STRATUS (R) CS STAT FLUOROMETRIC ANALYZER, STRATUS (R) CS CKMB TEST PAK
K984067 · Dade Behring, Inc. · JHX · Jan 13, 1999 · Clinical Chemistry
Device Facts
Record ID
K984067
Device Name
STRATUS (R) CS STAT FLUOROMETRIC ANALYZER, STRATUS (R) CS CKMB TEST PAK
Applicant
Dade Behring, Inc.
Product Code
JHX · Clinical Chemistry
Decision Date
Jan 13, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1215
Device Class
Class 2
Indications for Use
The Stratus® CS STAT Fluorometric Analyzer is a microprocessor-controlled instrument that measures analytes in body fluid for in vitro diagnostic use. The CKMB TestPak used on the Stratus® CS STAT Fluorometric Analyzer is an in vitro diagnostic product for the measurement of the MB isoenzyme of creatine kinase in heparinized plasma. Measurements of CKMB can be used as an aid in the diagnosing of acute myocardial infarction.
Device Story
Benchtop analyzer processes whole blood samples in collection tubes using specific TestPaks; performs automated pipetting, incubation, and fluorescence measurement. Uses Radial Partition Immunoassay (RPIA) technology: sample reacts with immobilized anti-CKMB antibody on glass fiber paper; enzyme-labeled antibody conjugate added to form sandwich; unbound label eluted via substrate wash. Microprocessor monitors reaction rate via front surface fluorescence; prints quantitative results. Used in clinical settings; operated by laboratory personnel. Provides rapid CKMB quantification to assist physicians in diagnosing acute myocardial infarction and muscle diseases.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing and QC data demonstrating that the modified QC frequency does not impact device performance.
Technological Characteristics
Benchtop fluorometric analyzer; microprocessor-controlled. TestPak utilizes plastic cartridge with glass fiber paper. Assay principle: two-site sandwich Radial Partition Immunoassay (RPIA). Detection: front surface fluorescence optical system. Connectivity: standalone with integrated printer. Software: modified to support updated QC frequency intervals.
Indications for Use
Indicated for patients requiring measurement of CKMB isoenzyme in heparinized plasma to aid in the diagnosis of acute myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
Regulatory Classification
Identification
A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
K981099 — STRATUS CS STAT FLUOROMETRIC ANALYZER AND STRATUS CS CKMB TESTPAK · Dade Behring, Inc. · May 26, 1998
K120326 — AQT90 FLEX CKMB TEST KIT, AQT90 FLEX LQC MULTI-CHECK, LEVELS 1-3, AQT90 FLEX CKMB CAL CARTRIDGE · Radiometer Medical Aps · Nov 21, 2012
K051650 — STRATUS CS ACUTE CARE MB ISOENZYME OF CREATINE KINASE (CKMB) TEST PAK; STRATUS CS ACUTE CARE CKMB TESTPAK. · Dade Behring, Inc. · Sep 8, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
DADE BEHRING INC. P.O. Box 6101 Newark, DE 19714
## Summary of Safety and Effectiveness Information
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| Submitter's Name: | Rebecca S. Ayash<br>Dade Behring Inc.<br>Building 500, Mailbox 514<br>P.O. Box 6101<br>Newark, DE 19714-6101 |
|----------------------|--------------------------------------------------------------------------------------------------------------|
| Date of Preparation: | 11/13/98 |
| Device Name: | Stratus® CS STAT Fluorometric Analyzer<br>Stratus® CS CKMB TestPak |
| Classification Name: | Fluorometer, for Clinical Use<br>Fluorometric Method, CPK or Isoenzymes |
| Predicate Device: | Stratus® CS STAT Fluorometric Analyzer (K981099)<br>Stratus® CS CKMB TestPak (K981099) |
Device Description: The Stratus® CS STAT Fluorometric Analyzer is a benchtop analyzer capable of processing up to four in vitro diagnostic tests per sample. An operator of the analyzer introduces a specimen collection tube filled with whole blood into the analyzer, along with the appropriate TestPaks for processing. The analyzer transfers and spins an aliquot of the sample, delivers the spun plasma and other self-contained reagents to the reaction area of the TestPak. reads the reaction rate via front surface fluorescence and prints out quantitative test results.
The CKMB TestPak consists of a plastic cartridge with five wells and a small square of glass fiber paper embedded in it. The method utilizes a two-site sandwich assay based upon solid phase Radial Partition Immunoassay (RPIA) technology. In this procedure, dendrimer linked monoclonal antibody is added to the center portion of a square piece of glass fiber paper in the CKMB TestPak. Sample is then added onto the paper where it reacts with the immobilized anti-CKMB antibody. After a short incubation, a conjugate consisting of enzyme-labeled antibody directed against a distinct antigenic site of the B subunit on the CKMB molecule is pipetted onto the reaction zone of the paper. Dunng this second incubation period, enzyme-labeled antibody reacts with the bound CKMB, forming an antibody-antibedy-antibody sandwich. The unbound labeled antibody is later eluted from the field of view of the Stratus® CS Analyzer by applying a substrate wash solution to the center of the reaction zone of the TestPak. By including substrate for the enzyme within the wash solution of enzyme activity occurs simultaneously with the wash. The enzymatic rate of the bound fraction increases directly with the concentration of CKMB in the sample. The reaction rate is measured by an optical system that monitors the reaction rate via front surface fluorescence. All data analysis functions are performed by the microprocessor within the analyzer.
These devices were previously cleared under K981099. Subsequently, the labeling of the TestPak and the analyzer software and Operator's Guide have been modified to change the recommended frequency of quality control from daily to weekly.
{1}------------------------------------------------
Intended Use: The Stratus® CS STAT Fluorometric Analyzer is a microprocessor-controlled instrument that measures analytes in body fluid for in vitro diagnostic use.
The CKMB TestPak used on the Stratus® CS STAT Fluorometric Analyzer is an in vitro diagnostic product for the measurement of the MB isoenzyme of creatine kinase in heparinized plasma. Measurements of CKMB can be used as an aid in the diagnosing of acute myocardial infarction.
Comparison to Predicate Device: The Stratus® CS STAT Fluorometric Analyzer and Stratus®CS CKMB TestPak differ from the devices cleared under K981099 only in the allowed QC frequency in analyzer software and the QC frequency recommendation in corresponding labeling. All other features of these devices are the same.
Comments on Substantial Equivalence: Both the analyzer and the CKMB TestPak have the same intended uses as the devices cleared under K981099. The modified devices are the same other than software and labeling changes to accommodate a recommended change in QC frequency.
Conclusion: The QC data collected supports the recommended change in QC frequency and indicates the change will not adversely affect performance of these devices.
Rebecca S. Ayash
Rebecca Regulatory Affairs and Compliance Manager Date: 11/13/98
{2}------------------------------------------------
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JAN 1 3 1999
Ms. Rebecca S. Ayash Regulatory Affairs and Compliance Manager Dade Behring, Inc. Building 500, Mailbox 514 P.O. Box 6101 Newark, Delaware 19714-6101
Re: K984067
> Trade Name: Stratus® CS STAT Fluorometric Analyzer, Stratus® CS CKMB TestPak Regulatory Class: II Product Code: JHX Dated: November 13, 1998 Received: November 16, 1998
Dear Ms. Ayash:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours.
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications Statement
Device Name: Stratus® CS STAT Fluorometric Analyzer
Indications for Use: The Stratus® CS STAT Fluorometric Analyzer is intended to duplicate manual analytical procedures by performing automatically various steps, such as pipetting, incubation and measuring fluorescence.
Device Name: Stratus® CS CKMB TestPak
Indications for Use: Measurements of CKMB are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
Rebecca S. Aycock
Rebecca S. Ayast Regulatory Affairs and Compliance Manager Date: 11/13/98
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
K984067
510(k) Number
Kinn Cooper
Division Sign-Off/ Office of Device Evaluation
prescription use
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.