STRATUS CS ACUTE CARE MB ISOENZYME OF CREATINE KINASE (CKMB) TEST PAK; STRATUS CS ACUTE CARE CKMB TESTPAK.
Applicant
Dade Behring, Inc.
Product Code
JHX · Clinical Chemistry
Decision Date
Sep 8, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1215
Device Class
Class 2
Indications for Use
The Stratus® CS Acute Care™ CKMB method is an in vitro diagnostic test for the measurement of the MB isoenzyme of creatine kinase (ATP:Creatine N-Phosphostransferase, E.C. No. 2.7.3.2) in heparinized plasma. Measurements of CKMB are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy. This method is for use by trained health care professionals in the clinical laboratory and point of care (POC) settings. The Stratus® CS Acute Care™ CKMB Calibrator (CKMB CalPak) is an in vitro diagnostic product intended to be used for calibration of the Stratus® CS Acute Care™ CKMB method. The Stratus® CS Acute Care™ CKMB Dilution Pak (CKMB DilPak) is an in vitro diagnostic product intended to be used in conjunction with the Acute Care™ CKMB TestPak for the measurement of samples with elevated levels of CKMB. The Stratus® CS Acute Care™ Troponin I method (cTnI) is an in vitro diagnostic assay for the measurement of cardiac troponin I in heparinized plasma. Cardiac troponin I measurements can be used as an aid in the diagnosis of acute myocardial infarction (AMI). Cardiac troponin I can also be used as an aid in the risk stratification of patients with acute coronary syndromes (ACS) with respect to their relative risk of mortality. This method is for use by trained health care professionals in the clinical laboratory and point of care (POC) settings. The Stratus® CS Acute Care™ Troponin I Calibrator (cTnI CalPak) is an in vitro diagnostic product intended to be used for calibration of the Stratus® CS Acute Care™ Troponin I method. The Stratus® CS Acute Care™ Troponin I Dilution Pak (cTnI DilPak) is an in vitro diagnostic product intended to be used in conjunction with the Acute Care™ cTnI TestPak for the measurement of samples with elevated levels of cardiac troponin I. The Stratus® CS Acute Care™ Myoglobin method (MYO) is an in vitro diagnostic assay for the measurement of myoglobin in heparinized plasma. Measurements of myoglobin aid in the rapid diagnosis of renal or heart disease, e.g. myocardial infarction. This method is for use by trained health care professionals in the clinical laboratory and point of care (POC) settings. The Stratus® CS Acute Care™ Myoglobin Calibrator (MYO CalPak) is an in vitro diagnostic product intended to be used for calibration of the Stratus® CS Acute Care™ myoglobin method. The Stratus® CS Acute Care™ Myoglobin Dilution Pak (MYO DilPak) is an in vitro diagnostic product intended to be used in conjunction with the Acute Care™ MYO TestPak for the measurement of samples with elevated Myoglobin levels.
Device Story
The Stratus® CS Acute Care™ system performs in vitro diagnostic immunoassays for CKMB, cTnI, and Myoglobin. The device uses solid-phase Radial Partition Immunoassay (RPIA) technology. Input samples (heparinized plasma) are applied to glass fiber paper containing immobilized antibodies. After incubation, enzyme-labeled monoclonal antibodies are added to form an antibody-antigen-labeled antibody sandwich. A substrate wash solution elutes unbound labeled antibody and initiates enzyme activity. An optical system monitors the reaction rate via front surface fluorescence. A microprocessor performs all data analysis. The device is intended for use by trained healthcare professionals in clinical laboratories and point-of-care (POC) settings. Results assist clinicians in diagnosing myocardial infarction, muscular dystrophy, and risk-stratifying acute coronary syndrome patients. The system includes specific TestPaks, CalPaks (calibrators), and DilPaks (diluents) for each analyte.
Clinical Evidence
Bench testing only. Method comparison and precision analysis were performed at three locations (clinical laboratory, Emergency Department, and Cardiac Care Unit) across two sites. Data demonstrated that precision and accuracy generated by non-laboratory personnel are comparable to those generated by laboratory personnel, supporting the expansion of the intended use to point-of-care settings.
Technological Characteristics
Two-site sandwich Radial Partition Immunoassay (RPIA). Reagents: alkaline phosphatase-conjugated Fab fragments, dendrimer-linked monoclonal antibodies, 4-methylumbelliferyl phosphate substrate. Form factor: TestPak cartridges for use on Stratus® CS STAT Fluorometric Analyzer. Energy: front-surface fluorescence optical detection. Connectivity: microprocessor-controlled analyzer. Sterilization: N/A (reagents).
Indications for Use
Indicated for trained healthcare professionals in clinical laboratory and POC settings to measure CKMB (myocardial infarction, muscular dystrophy), cardiac troponin I (AMI diagnosis, ACS risk stratification), and myoglobin (renal or heart disease, myocardial infarction) in heparinized plasma.
Regulatory Classification
Identification
A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
Predicate Devices
Stratus® CS CKMB TestPak and DilPak (k984067, k981099)
Stratus® CS MYO TestPak and DilPak (k981102, k984056)
Stratus® CS MYO CalPak (k981101)
Submission Summary (Full Text)
{1}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k051650
B. Purpose for Submission:
Modification to Indications for Use and Intended Use of Dade Behring Stratus® CS CKMB, troponin I and myoglobin assays to add: This method is for use by trained health care professionals in the clinical laboratory and point of care (POC) settings. CKMB and myoglobin assays renamed to Acute Care™ CKMB and Acute Care™ MYO.
C. Measurand:
MB isozyme of Creatine Kinase (CKMB)
Troponin I
Myoglobin
D. Type of Test:
Two-site sandwich immunoassay
E. Applicant:
Dade Behring Inc.
F. Proprietary and Established Names:
Stratus® CS Acute Care™ MB isoenzyme of creatine kinase (CKMB) TestPak
Stratus® CS Acute Care™ CKMB CalPak
Stratus® CS Acute Care™ CKMB DilPak
Stratus® CS Acute Care™ Troponin I (cTnI) TestPak
Stratus® CS Acute Care™ cTnI CalPak
Stratus® CS Acute Care™ cTnI DilPak
{2}
Stratus® CS Acute Care™ Myoglobin (MYO) TestPak
Stratus® CS Acute Care™ MYO CalPak
Stratus® CS Acute Care™ MYO DilPak
## G. Regulatory Information:
1. Regulation section:
- 21 CFR 862.1215, Creatine phosphokinase/creatine kinase or isoenzymes test system
- 21 CFR 866.5680, Myoglobin immunological test system
- 21 CFR 862.1150, Calibrator
2. Classification:
- Class II
3. Product code:
- JHX, Fluorometric method, CPK or isoenzymes
- MMI, Immunoassay Method, Troponin Subunit
- DDR, Myoglobin, antigen, antiserum, control
- JIT, Calibrator, Secondary
4. Panel:
- Chemistry (75)
- Immunology (82)
## H. Intended Use:
1. Intended use(s):
- See Indications for Use
2. Indication(s) for use:
- CKMB
{3}
The Stratus® CS Acute Care™ CKMB method is an in vitro diagnostic test for the measurement of the MB isoenzyme of creatine kinase (ATP:Creatine N-Phosphostransferase, E.C. No. 2.7.3.2) in heparinized plasma. Measurements of CKMB are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy. This method is for use by trained health care professionals in the clinical laboratory and point of care (POC) settings.
The Stratus® CS Acute Care™ CKMB Calibrator (CKMB CalPak) is an in vitro diagnostic product intended to be used for calibration of the Stratus® CS Acute Care™ CKMB method.
The Stratus® CS Acute Care™ CKMB Dilution Pak (CKMB DilPak) is an in vitro diagnostic product intended to be used in conjunction with the Acute Care™ CKMB TestPak for the measurement of samples with elevated levels of CKMB.
## Troponin I
The Stratus® CS Acute Care™ Troponin I method (cTnI) is an in vitro diagnostic assay for the measurement of cardiac troponin I in heparinized plasma. Cardiac troponin I measurements can be used as an aid in the diagnosis of acute myocardial infarction (AMI). Cardiac troponin I can also be used as an aid in the risk stratification of patients with acute coronary syndromes (ACS) with respect to their relative risk of mortality. This method is for use by trained health care professionals in the clinical laboratory and point of care (POC) settings.
The Stratus® CS Acute Care™ Troponin I Calibrator (cTnI CalPak) is an in vitro diagnostic product intended to be used for calibration of the Stratus® CS Acute Care™ Troponin I method.
The Stratus® CS Acute Care™ Troponin I Dilution Pak (cTnI DilPak) is an in vitro diagnostic product intended to be used in conjunction with the Acute Care™ cTnI TestPak for the measurement of samples with elevated levels of cardiac troponin I.
## Myoglobin
The Stratus® CS Acute Care™ Myoglobin method (MYO) is an in vitro diagnostic assay for the measurement of myoglobin in heparinized plasma. Measurements of myoglobin aid in the rapid diagnosis of renal or heart disease, e.g. myocardial infarction. This method is for use by trained health care professionals in the clinical laboratory and point of care (POC) settings.
The Stratus® CS Acute Care™ Myoglobin Calibrator (MYO CalPak) is an in vitro diagnostic product intended to be used for calibration of the Stratus® CS Acute Care™ myoglobin method.
{4}
The Stratus® CS Acute Care™ Myoglobin Dilution Pak (MYO DilPak) is an in vitro diagnostic product intended to be used in conjunction with the Acute Care™ MYO TestPak for the measurement of samples with elevated Myoglobin levels.
3. Special conditions for use statement(s):
Professional use only
4. Special instrument requirements:
Stratus® CS STAT Fluorometric Analyzer
I. Device Description:
Stratus® CS Acute Care™ MB isoenzyme of creatine kinase (CKMB) TestPak kit contains 100 TestPaks composed of the following reagents: Alkaline phosphatase (calf intestine enzyme) conjugated to anti-CKBB Fab (mouse monoclonal antibody), ACES buffer, sodium azide < 0.1 %, dendrimer linked CKMB antibody (mouse IgG monoclonal), 4-methylumbelliferyl phosphate, diethanolamine buffer.
Stratus® CS Acute Care™ CKMB CalPak is a refrigerated liquid product containing human heart CKMB in a buffered bovine protein matrix with stabilizers and preservative. The kit consists of five CalPaks at a single calibrator level. Each CalPak contains calibrator reagent in three wells.
Stratus® CS Acute Care™ CKMB DilPak is a refrigerated liquid product containing buffered human protein matrix with stabilizers and 0.2 % sodium azide. The kit consists of 5 DilPaks with diluent in one well.
Stratus® CS Acute Care™ Troponin I (cTnI) TestPak kit contains 100 TestPaks composed of the following reagents: Alkaline phosphatase conjugated (calf intestine enzyme) conjugated to anti-cardiac troponin I Fab (mouse monoclonal antibody), buffer, sodium azide < 0.1 %, dendrimer linked cardiac troponin I antibody (mouse IgG monoclonal), 4-methylumbelliferyl phosphate, diethanolamine buffer.
Stratus® CS Acute Care™ cTnI CalPak is a frozen liquid product containing native human troponin complex in a human serum base with < 0.1 % sodium azide. The kit consists of five CalPaks at a single calibrator level. Each CalPak contains calibrator reagent in three wells.
Stratus® CS Acute Care™ cTnI DilPak is a refrigerated liquid product containing phosphate buffer with preservatives including < 0.1 % sodium azide. The kit consists of 5 DilPaks with diluent in one well.
Stratus® CS Acute Care™ Myoglobin (MYO) TestPak kit contains 100 TestPaks composed of the following reagents: Alkaline phosphatase (calf intestine enzyme)
{5}
conjugated to anti-MYO Fab (mouse monoclonal antibody), ACES buffer, sodium azide < 0.1 %, dendrimer linked (mouse IgG monoclonal) MYO antibody, 4-methylumbelliferyl phosphate, diethanolamine buffer.
Stratus® CS Acute Care™ MYO CalPak is a refrigerated liquid product containing human heart myoglobin in a bovine albumin matrix with stabilizers and < 0.1 % sodium azide. The kit consists of five CalPaks at a single calibrator level. Each CalPak contains calibrator reagent in three wells.
Stratus® CS Acute Care™ MYO DilPak is a refrigerated liquid product containing buffered bovine protein matrix with stabilizers and 0.2 % sodium azide. The kit consists of 5 DilPaks with diluent in one well.
Human source material was tested and found to be negative/non-reactive for antibodies to HIV-1/2, HBsAg, and HCV.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Stratus® CS CKMB TestPak and DilPak
Stratus® CS CKMB CalPak
Stratus® CS cTnI Acute Care™ TestPak and DilPak
Stratus® CS cTnI CalPak
Stratus® CS MYO TestPak and DilPak
Stratus® CS MYO CalPak
2. Predicate 510(k) number(s):
Stratus® CS CKMB TestPak and DilPak k984067
k981099
Stratus® CS CKMB CalPak k981097
Stratus® CS cTnI Acute Care™ TestPak and DilPak k033487
k984093
k981098
{6}
6
Stratus® CS cTnI CalPak
k012233
k983722
k012233
Stratus® CS MYO TestPak and DilPak
k981102
k984056
Stratus® CS MYO CalPak
k981101
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Test principle for CKMB, cTnI and MYO assays | Same | Same |
| Performance characteristics for CKMB, cTnI and MYO assays | Same | Same |
| Composition and traceability of CKMB, cTnI and MYO calibrators | Same | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| CKMB, cTnI and MYO calibrators | Addition of Stratus® CS Acute Care to name | Not applicable |
| CKMB and MYO assays. | Addition of Stratus® CS Acute Care to name (Acute Care was added to cTnI assay name in k033487) | Not applicable |
| Indications for Use for CKMB, cTnI and MYO assays | Addition of: This method is for use by trained health care professionals in the clinical laboratory and point of care (POC) settings. | Not applicable |
| Reproducibility | Addition of reproducibility data from ED, CCU and Central Laboratory to labeling | Not applicable |
{7}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Correlation | Addition of POC data subset regression statistics to labeling | Not applicable |
## K. Standard/Guidance Document Referenced (if applicable):
CLSI EP5-T2
## L. Test Principle:
### CKMB
The Stratus® CS Acute Care™ CKMB method is a two-site sandwich assay based upon solid phase Radial Partition Immunoassay (RPIA) technology. Dendrimer linked monoclonal CKMB antibody is added to the center portion of a square piece of glass fiber paper in the CKMB TestPak. Sample is then added onto the paper where it reacts with the immobilized anti-CKMB antibody. After a short incubation, a conjugate consisting of enzyme labeled monoclonal antibody directed against a distinct antigenic site on the B subunit of the CKMB molecule is pipetted onto the reaction zone of the paper. During the second incubation period, enzyme-labeled antibody reacts with the bound CKMB, forming an antibody-antigen-labeled antibody sandwich. The unbound labeled antibody is later eluted from the field of view of the Stratus® CS analyzer by applying a substrate wash solution to the center of the reaction zone. By including substrate for the enzyme within the wash solution, initiation of enzyme activity occurs simultaneously with the wash. The enzymatic rate of the bound fraction increases directly with the concentration of CKMB in the sample. The reaction rate is measured by an optical system that monitors the reaction rate via front surface fluorescence. All data analysis functions are performed by the microprocessor within the analyzer.
### Troponin I (cTnI)
The Stratus® CS Acute Care™ Troponin I method is a two-site sandwich assay based upon solid phase Radial Partition Immunoassay (RPIA) technology. Dendrimer linked monoclonal antibody is added to the center portion of a square piece of glass fiber paper in the cTnI TestPak. This antibody recognizes a distinct antigenic site on the cardiac troponin I molecule. Sample is then added onto the paper where it reacts with the immobilized antibody. After a short incubation, a conjugate consisting of enzyme labeled monoclonal antibody directed against a second distinct antigenic site on the cardiac troponin I molecule is pipetted onto the reaction zone of the paper. During the second incubation period, enzyme-labeled antibody reacts with the bound cardiac troponin I, forming an antibody-antigen-labeled antibody sandwich. The unbound labeled antibody is later eluted from the field of view of the Stratus® CS analyzer by applying a substrate wash solution to the center of the reaction zone. By
{8}
including substrate for the enzyme within the wash solution, initiation of enzyme activity occurs simultaneously with the wash. The enzymatic rate of the bound fraction increases directly with the concentration of cardiac troponin I in the sample. The reaction rate is measured by an optical system that monitors the reaction rate via front surface fluorescence. All data analysis functions are performed by the microprocessor within the analyzer.
## Myoglobin (MYO)
The Stratus® CS Acute Care™ MYO method is a two-site sandwich assay based upon solid phase Radial Partition Immunoassay (RPIA) technology. Dendrimer linked monoclonal antibody is added to the center portion of a square piece of glass fiber paper in the MYO TestPak. This antibody recognizes a distinct antigenic site on the myoglobin molecule. Sample is then added onto the paper where it reacts with the immobilized anti-myoglobin antibody. After a short incubation, a conjugate consisting of enzyme labeled monoclonal antibody directed against a second distinct antigenic site on the myoglobin molecule is pipetted onto the reaction zone of the paper. During the second incubation period, enzyme-labeled antibody reacts with the bound myoglobin, forming an antibody-antigen-labeled antibody sandwich. The unbound labeled antibody is later eluted from the field of view of the Stratus® CS analyzer by applying a substrate wash solution to the center of the reaction zone. By including substrate for the enzyme within the wash solution, initiation of enzyme activity occurs simultaneously with the wash. The enzymatic rate of the bound fraction increases directly with the concentration of myoglobin in the sample. The reaction rate is measured by an optical system that monitors the reaction rate via front surface fluorescence. All data analysis functions are performed by the microprocessor within the analyzer.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
a. Precision/Reproducibility:
Previously established for predicate devices
b. Linearity/assay reportable range:
Previously established for predicate devices
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Previously established for predicate calibrators. The calibrators (Stratus® CS Acute Care™ CKMB CalPak, Stratus® CS Acute Care™ cTnI CalPak, and Stratus® CS Acute Care™ MYO CalPak) are identical to the predicate calibrators. Labeling changes reflect the name change in addition to minor
{9}
format changes.
d. Detection limit:
Previously established for predicate devices
e. Analytical specificity:
Previously established for predicate devices
f. Assay cut-off:
Previously established for predicate devices
2. Comparison studies:
a. Method comparison with predicate device:
Previously established for predicate devices
b. Matrix comparison:
Previously established for predicate devices
3. Clinical studies:
a. Clinical Sensitivity:
Previously established for predicate devices
b. Clinical specificity:
Previously established for predicate devices
c. Other clinical supportive data (when a. and b. are not applicable):
**POC Studies**
Method comparison and precision analyses were performed at two external evaluation sites: York Hospital in York, Pennsylvania and Western Pennsylvania Hospital in Pittsburgh, Pennsylvania. At each of the two sites, three Stratus® CS analyzers were installed, one in each of these three different locations: clinical laboratory (LAB), Emergency Department (ED) and Cardiac Care (CCU). The data supports use of these products by trained health professionals in the clinical laboratory and point of care (POC) settings. The following table represents the comparison between the POC sites vs. the
9
{10}
LAB site.
| | n | Slope | Intercept | r | Range of samples |
| --- | --- | --- | --- | --- | --- |
| CKMB | 147 | 0.95 | 0.56 | 0.990 | 0.3–123.9 ng/mL |
| cTnI | 149 | 0.99 | -0.04 | 0.990 | 0.0-43.3 ng/mL |
| MYO | 146 | 1.01 | 4.0 | 0.987 | 15-829 ng/mL |
The following table represents the $\%$ CV for total imprecision for the POC sites vs. the LAB site for CKMB. Specimens at each level were analyzed in quadruplicate for 5 runs.
| | LAB % CV | CCU | ED |
| --- | --- | --- | --- |
| Control Level 1 | 3.4 | 6.0 | 3.5 |
| Control Level 2 | 4.1 | 4.8 | 3.5 |
The following table represents the $\%$ CV for total imprecision for the POC sites vs. the LAB site for cTnI. Specimens at each level were analyzed in quadruplicate for 5 runs.
| | LAB % CV | CCU | ED |
| --- | --- | --- | --- |
| Control Level 1 | 4.1 | 4.5 | 3.6 |
| Control Level 2 | 3.7 | 4.2 | 3.4 |
The following table represents the $\%$ CV for total imprecision for the POC sites vs. the LAB site for MYO. Specimens at each level were analyzed in quadruplicate for 5 runs.
| | LAB % CV | CCU | ED |
| --- | --- | --- | --- |
| Control Level 1 | 4.2 | 6.9 | 5.6 |
| Control Level 2 | 4.6 | 5.2 | 4.8 |
# 4. Clinical cut-off:
Previously established for predicate devices
# 5. Expected values/Reference range:
Previously established for predicate devices
# N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
# O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.