The Nichols Advantage® Bio-Intact PTH (1-84) immunoassay is intended for use with the Nichols Advantage Specialty System to measure the levels of parathyroid hormone in serum, EDTA plasma and heparinized plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism. Measurements of parathyroid hormone levels with the Bio-Intact PTH (1-84) Immunoassay are also used as an aid in monitoring therapeutic intervention of secondary hyperparathyroidism that frequently occurs in chronic kidney disease. Assay results should be used in conjunction with other clinical data to assist the clinician in making individual patient management decisions.
Device Story
The Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay is a two-site chemiluminescence assay designed for use on the Nichols Advantage® Specialty System. It processes human serum, EDTA plasma, or heparinized plasma samples to quantify intact parathyroid hormone (PTH) levels. The assay utilizes acridinium ester-based chemiluminometric technology and streptavidin-coated magnetic particles for separation. The system performs a 30-minute incubation at 37°C. Clinicians use the resulting quantitative PTH measurements in conjunction with other clinical data to assist in the differential diagnosis of calcium metabolism disorders and to monitor secondary hyperparathyroidism in chronic kidney disease patients. The device is intended for professional use in clinical laboratory settings.
Clinical Evidence
Bench testing only. Precision evaluated per NCCLS EP5-A (total CV 5.1-10.9%). Linearity confirmed up to 1800 pg/mL. Functional sensitivity (LOQ) < 4.0 pg/mL. Method comparison with predicate (n=305) showed Pearson's r=0.97; dialysis patient comparison (n=3199) showed r=0.97. Matrix comparison (n=27) confirmed no significant difference between serum, EDTA plasma, and heparin plasma. Stability studies confirmed sample integrity under various refrigerated and room temperature conditions.
Technological Characteristics
Two-site immunochemiluminometric sandwich assay. Reagents: streptavidin-coated magnetic particles, acridinium ester-labeled goat polyclonal anti-PTH antibody, biotinylated anti-PTH antibody. Energy source: system luminometer. Form factor: cartridge-based reagents for automated analyzer. Connectivity: standalone instrument (Nichols Advantage® Specialty System). Standards: NCCLS EP5-A for precision.
Indications for Use
Indicated for patients requiring measurement of parathyroid hormone levels to aid in the differential diagnosis of hypercalcemia or hypocalcemia and to monitor therapeutic intervention of secondary hyperparathyroidism in chronic kidney disease.
Regulatory Classification
Identification
A parathyroid hormone test system is a device intended to measure the levels of parathyroid hormone in serum and plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism.
K040813 — NICHOLS ADVANTAGE CHEMILUMINESCENCE INTACT PARATHYROID HORMONE, MODEL 62-7022 · Nichols Institute Diagnostics · Jun 8, 2004
K013992 — NICHOLS ADVANTAGE CHEMILUMINESCENCE BIO-INTACT PARATHYROID HORMONE (1-84) IMMUNOASSAY · Nichols Institute Diagnostics · Dec 13, 2001
K022472 — PARATHYROID HORMONE TEST SYSTEM/NICHOLS ADVANTAGE QUICK-INTRAOPERATIVE BIO-INTACT PTH (1-84) · Nichols Institute Diagnostics · Oct 9, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
# 12.0 510(k) SUMMARY
# K033727
This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of CFR 807.92.
# 510(k) Number: Not Known
## 1. Name of Submitter, Contact Person and Date Summary Prepared:
Nichols Institute Diagnostics 1311 Calle Batido San Clemente, CA 92673 Phone: 949-940-7417 Fax: 949-940-7440
Contact Person: Robert L. Schmidt Date Prepared: November 25, 2003
### 2. Device Name
| Trade/Proprietary Name: | Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay |
|-------------------------|------------------------------------------------------|
| Common/Usual Name: | Bio-Intact PTH (1-84) Assay |
| Classification Name: | Radioimmunoassay, Parathyroid Hormone |
# 3. Predicate Device:
The device that is the subject of this submission is substantially equivalent to the Nichols Advantage® Bio-Intact (1-84) PTH Immunoassay (K013992; Cleared 12/31/01). This 510(k) is for labeling changes to the predicate device.
# 4. Device Description:
The Bio-Intact PTH (1-84) Assay is a two-site chemiluminescence assay for use with the Nichols Advantage® Specialty System.
#### 5. Intended Use
The Nichols Advantage® Bio-Intact PTH (1-84) immunoassay is intended for use with the Nichols Advantage Specialty System to measure the levels of parathyroid hormone in serum, EDTA plasma and heparinized plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism. Measurements of parathyroid hormone levels with the Bio-Intact PTH (1-84) Immunoassay are also used as an aid in monitoring therapeutic intervention of secondary hyperparathyroidism that frequently occurs in chronic kidney disease. Assay results should be
{1}------------------------------------------------
used in conjunction with other clinical data to assist the clinician in making individual patient management decisions.
# 6. Comparison to predicate device:
The Bio-Intact PTH (1-84) Assay is substantially equivalent to the Nichols Advantage Bio-Intact PTH(1-84) Immunoassay previously submitted as K013992. The following tables compare the Bio-Intact PTH (1-84) Assay (modified) with the predicate device, the Nichols Advantage Bio-Intact PTH (1-84) Immunoassay (K013992).
# Similarities:
- Intended Use: For the quantitative determination of intact PTH in human serum or EDTA . plasma.
- Both assays use specific antibodies to bind intact PTH. .
- Both assays can use human serum or EDTA plasma for the test sample. .
- Both assays use chemiluminometric technology based on acridinium esters. .
- Both assays use 150 microliters of sample. .
- Both assays use two point calibration every two weeks (maximum) of stored working . calibration curve; or when controls out of range.
- Both assays use Streptavidin-coated magnetic particles and streptavidin-biotin separation . technology.
- Both assays' total incubation period is 30 minutes at 37℃. .
- The sensitivity of both assays is 1.5 pg/mL. (analytical) and 4.0 pg/mL (functional). ♥
- . The throughput is the same for both assays at 180 test/hour.
- . The cross reactives are the same between assays.
- . The recovery is the same between assays.
- The parallelism is the same between assays. .
- The high dose hook effect is the same between assays. .
# Differences:
Intended Use: The Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay is intended for use with the Nichols Advantage® Specialty System to measure the levels of parathyroid hormone in serum, EDTA plasma and heparinized plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism. Measurements of parathyroid hormone levels with the Bio-Intact PTH (1-84) Immunoassay are also used as an aid in monitoring therapeutic intervention of secondary hyperparathyroidism that frequently occurs in chronic kidney disease. Assay results should be used in conjunction with other clinical data to assist the clinician in making individual patient management decisions.
- The modified device includes data on preanalytical variables for serum, EDTA plasma, ● heparinized plasma and EDTA whole blood.
{2}------------------------------------------------
| Performance Characteristics: | | | |
|--------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|
| FEATURE | Nichols Advantage®<br>Bio-Intact PTH (1-84)<br>Immunoassay<br>(K013992) | Bio-Intact PTH (1-84) Assay<br>(K033302)* | Bio-Intact PTH (1-84) Assay<br>(Modified) |
| Precision<br>within<br>Run | Not greater than 4% at<br>dose greater than 5<br>pg/mL | Not greater<br>than 6% at a<br>dose greater<br>than 5 pg/mL | Not greater<br>than 6% at a<br>dose greater<br>than 5 pg/mL |
| Precision<br>Total | Not greater than 9.5%<br>at dose greater than 34<br>pg/mL | Not greater<br>than 11% at a<br>dose greater<br>than 5 pg/mL | Not greater<br>than 11% at a<br>dose greater<br>than 5 pg/mL |
| *K033302 was submitted October 14, 2003 and is under review. | | | |
| Method Comparison 1 - Serum Analysis Nichols Advantage Intact PTH Assay Vs. Bio-Intact PTH (1-<br>84) Assay | | | |
| FEATURE | Nichols Advantage® Bio-Intact<br>PTH (1-84) Immunoassay<br>(K013992) | Bio-Intact PTH (1-84) Assay<br>(Modified) | |
| Sample Size: | 305 | 305 | |
| Range of<br>Results: | Nichols Advantage Intact PTH<br>Assay: 5.0 to 1387 pg/mL<br>Nichols Advantage Bio-Intact<br>PTH (1-84) Assay: 3.0 to 746<br>pg/mL | Nichols Advantage Intact PTH Assay: 5.0 to<br>1387 pg/mL<br>Nichols Advantage Bio-Intact PTH (1-84) Assay: 3.0<br>to 746 pg/mL | |
| Passing Bablok<br>Regression<br>Equation: | y = 0.66x - 0.6 | y = 0.66x - 0.6 | |
| Least Squares<br>Linear<br>Regression<br>Equation: | y = 0.60x + 4.2 | y = 0.60x + 4.2 | |
| Pearson's<br>Correlation<br>Coefficient (r): | 0.97 | 0.97 | |
| Method Comparison 2 - Renal Dialysis Samples Nichols Advantage Intact PTH Assay Vs. Bio-Intact<br>PTH (1-84) Assay | | | |
| FEATURE | Nichols Advantage® Bio-Intact<br>PTH (1-84) Immunoassay<br>(K013992) | Bio-Intact PTH (1-84) Assay<br>(Modified) | |
| Sample Size: | N/A | 3187 | |
| Range of results: | N/A | Nichols Advantage Intact PTH Assay:<br>7 - 1797pg/mL<br>Nichols Advantage Bio-Intact PTH (1-84) Assay 4-998 | |
| Least Squares Linear Regression Equation: | N/A | $Y = 0.52 x + 1.3$ | |
| Pearson's Correlation Coefficient (r): | N/A | 0.97 | |
{3}------------------------------------------------
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Public Health Service
APR 2 2 2004
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Robert L. Schmidt Director of Quality Nichols Institute Diagnostics 1311 Calle Batido San Clemente, CA 92673
k033727 Re:
Trade/Device Name: Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay Regulation Number: 21 CFR 862.1545 Regulation Name: Parathyroid hormone test system Regulatory Class: Class II Product Code: CEW Dated: March 18, 2004 Received: March 19, 2004
Dear Mr. Schmidt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, ITDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a dctermination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practicc requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{5}------------------------------------------------
#### Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Jean M. Cooper, MS, DVM.
Yean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT )
# INDICATIONS FOR USE STATEMENT
K033727 510(k) Number (if known):
Device Name: Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay
Indications For Use:
The Nichols Advantage® Bio-Intact PTH (1-84) Immunoassay is intended for use with the Nichols Advantage® Specialty System to measure the levels of parathyroid hormone in serum, EDTA plasma and heparinized plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism. Measurements of parathyroid hormone levels with the Bio-Intact PTH (1-84) Immunoassay are also used as an aid in monitoring therapeutic intervention of secondary hyperparathyroidism that frequently occurs in chronic kidney disease. Assay results should be used in conjunction with other clinical data to assist the clinician in making individual patient management decisions.
Prescription Use
(Per 21 CFR 801.109) ✓
OR
Over-The-Counter Use (Optional Format 1-2-96)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Albert C
---
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K033721
Page 6 of 119
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.