VITROS IMMUNODIAGNOSTIC PRODUCTS FOLATE VERIFIERS, VITROS IMMUNODIAGNOSTIC PRODUCTS VITAMIN B12 RANGE VERIFIERS

K990026 · Ortho-Clinical Diagnostics, Inc. · JJX · Jan 29, 1999 · Clinical Chemistry

Device Facts

Record IDK990026
Device NameVITROS IMMUNODIAGNOSTIC PRODUCTS FOLATE VERIFIERS, VITROS IMMUNODIAGNOSTIC PRODUCTS VITAMIN B12 RANGE VERIFIERS
ApplicantOrtho-Clinical Diagnostics, Inc.
Product CodeJJX · Clinical Chemistry
Decision DateJan 29, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Indications for Use

For in vitro use in verifying the calibration range of the VITROS Immunodiagnostic System when used for the quantitative measurement of folate in human serum, plasma (heparin) and whole blood. For in vitro use in verifying the calibration range of the VITROS Immunodiagnostic System when used for the quantitative measurement of vitamin B12 in human serum and plasma [EDTA or heparin].

Device Story

VITROS Immunodiagnostic Products Range Verifiers are liquid or freeze-dried human plasma/buffered matrix spiked with folate or vitamin B12. Used in clinical laboratory settings to verify the calibration range of the VITROS Immunodiagnostic System. The system utilizes luminescence-based immunoassay technology with coated microwells for solid-phase separation. The verifiers are provided in low and high levels to confirm system performance across the analytical measurement range. Healthcare providers use the output to ensure the accuracy and reliability of patient test results for folate and vitamin B12, facilitating appropriate clinical decision-making.

Clinical Evidence

No clinical data. Substantial equivalence is based on physical properties and intended use comparisons with the predicate device.

Technological Characteristics

Matrix: liquid or freeze-dried human plasma or buffered matrix spiked with analyte. Levels: low and high. Technology: luminescence-based immunoassay. System integration: dedicated for use with VITROS Immunodiagnostic System instrumentation.

Indications for Use

Indicated for in vitro verification of calibration ranges for the VITROS Immunodiagnostic System during quantitative measurement of folate (serum, plasma, whole blood) and vitamin B12 (serum, plasma).

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JAN 29 1999 # Chapter 1 - Summary Information # 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92 The assigned 510(k) number is: 3990034 __ ### 1. Submitter name, address, contact Ortho-Clinical Diagnostics, Inc. 100 Indigo Creek Drive Rochester, New York 14626-5101 (716) 453-3790 Contact Person: Anne Zavertnik Date 510(k) prepared: January 4, 1999 #### 2. Device Name Trade or Proprietary Name: VITROS Immunodiagnostic Products Folate Range Verifiers; VITROS Immunodiagnostic Products Vitamin B12 Range Verifiers. Common Name: Range Verifiers Classification Name: VITROS Range verifiers for use in verifying the calibration range of the VITROS Immunodiagnostic System when used for the measurement of immunoassays which include Folate and Vitamin B12. ### 3. Predicate Device The VITROS Immunodiagnostic Products Folate and Vitamin B12 Range Verifiers are substantially equivalent to VITROS Immunodiagnostic Products FSH Range Verifiers (K973517). ## 4. Device Description The VITROS Immunodiagnostic System uses luminescence as the signal in the quantitative and semi-quantitative determination of selected analytes in human body fluids, commonly serum, plasma and urine. Coated microwells are used as the solid phase separation system. The system is comprised of three main elements: - 1. The VITROS Immunodiagnostic Products (in this case VITROS Immunodiagnostic Products Reagent Pack, VITROS Immunodiagnostic Products Calibrators, which are combined by the VITROS Immunodiagnostic System to perform a VITROS assay). Continued on next page {1}------------------------------------------------ ## 510(k) Summary, Continued. - 2. The VITROS Immunodiagnostic System instrumentation, which provides automated use of the immunoassay kits. The VITROS Immunodiagnostic System was cleared for market by a separate 510(k) pre-market notification (K962919). - 3. Common reagents used by the VITROS System in each assay. The VITROS Immunodiagnostic Products Signal Reagent and VITROS Immunodiagnostic Products Universal Wash Reagent were cleared as part of the VITROS Immunodiagnostic Products Total T3 510(k) pre-market notification (K964310). The VITROS System and common reagents are dedicated specifically only for use with the VITROS Immunodiagnostic Products range of immunoassay products. #### 5. Device Intended Use For in vitro use in verifying the calibration range of the VITROS Immunodiagnostic System when used for the quantitative measurement of folate in human serum, plasma (heparin) and whole blood. For in vitro use in verifying the calibration range of the VITROS Immunodiagnostic System when used for the quantitative measurement of vitamin B12 in human serum and plasma [EDTA or heparin]. #### 6. Comparison to Predicate Device The VITROS Immunodiagnostic Products Range Verifiers are substantially equivalent to VITROS FSH Range Verifiers (predicate device), which was approved by FDA (K973517) for IVD use. Table 1 lists the similarities and differences of the device characteristics between the VITROS Folate Range Verifiers with the predicate device, VITROS FSH Range Verifiers. Continued on next page {2}------------------------------------------------ # 510(k) Summary, Continued Table 1 List of the assay characteristics | Device<br>Characteristic | VITROS FOLATE<br>Range Verifiers | Predicate<br>Device | |---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------| | Intended use | For use in verifying the<br>calibration range of the<br>VITROS<br>Immunodiagnostic System<br>when used for the<br>measurement of a<br>particular analyte (see page<br>11 for a list of analytes). | For use in verifying the<br>calibration range of the<br>VITROS<br>Immunodiagnostic<br>System when used for the<br>measurement of FSH. | | Matrix of Range Verifiers | A base matrix of liquid or<br>freeze-dried human plasma<br>or buffered matrix spiked<br>with analyte (see page 11<br>for full details). | A base matrix of freeze-<br>dried human plasma<br>spiked with human<br>pituitary FSH. | | Range Verifier levels | Low and high | Low and high | ## 7. Conclusions The data presented in the pre-market notification demonstrate that the VITROS Folate and Vitamin B12 Range Verifiers are substantially equivalent to the predicate device, for which there is FDA clearance. Equivalence was demonstrated by comparing the physical properties and intended uses of these devices with commercially available reagents. The data presented in the premarket notification provide a reasonable assurance that the VITROS Folate and Vitamin B12 Range Verifiers are safe and effective for the stated intended use. {3}------------------------------------------------ Public Health Service Image /page/3/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which consists of a stylized caduceus-like symbol with three human profiles facing right. The profiles are stacked on top of each other, creating a sense of depth. Encircling the emblem is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. JAN 29 1999 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Anne Zavertnik Regulatory Affairs Associate Ortho-Clinical Diagnostics Regulatory Affairs MC00882 100 Indigo Creek Drive Rochester, NY 14626-5101 - Re: K990026 Trade Name: VITROS Immunodiagnostic Products Folate Range Verifiers VITROS Immunodiagnostic Products Vitamin B12 Range Verifiers Regulatory Class: I Product Code: JJX Dated: January 4, 1999 Received: January 5, 1999 Dear Ms. Zavertnik: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Statement of Intended Use | 510(k) Number (if known): | K 990026 | |---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | VITROS Immunodiagnostic Products Range Verifiers | | Indications for Use: | For <i>in vitro</i> use in verifying the calibration range of the VITROS Immunodiagnostic System when used for the quantitative measurement of folate in human serum, plasma (heparin) and whole blood. | | | For <i>in vitro</i> use in verifying the calibration range of the VITROS Immunodiagnostic System when used for the quantitative measurement of vitamin B12 in human serum and plasma [EDTA or heparin]. | (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number: K990026 ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) . Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use<br>(Per 21 CFR 801.109) | <div> <img alt="Checkmark" src="checkmark.png"/> </div> | |------------------------------------------|---------------------------------------------------------| |------------------------------------------|---------------------------------------------------------| OR | Over-The-Counter Use<br>(Optional Format 1-2-96) | | |--------------------------------------------------|--| |--------------------------------------------------|--|
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...