K203549 · Nova Biomedical Corporation · CGA · Oct 18, 2022 · Clinical Chemistry
Device Facts
Record ID
K203549
Device Name
Nova Primary Glucose Analyzer System
Applicant
Nova Biomedical Corporation
Product Code
CGA · Clinical Chemistry
Decision Date
Oct 18, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Nova Primary Glucose Analyzer System is indicated for in vitro diagnostic use by healthcare professionals in clinical laboratory settings for the quantitative determination of Glucose in lithium heparinized venous whole blood and plasma. The measurement of Glucose is used in the diagnosis and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and pancreatic islet cell carcinoma.
Device Story
Nova Primary Glucose Analyzer is a portable laboratory instrument for quantitative glucose measurement in lithium heparinized venous whole blood or plasma. Device accepts 25 microliter samples via syringes, collection tubes, or cups. Principle of operation: glucose oxidase enzymatic reaction produces H2O2; platinum anode oxidizes H2O2 at 0.70V constant potential; generated current is proportional to glucose concentration. For whole blood, an integrated impedance-based conductivity detector automatically adjusts results to provide plasma-equivalent values. Operated by trained laboratory personnel in clinical settings. Output displayed on PCAP color touchscreen; results printed via onboard thermal printer. Device aids clinicians in diagnosing and managing carbohydrate metabolism disorders. Benefits include automated hematocrit correction for whole blood samples and modern user interface compared to legacy systems.
Clinical Evidence
Bench testing only. Precision evaluated with QC, venous whole blood, and plasma samples (n=20 replicates; run-to-run n=240/80). Linearity confirmed across 20–900 mg/dL range using 10 levels. Interference testing performed for 30+ substances. Method comparison against YSI 2300 (n=174 whole blood, n=170 plasma) showed high correlation (r > 0.99). No clinical studies performed.
Technological Characteristics
Glucose oxidase enzymatic sensor; platinum anode electrode; impedance-based conductivity detector for hematocrit correction. Dimensions: 19.0 x 19.0 x 6.0 inches. Power: 100-240 VAC, 47-63 Hz. Interface: PCAP color touchscreen. Connectivity: Optional wireless barcode scanner/keyboard via internal USB dongle. Traceable to isotope dilution mass spectrometry.
Indications for Use
Indicated for healthcare professionals in clinical laboratory settings for quantitative glucose determination in lithium heparinized venous whole blood and plasma; used for diagnosis and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
YSI Model 2300 Stat Glucose And L-lactate Analyzer (K891480)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K203549
B Applicant
Nova Biomedical Corporation
C Proprietary and Established Names
Nova Primary Glucose Analyzer System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| CGA | Class II | 21 CFR 862.1345 - Glucose Test System | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
New Device
B Measurand:
Glucose
C Type of Test:
Quantitative amperometric
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
Please see indication(s) for use.
B Indication(s) for Use:
The Nova Primary Glucose Analyzer System is indicated for in vitro diagnostic use by healthcare professionals in clinical laboratory settings for the quantitative determination of Glucose in lithium heparinized venous whole blood and plasma.
The measurement of Glucose is used in the diagnosis and treatment of carbohydrate metabolism disturbances including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and pancreatic islet cell carcinoma.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
Not for point-of-care use.
D Special Instrument Requirements:
Nova Primary Glucose Analyzer
IV Device/System Characteristics:
A Device Description:
The Nova Primary Glucose Analyzer System is a benchtop laboratory glucose analyzer that measures blood glucose levels in lithium heparinized venous whole blood or plasma and utilizes a glucose-oxidase-based sensor and membrane/cap assembly. The system is comprised of the following: the analyzer, a wireless keyboard, a glucose sensor, a glucose sensor membrane cap, a calibrator pack, thermal paper for an onboard printer, and an optional barcode scanner.
The Nova Primary Glucose Analyzer utilizes an integrated calibrator pack with an RFID data management tag to monitor pack expiration date and number of samples remaining. All calibrator lot information is programmed into the RFID tags.
B Principle of Operation:
Glucose measurement is based on the level of $\mathrm{H}_2\mathrm{O}_2$ produced during the enzymatic reaction between glucose and oxygen molecules in the presence of the glucose oxidase enzyme. At a constant potential of 0.70 volts, electroactive $\mathrm{H}_2\mathrm{O}_2$ is oxidized at the surface of the platinum anode. The current generated by the flow of electrons at the surface of the platinum electrode is proportional to the glucose concentration of the sample.
K203549 - Page 2 of 12
{2}
C Instrument Description Information:
1. Instrument Name:
Nova Primary Glucose Analyzer
2. Specimen Identification:
When running a sample, an operator enters a sample ID of up to 40 alpha-numeric characters using either (or a combination of) the keyboard or the optional wireless scanner, if desired. Previously analyzed sample and quality control results are retained indefinitely in the analyzer’s database and can be recalled for review, printed, and exported to a USB device as a .csv file for use in an offline spreadsheet.
3. Specimen Sampling and Handling:
The device is intended to be used with lithium heparin venous whole blood and plasma samples from syringes using a port on a venous line, blood collection tubes, or small cups. Once an analysis sequence has been initiated by the user, the device extends the sample probe to aspirate a volume of sample (minimum sample volume for analysis is 25 µL) from one of the appropriate containers listed above into the analyzer’s flow path. The sample probe then retracts into the device and the analysis starts. Once the sample measurement is complete, the device aspirates calibrator pack reagents through the sample probe to flush/wash the sample probe between sample analyses. The device then conducts a 1-point sensor calibration prior to each sample analysis to ensure no carry over between samples.
If using blood collection tubes, samples should be collected with minimal stasis, without the exercise of the arm. Samples should be well mixed before introduction into the analyzer and should be analyzed within 15 minutes of collection for glucose to minimize the clinical impact of glycolysis on the measured glucose result. Measurement delays greater than 15 minutes may impact the clinical accuracy of the glucose measurement. Plasma samples should be separated from contact with cells as soon as possible to a maximum time limit of 2 hours from the time of collection.
4. Calibration:
Automatic two-point calibration every 2 hours, automatic one-point calibration with samples.
5. Quality Control:
Two levels of quality control material can be monitored on the Nova Primary Glucose Analyzer System. Results that exceed the entered ranges are flagged for identification. Laboratories should perform the following minimum QC procedures (External Ampule QC) on each analyzer: (1) During every 24 hours of testing one normal and one abnormal level of control, (2) After performing system maintenance, follow good laboratory practice guideline for performing quality control analysis.
K203549 - Page 3 of 12
{3}
V Substantial Equivalence Information:
A Predicate Device Name(s):
YSI Model 2300 Stat Glucose And L-lactate Analyzer
B Predicate 510(k) Number(s):
K891480
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K203549 | K891480 |
| --- | --- | --- |
| Device Trade Name | Nova Primary Glucose Analyzer | YSI 2300 STAT PLUS Glucose and Lactate Analyzer |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The Nova Primary Glucose Analyzer System is indicated for the determination of glucose in venous whole blood and plasma. | Same |
| Intended Users | Trained laboratory personnel | Same |
| Enzyme | Glucose Oxidase | Same |
| Sample size requirement | 25 μL | Same |
| General Device Characteristic Differences | | |
| Measured Tests | Glucose | Glucose and Lactate |
| Acceptable Sample Types | Lithium Heparin Venous Whole Blood and Plasma | Whole Blood, Plasma, Serum |
| Physical Dimensions | 19.0 x 19.0 x 6.0 inches | 10.0 x 14.0 x 14.0 inches |
| Automatic Whole Blood Hematocrit Correction | Yes – Contains an impedance-based Conductivity Detector. Whole blood samples are automatically adjusted to provide plasma equivalent glucose values based on the conductivity of the sample. | No – Hematocrit is independently measured on the whole blood specimen then manually entered into YSI 2300 to calculate equivalent plasma glucose value. |
K203549 - Page 4 of 12
{4}
VI Standards/Guidance Documents Referenced:
- IEC 61010-1:2010+AMD1:2016 Safety Requirements For Electrical Equipment For Measurement, Control, And Laboratory Use - Part 1: General Requirements
- IEC 60601-1-2:2014 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Within-Run Precision:
Within-run precision testing was performed using two levels of quality control (QC) samples, venous whole blood (WB) samples, and plasma samples and analyzed in one run, with 20 replicates per run on 3 different analyzers. Seven different levels of samples concentrations were used to cover the measuring range of the device. Two lots of Nova Primary Calibrator Cartridges and two lots of glucose membranes were used over the course of the study. The results of one representative analyzer are shown below:
Within-Run Precision Results using QC Samples:
| n = 20 | QC Level 1 | QC Level 2 |
| --- | --- | --- |
| Mean (mg/dL) | 73 | 181 |
| SD | 1.9 | 1.4 |
| %CV | 2.5 | 0.8 |
Within-Run Precision Results using Venous Whole Blood Samples:
| n= 20 | Mean (mg/dL) | SD | %CV |
| --- | --- | --- | --- |
| Level 1 | 34 | 1.0 | 2.9 |
| Level 2 | 78 | 2.1 | 2.7 |
| Level 3 | 178 | 2.9 | 1.6 |
| Level 4 | 308 | 3.2 | 1.0 |
| Level 5 | 512 | 7.3 | 1.4 |
| Level 6 | 675 | 12.4 | 1.8 |
| Level 7 | 762 | 14.2 | 1.9 |
K203549 - Page 5 of 12
{5}
Within-Run Precision Results using Plasma Samples:
| n = 20 | Mean (mg/dL) | SD | %CV |
| --- | --- | --- | --- |
| Level 1 | 36 | 0.7 | 1.9 |
| Level 2 | 122 | 1.5 | 1.2 |
| Level 3 | 183 | 1.2 | 0.7 |
| Level 4 | 400 | 1.8 | 0.4 |
| Level 5 | 558 | 4.5 | 0.8 |
| Level 6 | 692 | 6.4 | 0.9 |
| Level 7 | 841 | 6.6 | 0.8 |
Run-to-Run Precision:
Run-to-Run Precision Results using QC Samples:
Run-to-run precision testing was performed using QC samples tested in duplicate, two runs per day for 20 days for a total of 40 runs. The summary results for the QC samples collected on 3 difference analyzers are shown below:
| QC Sample | Pooled Mean | N | Run-to-Run SD (Sr) | Run-to-Run %CV | Total imprecision SD (St) | Total imprecision %CV |
| --- | --- | --- | --- | --- | --- | --- |
| Level 1 | 74 | 240 | 1.3 | 1.8 | 2.0 | 2.7 |
| Level 2 | 198 | 240 | 2.3 | 1.2 | 3.9 | 2.0 |
Run-to-Run Precision Results using Venous Whole Blood Samples:
Venous whole blood samples were run in triplicate in 10 runs during a single day. Plasma samples were run in duplicate, two runs per day for 20 days for a total of 40 runs. Six different levels of samples concentrations were used to cover the measuring range of the device. Two lots of Nova Primary Calibrator Cartridges and two lots of glucose membranes were used over the course of the study. The results of one representative analyzer are shown below for whole blood samples:
| n = 30 | Mean (mg/dL) | SD | %CV |
| --- | --- | --- | --- |
| Level 1 | 37 | 1.2 | 3.3 |
| Level 2 | 80 | 1.6 | 2.0 |
| Level 3 | 190 | 3.1 | 1.6 |
K203549 - Page 6 of 12
{6}
| n = 30 | Mean (mg/dL) | SD | %CV |
| --- | --- | --- | --- |
| Level 4 | 370 | 7.2 | 1.9 |
| Level 5 | 547 | 14.2 | 2.6 |
| Level 6 | 656 | 13.3 | 2.0 |
Run-to-Run Precision Results using Plasma Samples:
The results of one representative analyzer are shown below for plasma samples:
| n = 80 | Mean (mg/dL) | SD | %CV |
| --- | --- | --- | --- |
| Level 1 | 40 | 1.3 | 3.2 |
| Level 2 | 93 | 2.9 | 3.1 |
| Level 3 | 198 | 6.1 | 3.1 |
| Level 4 | 408 | 13.4 | 3.3 |
| Level 5 | 576 | 19.2 | 3.3 |
| Level 6 | 781 | 26.2 | 3.4 |
2. Linearity:
Linearity of the Nova Primary Glucose Analyzer was evaluated using three (3) Nova Primary Glucose Analyzers and two (3) YSI 2300 Stat Plus Glucose and L-Lactate Analyzers (K891480). Two (2) lots of Nova Primary Calibrator Cartridges, and two (2) lots of glucose membranes were used over the course of the study.
Venous Whole Blood
Low and high pools were prepared from the pooled whole blood with glucose concentrations outside of the analytical measurement range. Samples of 10 different concentrations spanning beyond the claimed analytical range were made from the low and high pools using serial dilutions. Each blood level was analyzed in triplicate in random order on three (3) Nova Primary test analyzers and on the YSI 2300 analyzers. The results from each test analyzer were compared to the average of the test results from the YSI 2300 analyzers.
Plasma
Previously prepared whole blood samples were spun down to plasma. Each plasma level was analyzed in triplicate in random order on three (3) Nova Primary test analyzers and on the YSI 2300 analyzers. The results from each test analyzer were compared to the average of the test results from the YSI 2300 analyzers.
The results of the least squares linear regression analyses are shown below:
K203549 - Page 7 of 12
{7}
K203549 - Page 8 of 12
| Nova Primary Linearity – Whole Blood | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Claimed Measurement Range (mg/dL) | Analyzer | Total # of Levels | Specimen Range | Slope | Intercept | r |
| 20 – 900 | 1 | 10 | 0-976 | 1.0018 | -0.4476 | 0.9982 |
| | 2 | 10 | 0-967 | 0.9955 | -0.4923 | 0.9989 |
| | 3 | 10 | 0-970 | 0.9972 | 2.7854 | 0.9984 |
| Nova Primary Linearity – Plasma | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Claimed Measurement Range (mg/dL) | Analyzer | Total # of Levels | Specimen Range | Slope | Intercept | r |
| 20 – 900 | 1 | 10 | 0-998 | 1.0188 | 2.0841 | 0.9989 |
| | 2 | 10 | 0-1097 | 1.0026 | 1.9626 | 0.9944 |
| | 3 | 10 | 0-1016 | 1.0096 | 4.3566 | 0.9976 |
3. Analytical Specificity/Interference:
An interference study was performed using venous whole blood and plasma samples. The possible interfering substances were tested at two glucose concentrations (~ 80-100 mg/dL and ~ 200 mg/dL) and compared to a control sample.
The test substance was considered an interfering substance if the difference between the mean test value and the mean control value exceeded ±10%. If the difference between the test and control means was greater than the allowed bias, a dose response was performed.
The sponsor determined the following substances did not cause interference at the concentrations listed below:
| Nova Primary Whole Blood Interference | |
| --- | --- |
| Test Substance | Highest concentration tested that did not cause significant interference |
| Acetaminophen | 20 mg/dL |
| Acetoacetate | 20 mg/dL |
| Acetylsalicylic Acid | 65 mg/dL |
| Ammonium Chloride | 0.6 mg/dL |
| Ascorbic Acid | 50 mg/dL |
| Bilirubin | 20 mg/dL |
| Benzalkonium Chloride | 10 mg/dL |
| B-hydroxybutyrate | 21 mg/dL |
| Dobutamine | 2 mg/dL |
| Dopamine Hydrochloride | 0.1 mg/dL |
| Ethanol | 400 mg/dL |
{8}
K203549 - Page 9 of 12
| Nova Primary Whole Blood Interference | |
| --- | --- |
| Test Substance | Highest concentration tested that did not cause significant interference |
| Fluoride | 0.4 mg/dL |
| D-Galactose | 18 mg/dL |
| Glucosamine | 0.5 mg/dL |
| Glycolic Acid | 7.6 mg/dL |
| Hemoglobin | 200 mg/dL |
| High Hematocrit | 68% |
| Hydroxyurea | 0.8 mg/dL |
| Ibuprofen | 50 mg/dL |
| Intralipid | 1.0 % solution |
| Lactate | 60 mg/dL |
| Low Hematocrit | 18% |
| Maltose | 445 mg/dL |
| Mannose | 18 mg/dL |
| Pyruvate | 2.7 mg/dL |
| Salicylic Acid | 60 mg/dL |
| Sodium Citrate | 310 mg/dL |
| Sodium Oxalate | 500 mg/dL |
| Thiocyanate | 97.3 mg/dL |
| Xylose | 25 mg/dL |
| N-Acetylcysteine | 167 mg/dL |
| Glutathione | 92 mg/dL |
| Creatinine | 15 mg/dL |
| Cholesterol | 500 mg/dL |
| Methyl-DOPA | 2 mg/dL |
| Nova Primary Plasma Interference | |
| --- | --- |
| Test Substance | Highest concentration tested that did not cause significant interference |
| Acetaminophen | 20 mg/dL |
| Acetoacetate | 20 mg/dL |
| Acetylsalicylic Acid | 65 mg/dL |
| Ammonium Chloride | 0.6 mg/dL |
| Ascorbic Acid | 50 mg/dL |
| Bilirubin | 20 mg/dL |
| Benzalkonium Chloride | 10 mg/dL |
| B-hydroxybutyrate | 21 mg/dL |
| Dobutamine | 2 mg/dL |
| Dopamine Hydrochloride | 0.1 mg/dL |
| Ethanol | 400 mg/dL |
| Fluoride | 0.4 mg/dL |
| D-Galactose | 18 mg/dL |
| Glucosamine | 0.5 mg/dL |
{9}
| Nova Primary Plasma Interference | |
| --- | --- |
| Test Substance | Highest concentration tested that did not cause significant interference |
| Glycolic Acid | 7.6 mg/dL |
| Hemoglobin | 200 mg/dL |
| Hydroxyurea | 0.8 mg/dL |
| Ibuprofen | 50 mg/dL |
| Intralipid | 1.0 % solution |
| Lactate | 60 mg/dL |
| Maltose | 445 mg/dL |
| Mannose | 18 mg/dL |
| Pyruvate | 2.7 mg/dL |
| Salicylic Acid | 60 mg/dL |
| Sodium Citrate | 310 mg/dL |
| Sodium Oxalate | 500 mg/dL |
| Thiocyanate | 97.3 mg/dL |
| Xylose | 25 mg/dL |
| N-Acetylcysteine | 167 mg/dL |
| Glutathione | 92 mg/dL |
| Creatinine | 15 mg/dL |
| Cholesterol | 500 mg/dL |
| Methyl-DOPA | 2 mg/dL |
To mitigate the risk of false test results due to the presence of excess anticoagulant, the sponsor demonstrated no significant interference for samples containing up to 100 IU/mL lithium heparin.
4. **Assay Reportable Range:**
Glucose measuring range = 20 – 900 mg/dL
5. **Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):**
The Glucose Calibrators, Controls, and Standards are traceable to the National Institute of Standards and Technology (NIST) Standard SRM-917.
6. **Detection Limit:**
Detection capability studies of limit of blank, limit of detection, and limit of quantitation for glucose on the Nova Primary Glucose Analyzer System were conducted following the recommendations in CLSI EP17-A2. These studies supported the claim of 20 mg/dL as the lower end of the reportable range.
7. **Assay Cut-Off:**
Not applicable.
K203549 - Page 10 of 12
{10}
8. Accuracy (Instrument):
Please see section B.1. Method Comparison with Predicate Device for a discussion of the instrument's accuracy.
9. Carry-Over:
Not applicable.
B Comparison Studies:
1. Method Comparison with Predicate Device:
Venous Whole Blood
174 discarded lithium heparinized venous whole blood specimens (15 of which were altered) from hospital patients were analyzed in singlet on three Nova Primary Glucose Analyzers and two YSI 2300 Stat Plus Glucose and L-Lactate Analyzers (comparative method – K891480). Two lots of Nova Primary Calibrator Cartridges, and two lots of glucose membranes were used over the course of the study. The singlet result from the test analyzers was compared to the average of the test results from the comparative method.
| Whole Blood Method Comparison Summary – Nova Primary vs CM | | | | | |
| --- | --- | --- | --- | --- | --- |
| Analyzer | N | Range (mg/dL) | Slope | Intercept | r |
| 1 | 174 | 34-871 | 0.9927 | 0.3878 | 0.9992 |
| 2 | 174 | 34-871 | 1.0017 | -1.5404 | 0.9990 |
| 3 | 174 | 34-871 | 1.0083 | -3.7425 | 0.9989 |
Plasma
170 discarded lithium heparinized venous plasma specimens (15 of which were altered) from hospital patients were analyzed in singlet on three Nova Primary Glucose Analyzers (test analyzers) and two YSI 2300 Stat Plus Glucose and L-Lactate Analyzers (comparative method). Two lots of Nova Primary Calibrator Cartridges, and two lots of glucose membranes were used over the course of the study. The singlet result from the test analyzers was compared to the average of the test results from the comparative method.
| Plasma Method Comparison Summary – Nova Primary vs CM | | | | | |
| --- | --- | --- | --- | --- | --- |
| Analyzer | N | Range (mg/dL) | Slope | Intercept | r |
| 1 | 170 | 30-835 | 1.0088 | 1.1364 | 0.9992 |
| 2 | 170 | 30-835 | 0.9935 | 1.0018 | 0.9991 |
| 3 | 170 | 30-835 | 1.0002 | -0.2388 | 0.991 |
K203549 - Page 11 of 12
{11}
2. **Matrix Comparison:**
Not applicable. For use with lithium heparinized venous whole blood or plasma only.
### C Clinical Studies:
1. **Clinical Sensitivity:**
Not applicable.
2. **Clinical Specificity:**
Not applicable.
3. **Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):**
Not applicable.
### D Clinical Cut-Off:
Not applicable.
### E Expected Values/Reference Range:
Glucose (Serum, Fasting, Adult): 74 – 100 mg/dL
Reference cited:
Curtis, Carl A. and Bruns, David E., ed. 2015. Tietz Fundamentals of Clinical Chemistry, Saunders St. Louis, MO.
### F Other Supportive Instrument Performance Characteristics Data:
The sponsor provided information to demonstrate they have mitigated risks associated with changes in sample volume and operating conditions.
The sponsor provided information to demonstrate they have validated use of the device up to 10,000 ft.
### VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
### IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K203549 - Page 12 of 12