Safelock Disposable Blood Collection Set, Disposable Blood Collection Set
K151991 · Jiangyin Caina Technology Co., Ltd. · FPA · Dec 11, 2015 · General Hospital
Device Facts
Record ID
K151991
Device Name
Safelock Disposable Blood Collection Set, Disposable Blood Collection Set
Applicant
Jiangyin Caina Technology Co., Ltd.
Product Code
FPA · General Hospital
Decision Date
Dec 11, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The Disposable blood collection sets are intended to be used with vacuum blood collection tube for venipuncture to collect blood specimens from patients. The Safelock disposable blood collection sets are intended to be used with vacuum blood collection tube for venipuncture to collect blood specimens from patients, and the safety sheath is designed to prevent accidental needlesticks.
Device Story
Disposable blood collection sets; used for venipuncture to collect blood specimens into vacuum tubes. Safelock model includes safety sheath to reduce accidental needlesticks. Components include patient-end needle, double wing handle, flexing tube, and non-patient needle hub. Safelock model adds safety sheath and locking joint. Used in clinical settings by healthcare professionals. Safety mechanism activated by withdrawing needle handle backward to lock needle in sheath. Benefits include standardized blood collection and sharps injury prevention.
Clinical Evidence
No clinical study included. Simulated clinical study performed per FDA guidance 'Medical Device with Sharps Injury Prevention Feature' (2005) to evaluate safety mechanism; results met pre-established criteria. Bench testing included ISO 9626:1991 (needle tubing), ISO 10993 (biocompatibility), and ASTM standards for sterile barrier packaging (F88, F1140, F1929).
Technological Characteristics
Sterile, single-use blood collection sets. Materials conform to ISO 10993. Needle tubing meets ISO 9626:1991. Sterilization via Ethylene Oxide (EO). Features include safety sheath/locking joint for sharps protection. No electronic or software components.
Indications for Use
Indicated for venipuncture to collect blood specimens from patients using vacuum blood collection tubes.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
SURSHIELD™ Safety Winged Blood Collection Set (K031279)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure in profile, with three overlapping faces.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 11, 2015
Jiangyin Caina Technology Co., Ltd. c/o Ms. Diana Hong General Manager Mid-link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 CHINA
Re: K151991
Trade/Device Name: Safelock Disposable Blood Collection Set, Disposable Blood Collection Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: November 10, 2015 Received: November 13, 2015
## Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K151991
Device Name Safelock Disposable Blood Collection Set Disposable Blood Collection Set
Indications for Use (Describe)
The Disposable blood collection sets are intended to be used with vacuum blood collection tube for venipuncture to collect blood specimens from patients.
The Safelock disposable blood collection sets are intended to be used with vacuum blood collection tube for venipuncture to collect blood specimens from patients, and the safety sheath is designed to aid in the reduction of accidental needlesticks.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | |
| *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete | |
and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Exhibit # 2 Summary
This 510(k) Summary of is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number:_ K151991
- 1. Date of Preparation: 12/11/2015
- 2. Sponsor Identification
Jiangvin Caina Technology Co., Ltd. No.2, Taifu Road, Huashi Town, Jiangyin, Jiangsu, 214421, China
Establishment Registration Number: 3005670221
Contact Person: Jun Lu Position: General Manager Tel: +86-510-86205183 Fax: +86-510-86215183 Email: jun.lu@cainamedical.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Mr. Lee Fu (Alternative Contact Person)
### Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: Safelock disposable blood collection set Disposable blood collection set
Common Name: blood collection set
### Regulatory Information
Classification Name: Set, Administration, Intravascular Classification: II Product Code: FPA Regulation Number: CFR 880.5440 Review Panel: General Hospital
Intended Use Statement:
The Disposable blood collection sets are intended to be used with vacuum blood collection tube for venipuncture to collect blood specimens from patients.
The Safelock disposable blood collection sets are intended to be used with vacuum blood collection tube for venipuncture to collect blood specimens from patients, and the safety sheath is designed to prevent accidental needlesticks.
### Device Description
The proposed devices are provided sterile, single use. The proposed devices are intended to be used with vacuum blood collection tube for venipuncture to collect blood specimens from patients. It has two models, Safelock disposable blood collection set and Disposable blood collection set.
For disposable blood collection set, they consist nine pieces components: (1) needle protect cover (2) patient-end tube needle (3) double wing needle handle (4)flexing tube (5) connect A (6) non-patient needle hub (7) puncture needle (8) rubber sleeve (9) puncture needle protective cover.
For Safelock disposable blood collection sets, they consist twelve components: (1) needle protect cover (2) patient-end tube needle (3) safety needle handle (4) double wing (5) safety sheath (6) Locking joint (7) flexing tube (8) connect A (9) non-patient needle hub(10) puncture needle (11) rubber sleeve (12) puncture needle protect cover.
#### 5. Identification of Predicate Device
Predicate Device 1 510(k) Number: K020533 Product Name: EXEL Vaculet Blood Collection Set
{5}------------------------------------------------
Regulation No.: CFR 880.5440 Product Code: FPA
Predicate Device 2 510(k) Number: K031279 Product Name: SURSHIELD™ Safety Winged Blood Collection Set Regulation No.: CFR 880.5440 Product Code: FPA
#### 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The tests provided in this submission include:
Physical, Mechanical and Chemical Tests performed on the proposed device
ISO 9626:1991 AMENDMENT 1 2001 Stainless steel needle tubing for the manufacturing of medical devices
Sterile Barrier Packaging Testing performed on the proposed device:
| Seal strength | ASTM F88/F88-09 |
|-------------------|----------------------|
| Internal pressure | ASTM F1140/F1140M-13 |
| Dye penetration | ASTM F 1929-12 |
Sterilization and Shelf Life Testing performed on the proposed device:
| EO/ECH residue | ISO 10993-7:2008 |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Bacteria Endotoxin Limit | USP 37-NF 32 <85> |
| Shelf Life Evaluation | Physical, Mechanical, Chemical, Package and Sterility<br>Tests were performed on accelerated aging samples to<br>verify the claimed shelf life of the device |
Biocompatibility Testing:
The patient-contact materials of blood collection sets are identified and biocompatibility testing is performed, which include
| Cytotoxicity | ISO 10993-5: 2009 |
|---------------------------|--------------------|
| Intracutaneous Reactivity | ISO 10993-10: 2010 |
| Skin Sensitization | ISO 10993-10: 2010 |
| Acute Systemic Toxicity | ISO 10993-11:2006 |
{6}------------------------------------------------
Simulated Clinical Study performed on the proposed device:
A simulated clinical study was performed according to FDA Guidance for Industry and FDA Staff: Medical Device with Sharps Injury Prevention Feature, issued on August 9, 2005 to evaluate the safety mechanism of the proposed device. The results demonstrated that the proposed device met the pre-established criteria.
- Clinical Test Conclusion 7.
No clinical study is included in this submission.
- 8. Substantially Equivalent (SE) Comparison
| Item | Proposed device | Predicate Device 1 | Predicate Device 2 |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | K020533 | K031279 |
| Product | Disposable blood collection set<br>Safelock disposable blood collection set | EXEL Vaculet Blood Collection Set | SURSHIELDTM Safety Winged Blood Collection Set |
| Regulation No. | 880.5440 | 880.5440 | 880.5440 |
| Product Code | FPA | FPA | FPA |
| Indication for Use | The Disposable blood collection sets are intended to be used with vacuum blood collection tube for venipuncture to collect blood specimens from patients.<br>The Safelock disposable blood collection sets are intended to be used with vacuum blood collection tube for venipuncture to collect blood specimens from patients, and the safety sheath is designed to prevent accidental needlesticks. | This device is an integral part of blood collection (intravenously) used in connection with luer adapter for collection of whole blood into tubes | The TERUMO® SURSHIELDTM Safety Winged Blood Collection Set is a winged blood collection needle intended for venipuncture to collec blood specimens from patients. The TERUMO ® SURSHIELDTM Safety Winged Blood Collection Set is also indicated for intravenous administration of fluids after removing the attached luer adapter from the blood collection set connector and attaching a syringe, or other conipatible/appropriate device. This device may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused, and duration of therapy. Additionally, after withdraw of the needle from the patient's vein, the attached needle |
### Table 1 Comparison of between proposed device and predicate device
{7}------------------------------------------------
| | | | safety shield can be manually<br>activated to cover the needle<br>immediately after use to minimize<br>risk of accidental needlestick. |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The disposable blood collection<br>set and safelock disposable<br>blood collection set are intended<br>for collection of blood<br>specimens from patients. | EXEL Vaculet Blood Collection Set<br>is intended for collection of blood<br>specimens from patients. | SURSHIELD™ Safety Winged<br>Blood Collection Set is intended<br>for collection of blood specimens<br>from patients. |
| Feature | The needle is locked in safety<br>sheath by withdraw safety needle<br>handle backward | The device does not include safety<br>feature | The safety shield can be manually<br>activated to cover the needle after<br>use. |
| Sterile | Sterilized by EO | Sterilized by EO | Sterilized by EO |
| Single Use | Single Use | Single Use | Single Use |
| Performance | Complied with ISO 9626:1991,<br>AMENDMENT 1 2001 | unknown | unknown |
| Biocompatibility | Conform with ISO 10993 | Conform with ISO 10993 | Conform with ISO 10993 |
#### 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.