i-STAT Alinity System with i-STAT Sodium test
Device Facts
| Record ID | K153357 |
|---|---|
| Device Name | i-STAT Alinity System with i-STAT Sodium test |
| Applicant | Abbott Point of Care, Inc. |
| Product Code | JGS · Clinical Chemistry |
| Decision Date | Jul 8, 2016 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1665 |
| Device Class | Class 2 |
| Attributes | Real-World Evidence |
Real-World Evidence
| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
|---|---|---|---|---|---|
| K153357 · Jul 8, 2016 | i-STAT Alinity System with i-STAT Sodium test | Abbott Point of Care, Inc. | Left-over clinical patient samples from routine care | The sponsor conducted a method comparison study comparing the candidate device to the predicate device using 174 clinical whole blood samples, including those obtained from routine patient care, to demonstrate substantial equivalence. | Method comparison; Routine clinical care samples; Point-of-care |
Clinical Evidence
| Study Design | Population | Comparator | Key Endpoints |
|---|---|---|---|
| Method comparison study; Method comparison study using prospectively collected or left-over routine clinical care samples | 174 subjects (venous or arterial lithium heparin whole blood); Sample Size: 174; Number of Sites: 4 | i-STAT 1 Wireless Analyzer (predicate device) | Method comparison (Weighted Deming regression) |
Indications for Use
The i-STAT Alinity System with i-STAT Sodium test is intended for use in point of care or clinical laboratory settings. The i-STAT Alinity System with Sodium test is intended for the quantitative measurement of sodium in arterial and venous whole blood. Sodium measurements are used for monitoring electrolyte imbalances. For in vitro diagnostic use.
Device Story
Handheld in vitro diagnostic system for point-of-care or clinical laboratory use; operated by trained medical professionals. System components include instrument, rechargeable battery, base station, electronic simulator, control material, printer, and i-STAT test cartridges. Input: 2-3 drops of fresh arterial or venous whole blood applied to cartridge via syringe. Operation: Instrument interacts with cartridge to move fluid across biosensors; utilizes ion-selective electrode technology. Output: Quantitative sodium concentration displayed on color LCD touch screen within ~2 minutes. Data management: Wireless connectivity (802.11 a/b/g/n) and TCP/IP for record storage. Clinical utility: Enables rapid assessment of electrolyte status to guide clinical decision-making and patient management.
Clinical Evidence
Bench testing only. Precision evaluated via 20-day internal study and multi-day POC site studies using aqueous materials and whole blood (n=174). Linearity confirmed across 87-204 mmol/L range. Method comparison against predicate showed high correlation (r=0.999). Interference testing performed per CLSI EP07-A2; limitations identified for bromide and sodium thiosulfate.
Technological Characteristics
Ion-selective electrode (potentiometric) measurement. Thin-film microfabricated silicon biosensors. Lithium-ion rechargeable battery. Color LCD touchscreen. Wireless (802.11 a/b/g/n) and barcode scanner. Standalone handheld form factor. Calibration: 1-point on-board. Standards: CLSI EP05-A3, EP06-A, EP07-A2, EP09-A3, EP15-A3, EP17-A2, IEC 62366-1.
Indications for Use
Indicated for the quantitative measurement of sodium in arterial and venous whole blood for monitoring electrolyte imbalances in patients in point of care or clinical laboratory settings.
Regulatory Classification
Identification
A sodium test system is a device intended to measure sodium in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.
Predicate Devices
- i-STAT 1 Wireless Analyzer (k103195)