new strip for additional readings.
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?:
Yes ☐ X or No ☐
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission:
Yes ☐ or No ☐ X
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
3. Specimen Identification:
There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected.
4. Specimen Sampling and Handling:
This device is intended to be used with capillary whole blood from the finger, palm, upper arm, forearm, thigh, and calf. Since the whole blood sample is applied directly to the test strip there are no special handling or storage issues.
5. Calibration:
A code number is provided with each batch of test strips to calibrate the meter for that batch. The code number is associated with the meter by inserting a test strip and entering the code number to match that found on the test strip bottle. No further calibrations are required of the user.
6. Quality Control:
The sponsor has three levels of controls available for this meter with one level coming with the kit and the others being available through the distributor. When a test strip is inserted into the meter, a control can be run. An acceptable range for each control level is printed on the test strip vial label. The user is referred to a troubleshooting section in the control test instructions section of the owner’s manual to identify possible reasons control results fall outside these ranges.
{9}
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
None
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
10
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.