IMMULITE 2000; CORTISOL CALIBRATION VERIFICATION MATERIAL, FOLIC ACID CALIBRATION VERIFICATION MATERIAL, VITAMIN B12 CAL

K141444 · Siemens Healthcare Diagnostics, Inc. · JJX · Jul 9, 2014 · Clinical Chemistry

Device Facts

Record IDK141444
Device NameIMMULITE 2000; CORTISOL CALIBRATION VERIFICATION MATERIAL, FOLIC ACID CALIBRATION VERIFICATION MATERIAL, VITAMIN B12 CAL
ApplicantSiemens Healthcare Diagnostics, Inc.
Product CodeJJX · Clinical Chemistry
Decision DateJul 9, 2014
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 862.1660
Device ClassClass 1

Intended Use

The IMMULITE® Cortisol Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE Cortisol assay on the IMMULITE 2000 systems. The IMMULITE® Folic Acid Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE Folic Acid assay on the IMMULITE 2000 systems. The IMMULITE® Vitamin B12 Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of the IMMULITE Vitamin B12 assay on the IMMULITE 2000 systems.

Device Story

The IMMULITE 2000 Calibration Verification Materials (CVM) are multi-level, assayed quality control materials used to verify the calibration of specific assays on the IMMULITE 2000 system. The CVMs are provided as liquid (Cortisol) or lyophilized (Folic Acid, Vitamin B12) human serum or albumin-based matrices containing the respective analytes. The device is used in clinical laboratory settings by trained personnel. The healthcare provider runs the CVMs on the IMMULITE 2000 system; the instrument processes the samples and generates dose values based on the reference calibrator curve. The provider compares these recovered values against established target ranges to verify that the assay calibration remains accurate. This verification process ensures the reliability of patient sample testing, allowing for accurate clinical decision-making regarding patient diagnosis and monitoring.

Clinical Evidence

Bench testing only. Stability studies validated real-time shelf life and open-vial stability. Value assignment was validated using commercially available controls and patient serum samples. Acceptance criteria were based on guideline ranges (±2SD) and specific dose recovery limits for each analyte level.

Technological Characteristics

Multi-level (4 levels) calibration verification materials. Matrices: human serum (Cortisol) or human albumin (Folic Acid, Vitamin B12) with preservatives. Form: liquid or lyophilized. Storage: -20°C. Connectivity: used on IMMULITE 2000 systems. Stability: validated per CEN 13640.

Indications for Use

Indicated for in vitro diagnostic use to verify calibration of IMMULITE 2000 Cortisol, Folic Acid, and Vitamin B12 assays on IMMULITE 2000 systems. For prescription use only.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 1 ### 510(k) Summary Introduction: According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for determination of substantial equivalence. The assigned 510(k) Number: ______k1414444 | 1. Submitter<br>Mailing Address: | Siemens Healthcare Diagnostics Inc.<br>5210 Pacific Concourse Drive<br>Los Angeles, CA 90045 | |------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person:<br>Phone Number:<br>Fax Number:<br>E-mail Address:<br>Date Prepared: | Donna Velasquez<br>Regulatory Technical Specialist<br>(310)-645-8200 x7403<br>(310)-645-9999<br>donna.velasquez@siemens.com<br>May 30, 2014 | | 2. Device Name<br>Proprietary Name:<br>Measurand:<br>Type of Test:<br>Regulation Section:<br>Classification:<br>Products Code:<br>Panel: | IMMULITE® 2000 Cortisol Calibration Verification Material<br>Quality Control materials for IMMULITE® 2000 Cortisol assay<br>Calibration Verification Material (CVM) for IMMULITE® 2000<br>Cortisol assay<br>21 CFR 862.1660, Quality Control Material<br>Class I Reserved<br>JJX - Single (Specified) Analyte Controls (Assayed and<br>Unassayed)<br>Clinical Chemistry (75) | | 3. Predicate Device Name<br>Predicate 510(k) No: | IMMULITE® 2000 Total T3 Calibration Verification Material<br>(CVM)<br>K133124 | | 4. Device Description: | The IMMULITE® 2000 Cortisol Calibration Verification Material<br>(CVM) contains one set of four vials each 3mL. CVM1 contains<br>human serum with preservatives and CVM2, CVM3, and CVM4<br>contain various levels of cortisol in human serum<br>with<br>preservatives. | | 5. Intended Use:<br>Indication for Use: | See Indications for Use Statement below<br>The IMMULITE® Cortisol Calibration Verification Material<br>(CVM) is for in vitro diagnostic use in the verification of<br>calibration of the IMMULITE Cortisol assay on the IMMULITE<br>2000 systems | | Special Conditions for<br>Use Statement(s):<br>Special Instrument<br>Requirements: | For prescription use only<br>IMMULITE® 2000 Systems | {1}------------------------------------------------ #### 6. Technological Characteristics and Substantial Equivalence Comparison with Predicate: A comparison of the device features, intended use, and other · information demonstrates that the IMMULITE® 2000 Cortisol Calibration Verification Material (CVM) is substantially equivalent to the predicate device as summarized in Table 1. | SIMILARITIES | | | |-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Candidate Device<br>IMMULITE 2000 Cortisol CVM | Predicate Device<br>IMMULITE 2000 Total T3 CVM | | Intended<br>Use | The IMMULITE® Cortisol Calibration<br>Verification Material (CVM) is for in<br>vitro diagnostic use in the verification of<br>calibration of the IMMULITE Cortisol<br>assay on the IMMULITE 2000 systems. | The IMMULITE® Total T3<br>Calibration Verification Material<br>(CVM) is for in vitro diagnostic use in<br>the verification of calibration of the<br>IMMULITE Total T3 assay on the<br>IMMULITE 2000 systems | | Storage | -20°C | Same | | Form | Liquid | Same | | Stability | Stable unopened until the<br>expiration date | Same | | Levels | 4 | Same | | Matrix | Human Serum with preservatives | Same | | Use | Single Use Only | Same | | | DIFFERENCES | | | | Candidate Device<br>IMMULITE 2000 Cortisol CVM | Predicate Device<br>IMMULITE 2000 Total T4 CVM | | Analyte | Cortisol | Total T3 | Table 1: Substantial Equivalence Comparison ### 7. Non-Clinical Performance Testing Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the candidate device. #### Stability Summary: The stability study was conducted to validate real-time shelf life and open component (in-use or open vial) claim for the IMMULITE 2000 Cortisol Calibration Material (CVM) to ensure that it maintains optimal product performance on IMMULITE 2000 platforms throughout the established shelf life of the CVM. The Calibration Verification Materials are stable up to 14 months when stored at -20°C prior to opening and stable for 8 hours at ambient or room temperature (15-25°C) after opening. {2}------------------------------------------------ #### Stability Protocol Summary: The CVM study protocols are run as part of the calibrator stability testing. The stability calibrators/CVMs are run in duplicate (as a minimum) at the time points shown in Table 2, and the dose value is determined from the reference calibrator curve. | CVM Level | Time-Points<br>(Days) | | | | |-----------|-----------------------|-----|-----|-----| | LCOCVM1 | 1 | 182 | 365 | 425 | | LCOCVM2 | 1 | 182 | 365 | 425 | | LCOCVM3 | 1 | 182 | 365 | 425 | | LCOCVM4 | 1 | 182 | 365 | 425 | #### Table 2: Stability Protocol Summary For Open Component testing, the results were determined from a 2-point adjustment. Using IMMULITE 2000 Cortisol (L2KCO) kit lot 362, lot 090 CVMs were tested at 2-hourly intervals for up to 9 hours at ambient or room temperature (15-25°C) conditions. #### Stability Acceptance Criteria Summary: The Acceptance Criteria for the Cortisol Calibration Verification Material (CVM) are in 2 parts. Part 1 consists of guideline acceptance criteria which require dose value of stability calibrators/CVM to fall between ±20% of assigned dose for CVM level 2, ±15% of assigned dose for level 3 and ±19% of assigned dose for level 4. Part 2 review limits criteria which require dose value of the controls to be within 2 Standard Deviations (SD) of the control target value when generated from the stability calibrator curve. If the result is not within acceptable dose range of ±20% for level 2, ±15% of assigned dose for level 3 and within ±19% of assigned dose for level 4, then additional data review is conducted using part 2 criteria. ### Traceability: The IMMULITE Cortisol CVMs are traceable to an internal material which has been gravimetrically prepared. The CVMs are manufactured using qualified materials and measurement procedures. ### Value Assignment: The IMMULITE Cortisol CVMs are 4 level materials which are a subset of 8 level Cortisol calibrators. Calibrators are not commercialized but are used internally during manufacture and release testing of Cortisol reagents and two point adjustors. The IMMULITE calibrators and therefore CVMs are value assigned using assigned reference calibrators. The CVMs are manufactured using qualified materials and measurement procedures. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples, spiked patient samples and controls using the assigned CVM values. Six levels of commercially available controls and 25 patient serum samples, 3 spiked normal serum samples and 5 normal serum samples were used to validate calibrator/CVM value assignments. {3}------------------------------------------------ ### Expected Values/Target Values/Reference Range: Each CVM level was tested for a total of 27 replicates; 9 runs and 3 replicates per run, 3 different reagent kit lots and 8 IMMULITE 2000 systems. The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD). The target values are provided in the IMMULITE® 2000 CVM Calibration Material lotspecific value card. The expected assay range is 1 to 50 g/dL. The target values in Table 3 can be considered as guidelines. | Analyte target<br>levels | CVM Level | *Target<br>Mean<br>g/dL) | Standard<br>Deviation<br>(SD) | Guideline ±2SD<br>Range | | |--------------------------|--------------|--------------------------|-------------------------------|-------------------------|------| | | LCOCVM1 | 0.00 | - | 0. | 1.0 | | | LCOCVM2 | 4.54 | 0.455 | 3. | 5.45 | | | LCOCVM3 | 18.7 | 1.4 | 15 | 21.5 | | | LCOCVM4 | 54.5 | 5.2 | 44 | 64.9 | | Assay Range | 1 to 50 g/dL | | | | | #### Table 3: Analyte Target Range Values *Note: when CVMs are run by the customer, an actual value below and above the assay range will result when the customer programs the CVMs as calibration verifiers in the instrument software. If programmed as a patient or control, then the software will give values as < or > the assay lower and upper range. Each laboratory should establish their limits for acceptability based on methodology, clinical significance and medical decision levels of the test analyte. The representative, total precision tabulated in the respective assay instructions for use may be considered as one factor when establishing local, acceptable ranges. The values provided above may be considered as guidelines. Value assignment is lot specific. #### Standard/Guidance Documents Referenced: CEN 13640 Stability Testing of In Vitro Diagnostic Reagents Guidance for Industry - Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators Guidance for Industry and FDA Staff - Assayed and Unassayed Quality Control Material #### Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10 #### 8. Conclusion: The IMMULITE® 2000 Cortisol Calibration Verification Material is substantially equivalent to the predicate device intended for similar use. The substantial equivalence of the device is supported by the non-clinical testing and was found to be comparable and supports the claims of substantial equivalence, product safety and effectiveness. Based on the testing completed and the comparisons with predicate device, the IMMULITE® 2000 Cortisol Calibration Verification Material does not raise any new questions on safety and effectiveness and the results support a determination of substantial equivalence. {4}------------------------------------------------ ### 510(k) Summary Introduction: According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for determination of substantial equivalence. The assigned 510(k) Number: k141444 - 1. Submitter Mailing Address: Contact Person: Phone Number: Fax Number: E-mail Address: Date Prepared: - 2. Device Name Proprietary Name: Measurand: Type of Test: Regulation Section: Classification: Products Code: Panel: - 3. Predicate Device Name Predicate 510(k) No: - 4. Device Description: - 5. Intended Use: Indication for Use: Special Conditions for Use Statement(s): Special Instrument Requirements: Siemens Healthcare Diagnostics Inc. 5210 Pacific Concourse Drive Los Angeles. CA 90045 Donna Velasquez Regulatory Technical Specialist (310) 645-8200 x7403 (310) 645-9999 donna.velasquez@siemens.com May 30, 2014 IMMULITE® 2000 Folic Acid Calibration Verification Material Quality Control materials for IMMULITE® 2000 Folic Acid assay Calibration Verification Material (CVM) for IMMULITE® 2000 Folic Acid assay 21 CFR 862.1660, Quality Control Material Class I Reserved JJX - Single (Specified) Analyte Controls (Assayed and Unassaved) Clinical Chemistry (75) IMMULITE 2000 DHEA-SO4 Calibration Verification Material (CVM) K140258 The IMMULITE® 2000 Folic Acid Calibration Verification Material (CVM) contains one set of four vials, 3mL (CVM1) each. when reconstituted. CVM1 contains a lyophilized human albumin based matrix with preservatives. CVM2. CVM3. and CVM4 contain various levels of lyophilized folic acid in human albumin based matrix with preservatives. See Indications for Use Statement below The IMMULITE® Folic Acid Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE Folic Acid assay on the IMMULITE 2000 systems For prescription use only IMMULITE® 2000 Systems {5}------------------------------------------------ - 6. Technological Characteristics and Substantial Equivalence Comparison with Predicate: A comparison of the device features, intended use, and other information demonstrates that the IMMULITE® 2000 Folic Acid Calibration Verification Material (CVM) is substantially equivalent to the predicate device as summarized in Table 1. | SIMILARITIES | | | | | |-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--| | | Candidate Device<br>IMMULITE 2000 Folic Acid CVM | Predicate Device<br>IMMULITE 2000 DHEA-SO4<br>CVM | | | | Intended<br>Use | The IMMULITE® Folic Acid<br>Calibration Verification Material (CVM)<br>is for in vitro diagnostic use in the<br>verification of calibration of the<br>IMMULITE Folic Acid assay on the<br>IMMULITE 2000 systems | The IMMULITE® DHEA-SO4<br>Calibration Verification Material<br>(CVM) is for in vitro diagnostic use<br>in the verification of calibration of<br>the IMMULITE DHEA-SO4 assay<br>on the IMMULITE 2000 systems | | | | Form | Lyophilized | Same | | | | Levels | 4 | Same | | | | Stability | Stable unopened until the expiration date | Same | | | | Storage | -20°C | Same | | | | Use | Single Use Only | Same | | | | | Table 1: Substantial Equivalence Comparison | | | | |--|---------------------------------------------|--|--|--| |--|---------------------------------------------|--|--|--| | | DIFFERENCES | | |---------|--------------------------------------------------|---------------------------------------------------| | | Candidate Device<br>IMMULITE 2000 Folic Acid CVM | Predicate Device<br>IMMULITE 2000 DHEA-SO4<br>CVM | | Analyte | Folic Acid | DHEA-SO4 | | Matrix | Human Albumin with preservatives | Human Serum with Preservatives | ### 7. Non-Clinical Performance Testing Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the candidate device. #### Stability Summary: The stability study was conducted to validate real-time shelf life and open component (in-use or open vial) claim for the IMMULITE 2000 Folic Acid Calibration Verification Material (CVM) to ensure that it maintains optimal product performance on IMMULITE 2000 platforms throughout the established shelf life of the CVM. The Calibration Verification Materials are stable up to 3.5 years when stored at -20°C prior to opening, and for 8 hours at ambient or room temperature (15-25°C) after reconstitution. ### Stability Protocol Summary: 1 The CVM study protocols are run as part of the calibrator stability testing. The stability CVMs and reference CVMs are run in duplicate (as a minimum) at the time points shown in Table 2 and the dose value determined from the reference calibrator curve. {6}------------------------------------------------ | CVM Level | Time-Points<br>(Days) | | | | |-----------|-----------------------|-----|------|------| | LFOCVM1 | 1 | 912 | 1095 | 1280 | | LFOCVM2- | 1 | 912 | 1095 | 1280 | | LFOCVM3 | 1 | 912 | 1095 | 1280 | | LFOCVM4 | 1 | 912 | 1095 | 1280 | #### Table 2: Stability Protocol Summary For Open Component testing, the results were determined from a 2-point adjustment. Using IMMULITE 2000 Folic Acid (L2KFO) kit lot 446, lot 090B Folic Acid CVMs were tested at 2hourly intervals for up to 9 hours at ambient or room temperature (15-25°C) conditions. #### Stability Acceptance Criteria Summary: The Acceptance Criteria for the IMMULITE Folic Acid Calibration Verification Material (CVM) are in 2 parts. Part 1 consists of guideline acceptance criteria which require dose value of stability CVMs to fall between ±15% of the assigned dose for CVM level 2, within ±10% of the assigned dose for level 3 and within ±12% of assigned dose for CVM level 4. Part 2 review limits criteria which require dose value of the controls to be within 2 Standard Deviations (SD) of the control target value when generated from the stability calibrator curve. If the result is not within acceptable dose range of ±15% of the assigned dose for CVM level 2, within ±10% of the assigned dose for level 3 and within ±12% of assigned dose for CVM level 4 then additional data review is conducted using part 2 criteria. #### Traceability: The IMMULITE Folic Acid CVMs are traceable to an internal material which has been gravimetrically prepared. The CVMs are manufactured using qualified materials and measurement procedures. #### Value Assignment: The IMMULITE Folic Acid CVMs are 4 level materials which are a subset of 8 level Folic Acid calibrators. Calibrators are not commercialized but are used internally during manufacture and release testing of Folic Acid reagents and two point adjustors. The IMMULITE calibrators and therefore CVMs are value assigned using assigned reference calibrators. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. The CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples and controls using the assigned CVM values. The controls must fall within their target ranges. Six levels of commercially available controls and 14 patient serum samples were used to validate calibrator/CVM value assignments. ### Expected Values/Target Values/Reference Range: The CVMs are manufactured using qualified materials and measurement procedures. The Folic Acid CVMs were tested on 27 replicates in total comprised of 9 runs and 3 replicates per run on 6 IMMULITE 2000 systems and 3 different reagent kit lots. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. The CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples and controls using the assigned CVM values. The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD). The expected assay range is 1 to 24 ng/mL. The target values in Table 3 can be considered as guidelines. {7}------------------------------------------------ #### Table 3: Analyte Target Range Levels | Analyte target<br>levels | CVM Level | *Target<br>Mean<br>(ng/mL) | Standard<br>Deviation<br>(SD) | Guideline ±2SD Range<br>(ng/mL) | |--------------------------|---------------|----------------------------|-------------------------------|---------------------------------| | | LFOCVM1 | 0.00 | - | 0.00<br>1.0 | | | LFOCVM2 | 3.11 | 0.28 | 2.55<br>3.67 | | | LFOCVM3 | 12.4 | 0.75 | 10.9<br>13.9 | | | LFOCVM4 | 26.3 | 1.6 | 23.1<br>29.5 | | Assay Range | 1 to 24 ng/mL | | | | *Note: when CVMs are run by the customer, an actual value below and above the assay range will result when the customer programs the CVMs as calibration verifiers in the instrument software. If programmed as a patient or control, then the software will give values as < or > the assay lower and upper range. Each laboratory should establish their limits for acceptability based on methodology, clinical significance and medical decision levels of the test analyte. The representative, total precision tabulated in the respective assay instructions for use may be considered as one factor when establishing local, acceptable ranges. The values provided above may be considered as guidelines. Value assignment is lot specific. Standard/Guidance Documents Referenced: CEN 13640 Stability Testing of In Vitro Diagnostic Reagents Guidance for Industry - Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators Guidance for Industry and FDA Staff – Assayed and Unassayed Quality Control Material ### Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10 #### 8. Conclusion: The IMMULITE® 2000 Folic Acid Calibration Verification Material is substantially equivalent to the predicate device intended for similar use. The substantial equivalence of the device is supported by the non-clinical testing and was found to be comparable and supports the claims of substantial equivalence, product safety and effectiveness. Based on the testing completed and the comparisons with predicate device, the IMMULITE® 2000 Folic Acid Calibration Verification Material does not raise any new questions on safety and effectiveness and the results support a determination of substantial equivalence. {8}------------------------------------------------ ### 510(k) Summary Introduction: According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for determination of substantial equivalence. . The assigned 510(k) Number: k141444 | the assigned 510(k) Number: | K141444 | |------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 1. Submitter<br>Mailing Address: | Siemens Healthcare Diagnostics Inc.<br>5210 Pacific Concourse Drive<br>Los Angeles, CA 90045 | | Contact Person:<br>Phone Number:<br>Fax Number:<br>E-mail Address:<br>Date Prepared: | Donna Velasquez<br>Regulatory Technical Specialist<br>(310)-645-8200 x7403<br>(310)-645-9999<br>donna.velasquez@siemens.com<br>May 30, 2014 | | 2. Device Name<br>Proprietary Name:<br>Measurand:<br>Type of Test:<br>Regulation Section:<br>Classification:<br>Products Code:<br>Panel: | IMMULITE® 2000 Vitamin B12 Calibration Verification Material<br>Quality Control materials for IMMULITE 2000 Vitamin B12<br>Calibration Verification Material (CVM) for IMMULITE® 2000<br>Vitamin B12 assay<br>21 CFR 862.1660, Quality Control Material<br>Class I Reserved<br>JJX - Single (Specified) Analyte Controls (Assayed and Unassayed)<br>Clinical Chemistry (75) | | 3. Predicate Device Name<br>Predicate 510(k) No: | IMMULITE® 2000 DHEA-SO4 Calibration Verification Material<br>K140258 | | 4. Device Description: | IMMULITE® 2000 Vitamin B12 Calibration Verification Material<br>(CVM) contains one set of four vials, 3 mL each, when reconstituted.<br>CVM1 contains a lyophilized human albumin-based matrix with<br>preservatives. CVM2, CVM3, and CVM4 contain various levels of<br>lyophilized Vitamin B12 in human albumin-based matrix with<br>preservatives. | | 5. Intended Use:<br>Indication for Use: | See Indications for Use Statement below<br>The IMMULITE® Vitamin B12 Calibration Verification Material<br>(CVM) is for in vitro diagnostic use in the verification of the<br>IMMULITE Vitamin B12 assay on the IMMULITE 2000 systems | | Special Conditions for<br>Use Statement(s):<br>Special Instrument<br>Requirements: | For prescription use only<br>IMMULITE® 2000 Systems | {9}------------------------------------------------ - 6. Technological Characteristics A comparison of the device features, intended use, and other information demonstrates that the IMMULITE® 2000 Vitamin B12 and Substantial Equivalence Calibration Verification Material (CVM) is substantially equivalent Comparison with Predicate: to the predicate device, as summarized in Table 1. #### Table 1: Substantial Equivalence Comparison | SIMILARITIES | | | |-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Candidate Device<br>IMMULITE 2000 Vitamin B12 CVM | Predicate Device<br>IMMULITE 2000 DHEA-SO4 CVM | | Intended<br>Use | The IMMULITE® Vitamin B12<br>Calibration Verification Material<br>(CVM) is for in vitro diagnostic use in<br>the verification of calibration of the<br>IMMULITE Vitamin B12 assay on the<br>IMMULITE 2000 systems. | The IMMULITE® DHEA-SO4<br>Calibration Verification Material<br>(CVM) is for in vitro diagnostic use in<br>the verification of calibration of the<br>IMMULITE DHEA-SO4 assay on the<br>IMMULITE 2000 systems | | Form | Lyophilized | Same | | Stability | Stable unopened until the<br>expiration date | Same | | Storage | -20°C | Same | | Levels | 4 | Same | | Use | Single Use Only | Same | | | DIFFERENCES | | |---------|---------------------------------------------------|------------------------------------------------| | | Candidate Device<br>IMMULITE 2000 Vitamin B12 CVM | Predicate Device<br>IMMULITE 2000 DHEA-SO4 CVM | | Analyte | Vitamin B12 | DHEA-SO4 | | Matrix | Human Albumin with preservatives | Human Serum with preservatives | #### 7. Non-Clinical Performance Testing Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the candidate device. #### Stability Summary: The stability study was conducted to validate real-time shelf life and open component (in-use or open vial) claim for the IMMULITE 2000 Vitamin B12 CVM to ensure that it maintains optimal product performance on IMMULITE 2000 platforms throughout the established shelf life of the CVM. The Calibration Verification Materials are stable up to 69 months when stored at -20°C prior to opening and stable for 8 hours at room (ambient) temperature (15-25°C) after reconstitution. #### Stability Protocol Summary: The CVM study protocols are run as part of the calibrator stability testing. As summarized in Table 2, the testing was run in duplicate (at the minimum) and contains time-points at which the testing was performed. {10}------------------------------------------------ | CVM Level | Time-Points<br>(Months) | | | | |-----------|-------------------------|----|----|----| | LVBCVM1 | 0 | 67 | 68 | 69 | | LVBCVM2 | 0 | 67 | 68 | 69 | | LVBCVM3 | 0 | 67 | 68 | 69 | | LVBCVM4 | 0 | 67 | 68 | 69 | For Open Component testing, the results were determined from a 2-point adjustment. Using IMMULITE 2000 Vitamin B12 (L2KVB) kit lot 467, lot 108 Vitamin B12 CVMs were tested at 2hourly intervals for up to 9 hours at ambient or room temperature (15-25°C) conditions. #### Stability Acceptance Criteria Summary: The Acceptance Criteria for the IMMULITE Vitamin B12 Calibration Verification Material (CVM) are in 2 parts. Part 1 consists of guideline acceptance criteria which require dose value of stability calibrator/CVM to fall between ±10% of assigned dose for CVM level 2 and level 3 and within ±21% of assigned dose for CVM level 4. Part 2 review limits criteria which require dose value of the controls to be within 2SD of the control target value when generated from the stability calibrator curve. If the result is not within acceptable dose range of ±10% for level 3 and ±21% for level 4, then additional data review is conducted using part 2 criteria. #### Traceability: The IMMULITE Vitamin B12 CVMs are traceable to an internal standard. The CVMs are manufactured using qualified materials and measurement procedures. ### Value Assignment: The IMMULITE® 2000 Vitamin B12 CVMs are 4 level materials which are a subset of 8 level Vitamin B12 calibrators. Calibrators are not commercialized but are used internally during manufacture and release testing of Vitamin B12 reagents and two point adjustors. The IMMULITE calibrators and therefore CVMs are value assigned reference calibrators. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each instrument independently. CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples and controls using the assigned CVM values. Eight levels of commercially available controls and 21 patient serum samples were used to validate CVM value assignments. {11}------------------------------------------------ #### Expected Values/Target Values/Reference Range: The Vitamin B12 CVMs were tested on 24 replicates in total comprised of 12 runs and 2 replicates per run, 3 IMMULITE 2000 and 5 IMMULITE/IMMULITE 1000 systems and 4 different reagent kit lots. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples and controls using the assigned CVM values. The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD). The expected assay range is 150 to 1000 pg/mL. The target values in Table 3 can be considered as guidelines. | Analyte target<br>levels | Level | *Target Mean<br>(ng/mL) | Standard<br>Deviation<br>(SD) | Guideline ±2SD Range<br>(ng/mL) | |--------------------------|-------------------|-------------------------|-------------------------------|---------------------------------| | | LVBCVM1 | 0.00 | - | 0.00 - 150 | | | LVBCVM2 | 234 | 23.5 | 187 - 281 | | | LVBCVM3 | 579 | 34.5 | 510 - 648 | | | LVBCVM4 | 1144 | 194.5 | 755 - 1533 | | Assay Range | 150 to 1000 pg/mL | | | | #### Table 3: Analyte Target Range Levels *Note: when CVMs are run by the customer, an actual value below and above the assay range will result when the customer programs the CVMs as calibration verifiers in the instrument software. If programmed as a patient or control, then the software will give values as < or > the assay lower and upper range. Each laboratory should establish their limits for acceptability based on methodology, clinical significance and medical decision levels of the test analyte. The representative, total precision tabulated in the respective assay instructions for use may be considered as one factor when establishing local, acceptable ranges. The values provided above may be considered as guidelines. Value assignment is lot specific. Standard/Guidance Documents Referenced: CEN 13640 Stability Testing of In Vitro Diagnostic Reagents Guidance for Industry - Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators Guidance for Industry and FDA Staff - Assayed and Unassayed Quality Control Material ### Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10 #### Conclusion: 9. The IMMULITE® 2000 Vitamin B12 Calibration Material is substantially equivalent to the predicate device intended for similar use. The substantial equivalence of the device is supported by the non-clinical testing and was found to be comparable and supports the claims of substantial equivalence, product safety and effectiveness. Based on the testing completed and the comparisons with predicate device, the IMMULITE® 2000 Vitamin B12 Calibration Material does not raise any new questions on safety and effectiveness and the results support a determination of substantial equivalence. {12}------------------------------------------------ Image /page/12/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized caduceus, a symbol often associated with medicine and healthcare. The emblem is positioned to the right of the department's name, which is written in a circular arrangement around the emblem. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 9, 2014 SIEMENS HEALTHCARE DIAGNOSTICS, INC. C/O DONNA VELASQUEZ REGULATORY TECHNICAL SPECIALIST 5210 PACIFIC CONCURSE DRIVE LOS ANGELES CA 90045 Re: k141444 Trade/Device Name: IMMULITE® 2000 Cortisol Calibration Verification Material IMMULITE® 2000 Folic Acid Calibration Verification Material IMMULITE® 2000 Vitamin B12 Calibration Verification Material Regulation Number: 21 CFR 862.1660 Regulation Name: Quality Control Material (assayed and unassayed) Regulatory Class: I, Reserved Product Code: JJX Dated: June 3, 2014 Received: June 5, 2014 Dear Donna Velasquez: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {13}------------------------------------------------ Page 2 - Ms. Donna Velasquez device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resources/orYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Courtney H. Lias -S Courtney H. Lias Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {14}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration #### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Dale: January 31, 2017 See PRA Statement below. 510(k) Number (if known) k141444 Device Name IMMULITE® 2000 Cortiso! Calibration Verification Material IMMULITE® 2000 Folic Acid Calibration Verification Material IMMULITE® 2000 Vitamin B12 Calibration Verification Material #### Indications for Use (Describe) The IMMULITE® Corlisol Calibration Material is for in vitro diagnostic use in the verification of calibration of the IMMULITE Cortisol assay on the IMMULITE 2000 systems. The IMMULITE® Folic Acid Calibration Material is for in vitro diagnostic use in the verification of calibration of the IMMULITE Folic Acid assay on the IMMULITE 2000 systems. The IMMULITE® Vitanin B12 Calibration Material is for in vitro diagnostic use in the verification of calibration of the IMMULITE Vitamin B12 assay on the IMMULITE 2000 systems. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpant C) PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FDA USE ONLY 。 Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure) Yung W. Chan -S This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.4 The burden lime for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and mainlain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Orug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStall@lda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a curronly valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...