K230275 · Abbott Point of Care, Inc. · JGS · Oct 27, 2023 · Clinical Chemistry
Device Facts
Record ID
K230275
Device Name
i-STAT CG8+ cartridge with the i-STAT 1 System
Applicant
Abbott Point of Care, Inc.
Product Code
JGS · Clinical Chemistry
Decision Date
Oct 27, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1665
Device Class
Class 2
Indications for Use
The i-STAT CG8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of sodium and potassium in arterial or venous whole blood in point of care or clinical laboratory settings. The i-STAT CG8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of sodium in capillary whole blood in point of care or clinical laboratory settings. Sodium measurements are used for monitoring electrolyte imbalances. Potassium measurements are used in the diagnosis and monitoring of diseases and clinical conditions that manifest high and low potassium levels.
Device Story
The i-STAT CG8+ cartridge is a single-use, self-contained disposable unit used with the i-STAT 1 Analyzer. It processes 95 μL of whole blood (arterial, venous, or capillary) to quantify sodium and potassium levels. The cartridge contains ion-selective electrode (ISE) sensors, a fluid pouch with calibrator, and internal fluid channels. Operation involves applying a blood sample to the entry well; the analyzer performs all fluid movement and potentiometric measurements. The device calculates analyte concentration using the Nernst equation. Results are displayed on the i-STAT 1 Analyzer for clinicians to monitor electrolyte imbalances or diagnose related clinical conditions. It is intended for point-of-care or clinical laboratory use by healthcare professionals.
Clinical Evidence
Bench testing only. Precision/reproducibility evaluated per CLSI EP05-A3 (N=80-97 per level). Linearity assessed per CLSI EP06 (R² > 0.999). Interference testing performed per CLSI EP07. Method comparison study (N=185 venous, 157 arterial, 209 capillary) against predicate and epoc systems showed high agreement (Slope 1.00, R=0.89-1.00). No clinical studies required.
Technological Characteristics
Single-use disposable cartridge; ion-selective electrode (ISE) potentiometry; NIST SRM956 traceable. Measures potential difference between ion-selective and reference electrodes. Analyte concentration derived via Nernst equation. Compatible with i-STAT 1 Analyzer. Sample volume: 95 μL. Connectivity: i-STAT 1 system.
Indications for Use
Indicated for in vitro quantification of sodium and potassium in arterial, venous, or capillary whole blood for monitoring electrolyte imbalances and diagnosing/monitoring diseases manifesting high or low potassium levels in point-of-care or clinical laboratory settings.
Regulatory Classification
Identification
A sodium test system is a device intended to measure sodium in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.
Predicate Devices
i-STAT CHEM8+ cartridge with the i-STAT 1 System (K183688)
K153357 — i-STAT Alinity System with i-STAT Sodium test · Abbott Point of Care, Inc. · Jul 8, 2016
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K230275
B Applicant
Abbott Point of Care Inc.
C Proprietary and Established Names
i-STAT CG8+ cartridge with the i-STAT 1 System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JGS | Class II | 21 CFR 862.1665 - Sodium Test System | CH - Clinical Chemistry |
| CEM | Class II | 21 CFR 862.1600 - Potassium test system | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
Modification to a previously cleared device
B Measurand:
Sodium (Na) and Potassium (K)
C Type of Test:
Quantitative, Ion Specific Electrode (Potentiometric method)
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The i-STAT CG8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of sodium and potassium in arterial or venous whole blood in point of care or clinical laboratory settings.
The i-STAT CG8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of sodium in capillary whole blood in point of care or clinical laboratory settings.
Sodium measurements are used for monitoring electrolyte imbalances.
Potassium measurements are used in the diagnosis and monitoring of diseases and clinical conditions that manifest high and low potassium levels.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
For point of care or clinical laboratory setting.
D Special Instrument Requirements:
i-STAT 1 Analyzer
IV Device/System Characteristics:
A Device Description:
The candidate device is for the sodium and potassium tests on the i-STAT CG8+ (white) cartridge with the i-STAT 1 System. Additional analytes on the i-STAT CG8+ cartridge (white) with the i-STAT 1 System were cleared in K223710 (glucose), K230285 (pH, pCO2, pO2) and K230300 (ionized calcium and hematocrit).
The i-STAT CG8+ (white) cartridge contains sensors for the measurement of sodium and potassium, a fluid pouch containing aqueous calibrator, a sample entry well and snap closure, fluid channels, waste chamber, and the necessary mechanical features for controlled fluid movement within cartridge. It is a single-use disposable unit that is self-contained. The cartridge tests a sample consisting of two or three drops of whole blood (95 μL). All the test steps and fluid movement occur within the i-STAT CG8+ cartridges.
B Principle of Operation:
Sodium and Potassium tests on the i-STAT CG8+ (white) cartridge are measured by ion-selective electrode potentiometry, measurement of potential difference between the ion selective electrode and the reference electrode. In the calculation of concentration is related to potential through the Nernst equation.
K230275 - Page 2 of 13
{2}
V Substantial Equivalence Information:
A Predicate Device Name(s):
i-STAT CHEM8+ cartridge with the i-STAT 1 System
B Predicate 510(k) Number(s):
K183688
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K230275 | K183688 |
| --- | --- | --- |
| Device Trade Name | i-STAT CG8+ cartridge with the i-STAT 1 System | i-STAT CHEM8+ cartridge with the i-STAT 1 System |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Intended for in vitro quantification of sodium and potassium.
Sodium measurements are used for monitoring electrolyte imbalances.
Potassium measurements are used in the diagnosis and monitoring of diseases and clinical conditions that manifest high and low potassium levels. | Same |
| Reportable Range | Sodium: 100 -180 mmol/L
Potassium: 2.0-9.0 mmol/L | Same |
| General Device Characteristic Differences | | |
| Sample Type | Sodium and Potassium Test: arterial and venous whole blood with and without balanced heparin or lithium heparin anticoagulant.
Sodium Test: capillary | Sodium and Potassium Test: arterial and venous whole blood with lithium heparin anticoagulant |
K230275 - Page 3 of 13
{3}
| | whole blood with balanced heparin or lithium heparin anticoagulant. | |
| --- | --- | --- |
| Analytes | Sodium, Potassium | Sodium, Potassium, Chloride, Blood Urea Nitrogen |
VI Standards/Guidance Documents Referenced:
Clinical Laboratory Standards Institute (CLSI) EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline-3rd Edition
CLSI EP06: Evaluation of Linearity of Quantitative Measurement Procedures; 2nd Edition
CLSI EP07: Interference Testing in Clinical Chemistry; 3rd ed.
CLSI EP09c: Measurement Procedure Comparison and Bias Estimation using Patient Samples, 3rd ed.
CLSI EP 17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline
CLSI EP37: Supplement Tables for Interference Testing in Clinical Chemistry; 1st ed.
CLSI EP35: Assessment of Equivalence or Suitability of Specimen Types for Medical Laboratory Measurement Procedures; 1st ed.
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
i-STAT CG8+ cartridge (white) cartridge contains multiple assays. When a cartridge generated a star-out (non-reported result) for one assay, an additional cartridge was run to replace the star-out result, which produced additional test results for other assays. Therefore, the number of replicates per sample in the following precision/reproducibility studies may vary.
Internal within-site precision (aqueous control material)
i. 20-Day precision
K230275 - Page 4 of 13
{4}
A single site precision study for the i-STAT sodium and potassium tests in the i-STAT CG8+ cartridge (white) with the i-STAT 1 system was conducted following the recommendations in CLSI EP 05-A3. Several concentration levels of commercially available calibration verification (CalVer) samples were tested using one lot each of i-STAT CG8+ cartridges and ten i-STAT 1 Analyzers. Each sample was measured in duplicates per run, with two runs per day over 20-days by a minimum of two operators resulting in a total of 80 test results per level. The results are summarized below.
| Test | Level | N | Mean (mmol/L) | Repeatability | | Between-Run | | Between-Day | | Within-Laboratory | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Na | L 1 | 80 | 99.3 | 0.17 | 0.17 | 0.09 | 0.09 | 0.05 | 0.05 | 0.19 | 0.20 |
| | L 2 | 80 | 121.4 | 0.20 | 0.16 | 0.06 | 0.05 | 0.05 | 0.04 | 0.22 | 0.18 |
| | L 3 | 80 | 134.8 | 0.22 | 0.16 | 0.06 | 0.05 | 0.06 | 0.05 | 0.23 | 0.17 |
| | L 4 | 80 | 161.3 | 0.27 | 0.16 | 0.12 | 0.07 | 0.07 | 0.04 | 0.30 | 0.19 |
| K | L 1 | 80 | 2.09 | 0.008 | 0.37 | 0.003 | 0.16 | 0.001 | 0.04 | 0.009 | 0.41 |
| | L 2 | 80 | 2.87 | 0.007 | 0.25 | 0.002 | 0.09 | 0.002 | 0.08 | 0.008 | 0.28 |
| | L 3 | 80 | 3.76 | 0.012 | 0.31 | 0.004 | 0.10 | 0.006 | 0.15 | 0.014 | 0.36 |
| | L 4 | 80 | 6.41 | 0.021 | 0.33 | 0.008 | 0.12 | 0.012 | 0.18 | 0.025 | 0.39 |
| | L 5 | 80 | 7.99 | 0.027 | 0.33 | 0.010 | 0.13 | 0.014 | 0.18 | 0.032 | 0.40 |
# ii. Between-Lot Precision
The between-lot precision study for the i-STAT sodium and potassium test on the i-STAT CG 8+ cartridge (white) with the i-STAT 1 system was conducted using several concentration levels of commercially available calibration verification (CalVer) samples with three (3) lots of i-STAT CG8 cartridge (white). Each lot was tested in five (5) replicates over five (5) non-consecutive days for each sample. The results are summarized below:
| Test | Level | N | Mean (mmol/L) | Repeatability | | Between- lot | | Between-day | | Within- Laboratory | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Na | L1 | 76 | 100.1 | 0.22 | 0.22 | 0.32 | 0.32 | 0.03 | 0.03 | 0.39 | 0.39 |
| | L2 | 75 | 122.1 | 0.26 | 0.21 | 0.31 | 0.25 | 0.09 | 0.08 | 0.41 | 0.34 |
| | L3 | 75 | 135.0 | 0.34 | 0.25 | 0.17 | 0.13 | 0.09 | 0.06 | 0.39 | 0.29 |
| | L4 | 75 | 162.0 | 0.33 | 0.21 | 0.13 | 0.08 | 0.05 | 0.03 | 0.36 | 0.22 |
| K | L1 | 75 | 2.10 | 0.016 | 0.76 | 0.013 | 0.61 | 0.002 | 0.12 | 0.021 | 0.98 |
| | L2 | 75 | 2.86 | 0.013 | 0.47 | 0.012 | 0.43 | 0.002 | 0.08 | 0.018 | 0.64 |
| | L3 | 75 | 3.74 | 0.024 | 0.65 | 0.007 | 0.20 | 0.004 | 0.11 | 0.026 | 0.69 |
| | L4 | 75 | 6.40 | 0.050 | 0.79 | 0.017 | 0.26 | 0.009 | 0.13 | 0.054 | 0.84 |
| | L5 | 75 | 7.95 | 0.044 | 0.55 | 0.043 | 0.55 | 0.001 | 0.02 | 0.062 | 0.77 |
# iii. Between-Instrument Precision
The between-instrument precision study for the i-STAT sodium and potassium tests on the i-STAT CG8+ cartridge (white) was conducted using several concentration levels of commercially available calibration verification (CalVer) samples with three (3) i-STAT 1 analyzers. Each fluid level was tested in replicates of five (5) on one lot of i-STAT CG8+
K230275 - Page 5 of 13
{5}
cartridge (white) on each of the three i-STAT analyzers over five (5) non-consecutive days. The results are shown below.
| Test | Level | N | Mean (mmol/L) | Repeatability | | Between-Instrument | | Between-day | | Within- Laboratory | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Na | L1 | 76 | 99.7 | 0.16 | 0.16 | 0.04 | 0.04 | 0.02 | 0.02 | 0.17 | 0.17 |
| | L2 | 75 | 121.9 | 0.25 | 0.20 | 0.01 | 0.01 | 0.04 | 0.03 | 0.25 | 0.20 |
| | L3 | 76 | 134.8 | 0.22 | 0.17 | 0.08 | 0.06 | 0.03 | 0.02 | 0.24 | 0.18 |
| | L4 | 75 | 161.9 | 0.26 | 0.16 | 0.07 | 0.04 | 0.09 | 0.05 | 0.28 | 0.17 |
| K | L1 | 76 | 2.08 | 0.010 | 0.49 | 0.003 | 0.15 | 0.001 | 0.04 | 0.011 | 0.52 |
| | L2 | 75 | 2.84 | 0.011 | 0.40 | 0.005 | 0.17 | 0.003 | 0.12 | 0.013 | 0.46 |
| | L3 | 76 | 3.71 | 0.018 | 0.50 | 0.005 | 0.14 | 0.001 | 0.04 | 0.019 | 0.52 |
| | L4 | 75 | 6.38 | 0.030 | 0.46 | 0.010 | 0.16 | 0.005 | 0.07 | 0.032 | 0.50 |
| | L5 | 75 | 7.95 | 0.030 | 0.38 | 0.023 | 0.29 | 0.004 | 0.04 | 0.038 | 0.47 |
# Point-of-Care Precision (aqueous control material)
Multiple-day precision testing was performed at three (3) sites using a panel of several levels of calibration verification (CalVer) samples, containing different levels of each analyte. At each site, testing was performed once (1) per day by two (2) operators for five (5) days and each operator performed the test using one (1) lot of i-STAT CG8+ cartridges (white) on three (3) analyzers. The combined precision study results from all sites are summarized below:
| Test | Level | N | Mean (mmol/L) | Repeatability | | Between-Day | | Between-Operator | | Within-Site | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Na | L 1 | 91 | 100 0 | 0.29 | 0.29 | 0.12 | 0.12 | 0.08 | 0.08 | 0.32 | 0.32 |
| | L 2 | 90 | 121.9 | 0.28 | 0.23 | 0.04 | 0.04 | 0.00 | 0.00 | 0.29 | 0.23 |
| | L 3 | 97 | 134.9 | 0.34 | 0.25 | 0.00 | 0.00 | 0.06 | 0.04 | 0.34 | 0.25 |
| | L 4 | 90 | 161.2 | 0.46 | 0.29 | 0.00 | 0.00 | 0.00 | 0.00 | 0.46 | 0.29 |
| K | L 1 | 91 | 2.10 | 0.010 | 0.50 | 0.000 | 0.01 | 0.000 | 0.00 | 0.01 | 0.05 |
| | L 2 | 90 | 2.81 | 0.026 | 0.91 | 0.000 | 0.00 | 0.012 | 0.44 | 0.028 | 1.01 |
| | L 3 | 97 | 3.70 | 0.014 | 0.38 | 0.000 | 0.00 | 0.003 | 0.09 | 0.014 | 0.39 |
| | L 4 | 90 | 6.32 | 0.044 | 0.70 | 0.000 | 0.00 | 0.005 | 0.08 | 0.044 | 0.70 |
| | L 5 | 90 | 7.89 | 0.038 | 0.48 | 0.000 | 0.00 | 0.006 | 0.07 | 0.038 | 0.48 |
K230275 - Page 6 of 13
{6}
| Test | Level | N | Mean (mmol/L) | Between-Site | | Overall | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | %CV | SD | %CV |
| Na | L 1 | 91 | 100 | 0.00 | 0.00 | 0.32 | 0.32 |
| | L 2 | 90 | 121.9 | 0.00 | 0.00 | 0.29 | 0.23 |
| | L 3 | 97 | 134.9 | 0.03 | 0.02 | 0.34 | 0.26 |
| | L 4 | 90 | 161.2 | 0.21 | 0.13 | 0.51 | 0.32 |
| K | L 1 | 91 | 2.10 | 0.000 | 0.01 | 0.01 | 0.50 |
| | L 2 | 90 | 2.81 | 0.008 | 0.28 | 0.029 | 1.05 |
| | L 3 | 97 | 3.70 | 0.000 | 0.00 | 0.014 | 0.39 |
| | L 4 | 90 | 6.32 | 0.008 | 0.12 | 0.045 | 0.71 |
| | L 5 | 90 | 7.89 | 0.011 | 0.14 | 0.040 | 0.50 |
# Point-of-Care Precision (Whole Blood)
Whole blood precision for the i-STAT Sodium test in the i-STAT CG8+ (white) cartridge on the i-STAT 1 System was evaluated using arterial, venous, and capillary whole blood specimens collected with lithium heparin. Whole blood precision for the i-STAT Potassium tests in the i-STAT CG8+ (white) cartridge on the i-STAT 1 System was evaluated using arterial and venous whole blood specimens collected with lithium heparin. The whole blood precision was assessed using the duplicate test results collected across multiple point of care sites. The results of the whole blood precision are shown below. For each sample type, samples were grouped into subintervals based on their mean values. The precision results between sites were similar. The combined results from all sites are summarized below
| Test (units) | Sample Type | Sample Range (mmol/L) | N | Mean | SD | %CV |
| --- | --- | --- | --- | --- | --- | --- |
| Na (mmol/L) | Venous Whole Blood | 100-130 | 17 | 122.6 | 0.30 | 0.24 |
| | | >130-140 | 99 | 137.5 | 0.45 | 0.33 |
| | | >140-180 | 67 | 146.2 | 0.43 | 0.30 |
| | Arterial Whole Blood | 100-130 | 2 | 128.0 | 0.00 | 0.00 |
| | | >130-140 | 89 | 137.4 | 0.42 | 0.31 |
| | | >140-180 | 62 | 142.9 | 0.37 | 0.26 |
| | Capillary whole blood | 100-130 | 3 | 120.8 | 0.41 | 0.34 |
| | | >130-140 | 56 | 138.1 | 0.61 | 0.44 |
| | | >140-180 | 95 | 142.4 | 0.62 | 0.44 |
| K (mmol/L) | Venous Whole Blood | 2.0-3.5 | 27 | 3.22 | 0.036 | 1.12 |
| | | >3.5-5.0 | 135 | 4.12 | 0.038 | 0.92 |
| | | >5.0-9.0 | 19 | 6.49 | 0.032 | 0.50 |
| | Arterial Whole Blood | 2.0-3.5 | 23 | 3.21 | 0.021 | 0.65 |
| | | >3.5-5.0 | 124 | 4.11 | 0.032 | 0.79 |
| | | >5.0-9.0 | 6 | 5.67 | 0.041 | 0.72 |
K230275 - Page 7 of 13
{7}
K230275 - Page 8 of 13
2. Linearity:
The linearity study was designed based on the CLSI EP06 2nd ed guideline. Lithium heparin venous whole blood was obtained from a healthy subject and was altered to produce a high sample pool and a low sample pool. Samples of intermediate concentrations were prepared by intermixing the high and low pools. A total of eleven (11) samples were prepared for the linearity assessment with the concentrations spanning the claimed measuring range using five CG8+ cartridge lot. Each test sample was tested in replicates of two (2) per cartridge lot for a total of ten (10) results per level. An assessment of linearity was performed using weighted least squares linear regression. At each concentration level, the deviation from linearity was within ± 0.6% for i-STAT Sodium test and within ± 3% for i-STAT Potassium test. The regression parameters for the linearity study are summarized below:
| Analyte | Range tested (mmol/L) | Claimed Measuring Range (mmol/L) | Slope | Intercept | R² |
| --- | --- | --- | --- | --- | --- |
| Na | 91.3 – 209.8 | 100 - 180 mmol/L | 1.005 | -0.525 | 0.9996 |
| K | 1.79 – 10.04 | 2.0 – 9.0 mmol/L | 1.011 | 0.002 | 0.9994 |
3. Analytical Specificity/Interference:
The analytical specificity of the sodium and potassium assays on the i-STAT CG8+ (white) cartridge using the i-STAT 1 system was established by conducting interference testing following the recommendations in CLSI EP07-Edition 3. Interference from certain exogenous and endogenous substances was assessed using lithium heparin venous whole blood samples. A substance was identified as an interferent if the difference in the mean between the control and test sample was outside of the predefined allowable error:
For Sodium: ± 4 mmol/L.
For Potassium: ± 0.5 mmol/L.
The following table lists the concentrations of each substance at which no significant interference was found:
Sodium:
| Substance | Highest concentration at which no significant interference was observed |
| --- | --- |
| Acetaminophen | 15.6 mg/dL |
| N-Acetyl-L-Cysteine | 15.0 mg/dL |
| Acetylsalicylic Acid | 3.0 mg/dL |
| Ammonium Chloride | 10.7 mg/dL |
| Ascorbic Acid | 5.25 mg/dL |
| β-Hydroxybutyric Acid | 62.46 mg/dL |
| Bilirubin | 40 mg/dL |
| Calcium Chloride | 20 mg/dL |
{8}
Potassium:
| Substance | Highest concentration at which no significant interference was observed |
| --- | --- |
| Acetaminophen | 15.6 mg/dL |
| N-Acetyl-L-Cysteine | 15.0 mg/dL |
| Ammonium Chloride | 10.7 mg/dL |
| Ascorbic Acid | 5.25 mg/dL |
| β-Hydroxybutyric Acid | 62.46 mg/dL |
| Benzalkonium Chloride | 1.13 mg/dL |
| Unconjugated Bilirubin | 40 mg/dL |
| Calcium Chloride | 20 mg/dL |
| Cholesterol | 400 mg/dL |
| Hemoglobin | 1000 mg/dL |
| Lithium Bromide | 325.69 mg/dL (37.5 mmol/L) |
| Lithium Chloride | 13.6 mg/dL |
| Lithium Lactate | 90 mg/dL |
| Lithium Salicylate | 2.86 mg/dL |
| Magnesium Chloride | 10 mg/dL |
| Intralipid 20% | 3216 mg/dL |
| Sodium Thiosulfate | 264.04 mg/dL (16.7 mmol/L) |
| Triglyceride | 1500 mg/dL |
K230275 - Page 9 of 13
{9}
For those substances that on initial screening were found to interfere, dose response testing was conducted to establish the concentration limit below which no significant interference is expected. The results are given in the table below:
| Test | Substance | Concentration | Interference |
| --- | --- | --- | --- |
| Sodium | Cholesterol | ≥ 425mg/dL | Decreased results |
| | Nithiodote (Sodium Thiosulfate) | ≥ 2.7 mmol/L | Increased results |
## Limitations included in the labeling
- Bromide at 2.5 mmol/L is the peak plasma concentration associated with halothane anesthesia, in which bromide is released. Bromide may result in an increased rate of star outs (***).
- Nithiodote (Sodium Thiosulfate) is indicated for the treatment of acute cyanide poisoning. The journal article [1] titled “Falsely increased chloride and missed anion gap elevation during treatment with sodium thiosulfate” indicated that sodium thiosulfate could be used in the treatment of calciphylaxis indicating that the highest concentration likely to be seen in plasma [is] after infusion of a 12.5 g dose of sodium thiosulfate pentahydrate. Assuming that the 12.5 g dose of sodium thiosulfate pentahydrate is distributed in a typical blood volume of 5 L with a hematocrit of 40%, the peak sodium thiosulfate plasma concentration expected is 16.7 mmol/L.
[1] Wendroth Scott M. et.al. “Falsely increased chloride and missed anion gap elevation during treatment with sodium thiosulfate.” Clinica Chimica Acta 2014; 431:77-79.
4. **Assay Reportable Range:**
See section VII. A.2 Linearity.
5. **Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):**
### Traceability
The i-STAT sodium and potassium tests on the i-STAT CG8+ (white) cartridge are traceable to the U.S. National Institute of Standards and Technology (NIST) standard reference material NIST SRM956.
### Sample Stability
A study was conducted to verify the whole blood sample stability after sample collection with and without anticoagulant for the i-STAT sodium and potassium tests using the i-STAT CG8+ (white) cartridge on the i-STAT 1 Analyzer. The results support the sponsor’s sample stability claims. The sample stability claims for sodium and potassium tests are 30 minutes for (arterial, venous) samples with anticoagulant and 3 minutes for (arterial and venous) samples without anticoagulant. The sample stability claims for sodium is 3 minutes for capillary whole blood samples with anticoagulant.
K230275 - Page 10 of 13
{10}
K230275 - Page 11 of 13
6. Detection Limit:
Linearity studies were used to support the lower end of the measuring range for i-STAT sodium and potassium tests (see section VII. A.2 above). In addition, a limit of quantitation (LoQ) study for each analyte on the i-STAT CG8+(white) cartridge with the i-STAT 1 Analyzer was conducted following the recommendations in CLSI EP17-A2.
The LoQ was evaluated using lithium heparin venous whole blood altered to create four (4) samples with low sodium or potassium concentrations. Each sample was evaluated using two i-STAT CG8+ cartridge (white) lots. The LoQ was calculated for each of the two lots. The sponsor defined LoQ as the lowest concentration that met the pre-defined total error goal listed in the table below for each analyte. The sponsor based the claim on the lot with the highest LoQ estimate. The results are summarized in the table below.
| Analyte | Reportable Range | Total Error (TE) | LoQ |
| --- | --- | --- | --- |
| Na | 100 – 180 mmol/L | ≤ 4 mmol/L | 92 mmol/L |
| K | 2 – 9.0 mmol/L | ≤ 0.5 mmol/L | 1.6 mmol/L |
7. Assay Cut-Off:
Not applicable.
B Comparison Studies:
1. Method Comparison with Predicate Device:
The accuracy of the i-STAT sodium and potassium test with the i-STAT CG8+ cartridge (white) on the i-STAT 1 analyzer was evaluated by a method comparison study for agreement with the comparative methods. The study was conducted following the recommendations in CLSI EP09c-3rd ed. guideline.
A total of 185 lithium heparin arterial whole blood specimen and 157 lithium heparin venous whole blood (5% contrived venous blood samples) specimens collected across three (3) point of care sites were evaluated for sodium and potassium using the i-STAT CG8+ cartridge (white) and i-STAT CHEM8+ cartridge (comparative method) on the i-STAT 1 analyzer. The data were analyzed by Passing-Bablok regression analysis comparing the first replicate of the i-STAT CG8+ cartridge to the mean of duplicate results from the comparative method. Accuracy results for arterial and venous whole blood specimens for each analyte are shown in the table below.
{11}
| Analyte (Units) | Matrix | N | Concentration Range by Comparator (mmol/L) | Slope | Intercept | R | Claimed Range (mmol/L) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Na (mmol/L) | Venous | 185 | 103 – 178 | 1.00 | 0.00 | 1.00 | 100 – 180 |
| | Arterial | 157 | 128 – 175 | 1.00 | 0.00 | 0.99 | |
| K (mmol/L) | Venous | 183 | 2.4 – 8.8 | 1.00 | 0.00 | 1.00 | 2 – 9 |
| | Arterial | 157 | 2.4 – 6.3 | 1.00 | 0.00 | 1.00 | |
Capillary whole blood specimen (including 22 neonate samples) collected from skin punctures with balanced heparin capillary tubes from each study subject across multiple point of care sites were tested for Na using the i-STAT CG8+ cartridge (white) on the i-STAT 1 analyzer and the epoc Blood Analysis System (comparative method). A Passing-Bablok linear regression analysis was performed using the singlicate result from the i-STAT CG8+ cartridge (white) versus the singlicate result of the comparative method. Accuracy results for the capillary whole blood specimen are shown in the table below.
| Analyte (Units) | Matrix | N | Concentration Range by Comparator (mmol/L) | Slope | Intercept | R | Claimed Range (mmol/L) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Na (mmol/L) | Capillary blood (All samples) | 209 | 101 -172 | 1.00 | 0.00 | 0.98 | 100 – 180 |
| | Capillary blood (Native samples only) | 194 | 125-149 | 1.00 | 0.00 | 0.89 | |
Additionally, the systematic bias and predicted bias at medical decision levels for each analyte per sample type were evaluated and found substantially equivalent to previously cleared sodium and potassium test systems.
## 2. Matrix Comparison:
The sponsor has provided the information to support that the sodium and potassium in the i-STAT CG8+ cartridge (white) on the i-STAT 1 System can be performed using balanced heparin, lithium heparin and non-anticoagulant venous and arterial whole blood. The sodium assay can also be used for testing heparinized capillary whole blood.
## C Clinical Studies:
### 1. Clinical Sensitivity:
Not applicable
K230275 - Page 12 of 13
{12}
2. Clinical Specificity:
Not applicable
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
Expected values for the sodium and potassium assays on the i-STAT CG8+ cartridges are cited from literature*:
| Analyte | Units | Reference Range |
| --- | --- | --- |
| Sodium | mmol/L | 138-146 |
| Potassium | mmol/L | 3.5-4.9 |
*B.E. Statland, Clinical Decision Levels for Lab Tests (Oradell, NJ: Medical Economics Books, 1987).
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K230275 - Page 13 of 13
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.