BD Vacutainer Barricor Lithium Heparin Plasma Blood Collection Tube

K160657 · Becton, Dickinson and Company · JKA · Sep 23, 2016 · Clinical Chemistry

Device Facts

Record IDK160657
Device NameBD Vacutainer Barricor Lithium Heparin Plasma Blood Collection Tube
ApplicantBecton, Dickinson and Company
Product CodeJKA · Clinical Chemistry
Decision DateSep 23, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1675
Device ClassClass 2

Indications for Use

The BD Vacutainer® Barricor™ Lithium Heparin® Plasma Blood Collection Tubes (BD Barricor™ Tubes) are used to collect, separate, process, transport and store venous blood samples for use in chemistry determinations, therapeutic drug monitoring (TDM), and zinc testing in plasma for in vitro diagnostic use. It is used in settings where a venous blood sample is collected by a trained healthcare worker.

Device Story

Sterile, single-use, evacuated blood collection tube; collects venous blood via standard needle holder. Contains lithium heparin anticoagulant and mechanical separator (elastomer/high-density base). Centrifugation (4000 RCF, swing bucket) triggers separator movement; differential buoyancy stretches separator to allow cellular passage; separator returns to original shape post-centrifugation to form stable barrier between plasma and cells. Used in clinical/hospital settings by trained healthcare workers. Output is plasma sample for chemistry/TDM/zinc analysis. Enables stable, robust barrier formation; reduces fibrin/cellular contamination compared to gel-based tubes; facilitates accurate clinical laboratory testing.

Clinical Evidence

Clinical equivalence established per CLSI GP-34A. Six studies (570 subjects) compared BD Barricor™ to BD PST™ for routine/special chemistry; therapeutic drug monitoring compared BD Barricor™ to BD Serum tubes (705 subjects). Analytes tested across multiple platforms using Deming Regression. Results demonstrated clinical equivalence or acceptable performance (mean bias and 95% limits within Clinical Acceptance Limits). Repeatability, lot-to-lot, and tube-to-tube variation studies met acceptance criteria (SD ratio ≤ 2.0). Stability studies confirmed performance for up to 24 hours (room temp) and 7 days (refrigerated) for most analytes.

Technological Characteristics

Sterile, evacuated PET tube; 13x75mm or 13x100mm. Mechanical separator (elastomer/high-density base). Lithium heparin anticoagulant. Silicone-based lubricant. Hemogard™ safety closure. Centrifugation: 4000 RCF in swing bucket centrifuge. Standards: CLSI GP34-A, ISO 11137-1/2, ASTM D4169-14.

Indications for Use

Indicated for venous blood collection, separation, processing, transport, and storage for chemistry, therapeutic drug monitoring, and zinc testing in plasma. For use by trained healthcare workers in clinical settings.

Regulatory Classification

Identification

A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k160657 B. Purpose for Submission: New Device C. Measurand: Not applicable – blood collection system D. Type of Test: Not applicable E. Applicant: Becton, Dickinson and Company F. Proprietary and Established Names: BD Vacutainer® Barricor™ Lithium Heparin Plasma Blood Collection Tube G. Regulatory Information: 1. Regulation section: 21 CFR 862.1675 (Blood specimen collection devices) 2. Classification: Class II 3. Product code: JKA 4. Panel: 75 (Chemistry) {1} H. Intended Use: 1. Intended use(s): See Indication for use below 2. Indication(s) for use: The BD Vacutainer® Barricor™ Lithium Heparin® Plasma Blood Collection Tubes (BD Barricor™ Tubes) are used to collect, separate, process, transport and store venous blood samples for use in chemistry determinations, therapeutic drug monitoring (TDM), and zinc testing in plasma for in vitro diagnostic use. It is used in settings where a venous blood sample is collected by a trained healthcare worker. 3. Special conditions for use statement(s): Prescription use only. The BD Barricor™ Tube is not designed for use with fixed angle centrifuges. If spun in a fixed angle centrifuge, a barrier between the plasma and the cellular material will not be formed. The BD Barricor™ Tube is designed for use with swing bucket centrifuges. These blood collection tubes are not intended to be used to collect blood specimens for blood banking, immunohematology test, infectious disease, lithium, and Di (2-ethylhexyl) phthalate determination. The BD Barricor™ Tube is not designed for use with open blood collection systems (manual filling of tube with the BD Hemogard™ removed) due to the increased risk of exposure to blood borne pathogens. Blood should be collected directly into the tube or transfer devices should be used if blood is collected in a syringe. 4. Special instrument requirements: Specific analyzers used to evaluate the device are listed in the labeling and below in section M. 2. a. Method Comparison. I. Device Description: The BD Vacutainer® Barricor™ Lithium Heparin Plasma Blood Collection Tubes (BD Barricor™ Tube) are sterile (interior), single-use, evacuated blood collection tubes. The tubes are comprised of a plastic tube containing a mechanical separator (in place of gel), a low-zinc stopper and a plastic BD Hemogard™ color-coded lime green safety-engineered shield. The interior of the BD Barricor™ Tube is spray coated with lithium heparin anticoagulant. The tube stopper and mechanical separator are lubricated with silicone based surfactant. The tubes are available in 13 x 75mm and 13 x 100mm configurations with draw volumes from 3.0 to 5.5 mL. The BD Barricor™ Tube is designed to be compatible with current phlebotomy and clinical laboratory practice. It employs a novel separation technology, a mechanical separator, which remains stable in its initial position, to enable the 2 {2} blood to be filled via current methods and subsequently creates a stable, robust barrier during processing. The mechanical separator is comprised of two materials of different densities - an elastomer and a higher density base material. In its resting position, the diameter of the mechanical separator is greater than that of the tube. The resulting friction from this interface allows the separator to maintain its position and orientation prior to blood collection and permits filling of the tube. Under centrifugation, the force applied on the separator will correctly orient the separator and allows it to move within the tube. While immersed in the collected sample, the differential buoyancy of the two materials will stretch the separator enabling the passage of cellular content and appropriate positioning of the separator between the cell column and plasma sample. When centrifugation stops, the mechanical separator returns to its original shape to form a barrier between the plasma sample (at the top), which is subsequently available for analysis, and the sedimented cells below. # J. Substantial Equivalence Information: 1. Predicate device name(s): BD Vacutainer® Brand PST™ Plasma Separation Tube 2. Predicate $510(\mathrm{k})$ number(s): k954592 3. Comparison with predicate: | Similarities / Differences | | | | --- | --- | --- | | Item | BD Vacutainer® Barricor™ Tube (Candidate Device) k160657 | BD Vacutainer® PST™ Tube (Predicate Device) k954592 | | Intended Use | The BD Vacutainer® Barricor™ Lithium Heparin Plasma Blood Collection Tubes (BD Barricor™ Tubes) are used to collect, separate, process, transport and store venous blood samples in a closed tube system for clinical laboratory testing. It is used in settings where a venous blood sample is collected by a trained healthcare worker. | Same | | Clinical testing use for | Clinical chemistry, therapeutic drugs, and zinc testing | Clinical chemistry assays | | TUBE COMPARISON: | | | | Tube Dimension | 13 x 75 mm and 13 x 100 mm | Same | | Draw Volume | 3.0 – 5.5 mL | 4.0mL – 9.5 mL | | Closure | Hemogard™ safety closure | Same | | Tube Stopper | Halobutyl rubber – low zinc | Compression Molded Rubber | {3} | Similarities / Differences | | | | --- | --- | --- | | Item | BD Vacutainer® Barricor™ Tube (Candidate Device) k160657 | BD Vacutainer® PST™ Tube (Predicate Device) k954592 | | Lubricant | Silicone-based (Tube Stopper and Separator) | Silicone-based (Tube Stopper) | | Tube Material | Polyethylene Teraphthalate (PET) | Glass | | Barrier | Mechanical separator | Gel | | Clot Activator | None | None | | Anticoagulant | Lithium Heparin | Same | | Sterility | Sterile | Sterile | | Tube Shelf Life | 18 months at 4 – 25°C | 12 months at 4 – 25°C | | Shelf | Shrink-wrapped polystyrene tray | Printed shelf carton | | Case Level | Corrugated cardboard | Same | ## K. Standard/Guidance Document Referenced (if applicable): 1. CLSI GP34-A Validation and Verification of Tubes for Venous and Capillary Blood Specimens Collection 2. ISO 11137-1 Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices 3. ISO 11137-2 Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose 4. ASTM D4169-14 Standard Practice for Performance Testing of Shipping Containers and Systems ## L. Test Principle: The BD Vacutainer® Barricor™ Lithium Heparin Plasma Blood Collection Tube is intended to be placed inside any BD Vacutainer® Needle Holders of the standard size or an adapter of a blood collection system. Once the vein of the patient has been penetrated using a standard needle, center the collection tube in the holder and push the tube fully onto the needle, puncturing the stopper of the tube. Hold the tube in place to ensure complete vacuum draw. The tube uses a controlled vacuum to pull a specific volume of blood into the sterile interior of the tube. Once the pressure is equalized, the blood flow ceases and the tube is removed from the needle holder or needle. Immediately after the blood has been drawn, the tube is gently inverted 8 times to mix the blood with the anticoagulant. The BD Barricor™ Tubes are centrifuged at 4000 RCF (relative centrifuge force or g’s) for 3 minutes in a swing bucket centrifuge within 2 hours of collection. Once a complete barrier formed between the plasma (at the top of the tube) and the cellular components (at the bottom of the tube), the plasma portion of the sample is removed for various clinical laboratory testing. {4} # M. Performance Characteristics (if/when applicable): # 1. Analytical performance: # a. Precision/Reproducibility: Two studies were conducted to assess the repeatability (within tube), lot-to-lot variation (between lots), and tube-to-tube (between tube) variation in BD Barricor™ Tubes. The study designs included subjects from the clinical settings. Three lots of each of BD Barricor™ Tubes (candidate device), BD PST™ Plasma Tubes, and BD Serum Tubes were tested and evaluated. The first study was performed using samples collected from 35 subjects and evaluated the BD Barricor™ Tube performance for 15 representative routine and special chemistry analytes: Alanine Aminotransferase (ALT), Total Bilirubin (T Bili), Calcium (Ca), Chloride (Cl), Cortisol, Complement C3 (C3), Glucose, Lactate Dehydrogenase (LDH), Phosphorus (Phos), Potassium (K), Total Protein, Thyroxine (T4), Testosterone (Testo), Immunoglobulin G (IgG) and Prostate Specific Antigen (PSA). The second study assessed performance for 5 representative therapeutic drugs: Carbamazepine (CBZ), Acetaminophen (ACET), Digoxin (DIG), Phenytoin (PHT), and Vancomycin (VANCO). All samples were run in duplicate on the two instrument platforms, Roche cobas® 6000 and the Beckman UniCel DxC 680i. Both instrument platforms yielded similar results. One representative platform with precision results is summarized in the tables below (one table with $\% \mathrm{CV}$ , one table with SD). Table 1. Precision Summary (%CV) on Roche cobas® 6000 for representative routine and special chemistry analytes | Analyte/Unit | Mean | Variance Component | CV (%) | 95% Lower Confidence Limit | 95% Upper Confidence Limit | | --- | --- | --- | --- | --- | --- | | ALT U/L | 27.4 | Between Lots | 0.54% | <0.01% | 1.76% | | | | Between Tubes | 1.58% | 0.54% | 2.20% | | | | Within Tubes | 2.45% | 2.15% | 2.78% | | | | Total | 2.97% | 2.67% | 3.28% | | C3 mg/dL | 114.1 | Between Lots | <0.01% | <0.01% | 0.94% | | | | Between Tubes | 1.35% | 0.96% | 1.67% | | | | Within Tubes | 1.24% | 1.09% | 1.40% | | | | Total | 1.83% | 1.69% | 1.97% | | Ca mg/dL | 9.37 | Between Lots | <0.01% | <0.01% | 0.58% | | | | Between Tubes | 0.83% | 0.54% | 1.01% | | | | Within Tubes | 0.84% | 0.72% | 0.95% | | | | Total | 1.18% | 1.06% | 1.31% | {5} | Analyte/Unit | Mean | Variance Component | CV (%) | 95% Lower Confidence Limit | 95% Upper Confidence Limit | | --- | --- | --- | --- | --- | --- | | Cl mmol/L | 100.4 | Between Lots | <0.01% | <0.01% | 0.68% | | | | Between Tubes | 0.70% | 0.36% | 0.94% | | | | Within Tubes | 0.65% | 0.57% | 0.74% | | | | Total | 0.95% | 0.82% | 1.09% | | Cortisol μg/dL | 11.529 | Between Lots | <0.01% | <0.01% | 2.22% | | | | Between Tubes | 3.02% | 2.15% | 3.68% | | | | Within Tubes | 2.73% | 2.39% | 3.07% | | | | Total | 4.08% | 3.78% | 4.37% | | Glucose mg/dL | 97.3 | Between Lots | <0.01% | <0.01% | 0.99% | | | | Between Tubes | 1.49% | 1.02% | 1.80% | | | | Within Tubes | 1.11% | 0.97% | 1.24% | | | | Total | 1.86% | 1.63% | 2.05% | | IgG mg/dL | 951.1 | Between Lots | <0.01% | <0.01% | 0.94% | | | | Between Tubes | 1.48% | 1.13% | 1.77% | | | | Within Tubes | 1.00% | 0.88% | 1.12% | | | | Total | 1.79% | 1.62% | 1.96% | | K mmol/L | 4.07 | Between Lots | 3.76% | 2.67% | 5.02% | | | | Between Tubes | 2.34% | 1.78% | 2.95% | | | | Within Tubes | 0.84% | 0.74% | 0.95% | | | | Total | 4.51% | 3.96% | 5.11% | | LDH U/L | 158.6 | Between Lots | 0.46% | <0.01% | 1.78% | | | | Between Tubes | 2.42% | 1.89% | 2.84% | | | | Within Tubes | 1.06% | 0.93% | 1.20% | | | | Total | 2.68% | 2.45% | 2.93% | | Phos mg/dL | 3.13 | Between Lots | <0.01% | <0.01% | 0.93% | | | | Between Tubes | 1.35% | 0.97% | 1.61% | | | | Within Tubes | 1.11% | 0.97% | 1.24% | | | | Total | 1.75% | 1.62% | 1.87% | | Testo ≥1 ng/mL | 3.458 | Between Lots | <0.01% | <0.01% | 3.88% | | | | Between Tubes | 3.99% | 2.07% | 5.49% | | | | Within Tubes | 2.20% | 1.79% | 2.56% | | | | Total | 4.55% | 4.05% | 5.10% | | Total Protein g/dL | 7.34 | Between Lots | <0.01% | <0.01% | 0.97% | | | | Between Tubes | 1.48% | 1.11% | 1.76% | | | | Within Tubes | 1.19% | 1.04% | 1.34% | | | | Total | 1.90% | 1.65% | 2.15% | | PSA ≥3 ng/mL | 6.951 | Between Lots | 0.08% | <0.01% | 0.39% | | | | Between Tubes | 0.01% | <0.01% | 0.42% | | | | Within Tubes | 0.64% | 0.48% | 0.74% | | | | Total | 0.64% | 0.50% | 0.78% | {6} Table 2. Precision Summary (SD) on Roche cobas® 6000 for representative routine and special chemistry analytes | Analyte/Unit | Mean | Variance Component | SD | 95% Lower Confidence Limit | 95% Upper Confidence Limit | | --- | --- | --- | --- | --- | --- | | T Bili mg/dL | 0.51 | Between Lots | <0.0001 | <0.0001 | 0.01 | | | | Between Tubes | 0.02 | 0.01 | 0.02 | | | | Within Tubes | 0.03 | 0.02 | 0.03 | | | | Total | 0.03 | 0.03 | 0.03 | | Testo <1 ng/mL | 0.167 | Between Lots | <0.0001 | <0.0001 | 0.0054 | | | | Between Tubes | <0.0001 | <0.0001 | 0.0059 | | | | Within Tubes | 0.013 | 0.0112 | 0.0140 | | | | Total | 0.013 | 0.0119 | 0.0142 | | PSA <3 ng/mL | 0.5387 | Between Lots | 0.0018 | <0.0001 | 0.0093 | | | | Between Tubes | 0.0115 | 0.0081 | 0.0141 | | | | Within Tubes | 0.0057 | 0.0048 | 0.0067 | | | | Total | 0.0130 | 0.0112 | 0.0144 | Results of the comparator tube type, BD Vacutainer® PST™ Tube, were very similar to the candidate tube type, BD Vacutainer® Barricor™ Tube. Calculated SD for the BD Vacutainer® PST™ Tube was $&lt; 0.05\mathrm{mg / dL}$ for Total Bilirubin, Testosterone and PSA and $\%$ CVs for all other analytes were $&lt; 10\%$ . Table 3. Precision Summary (%CV) on Roche cobas® 6000 for 5 representative therapeutic drugs | Analyte/Unit | Mean | Variance Component | CV (%) | 95% Lower Confidence Limit | 95% Upper Confidence Limit | | --- | --- | --- | --- | --- | --- | | CBZ μg/mL | 5.514 | Between Lots | <0.005% | <0.005% | 1.21% | | | | Between Tubes | <0.005% | <0.005% | 1.43% | | | | Within Tubes | 3.24% | 2.84% | 3.52% | | | | Total | 3.24% | 2.81% | 3.72% | {7} | Analyte/Unit | Mean | Variance Component | CV (%) | 95% Lower Confidence Limit | 95% Upper Confidence Limit | | --- | --- | --- | --- | --- | --- | | ACET μg/mL | 7.59 | Between Lots | 0.16% | <0.005% | 1.37% | | | | Between Tubes | 1.03% | <0.005% | 1.77% | | | | Within Tubes | 2.91% | 2.51% | 3.24% | | | | Total | 3.09% | 2.72% | 3.40% | | DIG ng/mL | 1.418 | Between Lots | 0.63% | <0.005% | 1.26% | | | | Between Tubes | 0.78% | <0.005% | 1.33% | | | | Within Tubes | 1.75% | 1.53% | 1.94% | | | | Total | 2.01% | 1.81% | 2.21% | | PHT μg/mL | 13.45 | Between Lots | 0.34% | <0.005% | 1.12% | | | | Between Tubes | 0.20% | <0.005% | 1.19% | | | | Within Tubes | 2.65% | 2.31% | 2.90% | | | | Total | 2.68% | 2.44% | 2.92% | | VANCO μg/mL | 16.05 | Between Lots | 0.47% | <0.005% | 1.21% | | | | Between Tubes | 0.08% | <0.005% | 0.99% | | | | Within Tubes | 2.11% | 1.83% | 2.28% | | | | Total | 2.16% | 1.94% | 2.39% | Results of the comparator tube type, BD Serum Tube, were very similar to the candidate tube type, BD Vacutainer® Barricor™ Tube. b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): i. Shelf-life stability: Real Time stability testing of the BD Vacutainer® Barricor™ Tube showed that the tube is stable for 18 months when stored at 4 to $25^{\circ}\mathrm{C}$ . Stability study protocol and acceptance criteria has been reviewed and found to be acceptable. ii. Analyte stability: Multiple analyte stability studies were conducted to assess the analyte within tube stability for representative routine and special chemistry analytes and therapeutic drugs in BD Barricor™ Tubes. Stability studies were assessed for representative analytes at initial time (Time 0) and 24hrs or 48hrs with room temperature storage, followed by refrigerated storage for 3 days and 7 days post centrifugation. Please see the analytes assessed in the summary table below. {8} Table 4. Within Tube Stability in the BD Barricor™ Tube: Study, Analyte, and Time points | Study | Analyte | Time points tested | | --- | --- | --- | | #1 | Routine Chemistry: Albumin (ALB), Alkaline Phosphatase (ALKP), Alanine Aminotransferase (ALT), Amylase (AMY), Aspartate Aminotransferase (AST), Bilirubin –Total (TBIL),Blood Urea Nitrogen (BUN), Calcium (Ca), Carbon Dioxide (CO2), Chloride (Cl), Cholesterol (Chol), Creatine Kinase (CK), Creatinine (Creat), Gamma-glutamyltransferase (GGT), Glucose (GLU), High-Density Lipoprotein (HDL), Iron (Fe), Lactate Dehydrogenase (LDH), Lipase (Lip), Low-Density Lipoprotein(LDL), Magnesium (Mg), Phosphorus (Phos), Potassium (K), Sodium (Na), Total Protein (TP), Triglycerides (Trig), Uric Acid (UA) | 0hr, 24hrs, 3 days and 7 days | | | Special Chemistry: Complement C3 (C3), Cortisol (CORT), Ferritin (FERR), Folate, Follicle Stimulating Hormone (FSH), Free Triiodothyronine (Free T3), Free Thyroxine (Free T4), Haptoglobin (HPT), Immunoglobulin A (IgA), Immunoglobulin G (IgG), Immunoglobulin M (IgM),Luteinizing Hormone (LH), Testosterone, Thyroid Stimulating Hormone (TSH), Total Triiodothyronine (Total T3), Total Thyroxine (Total T4), Transferrin, Vitamin B12 (Vit B12) | 0hr, 24hrs, 3 days and 7 days | | #2 | Glucose (GLU) Carbon Dioxide (CO2) | 0 hr, 6 hrs and 12 hrs; 0 and 18 hrs | | #3 | C-Reactive Protein (CRP) | 0 hr, 24 hrs, 3 days and 7 days | | #4 | β Human Chorionic Gonadotropin (βhCG), Prostate Specific Antigen (PSA) | 0 hr and 24 hrs | | #5 | Creatine Kinase MB Fraction (CKMB) Troponin I(TnI) Troponin T(TnT) | 0 hr and 24 hrs | | #6 | Therapeutic Drugs: Acetaminophen (ACET), Carbamazepine (CBZ), Digoxin (DIG), Phenytoin (PHT), Salicylate (ASA), Valproic Acid (VPA), and Vancomycin (VANCO) | 0 hr, 48 hrs and 7 days | The results of the study support the sponsor's claim the within-tube stability was demonstrated in the BD Barricor™ Tube for up to 24 hrs with room temperature storage and up to 7 days of refrigerated storage for all routine and special chemistry analytes except Folate, Glucose and $\mathrm{CO}_{2}$ . {9} For Folate, stability was established for 24 hours, while Glucose and $\mathrm{CO}_{2}$ demonstrated stability for 18 hrs with room temperature storage. Within tube analyte stability was demonstrated for CRP for up to 24 hrs with room temperature storage and up to 7 days of refrigerated storage. $\beta \mathrm{hCG}$, PSA, CKMB, TnI, and TnT were only tested for 24 hrs at room temperature and demonstrated stability at that time point. Within tube analyte stability was demonstrated for all therapeutic drugs for up to 48 hours of storage at room temperature and up to 7 days of refrigerated storage. iii. Additional bench testing evaluated on the candidate device: Clinical and benchtop studies were conducted to evaluate the effect of various centrifugation conditions on sample quality and analyte stability in the BD Barricor™ Tubes at multiple time points. The study protocols were reviewed and performance was considered acceptable at the recommended centrifugation range. Additional studies were conducted to assess draw volume, Zinc testing, stopper closure assembly stability, stopper pull out force, and simulated ship testing, including both pre-use shipping testing and post-centrifugation (after blood collection and centrifugation) ship testing. Study protocols, acceptance criteria and results for these studies were provided and found to be acceptable. d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: To demonstrate comparable performance with the comparator device, apparently healthy subjects and patients admitted into hospitals with various diseases were used. Testing was performed at seven different clinical and hospital sites. All subjects in the comparative studies have blood samples collected into the BD Barricor™ tubes (candidate device) and the comparator tubes (BD PST™ (k945952) for chemistry and BD Serum tube (k960250) for TDM) at the same time. The specimens collected in 10 {10} BD Serum tube (for TDM) were allowed to clot, and the serum was removed for testing immediately after centrifugation according to the instructions provided in the labeling. Evaluations were performed on a selective common general chemistry analytes, immunology analytes, special chemistry analytes, cardiac markers, and therapeutic drug monitoring analytes on multiple instrument platforms. A total of 62 analytes (55 chemistry and 7 TDM) were evaluated and demonstrated comparable results between the candidate tube and the comparator tubes. A small number of spiked samples were also prepared to obtain abnormal low or high values for selected analytes. Summary of the results (Deming or Passing-Bablok regressions correlations with $95\%$ confidence intervals) on one representative platform with the analytes and instruments evaluated are provided in the tables below: Table 5. Deming regressions correlations with $95\%$ confidence intervals from one representative platform/study | Analyte | Instrument(s) | Intercept (95% CI) | Slope (95% CI) | | --- | --- | --- | --- | | Acetaminophen | 4, 15 | -0.76 (-1.47, -0.06) | 0.98 (0.91, 1.05) | | Alanine Aminotransferase | 6, 12 | -0.5 (-0.9, -0.2) | 1.0 (1.0, 1.0) | | Albumin | 6, 12 | 0.02 (-0.19, 0.23) | 0.99 (0.93, 1.05) | | Alkaline Phosphatase | 6, 12 | -0.1 (-1.2, 1.0) | 1.0 (1.0, 1.0) | | Amylase | 6, 12 | 0.2 (-0.6, 1.0) | 1.0 (1.0, 1.0) | | Aspartate Aminotransferase | 6, 12 | 0.0 (-0.5, 0.6) | 1.0 (1.0, 1.0) | | Bilirubin, Direct | 6, 12 | 0.01 (-0.02, 0.05) | 0.99 (0.97, 1) | | Bilirubin, Total | 6, 12 | -0.26 (-1.05, 0.53) | 1.00 (0.98, 1.01) | | Blood Urea Nitrogen | 6, 12 | -0.3 (-0.5, 0.0) | 1.0 (1.0, 1.0) | | Calcium | 6, 12 | -0.08 (-0.34, 0.19) | 1.00 (0.97, 1.02) | | Carbamazepine | 4, 15 | 0.414 (0.127, 0.700) | 0.935 (0.898, 0.972) | | Carbon Dioxide, Total | 6, 12 | -0.41 (-1.62, 0.80) | 1.04 (0.99, 1.08) | | Chloride | 6, 12 | 0.5 (-1.5, 2.6) | 1.0 (1.0, 1.0) | | Cholesterol | 6, 12 | 1.3 (-1.9, 4.4) | 1.0 (1.0, 1.0) | | Complement C3 | 1, 7 | 1.1 (-1.0, 3.2) | 1.0 (1.0, 1.0) | | Cortisol | 3, 11 | -0.083 (-0.212, 0.047) | 1.006 (0.992, 1.021) | | C-Reactive Protein | 2, 7 | -0.011 (-0.171, 0.149) | 0.988 (0.958, 1.017) | | Creatine Kinase, MB fraction | 5, 13 | -0.19 (-0.63, 0.25) | 1.03 (0.99, 1.07) | | Creatine Kinase, Total | 6, 14 | -3.5 (-11.3, 4.2) | 1.0 (1.0, 1.1) | | Creatinine | 6, 12 | 0.00 (-0.03, 0.03) | 1.00 (0.98, 1.02) | | Digoxin | 4, 15 | -0.039 (-0.090, 0.013) | 1.039 (0.992, 1.086) | | Ferritin | 8, 13 | -12.75 (-37.11, 11.61) | 1.06 (0.94, 1.19) | | Folate | 5, 8 | -0.284 (-0.981, 0.414) | 1.020 (0.975, 1.066) | | Free Thyroxine | 3, 15 | 0.0564 (0.0096, 0.1032) | 0.9333 (0.8859, 0.9807) | | Free Triiodothyronine | 3, 15 | 0.131 (0.035, 0.228) | 0.961 (0.927, 0.995) | | Gamma-glutamyltransferase | 6, 12 | -0.4 (-1.2, 0.4) | 1.0 (1.0, 1.0) | | Glucose | 6, 12 | -2.3 (-4.3, -0.4) | 1.0 (1.0, 1.0) | | Haptoglobin | 1, 7 | 0.3 (-0.7, 1.4) | 1.0 (1.0, 1.0) | | High Density Lipoprotein | 6, 12 | -0.1 (-0.4, 0.2) | 1.0 (1.0, 1.0) | | Immunoglobulin A | 1, 7 | 1.2 (-1.9, 4.3) | 1.0 (1.0, 1.0) | | Immunoglobulin G | 1, 7 | -16.7 (-44.8, 11.4) | 1.0 (1.0, 1.0) | | Immunoglobulin M | 1, 7 | 0.6 (-3.6, 4.8) | 1.0 (1.0, 1.0) | | Iron | 6, 12 | 1.9 (-2.0, 5.8) | 1.0 (0.9, 1.0) | {11} Table 6. Passing-Bablok regressions correlations with $95\%$ confidence intervals from one representative platform/study | Analyte | Instrument(s) | Intercept (95% CI) | Slope (95% CI) | | --- | --- | --- | --- | | Estradiol | 3, 15 | 0.76 (-0.97, 1.56) | 1.05 (1.03, 1.07) | | Follicle Stimulating Hormone | 3, 15 | -0.01 (-0.06, 0.03) | 1.0 (0.98, 1.01) | | Human Chorionic Gonadotropin | 3, 11 | 0.13 (-0.04, 1.02) | 1.0 (0.99, 1.01) | | Luteinizing Hormone | 3, 15 | 0.01 (0, 0.01) | 0.99 (0.98, 1.0) | | Progesterone | 3, 11 | -0.07 (-0.08, -0.03) | 1.03 (1, 1.06) | | Prostate Specific Antigen | 10, 11 | -0.01 (-0.02, 0.0) | 1.0 (0.99, 1.02) | | Testosterone | 3, 11 | -0.74 (-1.65, 0.03) | 1.04 (1.02, 1.06) | | Troponin I | 5, 13 | 0.0 (0.0, 0.0) | 1.01 (1, 1.03) | | Troponin T | 11 | 0.0 (0.0, 0.0) | 0.97 (0.93, 1.0) | Instruments: 1. Abbott ARCHITECT C4000 2. Abbott ARCHITECT C8000 3. Abbott ARCHITECT i1000SR 4. Abbott ARCHITECT® ci4100 5. Beckman Coulter Access 2 6. Beckman Coulter AU680 7. Beckman Coulter UniCel® DxC 660i 8. Beckman Coulter UniCel® DxI 800 9. Ortho Clinical Diagnostics Vitros® 5600 {12} 10. Ortho Clinical Diagnostics Vitros® ECi 11. Roche cobas® e411 12. Roche MODULAR® Analytics 13. Siemens ADVIA Centaur® XP 14. Siemens Dimension RxL 15. Siemens Dimension Vista® 1500 The sponsor stated the following: “Whenever changing any manufacturer’s blood collection tube, type, size, handling, processing or storage condition for a particular laboratory assay, the laboratory personnel should review the storage conditions for a particular laboratory assay, the laboratory personnel should review the tube manufacturer’s data and their own data to establish/verify the reference range for specific instrument/reagent system. Based on such information, the laboratory can then decide if changes are appropriate.” Summary of all the studies: Study 1: A total of 90 adult subjects participated in a study from a hospital site (site A) to compare BD Barricor™ with the BD PST™ tubes. Chemistry results were generated for 27 general chemistry analytes on the Roche Modular® Analytics. Study 2: A total of 85 adult subjects participated in a study from a hospital site (site B). Chemistry results were generated for 29 routine chemistry on the Beckman Coulter AU680 and 3 special chemistry analytes on the Beckman Coulter Unicel® DxI instruments. Study 3: A total of 103 adult subjects participated in a study from a clinical site (site C). Chemistry results were generated for 5 special chemistry analytes on the Abbott ARCHITECT C4000, 2 special chemistry analytes on the Abbott ARCHITECT C8000 and 7 special chemistry analytes on the Beckman Coulter DxC 660i. Study 4: A total of 96 adult subjects participated in a study from a hospital site (site D). Chemistry results were generated for 12 special chemistry analytes on the Abbott ARCHITECT i1000SR, 1 special chemistry analyte on the Roche cobas e411 and 1 special chemistry analyte on the Roche Modular® Analyzer. Study 5: A total of 121 adult subjects participated in a study from a clinical site (site E). Chemistry results were generated for 7 special chemistry analytes on the Siemens Dimension Vista® 1500. Study 6: A total of 88 adult subjects participated in a study from a clinical site (site F). Chemistry results were generated for 1 special chemistry analyte on the Siemens Dimension RxL, Ortho Clinical Diagnostics Vitros® ECi and Roche cobas e411; and 4 special chemistry analytes on the Beckman Coulter Access® 2 and the Siemens ADVIA Centaur® XP. 13 {13} Study 7: A total of 722 adult subjects participated in a study from a hospital site (site G, recruitment was also performed at 7 additional sites). Chemistry results were generated for 7 representative therapeutic drugs on the Abbott ARCHITECT and the Siemens Dimension Vista®, in addition one therapeutic drug was tested on the Ortho Clinical Diagnostics Vitros® 5600. The sponsor performed a meta-analysis to evaluate plasma quality in the BD Barricor™ Tube compared to the BD PST™ Tube based on visual observations for barrier formation, hemolysis, fibrin mass, fibrin strand, and stopper fibrin ring. Study protocols, acceptance criteria and results for these studies were provided and found to be acceptable. b. Matrix comparison: Not applicable. These tubes are for plasma only. 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Not applicable N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 14
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...