NICHOLS ADVANTAGE CHEMILUMINESCENCE BIO-INTACT PARATHYROID HORMONE (1-84) IMMUNOASSAY

K013992 · Nichols Institute Diagnostics · CEW · Dec 13, 2001 · Clinical Chemistry

Device Facts

Record IDK013992
Device NameNICHOLS ADVANTAGE CHEMILUMINESCENCE BIO-INTACT PARATHYROID HORMONE (1-84) IMMUNOASSAY
ApplicantNichols Institute Diagnostics
Product CodeCEW · Clinical Chemistry
Decision DateDec 13, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1545
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The Nichols Advantage® Bio-Intact PTH (1-84) immunometric assay is intended for use with the Nichols Advantage® Specialty System to measure the levels of parathyroid hormone in serum and EDTA plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism. Assay results should be used in conjunction with other clinical data to assist the clinician in making individual patient management decisions.

Device Story

In vitro diagnostic immunoassay for quantitative measurement of intact parathyroid hormone (1-84) in patient serum or EDTA plasma. Operates on Nichols Advantage® Specialty System; utilizes two-site immuno-chemiluminometric principle. Process: patient sample added to cuvette with biotin-labeled capture antibody and acridinium ester-labeled detection antibody; streptavidin-coated magnetic particles capture sandwich complex. System performs automated wash to remove unbound components; luminometer measures chemiluminescence (Relative Light Units). Output: PTH concentration proportional to light intensity. Used in clinical laboratory settings by trained personnel. Results assist clinicians in differential diagnosis of calcium metabolism disorders (hypercalcemia/hypocalcemia) and patient management.

Clinical Evidence

Bench testing only. Precision: Within-run <4%, Total <8.5% at >5 pg/mL. Limit of Detection: ≤1.5 pg/mL. Recovery: 93-103%; Parallelism: 92-111%. No High Dose Hook Effect up to 100,000 pg/mL. Comparative study (n=305) against predicate Nichols Intact PTH Immunoassay showed strong correlation (r=0.97).

Technological Characteristics

Two-site immuno-chemiluminometric assay. Reagents: two goat polyclonal antibodies (biotin-labeled capture, acridinium ester-labeled detection), streptavidin-coated magnetic particles. Energy source: luminometer-based chemiluminescence detection. Platform: Nichols Advantage® Specialty System. Throughput: ~180 results/hour.

Indications for Use

Indicated for patients requiring measurement of parathyroid hormone (PTH) levels in serum or EDTA plasma to assist in the differential diagnosis of hypercalcemia and hypocalcemia resulting from calcium metabolism disorders.

Regulatory Classification

Identification

A parathyroid hormone test system is a device intended to measure the levels of parathyroid hormone in serum and plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510(k) Submission for # Nichols Advantage® Chemiluminescence Bio-Intact PTH (1-84) Immunoassay Nichols Institute Diagnostics, Inc. November 5, 2001 Kol3992 # SUMMARY STATEMENT OF SAFETY AND EFFECTIVENESS The sponsor, Nichols Institute Diagnostics, Inc., has developed, manufactured and tested under current Good Manufacturing Practices/Quality System Regulations (cGMP/QSR) an in vitro diagnostic (IVD) device to quantitatively assay for parathyroid hormone (PTH) in serum and EDTA plasma. The trade name is: Nichols Advantage® Chemiluminescence Bio-Intact Parathyroid Hormone (1-84) Immunoassay, having a common name of: Bio-Intact PTH (1-84) Immunoassay. This is a Class II in vitro diagnostic (IVD) medical device, as per 21 CFR 862.1545 with a Product Code of 75 CEW. This Bio-Intact PTH (1-84) Immunoassay is intended for use with the Nichols Advantage® Specialty System for measurement of parathyroid hormone levels in patient serum (preferred) or plasma, for differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism. The Nichols Advantage® Chemiluminescence Bio-Intact PTH (1-84) Immunoassay is a two-site immuno-chemiluminometric IVD device with sufficient reagents for 100 tests, that are performed entirely on the Nichols Advantage® Specialty System (K961142; cleared February 18, 1997). Two goat polyclonal antibodies directed at epitopes on intact human PTH are used. One antibody ("capture reagent") is chemically labeled with biotin, while the second antibody ("detection reagent") is chemically labeled with acridinium ester for subsequent quantitative measurements. A sample of patient serum (preferred) or plasma is added to an assay cuvette, followed by addition of the two goat anti-PTH antibodies and the streptaviding outed magnetic particles. The reaction mixture is allowed to incubate for 30 minutes at 37°C. Because of the high affinity interaction between biotin-labeled antibody and streptavidin, the sandwich complex is captured onto the streptavidin-coated magnetic particles. The captured complex bound to the magnetic particle passes to the Nichols Specialty System for a wash to remove unbound components and/or patient substances. The washed captured complex bound to the magnetic particles within the cuvette wells are analyzed Nichols Specialty System's luminometer via the automatic injection of reagents that initiate the acridinium-based chemiluminescence quantitative reaction. The light is measured by the Nichols Specially System's luminometer and expressed as Relative Light Units (RLU). The amount of PTH bound-labeled antibody is directly proportional to the concentration of intact PTH in the serum or plasma sample. In-house testing of the Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay demonstrated Within-Run precision not greater than 4% at a dose higher than 5 pg PTH/mL, and a Total Precision of not greater than 8.5% at a dose higher than 5 pg PTH/mL, with a throughput of about 180 results per hour. The Limit of Detection for this Bio-Intact (1-84) PTH Immunoassay was estimated to be at or below 1.5 po/mL. with Recovery ranging from 93% to 103%, and Parallelism at 92% to 111%. No High Dose Hook Effect was observed up to 100,000 pg PTH/mL. The PTH fragment 7-84 did not interfere in this assay at a dose up to 3000 pg/mL; PTH fragments 39-68, 53-84, 44-68, 39-84 gave minimal interference. Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay (y) was compared to predicate Nichols Intact PTH Immunoassay (x), previously FDA cleared and currently marketed for Intact PTH assay. Three hundred five (305) patient serum samples were assayed by both methods using insert directions without modifications. The range of values observed with the predicate Intact PTH Immunoassay was 5.0 to 1387 pg/mL, while the range with the Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay was 3.0 to 746 pg/mL. Passing Bablok regression analysis of these data yielded an equation of y = 0.66x - 0.6 (95% confidence intervals of the slope and intercept were 0.64 to 0.68, and -1.1 to -0.2 respectively). Pearson's correlation coefficient (r) of the paired data was 0.97 (95% confidence interval was 0.96 to 0.98), demonstrating that this Bio-Intact PTH (1-84) Immunoassay provided essentially equivalent results to an approved predicate Intact PTH (1-84) Immunoassay. Additional information on this Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay submission may be obtained by contacting Mr. Robert Schmidt, Vice President of Quality Systems, Nichols Institute Diagnostics, Inc. by telephone at 1-949-240-5417 or by telefax at 1-949-240-5271. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with three curved lines representing its wings. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Nichols Institute Diagnostics c/o Raymond Wilson, Pharm.D. California Department of Health Medical Device Safety Section Food and Drug Branch P.O. Box 942732 (MS-357) Sacramento, CA 94234 DEC 1 3 2001 Re: k013992 Trade/Device Name: Nichols Advantage® Chemiluminescence Bio-Intact PTH (1-84) Immunoassay Regulation Number: 21 CFR 862.1545 Regulation Name: Parathyroid hormone test system Regulatory Class: Class II Product Code: CEW Dated: November 29, 2001 Received: December 4, 2001 Dear Dr. Wilson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your 510(k) premarket I his letter will and w you do organ substantial equivalence of your device to a legally marketed nonification. THE I Driving of incation for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 If you desire specific active diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, (201) 594-1566. Practicely) , 1981) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small information on your respond and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsma/dsmamain.html". Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ #### INDICATIONS FOR USE STATEMENT 4.0 ### INDICATIONS FOR USE STATEMENT 510(k) Number (if known): Device Name: Nichols Advantage® Chemiluminescence Bio-Intact PTH (1-84) Immunoassay Indications For Use: The Nichols Advantage® Bio-Intact PTH (1-84) immunometric assay is intended for use with the Nichols Advantage® Specialty System to measure the levels of parathyroid hormone in serum and EDTA plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism. Assay results should be used in conjunction with other clinical data to assist the clinician in making individual patient management decisions. Shoren C. Grotte for Jean Cooper --- (Division Sign-Off) ivision of Clinical Laboratory Devices 510(k) Number (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ## Concurrence of CDRH, Office of Device Evaluation (ODE) ﺳ Prescription Use OR Over-The-Counter Use (Optional Format 1-2-96) (Per 21 CFR 801.109)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...