STA - FREE PS CALIBRATOR

K092113 · Diagnostica Stago, Inc. · JIX · Nov 13, 2009 · Clinical Chemistry

Device Facts

Record IDK092113
Device NameSTA - FREE PS CALIBRATOR
ApplicantDiagnostica Stago, Inc.
Product CodeJIX · Clinical Chemistry
Decision DateNov 13, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1150
Device ClassClass 2

Indications for Use

STA® - Free PS Calibrator kit is a set of calibrator plasmas intended for use with analyzers of the STA® line suitable to these reagents for the calibration of free protein S assays by the immuno-turbidimetric method, STA® - Liatest® Free PS.

Device Story

STA® - Free PS Calibrator kit consists of four lyophilized human plasma vials containing defined quantities of free protein S, buffer, and stabilizers. Used in clinical laboratory settings with STA® line IVD analyzers (STA-R®, STA Compact®) to establish calibration curves for immuno-turbidimetric free protein S assays. Reconstituted with water by laboratory technicians; loaded onto analyzers to enable quantitative measurement of free protein S in patient samples. Output is a calibration curve used by the analyzer to calculate patient protein S levels, aiding in the assessment of congenital or acquired protein S deficiency and associated thrombo-embolism risk.

Clinical Evidence

No clinical data. Bench testing only. Precision/reproducibility studies performed across multiple lots and instruments (STA-R, STA Compact) showed total CVs ranging from 1.7% to 7.6%. Stability confirmed for 4 hours reconstituted on-board and 18 months lyophilized at 2-8°C.

Technological Characteristics

Lyophilized human plasma; contains buffer and stabilizers. Immuno-turbidimetric calibration principle. Designed for use with STA® line IVD analyzers. Traceable to NIBSC International Standard 03/228.

Indications for Use

Indicated for use as calibrator plasmas for the calibration of free protein S assays on STA® line IVD analyzers. For prescription use only.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k092113 B. Purpose for Submission: New device C. Measurand: Free Protein S D. Type of Test: Quantitative immuno-turbidimetric method E. Applicant: DIAGNOSTICA STAGO, INC. F. Proprietary and Established Names: STA® - Free PS Calibrator G. Regulatory Information: 1. Regulation section: 21 CFR 862.1150, Calibrator 2. Classification: Class II 3. Product code: JIX, Calibrator, Multi-Analyte Mixture 4. Panel: Hematology (81) H. Intended Use: 1. Intended use(s): STA® - Free PS Calibrator kit is a set of calibrator plasmas intended for use with analyzers of the STA® line suitable to these reagents for the calibration of free protein S assays by the immuno-turbidimetric method, STA® - Liatest® Free PS. 2. Indication(s) for use: Same as Intended Use 3. Special conditions for use statement(s): For Prescription Use Only 4. Special instrument requirements: STA® - Free PS Calibrator kit is for use with the STA® - Liatest® Free PS kit and STA® product line of IVD analyzers (STA-R® and STA Compact®). I. Device Description: The STA® - Free PS Calibrator kit is a set of lyophilized human plasmas used to create the calibration curve on an IVD instrument performing the immuno-turbidimetric method for free protein S assay, STA® - Liatest® Free PS. Such IVD instruments being the STA® product line of medical device analyzers such as STA-R® and STA Compact®. The calibrator kit consists of four (4) calibrators. Each calibrator is composed of lyophilized human plasma containing a well defined quantity of free protein S, buffer, and stabilizers. {1} 2 # Substantial Equivalence Information: 1. Predicate device name(s): HemosIL™ Calibration Plasma 2. Predicate K number(s): k041905 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | STA® - Free PS Calibrator | HemosIL™ Calibration Plasma | | Intended Use | STA® - Free PS Calibrator kit is a set of calibrator plasmas intended for use with analyzers of the STA® line suitable to these reagents for the calibration of free protein S assays by the immuno-turbidimetric method, STA® - Liatest® Free PS | HemosIL Calibration Plasma is intended for the calibration of coagulation assays on IL and ELECTRA coagulation systems. | | Parameters | Free Protein S | Protein S | | Design | Lyophilized human plasma reconstituted with water used in IVD analyzers | Same | | Storage | 2-8°C | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Instrumentation | STA-R® and STA Compact® | IL and ELECTRA coagulation systems. | | Parameters | | Antithrombin, Single Factors, Fibrinogen, Plasmin Inhibitor, Plasminogen, von Willebrand Factor, Protein C, PT, APTT, TT | # K. Standard/Guidance Document Referenced (if applicable): Not applicable # L. Test Principle: Protein S is a physiological inhibitor of the coagulation process. It acts as the cofactor of activated protein C. The congenital or acquired deficiency of protein S increases the risk of thrombo-embolism. The STA® - Free PS Calibrator kit is a set of lyophilized human plasmas used to create the calibration curve on an IVD instrument performing the immuno-turbidimetric method for free protein S assays. # M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: The free protein S level determinations were performed in eight different series on each calibrator level with the same instrument (STA-R®) using two different lots of STA® - Liatest® Free PS kit. Concurrently, the free protein S level determinations were also performed in eight different series on each {2} calibrator level with two different instruments (STA Compact® and STA-R®) using one lot of STA® - Liatest® Free PS kit. All tests were run within two to three weeks. A target value was established for each calibrator level; Reagent 1: 14%, Reagent 2: 29%, Reagent 3: 57%, Reagent 4: 95%. Total CV% for both studies is as follows: Inter-Assay Lot Reproducibility Data for STA® - Free® PS Calibrator | STA® - Liatest® Free PS | Free Protein S Level (%) | | | | | --- | --- | --- | --- | --- | | | Reagent 1 | Reagent 2 | Reagent 3 | Reagent 4 | | Lot 1 CV (%) | 6.7 | 3.7 | 1.7 | 2.6 | | Lot 2 CV (%) | 5.0 | 3.7 | 2.5 | 1.9 | Inter-Instrument Reproducibility Data for STA® - Free® PS Calibrator | Instrument | Free Protein S Level (%) | | | | | --- | --- | --- | --- | --- | | | Reagent 1 | Reagent 2 | Reagent 3 | Reagent 4 | | STA-R® CV (%) | 6.7 | 3.7 | 1.7 | 2.6 | | STA Compact® CV (%) | 7.6 | 3.6 | 2.8 | 2.6 | b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): Value assignment: For each calibrator level, the free protein S value corresponds to the mean of a minimum of 24 determinations performed with a minimum of 3 instruments, using at least two different lots of STA® - Liatest® Free PS. The determinations are carried out by a minimum of two laboratory technicians. Traceability: The STA® - Free PS Calibrators are determined against a secondary standard, International Standard 03/228 for free protein S established in 2006 by the National Institute for Biological Standards and Control (NIBSC). Stability: Three different lots of calibrators (Reagent 1, 2, 3, and 4) were used in the stability studies. Three vials each of Reagent 1, 2, 3, and 4 were reconstituted and stored for 4 hours on both analyzers (STA Compact and STA-R). In addition, freshly reconstituted reagents from the same three lots were run using the STA® - Liatest® Free PS assay. The reconstituted stability study result supports the claim of remaining stable for 4 hours on board the STA line of IVD analyzers. In the lyophilized state, the calibrator plasmas remain stable for a duration of 18 months from the date of manufacture when stored at 2-8°C. d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: {3} Not applicable b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Expected values are manufacturing lot specific. An Assay Value Insert (AVI) is provided with each box of calibrators. Target values are as follows: Reagent 1: 15%, Reagent 2: 31%, Reagent 3: 61%, Reagent 4: 101%. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 4
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...