BAYER DIAGNOSTICS ACS:180 AND ADVIA CENTAUR TROPONIN I ASSAY
Applicant
Bayer Corp.
Product Code
MMI · Clinical Chemistry
Decision Date
Dec 6, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1215
Device Class
Class 2
Indications for Use
For the quantitative determination of cardiac troponin I in serum or heparinized plasma. using the Bayer Diagnostics ACS:180® (ADVIA® Centaur™ ) Automated Chemiluminescence Systems. Cardiac troponin I determinations aid in the diagnosis of acute myocardial infarction and in the risk stratification of patients with non-ST segment elevation acute coronary syndromes with respect to relative risk of mortality, myocardial infarction or increased probability of ischemic events requiring urgent revascularization procedures.
Device Story
The Bayer Diagnostics ACS:180 and ADVIA Centaur Troponin I Immunoassay is an in vitro diagnostic test used to measure cardiac troponin I (cTnl) levels in serum or heparinized plasma. The assay utilizes chemiluminescence technology to quantify cTnl, a protein released into the blood during myocardial injury. The device is intended for use in clinical laboratory settings by trained personnel. By providing quantitative cTnl measurements, the assay assists clinicians in diagnosing acute myocardial infarction and stratifying the risk of mortality or ischemic events in patients presenting with non-ST segment elevation acute coronary syndromes. Higher cTnl levels correlate with increased myocardial damage and higher risk, enabling earlier clinical intervention and improved patient management.
Clinical Evidence
Clinical performance was evaluated in a multicenter study of 681 patients with unstable angina or non-Q wave myocardial infarction. Patients with cTnl levels ≥ 0.10 ng/mL (n=444) showed higher risk of morbidity and mortality compared to those with levels < 0.10 ng/mL (n=237). Outcomes monitored up to 43 days included death, confirmed myocardial infarction, and urgent revascularization. Analytical sensitivity was established at 0.10 ng/mL, defined as the concentration corresponding to two standard deviations above the mean of 20 zero-standard replicates.
Technological Characteristics
The device is an automated chemiluminescence immunoassay system. It utilizes specific polyclonal and monoclonal antibodies to detect cardiac troponin I. The assay is designed for use on the ACS:180 and ADVIA Centaur platforms. It provides quantitative results up to 50 ng/mL with an analytical sensitivity of 0.10 ng/mL. The system is intended for professional use in clinical laboratories.
Indications for Use
Indicated for quantitative determination of cardiac troponin I in serum or heparinized plasma to aid in diagnosis of acute myocardial infarction and risk stratification of patients with non-ST segment elevation acute coronary syndromes regarding mortality, myocardial infarction, or need for urgent revascularization.
Regulatory Classification
Identification
A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
Predicate Devices
Bayer Diagnostics ACS:180 Troponin I Immunoassay
Related Devices
K980528 — CHIRON DIAGNOSTICS ACS:180 TROPONIN I ASSAY (CTNI) · Chiron Diagnostics Corp. · Mar 3, 1998
K053020 — TNL-ULTRA ASSAY FOR THE ADVIA CENTAUR SYSTEM · Bayer Healthcare, LLC · Dec 30, 2005
K062838 — VITROS TROPONIN I ES ASSAY, INCLUDING REAGENT PACK, CALIBRATORS AND RANGE VERIFIERS, MODELS 680 2301, 2302 AND 2303 · Ortho-Clinical Diagnostics · Dec 19, 2006
K171274 — ADVIA Centaur High-Sensitivity Troponin I (TNIH) · Siemens Healthcare Diagnostics, Inc. · Jul 12, 2018
K021814 — ACCUTNI AND ACCUTNI CALIBRATORS ON THE ACCESS IMMUNOASSAY SYSTEMS, MODEL #33340, 33345 · Beckman Coulter, Inc. · Sep 27, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a handwritten string of characters. The string is "K 993309". The characters are written in a cursive style, with the numbers being more rounded. The background is white.
DEC - 6 1999
### Summary of Safety and Effectiveness
As required by 21 CFR 807.92, the following 510(k) Summary is provided:
### 1. Submitters Information
| Contact person: | William J. Pignato<br>Director of Regulatory Affairs |
|------------------------|------------------------------------------------------------------------|
| Address: | Bayer Diagnostics Corporation<br>63 North Street<br>Medfield, MA 02052 |
| Phone<br>Fax | (508) 359-3825<br>(508) 359-3356 |
| Date Summary Prepared: | September 30, 1999 |
#### 2. Device Information
| Proprietary Name: | Bayer Diagnostics ACS:180 & ADVIA Centaur<br>Troponin I Immunoassay |
|------------------------|---------------------------------------------------------------------|
| Common Name: | Troponin I Immunological test system |
| Device Classification: | Class II |
#### 3. Predicate Device Information
Name: Bayer Diagnostics ACS:180 Troponin I Immunoassay
Manufacturer: Bayer Diagnostics Corporation
### 4. Device Description
Troponin is a structural protein which regulates the contraction of striated muscle. It consists of three subunits which are located periodically along the thin filament of the myofibrils. Troponin C binds calcium, troponin T attaches to tropomyosin on the thin filament, and troponin I inhibits actomyosin ATPase.
Troponin I (Tnl), an inhibitory protein of the troponin-tropomyosin complex, exists in three distinct isoforms: cardiac muscle, slow-twitch skeletal muscle, and fast-twitch, skeletal muscle. The cardiac form (cTnl) is further unique having 31 additional amino acid residues on its N-terminal, not present in the skeletal forms, which allows for specific polyclonal and monoclonal antibody development. The cardiac
{1}------------------------------------------------
specificity of this isoform improves the accuracy of diagnosis in patients with acute or chronic skeletal muscle injury and possible concomitant myocardial injury, and is the basis for its selection as a cardiac marker in the diagnosis of acute myocardial infarction.
Because of its cardiac specificity and sensitivity, cTnl has been used as a reliable marker in the diagnosis of perioperative myocardial infarction in patients undergoing cardiac surgery. It also provides useful information for evaluating patients with unstable angina, a condition implying increased risk of myocardial infarction and sudden death. Unstable angina patients with minimal values of cTnl are predicted to have a higher risk of short-term mortality. As the cTnl value progressively increases, the risk of mortality increases, presumably because the amount of myocardial damage also increases. Measuring cTnl provides useful prognostic information for evaluating patients with unstable angina, permitting early identification of patients at increased risk of death.
### 5. Statement of Intended Use
For the quantitative determination of cardiac troponin I in serum or heparinized plasma. using the Bayer Diagnostics ACS:180® (ADVIA® Centaur™ ) Automated Chemiluminescence Systems.
Cardiac troponin I determinations aid in the diagnosis of acute myocardial infarction and in the risk stratification of patients with non-ST segment elevation acute coronary syndromes with respect to relative risk of mortality, myocardial infarction or increased probability of ischemic events requiring urgent revascularization procedures.
### 6. Performance Characteristics
### Sensitivity and Assay Reportable Range
The ACS:180 cTnl assay measures cTnl concentrations up to 50 ng/mL (µg/L) with a minimum detectable concentration (analytical sensitivity) of 0.10 ng/mL (ug/L). Analytical sensitivity is defined as the concentration of cTnl that corresponds to the RLUs that are two standard deviations greater than the mean RLUs of 20 replicate determinations of the cTnl zero standard.
### Method Comparison
Method comparative studies of the ACS:180 cTnl assav and ADVIA Centaur cTnl assay gave the following relationship when analyzed by least squares linear regression. A total of 381 samples in the range of 0.17 to 49.47 ng/mL were tested.
{2}------------------------------------------------
ADVIA Centaur cTnl= 1.00 (ACS:180 cTnl) +0.00, r = 0.99
# 7. Clinical Performance Characteristics
## Assessment of Patients with Acute Coronary Syndromes
In a multicenter substudy of a clinical trial designed to assess the efficacy of low molecular weight heparin in the treatment of unstable angina and non-Q wave myocardial infarction, blood specimens from 681 patients were analyzed for cardiac troponin-1. The 444 patients with levels of cardiac troponin I of at least 0.10 ng/mL had higher risk of morbidity and mortality than the 237 patients whose levels were less than 0.10 ng/mL. Cardiac troponin I levels for risk assessment were determined within 24 hours after initial patient presentation with clinical and electrocardiographic symptoms of unstable angina or non-ST segment elevation myocardial infarction. Morbidity and mortality criterion included death, development of confirmed myocardial infarction after initial hospital presentation, and the need for urgent revascularization procedures as necessitated by recurrent episodes of angina. Clinical outcomes were monitored up to 43 days from hospital presentation.
## 8. Conclusions:
The Bayer ACS:180 and ADVIA Centaur cardiac troponin I assays have demonstrated the ability to act as an aid in the risk stratification of patients with acute coronary syndromes in addition to the approved indication of aiding in the diagnosis of acute myocardial infarction. The change in Minimal Detectable Concentration is verified by repeat analysis of the zero standard and determination of the concentration that corresponds to two standard deviations of the mean signal of that standard.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC - 6 1999
Mr. William J. Pignato Director of Regulatory Affairs Bayer Corporation Business Group Diagnostics 63 North Street Medfield, Massachusetts 02052-1688
Re: K993309
Trade Name: Bayer Diagnostics ACS:180 and ADVIA Centaur Troponin I Assay Regulatory Class: II Product Code: MMI Dated: September 30, 1999 Received: October 4, 1999
Dear Mr. Pignato:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
#### Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): _
Device Name: Bayer Diagnostics ACS:180 and ADVIA Centaur Troponin I Assay
14993309
# Indications for Use:
For the quantitative determination of cardiac troponin I in serum or heparinized plasma using the Bayer Diagnostics ACS:180® and ADVIA® Centaur™ Automated Chemiluminescent Systems. Cardiac troponin I determinations aid in the diagnosis of acute myocardial infarction and in the risk stratification of patients with non-ST segment elevation acute coronary syndromes with respect to relative risk of mortality, myocardial infarction or increased probability of ischemic events requiring urgent revascularization procedures.
(PLEASE DO NOT WRITE BELOW THIS LINE--CONTINUE ON ANOTHER PAGE, IF NEEDED)
| Prescription Use<br>(Per 21 CFR 801.109) | <div style="text-align:center;">OR</div> | Over-The-Counter Use<br>(Optional Format 1-2-96) |
|------------------------------------------|------------------------------------------|--------------------------------------------------|
|------------------------------------------|------------------------------------------|--------------------------------------------------|
(Division Sign-Off)
Division of Clinical Laboratory Devices
| 510(k) Number | 993309 |
|---------------|--------|
|---------------|--------|
Concurrence of CDRH, Office of Device Evaluation (ODE)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.