ONE TOUCH ULTRA AND ONE TOUCH INDUO BLOOD GLUCOSE MONITORING SYSTEMS

K024194 · Lifescan, Inc. · NBW · Feb 7, 2003 · Clinical Chemistry

Device Facts

Record IDK024194
Device NameONE TOUCH ULTRA AND ONE TOUCH INDUO BLOOD GLUCOSE MONITORING SYSTEMS
ApplicantLifescan, Inc.
Product CodeNBW · Clinical Chemistry
Decision DateFeb 7, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2

Indications for Use

The One Touch® Ultra® System is intended to be used for quantitative measurement of glucose in fresh capillary whole blood. The One Touch® Ultra® System is intended for use outside the body (in vitro diagnostic use) by diabetics at home as an aid to monitor the effectiveness of diabetes control. The One Touch® InDuo™ Blood Glucose Meter is intended to be used for quantitative measurement of glucose in fresh capillary whole blood. The InDuo™ Blood Glucose Meter is intended for use outside the body (in vitro diagnostic use) by diabetics at home as an aid to monitor the effectiveness of diabetes control. The One Touch® InDuo™ Blood Glucose Meter also functions as the cap for the InDuo™ Insulin Doser. The two devices fit together to form a single unit for user convenience.

Device Story

System measures glucose in fresh capillary whole blood; utilizes One Touch Ultra test strips and control solution; includes lancing device and lancets. InDuo model integrates meter as cap for insulin doser. Used by patients at home for diabetes management. Meter processes blood sample to provide quantitative glucose reading; aids patients in monitoring glycemic control. No changes to fundamental technology or operating principles in this modification.

Clinical Evidence

No clinical data provided; bench testing only. Performance characteristics verified as unchanged from predicate devices.

Technological Characteristics

In vitro diagnostic glucose monitoring system. Uses capillary whole blood samples. Meter-based measurement. Software-controlled. No changes to fundamental scientific technology or materials.

Indications for Use

Indicated for diabetics at home for quantitative measurement of glucose in fresh capillary whole blood to monitor diabetes control effectiveness.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # FEB 0 7 2003 ## 510(k) Summary | Submitter | LifeScan, Inc.<br>1000 Gibraltar Drive<br>Milpitas, CA 95035<br>Contact: Mary Ellen Holden<br>Date Prepared: December 19, 2002 | |------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | One Touch® Ultra® Blood Glucose Monitoring System<br>Common name: Glucose test system<br>Classification: Blood Glucose Meters and Test Strips are Class II devices (21 CFR Section 862.1345, Glucose Monitor) | | Predicate Device | One Touch® Ultra® Blood Glucose Monitoring System | #### Device Description The One Touch Ultra System consists of the One Touch Ultra Meter. One Touch Ultra Test Strips. One Touch Ultra Control Solution, UltraSoft Lancing Device, UltraClear Cap and UltraSoft lancets. The One Touch Ultra meter, when used with the One Touch Ultra Blood Glucose Test Strips, quantitatively measures glucose in capillary whole blood. The One Touch Ultra Control Solution verifies the performance of the One Touch Ultra Blood Glucose Test Strips. #### Intended Use The One Touch Ultra System is intended to be used for quantitative measurement of glucose in fresh capillary whole blood. The One Touch Ultra System is intended for use outside the body (in vitro diagnostic use) by diabetics at home as an aid to monitor the effectiveness of diabetes control. #### Comparison to Predicate Device System The modifications to the device system encompass meter software and labeling changes. There has been no change to the intended use, fundamental scientific technology, physical design, operating principles, functionality or material composition of the device systems. #### Technological Characteristics There has been no change to the fundamental scientific technology. The meter software changes have been verified. #### Summary of Performance Characteristics There has been no change to the performance characteristics of the device system. #### Conclusion The modified One Touch Ultra Blood Glucose Monitoring System is substantially equivalent to the predicate device system. LifeScan, Inc. {1}------------------------------------------------ # 510(k) Summary | Submitter | LifeScan, Inc.<br>1000 Gibraltar Drive<br>Milpitas, CA 95035<br>Contact: Mary Ellen Holden<br>Date Prepared: December 19, 2002 | |------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | One Touch® InDuo™ Blood Glucose Monitoring System<br>Common name: Glucose test system<br>Classification: Blood Glucose Meters and Test Strips are Class II<br>devices (21 CFR Section 862.1345, Glucose Monitor) | | Predicate Device | One Touch® InDuo™ Blood Glucose Monitoring System | #### Device Description The One Touch InDuo System consists of the One Touch InDuo Meter (which also functions as a cap for the InDuo Insulin Doser), One Touch Ultra Test Strips, One Touch Ultra Control Solution, UltraSoft Lancing Device, UltraClear Cap and UltraSoft lancets. The One Touch InDuo meter, when used with the One Touch Ultra Blood Glucose Test Strips, quantitatively measures glucose in capillary whole blood. The One Touch Ultra Control Solution verifies the performance of the One Touch Ultra Blood Glucose Test Strips. #### Intended Use The InDuo Blood Glucose Meter is intended to be used for quantitative measurement of glucose in fresh capillary whole blood. The InDuo Blood Glucose Meter is intended for use outside the body (in vitro diagnostic use) by diabetics at home as an aid to monitor the effectiveness of diabetes control. The InDuo Blood Glucose Meter also functions as the cap for the InDuo Insulin Doser. The two devices fit together to form a single unit for user convenience. #### Comparison to Predicate Device System The modifications to the device system encompass meter software and labeling changes. There has been no change to the intended use, fundamental scientific technology, physical design, operating principles, functionality or material composition of the device systems. #### Technological Characteristics There has been no change to the fundamental scientific technology. The meter software changes have been verified. #### Summary of Performance Characteristics There has been no change to the performance characteristics of the device system. #### Conclusion The modified One Touch InDuo Blood Glucose Monitoring System is substantially equivalent to the predicate device system. LifeScan, Inc. Special 510(k) - One Touch® Ultra® and One Touch® InDuo™ Blood Glucose Monitoring Systems {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the USA. The seal features the department's name encircling a symbol. The symbol is a stylized representation of a human figure, with three overlapping profiles suggesting a sense of community and care. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 FEB 0 7 2003 Ms. Mary Ellen Holden Senior Regulatory Submissions Specialist LifeScan, Inc. 1000 Gibraltar Drive Milpitas, CA 95035 Re: k024194 Trade/Device Name: One Touch® InDuo™ Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: NBW Dated: December 19, 2002 Received: December 20, 2002 Dear Ms. Holden: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use Statement 510(k) Number: K024194 ONE TOUCH® ULTRA® Blood Glucose Monitoring System Device Name: #### Indications for Use: The One Touch® Ultra® System is intended to be used for quantitative measurement of glucose in fresh capillary whole blood. The One Touch® Ultra® System is intended for use outside the body (in vitro diagnostic use) by diabetics at home as an aid to monitor the effectiveness of diabetes control. lea lemon (Division Sign-Off Division of Clinic Laboratory Devices 510(k) Number K024194 Concurrence of CDRH, Office of Device Evaluation Over-the-Counter Use Prescription Use (Per 21 CFR 801.109) OR LifeScan, Inc. Special 510(k) - One Touch® Ultra® and One Touch® InDuo™ Blood Glucose Monitoring Systems {5}------------------------------------------------ ### Indications for Use Statement 510(k) Number: < 024 194 Device Name: ONE TOUCH® INDUO™ Blood Glucose Monitoring System Indications for Use: The One Touch® InDuo™ Blood Glucose Meter is intended to be used for quantitative measurement of glucose in fresh capillary whole blood. The InDuo™ Blood Glucose Meter is intended for use outside the body (in vitro diagnostic use) by diabetics at home as an aid to monitor the effectiveness of diabetes control. The One Touch® InDuo™ Blood Glucose Meter also functions as the cap for the InDuo™ Insulin Doser. The two devices fit together to form a single unit for user convenience. Division Sign-Off) Division of Clinical Laboratory Devices K027194 Concurrence of CDRH, Office of Device Evaluation Over-the-Counter Use Prescription Use (Per 21 CFR 801.109) OR Special 510(k) - One Touch® Ultra® and One Touch® InDuo™ Blood Glucose Monitoring Systems
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...