BD VACUTAINER RAPID SERUM TUBE PLUS BLOOD COLLECTION TUBE, MODEL 368771
K091292 · Becton, Dickinson and Company (BD) · JKA · Aug 25, 2009 · Clinical Chemistry
Device Facts
Record ID
K091292
Device Name
BD VACUTAINER RAPID SERUM TUBE PLUS BLOOD COLLECTION TUBE, MODEL 368771
Applicant
Becton, Dickinson and Company (BD)
Product Code
JKA · Clinical Chemistry
Decision Date
Aug 25, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1675
Device Class
Class 2
Indications for Use
The BD Vacutainer® Rapid Serum Tube Plus Blood Collection Tube (BD RST) is a single use tube used to collect, transport, separate, and process venous blood specimens to obtain serum for chemistry determinations for in vitro diagnostic use. It is used in settings where a venous blood sample is collected by a trained healthcare worker. The BD RST is not recommended for patients on heparin therapy, direct thrombin inhibitor therapy or with Factor I deficiency.
Device Story
BD Vacutainer® Rapid Serum Tube Plus (BD RST) is a sterile, single-use, evacuated plastic blood collection tube with rubber stopper. Designed for rapid serum preparation; clots blood specimens in 5 minutes to facilitate faster turnaround times for chemistry, serology, and immunology testing. Used by trained healthcare workers in clinical settings for venous blood collection. Operates as a closed system for specimen collection, transport, and processing. Healthcare providers use the resulting serum for diagnostic determinations. Benefits include reduced processing time compared to standard serum tubes.
Clinical Evidence
Clinical testing compared blood collected in BD RST versus the predicate BD SST™ tube. Testing included a battery of routine and special chemistry analytes, as well as selected serology (anti-CMV IgG, anti-CMV-IgM, CRP) and immunology (C3, C4, IgG1, IgM, Rf) analytes. Results demonstrated substantial equivalence in performance. The 5-minute clotting time was successfully verified.
Technological Characteristics
Sterile, single-use, evacuated plastic tube with rubber stopper. Closed system design. Additive-based rapid clotting mechanism (5 minutes).
Indications for Use
Indicated for venous blood specimen collection, transport, separation, and processing to obtain serum for chemistry determinations in patients requiring diagnostic testing. Not for patients on heparin therapy, direct thrombin inhibitor therapy, or with Factor I deficiency.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
Predicate Devices
BD Vacutainer® SST™ Plus Blood Collection tube (BD SST™)
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k091292
B. Purpose for Submission:
New device
C. Measurand:
Not applicable - blood collection system
D. Type of Test:
Not applicable
E. Applicant:
Becton, Dickinson and Company
F. Proprietary and Established Names:
BD Vacutainer® Rapid Serum Tube Plus Blood Collection Tube
G. Regulatory Information:
1. Regulation section:
21CFR 862.1675 (Blood specimen collection devices)
2. Classification:
Class II
3. Product code:
JKA
4. Panel:
75 (Chemistry)
{1}
H. Intended Use:
1. Intended use(s):
See Indication for use below.
2. Indication(s) for use:
The BD Vacutainer® Rapid Serum Tube Plus Blood Collection Tube (BD RST) is a single use tube used to collect, transport, separate, and process venous blood specimens to obtain serum for chemistry determinations for in vitro diagnostic use. It is used in settings where a venous blood sample is collected by a trained healthcare worker. The BD RST is not recommended for patients on heparin therapy, direct thrombin inhibitor therapy or with Factor I deficiency.
3. Special conditions for use statement(s):
Prescription Use only.
These blood collection tubes are not intended to be used to collect blood specimens for blood banking, infectious disease, therapeutic drug monitoring, and Di(2-ethylhexyl) phthalate determinations.
4. Special instrument requirements:
Specific analyzers used to evaluate the device are listed in the labeling and below in section M. 2. a. method comparison.
I. Device Description:
The BD Vacutainer® Rapid Serum Tube Plus Blood Collection Tube (BD RST) a 13 x 100 mm, 4.0 mL, plastic evacuated tube with an orange conventional stopper. The interior of the tube wall is coated with thrombin to promote rapid clotting. The tube contains a gel barrier polymer at the tube bottom. The density of this material causes it to move upward during centrifugation to the serum-clot interface, where it forms a barrier separating serum from the clot. Serum may be aspirated directly from the collection tube, eliminating the need to transfer to another container. The tube interior is sterile and the product is for single use only.
J. Substantial Equivalence Information:
1. Predicate device name(s):
BD Vacutainer® SST™ Plus Blood Collection Tube
2. Predicate 510(k) number(s):
{2}
k023075
## 3. Comparison with predicate:
Similarities and Differences between the candidate device and the predicate device
| Items | BD Vacutainer® RST Plus Tube (Candidate device) | BD Vacutainer® SST™ Plus Tube (Predicate device) |
| --- | --- | --- |
| Intended use | Single use tube used to collect, separate, transport, and process venous blood specimens to obtain serum for chemistry determinations for in vitro diagnostic use. It is used in settings where a venous blood sample is collected by a trained healthcare worker. The BD RST is not recommended for patients on heparin therapy, direct thrombin inhibitor therapy or with Factor I deficiency. | Single use tube used to collect, separate, transport, and process venous blood specimens to obtain serum for chemistry determinations for in vitro diagnostic use. It is used in a clinical laboratory setting. The BD SST™ has additional claims for routine blood banking and infectious disease testing, as well as therapeutic drug monitoring (TDM). |
| TUBE COMPARISON | | |
| Tube Dimension | 13 x 100 mm | • 13 x 75 mm
• 13 x 100 mm
• 16 x 100 mm |
| Draw Volume | 4.0 mL | 3.0 mL – 10.0 mL |
| Closure | Conventional rubber closure | Conventional rubber closure and BD Hemogard™ closure |
| Clot Activator | Thrombin | Silica |
| Clotting Time | 5 minutes | 30 minutes |
| Tube Shelf Life | 12 months at 4 – 25°C | Same |
| Tube Sterility | Sterile | Same |
| PACKAGING COMPARISON | | |
| Shelf | Plastic film barrier bag in cardboard shelf carton | Shrink-wrapped EPS tray |
| Case | Corrugated cardboard | Same |
## K. Standard/Guidance Document Referenced (if applicable):
1. EN 00552 Sterilization of Medical Devices - Validation and Routine Control of Sterilization by Irradiation (1994)
2. EN 00980 Graphical symbols for use in the labeling of medical devices (2003)
3. EN 14820 Single-Use Containers for Human Venous Blood Specimen Collection (2004)
4. ISO 11137-1 Sterilization of Health Care Products - Requirements for the Validation and Routine Control - Radiation Sterilization (2006)
{3}
5. ISO 14001 Environmental Management Systems - Requirements with Guidance for Use (2004)
6. ISO 15223 Medical devices - Symbols to be used with medical device labels, labeling and information to be supplied (2000)
7. JIS T 3233 Evacuated Single - Use Containers for Venous Blood Specimen Collection (except color coding) (2005)
8. JIS T 14971 Medical Devices - Application of Risk Management to Medical Devices (2003)
L. Test Principle:
The BD Vacutainer® Rapid Serum Tube (RST) is intended to be placed inside either a holder or an adapter of a blood collection system. Once the vein of the patient has been penetrated using a standard needle, the tube is pushed fully into the needle holder so that the non-patient's end of the needle pierces the rubber septum of the stopper of the tube. The tube uses a controlled vacuum to pull a specific volume of blood into the sterile interior of the tube. The pressure differential caused the venous blood to flow into the tube. Once pressure is equalized, the blood flow ceases and the tube is withdrawn from the needle holder or needle. After blood has been drawn, the tube shall be immediately inverted gently for 5 to 6 times to mix the blood with the additives, and then allowed to clot and stand for 5 minutes. Once clotting is complete, the tube is to be centrifuged for 10 minutes, at a minimum g force of 1500g to initiate movement of gel and to obtain separation of serum.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision/repeatability studies were performed using fifty-three apparently healthy subjects in an internal testing site. Each subject has a venous blood collection into six blood collection tubes: 3 BD SST tubes (predicate device) with three different lot numbers and 3 BD RST tubes (candidate device) with three different lot numbers. Each tube was tested in duplicates for each of the 12 analytes and on multiple different instrument platforms. Results are summarized in the tables below:
Table 1 through Table 6 give the estimates from the variance components analysis for the BD RST tube type: the within-tube, within-lot, between-lot and total coefficient of variation (%CV) or standard deviation (SD) along with 95% confidence limits.
{4}
Table 1. Precision Summary (SD) for Roche Integra® 800
| Analyte/Unit | Tube Type | Mean | Variance Component | SD | 95% Lower Confidence Limit (LCL) | 95% Upper Confidence Limit (UCL) |
| --- | --- | --- | --- | --- | --- | --- |
| ALT U/L | BD RST | 27.12 | Between Lot | 0.26 | 0.10 | 0.36 |
| | | | Between Tubes, Within Lot | 0.11 | <0.005 | 0.24 |
| | | | Within Tubes | 0.44 | 0.39 | 0.50 |
| | | | Total | 0.52 | 0.40 | 0.66 |
| TBIL Mg/dL | BD RST | 8.60 | Between Lot | 0.18 | <0.005 | 0.27 |
| | | | Between Tubes, Within Lot | 0.06 | <0.005 | 0.21 |
| | | | Within Tubes | 0.43 | 0.38 | 0.49 |
| | | | Total | 0.47 | 0.38 | 0.60 |
Table 2. Precision Summary (%CV) for Roche Integra® 800
| Analyte/Unit | Tube Type | Mean | Variance Component | CV(%) | 95% LCL | 95% UCL |
| --- | --- | --- | --- | --- | --- | --- |
| Ca Mg/dL | BD RST | 2.351 | Between Lot | 0.65% | 0.35% | 0.87% |
| | | | Between Tubes, Within Lot | <0.005% | <0.005% | <0.005% |
| | | | Within Tubes | 1.31% | 1.17% | 1.50% |
| | | | Total | 1.47% | 1.22% | 1.73% |
| Creat Mg/dL | BD RST | 68.54 | Between Lot | 0.77% | <0.005% | 1.17% |
| | | | Between Tubes, Within Lot | <0.005% | <0.005% | 0.69% |
| | | | Within Tubes | 1.79% | 1.59% | 2.05% |
| | | | Total | 1.95% | 1.59% | 2.45% |
| Gluc Mg/dL | BD RST | 6.538 | Between Lot | 1.15% | 0.70% | 1.55% |
| | | | Between Tubes, Within Lot | <0.005% | <0.005% | <0.005% |
| | | | Within Tubes | 1.64% | 1.45% | 1.88% |
| | | | Total | 2.01% | 1.61% | 2.44% |
| IgG Mg/dL | BD RST | 10.53 | Between Lot | 1.11% | 0.62% | 1.52% |
| | | | Between Tubes, Within Lot | <0.005% | <0.005% | 0.54% |
| | | | Within Tubes | 1.53% | 1.35% | 1.75% |
| | | | Total | 1.89% | 1.49% | 2.38% |
| K mmol/L | BD RST | 4.173 | Between Lot | 2.70% | 1.90% | 3.30% |
| | | | Between Tubes, Within Lot | 1.02% | 0.65% | 1.29% |
| | | | Within Tubes | 1.12% | 1.00% | 1.29% |
| | | | Total | 3.09% | 2.24% | 3.77% |
| Mg Mg/dL | BD RST | 0.879 | Between Lot | 0.66% | 0.32% | 0.90% |
| | | | Between Tubes, Within Lot | <0.005% | <0.005% | <0.005% |
| | | | Within Tubes | 1.73% | 1.53% | 1.98% |
| | | | Total | 1.85% | 1.57% | 2.17% |
| Phos Mg/dL | BD RST | 1.208 | Between Lot | 1.08% | 0.66% | 1.46% |
| | | | Between Tubes, Within Lot | <0.005% | <0.005% | <0.005% |
| | | | Within Tubes | 1.44% | 1.28% | 1.64% |
| | | | Total | 1.80% | 1.44% | 2.20% |
| TP g/dL | BD RST | 73.37 | Between Lot | 0.98% | 0.53% | 1.30% |
| | | | Between Tubes, Within Lot | <0.005% | <0.005% | <0.005% |
| | | | Within Tubes | 2.17% | 1.93% | 2.48% |
| | | | Total | 2.38% | 2.00% | 2.80% |
{5}
Table 3. Precision Summary (SD) for Ortho Clinical Diagnostics Vitros® 250
| Analyte/Unit | Tube Type | Mean | Variance Component | SD | SD 95% LCL | SD 95% UCL |
| --- | --- | --- | --- | --- | --- | --- |
| ALT U/L | BD RST | 31.9 | Between Lot | 0.09 | <0.005% | 0.38 |
| | | | Between Tubes, Within Lot | 0.35 | <0.005% | 0.95 |
| | | | Within Tubes | 1.85 | 1.64 | 2.12 |
| | | | Total | 1.88 | 1.64 | 2.35 |
| TBIL mg/dL | BD RST | 0.47 | Between Lot | <0.005% | <0.005% | 0.02 |
| | | | Between Tubes, Within Lot | 0.01 | <0.005% | 0.02 |
| | | | Within Tubes | 0.04 | 0.03 | 0.04 |
| | | | Total | 0.04 | 0.03 | 0.05 |
Table 4. Precision Summary (%CV) for Ortho Clinical Diagnostics Vitros® 250
| Analyte/Unit | Tube Type | Mean | Variance Component | CV(%) | 95% LCL | 95% UCL |
| --- | --- | --- | --- | --- | --- | --- |
| Ca mg/dL | BD RST | 9.86 | Between Lot | 0.60% | 0.20% | 0.92% |
| | | | Between Tubes, Within Lot | 0.44% | 0.27% | 0.57% |
| | | | Within Tubes | 0.52% | 0.46% | 0.59% |
| | | | Total | 0.91% | 0.57% | 1.23% |
| Creat mg/dL | BD RST | 0.94 | Between Lot | 1.42% | <0.005% | 2.11% |
| | | | Between Tubes, Within Lot | 1.68% | 0.62% | 2.30% |
| | | | Within Tubes | 2.66% | 2.36% | 3.04% |
| | | | Total | 3.45% | 2.44% | 4.36% |
| Gluc mg/dL | BD RST | 119 | Between Lot | 0.91% | 0.41% | 1.30% |
| | | | Between Tubes, Within Lot | 0.93% | 0.71% | 1.11% |
| | | | Within Tubes | 0.53% | 0.47% | 0.60% |
| | | | Total | 1.41% | 0.94% | 1.82% |
| K mmol/L | BD RST | 4.32 | Between Lot | 2.47% | 1.67% | 3.08% |
| | | | Between Tubes, Within Lot | 1.39% | 1.03% | 1.66% |
| | | | Within Tubes | 0.93% | 0.83% | 1.07% |
| | | | Total | 2.98% | 2.13% | 3.66% |
| Mg mg/dL | BD RST | 1.9 | Between Lot | 1.09% | <0.005% | 1.70% |
| | | | Between Tubes, Within Lot | 0.96% | 0.32% | 1.32% |
| | | | Within Tubes | 1.56% | 1.38% | 1.78% |
| | | | Total | 2.13% | 1.42% | 2.79% |
| Phos mg/dL | BD RST | 4.14 | Between Lot | 0.50% | <0.005% | 0.74% |
| | | | Between Tubes, Within Lot | 0.54% | 0.28% | 0.71% |
| | | | Within Tubes | 0.73% | 0.65% | 0.84% |
| | | | Total | 1.04% | 0.71% | 1.32% |
| TP g/dL | BD RST | 7.59 | Between Lot | 1.20% | 0.75% | 1.61% |
| | | | Between Tubes, Within Lot | 0.54% | 0.19% | 0.75% |
| | | | Within Tubes | 0.87% | 0.77% | 1.00% |
| | | | Total | 1.58% | 1.09% | 2.03% |
{6}
Table 5. Precision Summary (%CV) for Beckman Coulter Access® 2
| Analyte/Unit | Tube Type | Mean | Variance Component | CV(%) | 95% LCL | 95% UCL |
| --- | --- | --- | --- | --- | --- | --- |
| Cortisol μg/dL | BD RST | 10.138 | Between Lot | 3.16% | 1.93% | 4.05% |
| | | | Between Tubes, Within Lot | 0.54% | <0.005% | 1.43% |
| | | | Within Tubes | 2.74% | 2.42% | 3.15% |
| | | | Total | 4.25% | 3.11% | 5.37% |
| TSH μIU/mL | BD RST | 1.441 | Between Lot | <0.005% | <0.005% | 2.74% |
| | | | Between Tubes, Within Lot | 5.77% | 4.46% | 6.84% |
| | | | Within Tubes | 2.56% | 2.27% | 2.94% |
| | | | Total | 6.32% | 5.00% | 7.93% |
Table 6. Precision Summary (%CV) for Bayer ADVIA Centaur®
| Analyte/Unit | Tube Type | Mean | Variance Component | CV(%) | 95% LCL | 95% UCL |
| --- | --- | --- | --- | --- | --- | --- |
| Cortisol μg/dL | BD RST | 9.965 | Between Lot | <0.005% | <0.005% | 1.21% |
| | | | Between Tubes, Within Lot | 0.26% | <0.005% | 2.11% |
| | | | Within Tubes | 4.45% | 3.93% | 5.13% |
| | | | Total | 4.46% | 3.93% | 5.71% |
| TSH μIU/mL | BD RST | 1.528 | Between Lot | <0.005% | <0.005% | 1.92% |
| | | | Between Tubes, Within Lot | 2.34% | <0.005% | 3.48% |
| | | | Within Tubes | 4.65% | 4.11% | 5.36% |
| | | | Total | 5.24% | 4.11% | 6.75% |
Results of the predicate tube type, BD SST, were very similar to the candidate tube type, BD RST. Calculated SD for the BD SST tube was $< 1$ U/L for ALT and $< 0.6$ mg/dL for T.Bili. and $\%$ CVs for all other analytes are $< 10\%$ .
b. Linearity/assay reportable range:
Not applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Real Time stability testing of the BD Rapid Serum Tubes showed that the tube is stable for 12 months when stored at $4^{\circ}$ to $25^{\circ}\mathrm{C}$ .
To demonstrate analyte stability in the BD RST tubes at room temperature, analyte stability studies were performed at 0 hrs and 24 hrs after collecting blood into the BD RST and SST tubes. Studies are summarized below:
1. Study 1: 35 apparently healthy adults were enrolled internally at the manufacturer facility. Testing was performed internally on-site and at a reference laboratory.
{7}
A. 31 analytes (ALB, ALK, ALT, AST, Amy, BUN, Ca, Chol, Cl, CK, CO2, Creat, DBIL, Fe, Ferritin, GGT, Gluc, HDL, K, LD, LDL, Lip, Mg, Na, Phos, TBIL, TP, Transferrin, TRIG, UA, UIBC) were tested on the Roche Cobas Integra 800.
B. 12 analytes (Cortisol, Ferritin, Folate, FSH, Free T3, Free T4, LH, TSH, Testosterone, Total T3, Total T4, and Vitamin B12) were tested on the Bayer ADVIA Centaur.
2. Study 2: 84 adult subjects (43 for cardiac marker and 41 for female hormone) were recruited and tested in a hospital in New York. 6 analytes (hCG, Estradiol, Progesterone, CKMB, Myoglobin, and Troponin I) were tested on the Beckman Coulter Access 2.
3. Study 3: 78 adult patients were enrolled in an external hospital. A total of 8 analytes were tested. C3, C4, CRP, IgG, IgM, and RF were tested on Beckman Coulter UniCel DxC 800 analyzer. Anti-CMV IgG and Anti-CMV IgM were tested on the Biomerieux VIDAS analyzer.
4. Study 4: 42 healthy adult patients were tested internally for LDL, Triglycerides, and Rheumatoid factor for additional time-points stability study. Testing was performed at 0, 6, 12 and 24 hours after blood have been collected and separated from the centrifuge.
The results of the study support the sponsor's claimed that the BD RST tubes have analyte stability for 24 hours at room temperature except for LDL (19 hours) and triglycerides (5 hours). This information is provided in the sponsor's package insert. Acceptance criteria for analyte stability on a representative platform was provided and found to be acceptable.
d. Detection limit:
Not applicable.
e. Analytical specificity:
Not applicable.
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
{8}
To demonstrate comparable performance with the predicate device, apparently healthy subjects and patients admitted into hospitals with various diseases were used. Testing was performed both internally and externally at 4 different clinical sites. All subjects in the comparative studies have blood samples collected into the BD RST (candidate device) and the BD SST (predicate tubes) at the same time. The specimens were allowed to clot, and the serum was removed for testing immediately after centrifugation according to the instructions provided in the labeling. Evaluations were performed on a selective common chemistry analytes, immunology analytes, and serological analytes on certain instrument platforms. A total of fifty-seven analytes were evaluated and demonstrated comparable results between the BD RST tubes and the BD SST tubes. A small number of spiked samples were used to supplement the analyte ranges. Summary of the results of one representative study on one representative platform with the analytes and instruments evaluated are provided in the tables below:
Table 8. Deming regressions correlations with 95% confidence intervals from one representative platform/study:
| Analyte | Instrument(s) | Slope (95% CI) | Intercept (95% CI) |
| --- | --- | --- | --- |
| Alanine Aminotransferase | 8, 10, 11, 12 | 1.01 (0.99, 1.03) | -0.05 (-0.12, 0.03) |
| Albumin | 8, 11 | 1.00 (0.84, 1.16) | 0.00 (-0.25, 0.24) |
| Alkaline Phosphatase | 8, 11 | 1.02 (0.98, 1.06) | -0.09 (-0.26, 0.08) |
| Amylase | 8, 11 | 1.00 (0.96, 1.04) | -0.01 (-0.15, 0.13) |
| Aspartate Aminotransferase | 8, 11 | 1.03 (0.84, 1.22) | -0.10 (-0.66, 0.45) |
| Bilirubin, Direct | 8, 11 | 1.008 (0.923, 1.092) | 0.002 (-0.01, 0.013) |
| Bilirubin, Total | 8, 10, 11, 12 | 0.99 (0.99, 1.0) | 0.00 (-0.01, 0.01) |
| Blood Urea Nitrogen | 8, 11 | 0.90 (0.8, 1.0) | 1.3 (0.3, 2.4) |
| C-Reactive Protein | 4, 10 | 1.00 (0.99, 1.01) | -0.02 (-0.04, 0.01) |
| Calcium | 8, 10, 11, 12 | 0.99 (0.97, 1.01) | 0.07 (-0.13, 0.27) |
| Carbon Dioxide, Total | 8, 11 | 0.97 (0.81, 1.14) | 0.11 (-0.41, 0.62) |
| Chloride | 8, 11 | 1.00 (0.9, 1.2) | -4.1 (-18.3, 10.1) |
| Cholesterol | 8, 11 | 0.97 (0.92, 1.02) | 0.16 (-0.13, 0.44) |
| Complement C3 | 4, 10 | 1.01 (1.0, 1.02) | -0.04 (-0.1, 0.02) |
| Complement C4 | 4, 10 | 1.00 (0.98, 1.01) | 0.00 (-0.05, 0.05) |
| Cortisol | 2, 6 | 1.00 (0.94, 1.06) | -0.02 (-0.16, 0.12) |
| Creatine Kinase- MB fraction | 1, 3 | 1.00 (0.99, 1.02) | 0.01 (-0.03, 0.05) |
| Creatine Kinase, Total | 8, 11 | 1.00 (0.99, 1.01) | -0.02 (-0.08, 0.05) |
| Creatinine | 8, 10, 11, 12 | 1.001 (0.996, 1.007) | 0.013 (-0.013, 0.04) |
| Cytomegalovirus Antibodies, IgG | 5, 7 | See table 9. below | See table 9. below |
| Cytomegalovirus Antibodies, IgM | 5, 7 | See table 10. below | See table 10. below |
| Estradiol | 3, 9 | 1.00 (0.97, 1.02) | 0.02 (-0.17, 0.22) |
| Ferritin | 2, 11 | 1.01 (0.99, 1.02) | -0.04 (-0.1, 0.02) |
| Folate | 2, 3 | 0.947 (0.920, 0.975) | 0.153 (0.070, 0.236) |
| Follicle Stimulating Hormone | 2, 3, 6 | 1.021 (0.998, 1.044) | -0.042 (-0.104, 0.020) |
{9}
Table 9. Concordance Table: anti-CMV IgG (Biomerieux VIDAS®)
| | | BD RST | | | All |
| --- | --- | --- | --- | --- | --- |
| | | Negative (<4) | Equivocal (≥4 to <6) | Positive (≥6) | |
| BD SST™ | Negative (<4) | 31 | 1 | 0 | 32 |
| | Equivocal (≥4 to <6) | 0 | 0 | 0 | 0 |
| | Positive (≥6) | 0 | 1 | 71 | 72 |
| All | | 31 | 2 | 71 | 104 |
{10}
Table 10. Concordance Table: anti-CMV IgM (Biomerieux VIDAS®)
| | | BD RST | | | All |
| --- | --- | --- | --- | --- | --- |
| | | Negative (<0.7) | Equivocal (≥0.7 to <0.9) | Positive (≥0.9) | |
| BD SST™ | Negative (<0.7) | 99 | 1 | 0 | 100 |
| | Equivocal (≥0.7 to <0.9) | 2 | 1 | 0 | 3 |
| | Positive (≥0.9) | 0 | 1 | 0 | 1 |
| All | | 101 | 3 | 0 | 104 |
Instrument(s):
1- Abbott Axsym
2- Bayer ADVIA Centaur
3- Beckman Coulter Access 2
4- Beckman Coulter UniCel DxC 800
5- Biomerieux Vidas
6- DPC Immulite 1000
7- Dynex DS2/Wampole Laboratories
8- Olympus AU5200TM / AU5400TM
9- Ortho Clinical Diagnostics VITROS Q ECi
10- Ortho Clinical Diagnostics VITROS 5,1 FS
11- Roche COBAS Integra 800
12- Roche Modular
Summary of all the studies:
Study 1: 35 apparently healthy adults were enrolled internally at the manufacturer facility. Testing was performed internally on-site and at a reference laboratory. Chemistry results were generated for 23 routine chemistry analytes on an Olympus AU5200/AU5400 instrument and 5 immunoassay analytes on a DPC Immulite 1000 instrument.
Study 2: 32 apparently healthy adults were enrolled internally at the manufacturer facility. Testing was performed internally on-site and at a reference laboratory. Chemistry results were generated for 8 routine chemistry analytes on an Olympus AU5200/AU5400 instrument and 31 routine chemistry analytes on a Roche COBAS Integra 800 instrument.
Study 3: 42 apparently healthy adults were enrolled internally at the manufacturer facility. Testing was performed internally on-site. Chemistry results were generated for 1 special chemistry analyte on a DPC Immulite 1000 instrument, 10 special chemistry analytes on a Beckman Coulter Access 2 instrument and 12 special chemistry analytes on a Bayer ADVIA Centaur instrument.
{11}
Study 4: 84 adults' subjects (43 for cardiac marker and 41 for female hormone) were recruited and tested in a hospital setting. Testing was performed in that hospital laboratory. Chemistry results were generated for 3 female hormones on a Beckman Coulter Access 2 instrument and an Ortho clinical Diagnostics VITROS ECi instrument. 3 cardiac markers results were generated on a Beckman Coulter Access 2 instrument and on an Abbott Axsym instrument.
Study 5: 86 adults' subjects were enrolled in a hospital setting with various diagnoses and disease states to cover a wide range for the analytes to be tested. 10 chemistry analytes (ALT, calcium, creatinine, glucose, potassium, LDL cholesterol, magnesium, phosphorus, total bilirubin, and total protein) were tested on the Roche Modular analyzer.
Study 6: 78 adults' subjects were enrolled in a hospital setting with various diagnoses and disease states to cover a wide range for the analytes to be tested. Two serological analytes (anti-CNV IgG and IgM) were tested on the BioMerieux and Dynex DS2 analyzers. Two immunoassay analytes (TSH and Troponin) were tested on the Abbott Axsym and Beckman Access analyzers. Six immunological analytes (C3, C4, CRP, IgG, IgM, and RF) were tested on the Beckman UniCel DxC 800 analyzer.
Study 7: 82 adults' subjects were enrolled in a hospital setting and 19 adults' subjects were enrolled in a different hospital setting with various diagnoses and disease states to cover a wide range for the analytes to be tested. Six immunological analytes and 10 general chemistry analytes were tested on the Ortho-Diagnostics Vitros analyzer.
b. Matrix comparison:
Not applicable. These blood collection tubes are for serum only.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
{12}
5. Expected values/Reference range:
Not applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
13
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.