The YSI 2900C Biochemistry Analyzer is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of venous whole blood samples, fingerstick capillary whole blood samples, serum and plasma in the laboratory. For in vitro diagnostic use. Glucose measurements from the YSI 2900C Biochemistry Analyzer are used in the diagnosis and treatment of carbohydrate metabolism disorders, including diabetes mellitus and idiopathic hypoglycemia, and of pancreatic islet cell tumors. The YSI 2900C Biochemistry Analyzer is not intended for use in the screening or quantitative analysis on neonates. The YSI 2900C Biochemistry Analyzer is not intended for point of care (POC) use. This test is for prescription use only.
Device Story
Laboratory-based semi-automated biochemistry analyzer; measures glucose concentrations in venous/capillary whole blood, serum, and plasma. Input: patient samples via robotic sipper/injector from vials/tubes. Principle: amperometric assay using glucose oxidase membrane; glucose reacts with oxygen to produce H2O2; H2O2 oxidized at platinum anode; current proportional to glucose concentration. Operation: trained medical professionals; barcode or touchscreen sample ID; automatic self-calibration every 5 samples or 15 minutes. Output: quantitative glucose concentration displayed on GUI. Clinical utility: aids diagnosis/treatment of carbohydrate metabolism disorders. Benefits: provides accurate, automated glucose quantification in laboratory settings.
Clinical Evidence
Bench testing only. Precision study (CLSI EP05-A3) across 20 days (n=80/level) showed repeatability %CV 0.6-12.9% and within-lab %CV 1.3-16.88%. Linearity confirmed 5-900 mg/dL range. Method comparison (CLSI EP09c) against predicate using 111 whole blood, 101 plasma, 101 serum, and 111 capillary samples showed high correlation (Pearson's r > 0.99). Interference testing (CLSI EP07) identified D-xylose and hematocrit (<20% or >60%) as potential interferents.
Indicated for trained medical professionals to perform quantitative glucose testing on venous whole blood, fingerstick capillary whole blood, serum, and plasma in laboratory settings. Used for diagnosis and treatment of carbohydrate metabolism disorders (diabetes mellitus, idiopathic hypoglycemia) and pancreatic islet cell tumors. Contraindicated for neonatal screening/analysis and point-of-care use.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
{1}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
# ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K210933
B Applicant
YSI Incorporated
C Proprietary and Established Names
YSI 2900C Biochemistry Analyzer
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| CGA | Class II | 21 CFR 862.1345 - Glucose Test System | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
New Device
B Measurand:
Glucose
C Type of Test:
Quantitative amperometric assay – glucose oxidase
## III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
K210933 - Page 2 of 10
## B Indication(s) for Use:
The YSI 2900C Biochemistry Analyzer is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of venous whole blood samples, fingerstick capillary whole blood samples, serum and plasma in the laboratory.
For in vitro diagnostic use.
Glucose measurements from the YSI 2900C Biochemistry Analyzer are used in the diagnosis and treatment of carbohydrate metabolism disorders, including diabetes mellitus and idiopathic hypoglycemia, and of pancreatic islet cell tumors. The YSI 2900C Biochemistry Analyzer is not intended for use in the screening or quantitative analysis on neonates. The YSI 2900C Biochemistry Analyzer is not intended for point of care (POC) use.
This test is for prescription use only.
## C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
Not for Point-of-Care Use
## D Special Instrument Requirements:
YSI 2900C Biochemistry Analyzer
## IV Device/System Characteristics:
### A Device Description:
The YSI 2900C Biochemistry Analyzer, is a laboratory instrument for determining glucose concentrations in human venous whole blood (dipotassium (K2EDTA), tripotassium EDTA (K3EDTA), sodium fluoride/potassium oxalate, lithium heparin, and sodium heparin), venous K2EDTA plasma, fingerstick capillary K3EDTA whole blood, and serum samples. The YSI 2900C Biochemistry Analyzer is a semi-automated electronic device that incorporates fluidics for sampling, calibrating, and flushing, contains a membrane-immobilized enzyme-coupled electrochemical detection system, and has a graphical user interface for control of the instrument and access to data.
The instrument contains two stations, one to accommodate plates or racks for batch sampling and a second for manual sampling from tubes or syringes.
Assay components are provided to the customer individually and are listed below.
{3}
Materials Required:
- YSI 2357 Buffer Concentrate Kit
- YSI 2747 D-Glucose Calibrator Standard
- YSI 1531 D-Glucose Linearity Standard
- YSI 2363 Potassium Ferrocyanide
- YSI 2365 Glucose (Dextrose) Membrane Kit
- YSI 2392 NaCl Solution
B Principle of Operation:
Glucose measurement is based on the level of H₂O₂ produced during the enzymatic reaction between glucose and oxygen molecules in the presence of the glucose oxidase enzyme. At a constant potential, electroactive H₂O₂ is oxidized at the surface of the platinum anode. The current generated by the flow of electrons at the surface of the platinum electrode is proportional to the glucose concentration of the sample.
C Instrument Description Information:
1. Instrument Name:
YSI 2900C Biochemistry Analyzer
2. Specimen Identification:
Specimen may be identified by scanning a barcode or by manually entering the information via the instrument touchscreen.
3. Specimen Sampling and Handling:
A sample, contained in vials or tubes of various forms, is placed in a configured holder on the instrument. A robotic sampling sipper/injector travels to the sample and withdraws a sample from it. The glucose containing sample is transferred to and injected into a reaction chamber. Within the reaction chamber the sample is mixed to homogeneity with the co-contained system buffer.
4. Calibration:
The YSI 2900C Biochemistry Analyzer’s default setting is automatic self-calibration of the enzyme sensors every five samples or every fifteen minutes to a known concentration.
5. Quality Control:
To ensure that the YSI 2900C Biochemistry Analyzer is operating properly, two levels of quality control should be used and daily operational checks should be performed before running samples as described in the operation manual.
K210933 - Page 3 of 10
{4}
V Substantial Equivalence Information:
A Predicate Device Name(s): YSI MODEL 2300 STAT PLUS
B Predicate 510(k) Number(s): K913806
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K210933 | K913806 |
| --- | --- | --- |
| Device Trade Name | YSI 2900C Biochemistry Analyzer | YSI MODEL 2300 STAT PLUS |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Measurement of glucose in human whole blood, plasma and serum. | Same |
| Methodology | Amperometric | Same |
| General Device Characteristic Differences | | |
| Sipper Pump | High-Precision stepper motor driving precision screw piston with check valve. | Stepper motor driving precision screw piston w/ check valve. |
VI Standards/Guidance Documents Referenced:
| Standards Organization | Standards Designation Number and Date | Title |
| --- | --- | --- |
| IEC | 61010-1 Edition 3. 2017-01 | Safety requirements for electrical equipment for measurement control and laboratory use – Part 1: General Requirements |
| CLSI | CLSI AUTO 11-A2 | Information Technology Security of In Vitro Diagnostic Instruments and Software Systems |
| IEC | 60601-1-2 Edition 4.0 2014-02 | Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests |
| ISO | 14971 Third Edition 2019-12 | Medical Devices – Application of Risk Management to Medical Devices |
| CLSI | EP05-A3 Third Edition | Evaluation of Precision of Quantitative Measurement Procedure |
K210933 - Page 4 of 10
{5}
| Standards Organization | Standards Designation Number and Date | Title |
| --- | --- | --- |
| CLSI | EP06 Second Edition | Evaluation of Linearity of Quantitative Measurement Procedures |
| CLSI | EP07 Third Edition | Interference Testing in Clinical Chemistry |
| CLSI | EP09c Third Edition | Measurement Procedure Comparison and Bias Estimation Using Patient Samples |
| CLSI | EP37 First Edition | Supplemental Tables for Interference Testing in Clinical Chemistry |
| CLSI | EP17-A2 Second Edition | Evaluation of Detection Capability or Clinical Laboratory Measurement Procedures |
| CLSI | EP35 First Edition | Assessment of Equivalence or Suitability of Specimen Types for Medical Laboratory Measurement Procedures |
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
A precision study was conducted following the recommendations in the CLSI EP05- A3 guideline, third edition. Glucose levels at approximately 10, 50, 180, 400, 550, 700 and 900 mg/dL in sample pools were measured in duplicate per run with two runs per day for 20 days resulting in a total of 80 measurements per level.
Representative precision at glucose levels approximately 10-900 mg/Dl:
K210933 - Page 5 of 10
{6}
| Sample Type | Mean (mg/dL) | Repeatability | | Within-Lab | |
| --- | --- | --- | --- | --- | --- |
| | | SD | %CV | SD | %CV |
| Whole Blood | 8.81 | 1.14 | 12.9 | 1.49 | 16.88 |
| Whole Blood | 50.24 | 0.88 | 1.7 | 1.12 | 2.23 |
| Whole Blood | 184.77 | 2.04 | 1.11 | 2.85 | 1.54 |
| Whole Blood | 400.94 | 5.02 | 1.3 | 6.68 | 1.67 |
| Whole Blood | 536.65 | 5.47 | 1.0 | 8.77 | 1.63 |
| Whole Blood | 724.71 | 9.54 | 1.3 | 12.71 | 1.75 |
| Whole Blood | 851.68 | 12.62 | 1.5 | 15.80 | 1.86 |
| Plasma | 16.45 | 1.03 | 6.2 | 1.10 | 6.67 |
| Plasma | 58.70 | 0.69 | 1.2 | 1.36 | 2.32 |
| Plasma | 187.66 | 2.52 | 1.3 | 3.26 | 1.74 |
| Plasma | 404.29 | 2.52 | 0.6 | 6.59 | 1.63 |
| Plasma | 542.66 | 3.49 | 0.6 | 10.21 | 1.88 |
| Plasma | 676.03 | 5.32 | 0.8 | 10.75 | 1.59 |
| Plasma | 855.13 | 5.13 | 0.6 | 11.15 | 1.30 |
2. Linearity:
Linearity testing was performed in accordance with CLSI EP06 2nd Edition. The dilution series contained whole blood and plasma samples spanning concentrations throughout the measuring range (1.9 to 998 mg/dL). The maximum % deviation from linearity observed within the claimed measuring range was less than 10%. The results of the linearity study support the claimed measuring range of 5-900 mg/dL.
3. Analytical Specificity/Interference:
The analytical specificity of the candidate device was established following the recommendations in CLSI EP07 guideline, third edition. Interference from certain exogenous and endogenous substances was assessed using venous whole blood samples at two glucose concentrations: low (40 mg/dL) and high (200 mg/dL). Control samples without interferent were included with each target glucose level. Five replicate measurements were obtained for each glucose level and control using three analyzers and three glucose membrane lots. The following table lists the target glucose and substance concentrations at which no significant interference was found.
K210933 - Page 6 of 10
{7}
| Substance | Highest Concentration (mg/dL) that Does Not Cause Interference |
| --- | --- |
| Acetaminophen | 20 |
| Ascorbic Acid | 6 |
| Maltose | 480 |
| Xylose | 50 |
| Sorbitol | 0.09 |
| Heparin Sodium | 500 IU/dL |
| Creatinine | 15 |
| Hemoglobin | 1000 |
| Tolbutamide | 72 |
| Galactose | 60 |
| Ibuprofen | 50 |
| EDTA | 0.1 |
| Triglycerides | 1500 |
| Sodium | 1050 |
| Gentisic Acid | 1.8 |
| Salicylic Acid | 60 |
| Dopamine | 0.09 |
| Maltitol | 0.09 |
| L-Methyldopa | 2 |
| Mannitol | 1800 |
| L-Dopa | 1500 |
| Isomalt | 0.09 |
| Lactitol | 0.09 |
| Xylitol | 0.09 |
| Tolazamide | 9 |
| Uric Acid | 23.5 |
| Conjugated Bilirubin | 50 |
| Unconjugated Bilirubin | 40 |
| Glutathione, Reduced | 500 |
| Cholesterol | 500 |
D-xylose at concentrations higher than 100 mg/dL have significant interference and a warning was included in the sponsor's labeling as follows:
"Patients undergoing a D-Xylose Absorption Test and their treating clinicians should be aware that concurrent testing for D-Glucose in the YSI 2900C Biochemical Analyzer may show a falsely elevated D-Glucose reading. D-Xylose is a substrate for the glucose oxidase enzyme in the sensing membrane and causes a positive interference. For example, at a Glucose concentration of 60mg/dL and Xylose concentration of 600mg/dL will produce a +12% bias.
K210933 - Page 7 of 10
{8}
Hyperglycemia may be incorrectly indicated and treatments based on those false results may be harmful to the patient.
Hypoglycemia may be incorrectly nonindicated and treatments based on those false results may be harmful to the patient.
Any D-Glucose results from the YSI instrument should not be used until the administered D-Xylose is known to be cleared from the patient.”
Hematocrit at concentrations higher than 60% and lower than 20% may cause interference as described in the sponsor’s labeling as follows:
“Hematocrit causes some varying proportional bias in the results. Elevated bias, up to 10%, may occur at hematocrit levels at 20 or below. Depressed bias, up to 10%, may occur at hematocrit levels at 60 and higher.”
4. Assay Reportable Range:
Glucose measuring range = 5.0 – 900 mg/dL
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
The candidate device’s standards are traceable to U.S. National Institute of Standards and Technology (NIST) glucose standard reference material SRM 917c.
6. Detection Limit:
Detection capability studies of blank (LoB), limit of detection (LoD) and limit of quantitation (LoQ) for glucose on the YSI 2900C Biochemistry Analyzer were conducted following the recommendations in CLSI EP17-A2 using whole blood and plasma samples. The results from all studies are summarized in the table below.
LOB, LOD and LOQ Results
| Sample | LOB (mg/dL) | LOD (mg/dL) | LOQ (mg/dL) |
| --- | --- | --- | --- |
| Whole blood | 0.24 | 1.69 | 4.00 |
| Plasma | 0.48 | 1.91 | 3.50 |
7. Assay Cut-Off:
Not Applicable.
K210933 - Page 8 of 10
{9}
8. Accuracy (Instrument):
Refer to section B.1. Method Comparison Studies.
9. Carry-Over:
The sponsor provided adequate information to support their device is not subject to carryover.
B Comparison Studies:
1. Method Comparison with Predicate Device:
A method comparison study was conducted following CLSI EP09c 3rd Edition. Three instruments/membrane lots were used for the candidate device and three analyzers for the comparator (i.e., the predicate). Testing was conducted at one site on serum, K2EDTA venous whole blood, K2EDTA venous plasma, and K3EDTA fingerstick whole blood capillary samples. Less than 10% of samples tested were contrived.
Weighted Deming regression results are presented below:
| Sample Type | n | Range Tested (mg/dL) | Slope (95% CI) | Intercept (95%CI) | Pearson's r |
| --- | --- | --- | --- | --- | --- |
| Whole Blood samples | 111 | 8.8-896 | 0.984 (0.975–0.944) | 0.415 (-0.427 – 1.258) | 0.99962 |
| Plasma samples | 101 | 8.54-940 | 0.982 (0.973–0.992) | 0.069 (-0.887 – 1.026) | 0.99976 |
| Serum samples | 101 | 9.82-910 | 0.988 (0.985–0.992) | -0.231 (-0.531 – 0.070) | 0.99975 |
| Capillary samples | 111 | 63.7-389 | 1.004 (0.979-1.029) | -2.034 (-4.525-0.457) | 0.99466 |
2. Matrix Comparison:
The sponsor has provided information to support that the glucose evaluated on the YSI 2900C Biochemistry Analyzer can be performed using K2EDTA, K3EDTA, lithium heparin, sodium heparin, and sodium fluoride/potassium oxalate on venous whole blood. K3EDTA is the only anticoagulant that should be used for capillary fingerstick whole blood and K2EDTA is the only anticoagulant that should be used for venous plasma samples.
K210933 - Page 9 of 10
{10}
K210933 - Page 10 of 10
C Clinical Studies:
1. Clinical Sensitivity:
Not Applicable.
2. Clinical Specificity:
Not Applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not Applicable.
D Clinical Cut-Off:
Not Applicable.
E Expected Values/Reference Range:
Normal (non-diabetic) adult fasting plasma glucose: Less than 100 mg/dL (5.55 mmol/L) and less than 140 mg/dL (7.77 mmol/L) 1-2 hours after meals.
Reference: ElSayed NA, Aleppo G, Aroda VR, et al, American Diabetes Association. 2. Classification and diagnosis of diabetes: Standards of Care in Diabetes-2023. Diabetes Care 2023; 46(Suppl. 1):S19-S40
F Other Supportive Instrument Performance Characteristics Data:
Not Applicable.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.