EML105 ELECTROLYTE METABOLITE ANALYZER
Device Facts
| Record ID | K962334 |
|---|---|
| Device Name | EML105 ELECTROLYTE METABOLITE ANALYZER |
| Applicant | Radiometer America, Inc. |
| Product Code | JGS · Clinical Chemistry |
| Decision Date | Oct 24, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1665 |
| Device Class | Class 2 |
Indications for Use
EML105 is a stand alone, electrolyte metabolite analyzer that measures Na⁺, K⁺, Ca²⁺, Cl⁻, glucose and lactate on whole blood and urine. It can be interfaced with an ABL analyzer 500/510/520 to become an ABL System 605/615/625 for the measurement of pH/Blood gases, electrolytes, metabolites and co-oximetry.
Device Story
Automated electrolyte metabolite analyzer; measures Na⁺, K⁺, Ca²⁺, Cl⁻, glucose, lactate in whole blood/urine; analysis time ~54 seconds. Standalone unit (40x35x40 cm, 22 kg) or integrated with ABL 500/510/520 analyzers to form ABL System 605/615/625. Used in clinical settings; operated by laboratory personnel. Provides quantitative analyte concentrations to assist clinicians in metabolic/electrolyte status assessment.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Automated electrolyte/metabolite analyzer; ion-specific electrode technology; dimensions 40cm x 35cm x 40cm; weight 22kg; standalone or networked via interface with ABL 500/510/520 series.
Indications for Use
Indicated for measurement of Na⁺, K⁺, Ca²⁺, Cl⁻, glucose, and lactate in whole blood and urine samples.
Regulatory Classification
Identification
A sodium test system is a device intended to measure sodium in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.
Predicate Devices
- EML105 (K952621)
- YSI 2300
Reference Devices
- ABL analyzer 500/510/520
Related Devices
- K241037 — ABL90 FLEX PLUS System · Radiometer Medicals Aps · Jan 14, 2025
- K974818 — ABL555 BLOOD GAS, ELECTROLYTE & METABOLITE SYSTEM · Radiometer America, Inc. · Jan 16, 1998
- K973367 — ABL555 BLOOD GAS, ELECTROLYTE & METABOLITE SYSTEM · Radiometer America, Inc. · Oct 6, 1997
- K980130 — ABL700 SERIES BLOOD GAS , CO-OXIMETRY, ELECTROLYTE & METABOLITE SYSTEM · Radiometer America, Inc. · Feb 9, 1998