The Eclipse Blood Collection Set with holder is intended to be used with vacuum blood collection tube for the collection of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.
Device Story
Eclipse Blood Collection Set is a winged venipuncture needle with flexible tubing and integrated safety shield; used by healthcare professionals to collect venous blood samples into vacuum-based collection vials. Device includes a blood collection holder. Safety shield is manually activated post-draw to cover the needle, reducing risk of accidental needle stick injury. Device is sterile, single-use, and manufactured from stainless steel, polypropylene, PVC, and isoprene rubber. Performance verified through biocompatibility, mechanical, and sterilization testing. No clinical data required.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993), mechanical performance (ISO 9626, ISO 7864, ISO 80369), and sterilization/shelf-life validation (ISO 11135, ASTM standards).
Technological Characteristics
Winged needle set; materials: stainless steel (needle), polypropylene (shield/holder), PVC (tubing), isoprene (rubber sleeve), MABS (hub), polydimethylsiloxane (lubricant). Sterilization: Ethylene Oxide (EtO) to SAL 10^-6. Standards: ISO 10993 (biocompatibility), ISO 9626/7864 (needles), ISO 80369-7/20 (connectors), ISO 11135 (sterilization), ASTM F1980/F1886/F88/F2096 (packaging).
Indications for Use
Indicated for the collection of venous blood using vacuum blood collection tubes by licensed healthcare practitioners or individuals directed by practitioners.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
Safelock Disposable Blood Collection Set (K151991)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the agency's name, "U.S. Food & Drug Administration."
October 26, 2022
Eclipse Medcorp LLC Julie Summerville Sr. Director of Product Management 5916 Stone Creek Drive Suite #120 The Colony, Texas 75056
Re: K213690
Trade/Device Name: Eclipse Blood Collection Set Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA Dated: September 27, 2022 Received: September 27, 2022
Dear Julie Summerville:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
{1}------------------------------------------------
803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
David Wolloscheck, Ph.D. For Paval Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213690
Device Name Eclipse Blood Collection Set
Indications for Use (Describe)
The Eclipse Blood Collection Set with holder is intended to be used with vacuum blood collection tube for the collection of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
This 510K Summary of safety and effectiveness for the Eclipse Blood Collection Set is submitted in accordance with the requirements of the SMDA 1990, 21 CFR 807.92(c) and following guidance concerning the organization and content of a 510K summary.
| Applicant: | Eclipse MedCorp, LLC |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 5916 Stone Creek Drive<br>Suite 120<br>The Colony, TX 75056 USA |
| Establishment Number: | 3009032449 |
| Contact Person: | Julie Summerville |
| Telephone:<br>Email: | 972-380-2911<br>jsummerville@eclipsemed.com |
| Preparation Date: | October 26, 2022 |
| Device Trade Name: | Eclipse Blood Collection Set |
| Common Name: | Blood Collection Serum Separators, Systems, Vials, Tubes |
| Regulation Name: | Blood specimen collection device |
| Regulation Number: | 21 CFR 862.1675 (Product Code: JKA) |
| Predicate Device: | K200027 Blood Collection Needle |
| Reference Device: | K151991 Safelock Disposable Blood Collection Set<br>The reference device is included in this submission to<br>demonstrate compliance to ISO 23908:2011 Sharps injury<br>protection-Requirements and test methods. The sharps<br>safety protection for the Eclipse Blood Collection Set is<br>identical to the Safelock Disposable Blood Collection Set. |
| Regulatory Class: | Class II Prescription |
| Device Description: | The Eclipse Blood Collection Set is a winged blood collection<br>needle with flexible tubing intended for venipuncture to<br>obtain blood samples. It is provided with a safety shield for<br>covering the used venipuncture needle prior to disposal to aid<br>in the prevention of needle stick injury if manually activated |
| | after the blood draw. For blood collection, the set also<br>includes a blood collection holder for connection to vacuum-<br>based collection vials |
| Indications for Use: | The Eclipse Blood Collection Set with holder is intended to be<br>used with vacuum blood collection tube for the collection of<br>venous blood. The safety shield is intended to aid in the<br>protection against accidental needle stick injury. |
| Performance Testing<br>Support: | The following performance data was provided in support of<br>the substantial equivalence determination: |
{4}------------------------------------------------
The Eclipse Blood Collection Set was tested for validation and verification of functions based on risk analysis and the results passed predetermined acceptance criteria.
# Biocompatibility:
Biocompatibility was evaluated on the final, finished device per the FDA guidance titled Use of International Standard ISO 10993-1 "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" – Guidance for Industry and Food and Drug Administration Staff issued in September 2020 with the following endpoints:
- Cytotoxicity (ISO 10993-5:2009)
- Sensitization (ISO 10993-10:2010) ●
- Irritation (ISO 10993-10:2010)
- Acute Systemic Toxicity (ISO 10993-11:2017) ●
- Material Mediated Pyrogenicity (ISO 10993-11:2017, USP 41 NF 36:2018, <151> Pyrogen Test
- Hemocompatibility (Coagulation, Platelets and Hemolysis) (ISO 10993-4:2017)
# Performance Data:
Performance testing was evaluation per the International Standard ISO 80369-7:2016 and ISO 80369-20:2015.
# Sterilization and Shelf Life:
Ethylene oxide sterilization per ISO 11135-1:2014, ISO11737-1:2018; ISO 11737-2: 2009; ISO 10993-7:2008. The Sterility Assurance Level (SAL) is 10-6.
Sterilization, Shelf Life/Package Integrity in accordance with the following standards: ASTM-F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices; ASTM-F1886-16 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection; ASTM-F88/F88M-15, Standard Test Method for Seal Strength of Flexible Barrier Materials; ANSI/AAMI/ISO 11607-1: 2019, Packaging for Terminally Sterilized Medical Devices Part 1: Requirements for Materials, Sterile Barrier Systems and Packaging Systems; ASTM F2096-11 (2019), Standard Test Method for Detecting Gross Leaks in Porous Medical Packaging by Internal Pressurization (Bubble Leak); Real time aging studies are being conducted ambient conditioned storage temperatures for a period of 730 days (2 years). Upon completion of real time aging, half of
{5}------------------------------------------------
the samples will undergo label inspection for overall adhesions and legibility, and seal strength tests per ASTM F88/F88M-15. The other half of the samples will undergo visual inspection per ASTM F1886-16 and bubble leak testing per ASTM F2096-11.
Clinical Study: Not applicable
# Technological Characteristics and Comparison to the Predicate Device:
| | Subject Device:<br>Eclipse Blood<br>Collection Set | Predicate Device:<br>Safety Blood<br>Collection Needle<br>with/without Holder | Comparison |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | K213690 | K200027 | N/A |
| Manufacturer | Eclipse MedCorp, LLC<br>The Colony, TX, U.S.A | Jiangsu Caina Medical<br>Co., Ltd.<br>Jiangsu, China | N/A |
| Device Class | Class II | Class II | Same |
| Product Code | JKA | JKA | Same |
| Regulation Number | 21 CFR 862.1675 | 21 CFR 862.1675 | Same |
| Regulation Name | Blood specimen<br>collection device | Blood specimen<br>collection device | Same |
| Indications for Use /<br>Intended Use | The Eclipse Blood<br>Collection Set with<br>holder is intended to be<br>used with vacuum blood<br>collection tube for the<br>collection of venous<br>blood. The safety shield<br>is intended to aid in the<br>protection against<br>accidental needle stick<br>injury. | The Safety Blood<br>Collection Needle<br>with/without Holder is<br>intended to be used with<br>vacuum blood collection<br>tube for the collection of<br>venous blood. The safety<br>shield is intended to aid<br>in the protection against<br>accidental needle stick<br>injury.<br>The Luer access device-<br>holder with preattached<br>multiple sample adapter<br>is a sterile, non-invasive<br>device used to connect<br>devices with male or<br>female luer connectors to<br>blood collection tubes for<br>the collection of blood | Different<br>The Eclipse Blood<br>Collection set does<br>not include the<br>preattached adapter<br>therefore this is<br>excluded from the<br>indication for use<br>for the subject<br>device. |
| | Subject Device:<br>Eclipse Blood<br>Collection Set | Predicate Device:<br>Safety Blood<br>Collection Needle<br>with/without Holder | Comparison |
| Intended Users | Rx Only: Licensed<br>healthcare<br>practitioners or<br>individuals directed by<br>practitioners | Rx Only: Licensed<br>healthcare<br>practitioners or<br>individuals directed by<br>practitioners | Same |
| Needle Gauge for<br>Butterfly Needle | 23G | 20G - 23G | Different |
| Needle Length for<br>the Butterfly Needle | 19.5mm | 25.4mm<br>31.75mm<br>38.1mm | Different |
| Needle Gauge for<br>Tube Holder Needle | 21G | 18G-27G | Different |
| Needle Length for<br>Tube Holder Needle | 23.8mm with the<br>exposed portion<br>15mm | 23.8mm with the<br>exposed portion 15mm | Same |
| Tube Holder volume | 22ml | 22ml | Same |
| Tube Holder<br>function | Hold tube in place<br>during blood draw | Hold tube in place<br>during blood draw | Same |
| Adapter | Luer | Luer | Same |
| Needle Safety Shield | Protective sliding cap | Protective hinge shield | Different but same<br>as reference<br>device |
| Configuration and<br>Materials | Protective Cover of<br>Butterfly Needle –<br>Polypropylene<br>Rubber Sleeve – Isoprene<br>Rubber | Non-patient Needle Cap<br>– Polypropylene<br>Rubber Sleeve - Case<br>Gather Isoprene<br>Rubber | Same |
| | Patient Needle Tube –<br>Stainless Steel | Patient Needle Tube –<br>Stainless Steel | |
| | Tube Holder Needle –<br>Stainless Steel | Needle Tube – Stainless<br>Steel | |
| | Luer lock Male Hub –<br>MABS | Luer Lock Male Hub –<br>MABS | |
| | Tubing – PVC | Tubing – PVC | |
| | Safety Shield –<br>Polypropylene | Safety Shield –<br>Polypropylene | |
| | Subject Device:<br>Eclipse Blood<br>Collection Set | Predicate Device:<br>Safety Blood<br>Collection Needle<br>with/without Holder | Comparison |
| | *Tube Holder –<br>Polypropylene | Tube Holder –<br>Polypropylene | |
| | Lubricant -<br>Polydimethylsiloxane | Lubricant -<br>Polydimethylsiloxane | |
| | *NOTE: for the Eclipse<br>device this is also called<br>the Transfer Device | | |
| Sterilization Method | Ethylene Oxide (EtO)<br>Gas Sterilization | Ethylene Oxide (EtO)<br>Gas Sterilization | Same |
| Sterility | Meets the SAL of 10-6<br>per ISO 11135-2014 | Meets the SAL of 10-6<br>per ISO 11135-2014 | Same |
| Packaging | Tyvek Pouch | Tyvek Pouch | Same |
| Use | Single Use | Single Use | Same |
| Performance | Complies with: | Complies with: | |
| | ISO 9626 | ISO 9626 | |
| | ISO 7864 | ISO 7864 | Same |
| | ISO 80369-7 | ISO 80369-7 | |
| Biocompatibility | Tested to ISO 10993 | Tested to ISO 10993 | Same |
| Shelf Life | 2 years | 2 years | Same |
| Endotoxin Limit | <20 EU per device | <20 EU per device | Same |
{6}------------------------------------------------
{7}------------------------------------------------
# Discussion of Differences:
Needle gauge for the butterfly needle: The subject device is 23G which is a subset of the dimension for the predicate device, which is 20G – 23G. The subject device has only the 23G needle, but this size is previously cleared in the predicate device, therefore, it presents no new concerns for safety or efficacy. Therefore, the subject device needle gauge is the same as the predicate device. Performance testing was conducted per ISO 9626: 2016 Stainless Steel Needle tubing for the manufacture of medical devices and ISO 7864:2016 Sterile Hypodermic Needles for Single Use and verify the proposed device met all design specifications. There are no safety or efficacy concerns related to the use of the 23G needle.
Needle length for butterfly needle: The subject device length is 19.5mm. This is shorter than the lengths for the predicate device. According to peer reviewed published information butterfly needle length is typically between ½ to ¾ inches (12.7mm-19.05mm). This length is desirable because it can be inserted at a shallow angle and is therefore easier to use. The difference between 19.05mm (common size) and the subject device length of 19.50 is 0.45mm (0.017 inches) and is not significant Performance testing was conducted to the FDA consensus
{8}------------------------------------------------
standards and verify the proposed device met all design specifications. The difference in length for the subject device and predicate device does not raise any new safety or efficacy concerns. Performance testing was conducted per ISO 9626: 2016 Stainless Steel Needle tubing for the manufacture of medical devices and ISO 7864:2016 Sterile Hypodermic Needles for Single Use and verify the proposed device met all design specifications. There are no safety or efficacy concerns related to the use of the 23G needle
Needle gauge for the needle within the tube holder: The 21G needle for the tube holder is a subset of the cleared sizes for the predicate device. The subject device is the same as the predicate device. Performance testing was conducted per ISO 9626: 2016 Stainless Steel Needle tubing for the manufacture of medical devices and ISO 7864:2016 Sterile Hypodermic Needles for Single Use and verify the proposed device met all design specifications. There are no safety or efficacy concerns related to the use of the 23G needle
# Substantial Equivalence:
The Eclipse Blood Collection Set is substantially equivalent to the Jiangsu Caina Safety Blood Collection Needle with/without Holder predicate devices are under the same product code (JKA) and regulation number, both have the similar intended use/indication for use, similar lengths and gauges of needles, same materials, packaging and sterilization method. The only technological differences are the indications for use are with one exception. The Eclipse Blood Collection Set does not include a pre-attached adapter. Therefore, the following indication for use is not included: "The luer access device- holder with pre-attached multiple sample adapter is sterile, non-invasive device used to connect devices with male or female luer connectors to blood collection tubes for the collection of blood." The needle gauge is identical to the predicate, although both the butterfly needle within the tube holder differ in length between the proposed device and the predicate device; however, this difference is just in dimension. The difference in lengths is minor and does not raise new issues of safety or effectiveness and performance testing demonstrates the subject device can perform its intended function.
# Conclusion:
The Eclipse Blood Collection Set is considered to be substantially equivalent to the predicate device based on the intended use, technological characteristics, and the results of device testing submitted.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.