The ADVIA Centaur FT4 Immunoassay is for in vitro diagnostic use in the quantitative determination of free thyroxine (FT4) in serum or plasma (heparinized or EDTA) using the ADVIA Centaur and ADVIA Centaur XP Systems. Measurements of free thyroxine are used in the diagnosis and treatment of thyroid diseases.
Device Story
ADVIA Centaur FT4 is a competitive chemiluminescent immunoassay for quantitative FT4 measurement in serum/plasma. Patient sample competes with acridinium ester-labeled T4 for biotinylated polyclonal rabbit anti-T4 antibody bound to avidin-coated paramagnetic particles. Inverse relationship exists between FT4 concentration and relative light units. Used on ADVIA Centaur and ADVIA Centaur XP systems in clinical laboratory settings by trained technicians. Output is a numerical FT4 concentration (ng/dL) used by clinicians to assess thyroid function and guide treatment decisions. Benefits include rapid, automated diagnostic testing for thyroid disease management.
Clinical Evidence
Bench testing only. Precision study (n=81 per sample) showed total CVs ranging from 3.98% to 11.2%. Linearity verified across 0.0825–12.5 ng/dL range. Method comparison with predicate (n=283) yielded y = 0.973x + 0.016, r^2 = 0.995. Matrix comparison (n=91 heparinized plasma, n=73 K2EDTA plasma) showed high correlation (r=0.997 and r=0.998 respectively) with serum. Analytical specificity studies confirmed no significant interference from endogenous substances or common drugs (propylthiouracil, methimazole, etc.) at tested levels.
Technological Characteristics
Competitive chemiluminescent immunoassay. Reagents: acridinium ester-labeled T4 and biotinylated polyclonal rabbit anti-T4 antibody on avidin-coupled paramagnetic particles. Assay range: 0.1–12.0 ng/dL. Automated system-based processing. Standardized against USP internal standards.
Indications for Use
Indicated for the quantitative determination of free thyroxine (FT4) in human serum or plasma (heparinized or EDTA) to aid in the diagnosis and treatment of thyroid diseases.
Regulatory Classification
Identification
A free thyroxine test system is a device intended to measure free (not protein bound) thyroxine (thyroid hormone) in serum or plasma. Levels of free thyroxine in plasma are thought to reflect the amount of thyroxine hormone available to the cells and may therefore determine the clinical metabolic status of thyroxine. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k080167
B. Purpose for Submission:
New Device
C. Measurand:
Free thyroxine (FT4)
D. Type of Test:
Quantitative, Chemiluminescent Immunoassay
E. Applicant:
Siemens Healthcare Diagnostics Inc.
F. Proprietary and Established Names:
ADVIA Centaur FT4 Immunoassay
G. Regulatory Information:
1. Regulation section:
21CFR §862.1695-Free thyroxine test system
2. Classification:
Class II
3. Product code:
CEC
4. Panel:
Chemistry (75)
H. Intended Use:
1. Intended use(s):
The ADVIA Centaur FT4 Immunoassay is for in vitro diagnostic use in the quantitative determination of free thyroxine (FT4) in serum or plasma (heparinized or EDTA) using the ADVIA Centaur and ADVIA Centaur XP Systems. Measurements of free thyroxine are used in the diagnosis and treatment of thyroid diseases.
2. Indication(s) for use:
See intended use above.
{1}
2
3. Special conditions for use statement(s):
For prescription use only
The sponsor’s stated limitations in the labeling are as follow:
a. The anticonvulsant drug phenytoin may interfere with total and free T4 levels due to competition for TBG binding sites.
b. FT4 values may be decreased in patients with non-thyroidal conditions and in patients taking carbamazepine.
c. Performance of this assay has not been established with neonatal specimens.
d. Heterophilic antibodies in human serum can react with reagent immunoglobulins, interfering with in vitro immunoassays. Patients routinely exposed to animals or to animal serum products can be prone to this interference and anomalous values may be observed. Additional information may be required for diagnosis.
e. Thyroid auto antibodies in human serum may interfere and cause falsely elevated FT4 results.
4. Special instrument requirements:
ADVIA Centaur and ADVIA Centaur XP Systems
I. Device Description:
The ADVIA Centaur Free T4 (FT4) Immunoassays are adaptations of the ACS:180, ADVIA Centaur Free T4 (FrT4) Immunoassays. The ADVIA Centaur Free T4 (FT4) Immunoassay consists of two reagents. The Lite Reagent is an acridinium ester labeled T4 derivative. It is diluted to a concentration of about 0.2 ug/mL in sodium barbital buffer (1.03%) with protein stabilizers, EDTA and sodium azide (<0.1%). The solid phase is a biotinylated polyclonal rabbit anti-T4 antibody (0.525 ug/mL) that is coupled to avidin covalently linked to paramagnetic particles. The particles are diluted in sodium barbital buffer (1.03%) with protein stabilizers, EDTA and sodium azide (<0.1%).
J. Substantial Equivalence Information:
1. Predicate device name(s):
ACS:180 FrT4 Immunoassay
2. Predicate K number(s):
k961510
3. Comparison with predicate:
{2}
| Similarities | | |
| --- | --- | --- |
| Item | New Device ADVIA Centaur FT4 Assay | Predicate ACS:180 FrT4 Assay |
| Intended Use | For in vitro diagnostic use in the quantitative determination of free thyroxine (FT4) in serum or plasma | For in vitro diagnostic use in the quantitative determination of free thyroxine (FT4) in serum |
| Assay Method | Competitive immunoassay | Competitive immunoassay |
| Detection | Chemiluminescence | Chemiluminescence |
| Assay Range | 0.1 – 12.0 ng/dL | 0.1 – 12.0 ng/dL |
| Calibrators | Calibrator A | Calibrator A |
| Calibration | 2 point | 2 point |
| Reagents | Two liquid reagents, ready to use | Two liquid reagents, ready to use |
| Lite Reagent | Acridium ester labeled T4 | Acridium ester labeled T4 |
| Expected Values (ng/dL) | Euthyroid 0.89 – 1.76
Hypothyroid < 0.89
Hyperthyroid >1.76 | Euthyroid 0.89 – 1.76
Hypothyroid < 0.89
Hyperthyroid >1.76 |
| Standardization | Internal Standard (USP) | Internal Standard (USP) |
| Differences | | |
| --- | --- | --- |
| Item | New Device ADVIA Centaur FT4 Assay | Predicate ACS:180 FrT4 Assay |
| Specimen Type | Human serum or plasma (heparinized or EDTA) | Human serum |
| Solid Phase Reagent | Biotin-labeled polyclonal anti-T4 bound to avidin paramagnetic particles | Polyclonal anti-T4 bound to paramagnetic particles |
# K. Standard/Guidance Document referenced (if applicable):
CLSI EP17-A – Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline
CLSI EP05-2A – Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline- $2^{\mathrm{nd}}$ Edition
# L. Test Principle:
The FT4 immunoassays are competitive immunoassays using direct chemiluminescent technology. FT4 in the patient sample competes with acridinium ester labeled T4 in the Lite Reagent for a limited amount of biotinylated polyclonal rabbit-anti T4 that is coupled to avidin covalently linked to paramagnetic particles. There is an inverse relationship between the concentration of free thyroxine in the sample and relative light units.
{3}
The ADVIA Centaur FT4 assay utilizes a factory set master curve generated by the FT4 master curve standards. The master curve values are contained on the master curve card provided with each kit. The barcode reader or keyboard is used to enter the master curve values on the system. The 2 calibrators in the kit are assayed when the lot is first used or after expiration of the calibrator interval. If the calibration run is valid as determined by prearranged parameters, the values are stored and used to "normalize" test values to the master curve.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
An internal precision protocol, using serum and control samples, based on a modification of CLSI EP5-2A, was used: A precision study was done with 27 runs over 16 days with three reagent lots on four ADVIA Centaur systems, 1-4 runs per day, $n = 3$ per run for a total of $n = 81$ for each sample. Coefficient of Variation was calculated as shown below:
| | | Within-run | | Total | |
| --- | --- | --- | --- | --- | --- |
| Specimen | Level (ng/dL) | SD | CV (%) | SD | CV (%) |
| Serum 1 | 0.37 | 0.0300 | 8.11 | 0.0414 | 11.2 |
| Control 1 | 0.71 | 0.0233 | 3.28 | 0.0344 | 4.85 |
| Control 2 | 1.31 | 0.0449 | 3.43 | 0.0521 | 3.98 |
| Control 3 | 3.35 | 0.0891 | 2.66 | 0.1363 | 4.07 |
| Serum 2 | 5.32 | 0.1569 | 2.95 | 0.2139 | 4.02 |
| Serum 3 | 9.68 | 0.4327 | 4.47 | 0.6321 | 6.53 |
An additional study using low level patient samples pool was assayed as part of the study. This sample was tested over 4 days, 20 replicates per day on 1 system. The precision of this low sample was comparable to the low samples in the other studies. The results are presented in the table below;
| | | Within-run | | Total | |
| --- | --- | --- | --- | --- | --- |
| Specimen | Level (ng/dL) | SD | CV (%) | SD | CV (%) |
| Serum Pool | 0.206 | 0.0214 | 10.4 | 0.0214 | 10.4 |
# b. Linearity/assay reportable range:
Twelve samples prepared in a commercial human serum matrix containing USP material at various concentrations were tested. The FT4 concentrations of the samples tested ranged from 0 to $12.50\mathrm{ng / dL}$ (Expected Values). The recovery of each sample was measured using 5 replicates at each level.
| Expected ng/dL | Mean % Recovery |
| --- | --- |
| 0 | none |
| 0.0825 | 107 |
| 0.165 | 97 |
{4}
| 0.33 | 102 |
| --- | --- |
| 0.872 | 107 |
| 1.92 | 101 |
| 3.4 | 106 |
| 5.81 | 99.6 |
| 7.48 | 94.7 |
| 9.16 | 93 |
| 10.8 | 99.3 |
| 12.5 | 99.8 |
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The ADVIA Centaur FT4 assay is traceable to an internal standard manufactured using U.S.P. (United States Pharmacopeia) material. Assigned values of calibrators and ranges of Ligand Plus controls are traceable to this standardization.
On-board stability experiments showed that the reagent is stable for 28 days at $2 - 8^{\circ}\mathrm{C}$ , and the sponsor requires that it is calibrated every 7 days.
The calibrator and control materials were cleared in previous submissions as follows:
Calibrator A (k970539), Ligand Plus Control (k030452).
d. Detection limit:
Sponsor determined the Limit of Blank (LoB) and the Limit of Detection (LoD) for the ADVIA Centaur FT4 by assaying a zero sample and a low sample (pool of patient samples) in the following manner:
Eighty replicates of each sample were tested in 4 runs over 4 days, with 20 replicates per run. The results of the study are shown below.
| Product | Number of observations | Mean (ng/dL) | Total SD (ng/dL) | LoB (ng/dL) | LoD ng/dL) |
| --- | --- | --- | --- | --- | --- |
| Blank | 80 | 0.0323 | 0.0258 | 0.08 | 0.12 |
| Low sample #1 | 80 | 0.206 | 0.0214 | | |
| Low sample #2 | 80 | 0.405 | 0.0268 | | |
$\mathrm{LoB} =$ Mean of Blank $^+$ (1.645 x SD of Blank)
$\mathrm{LoB} = 0.0323 + (1.645 \times 0.0258) = \underline{0.075 \mathrm{ng} / \mathrm{dL}}$
$\mathrm{LoD} =$ Mean of Blank $^+$ (1.645 x SD of Blank) $^+$ (1.645 x pooled SD of Low samples)
$\mathrm{LoD} = 0.0323 + (1.645 \times 0.0258) + (1.645 \times 0.0242) = \underline{0.12 \mathrm{ng} / \mathrm{dL}}$
{5}
# e. Analytical specificity:
# i.) Cross-reactivity studies:
The cross-reactivity of the ADVIA Centaur FT4 assay with a substance has been expressed as the ratio of:
- The amount of T4 required to displace $50\%$ of the maximally bound labeled T4 from the anti-T4 antibody, and
- The amount of the cross-reactant to give the same $50\%$ displacement -
| Cross Reactant | Levels Tested (mg/mL) | % Cross Reactivity |
| --- | --- | --- |
| L-Triiodothyronine | 0.0001, 0.0005, 0.001, 0.005, 0.010 | <0.02% |
| Diiodotyrosine | 0.001, 0.005, 0.010, 0.050, 0.100, 1.00 | <0.02% |
| Monoiodotyrosine | 0.001, 0.005, 0.010, 0.050, 0.100, 1.00 | <0.02% |
| 3,5-Diodo-L-thyronine | 0.0001, 0.001, 0.005, 0.010, 0.050, 0.100 | <0.02% |
| Reverse Triiodothyronine (rT3) | 0.0001, 0.0005, 0.001, 0.005, 0.015 | <0.02% |
# ii.) Endogenous substances:
The effect of albumin (up to $4.9\mathrm{g / dL}$ ), hemoglobin (up to $300\mathrm{mg / dL}$ ), lipemic (up to $1000\mathrm{mg / dL}$ of triglyceride) and icteric (up to $20\mathrm{mg / dL}$ of free and conjugated bilirubin) was evaluated by testing patient samples with two levels of FT4 (0.91 ng/dL and 1.40 ng/dL) and comparing the results to the Predicate Device on the ACS:180 and ADVIA Centaur. All the studies met the sponsor's internal acceptance criteria and no significant interference was seen with the above tested compounds at the levels tested. Less than $10\%$ interferences are considered to be not significant by the sponsor.
# iii.) Drug interferences:
Known amounts of propylthiouracil (4.0 mg/dL), methimazole (0.40 mg/dL), phenylbutazone (15.0 mg/dL), phenytoin (4.00 mg/dL), sodium salicylate (50.0 mg/dL) and aspirin (50.0 mg/dL) were added to serum samples containing 0.91 to 6.10 ng/dL of FT4 and tested in the ADVIA Centaur FT4 assay. The percent change to the dose by the addition of the substance was determined. Phenylbutazone was found to result in an average % change of 6.38%, while all others resulted in changes below 0.2 and 4%. All results passed the sponsor's internal acceptance criteria of ≤10% (average % change) as an acceptable level of interference.
# f. Assay cut-off:
Not applicable.
# 2. Comparison studies:
{6}
a. Method comparison with predicate device:
Correlation studies were performed with the ADVIA Centaur FT4 assay using the predicate ACS:180 FrT4 immunoassay as the comparative method. The correlation studies were done over 10 days using 237 patient serum samples.
The sample distribution is described below.
| Sample Type (natural samples) | Number Tested Sample Range (~ 0.14 – 11.14 ng/dL) |
| --- | --- |
| Normal | 132 (~ 0.70 – 2.12 ng/dL) |
| Hypothyroid* | 80 (0.14 – 2.04 ng/dL) (5 samples below 0.20 ng/dL) |
| Hyperthyroid* | 71 (~ 0.99 – 11.14 ng/dL) |
| Total | 283 |
*Internally classified based on vendor's TSH Values - not clinically confirmed
Samples ranged from FT4 concentrations of $0.14\mathrm{ng / dL}$ to $11.14\mathrm{ng / dL}$ . Data were plotted as single values, and the regression analysis yielded the following:
$\mathrm{y} = 0.973\mathrm{x} + 0.016$ with an $\mathbf{r}^2$ value of 0.995.
b. Matrix comparison:
The serum/plasma equivalence study was carried out using lithium heparin and $\mathrm{K}_2\mathrm{EDTA}$ plasma samples and their corresponding serum samples. Samples were collected from self-described healthy volunteers. 91 matched heparinized plasma samples were compared with serum with T4 values ranging from $0.17 - 11.4\mathrm{ng / dL}$ . Samples include natural samples, and spiked in order to cover the hard to find sample range. Mean $\%$ bias was found to be $4.33\%$ . Linear regression of the data yielded the following values: $y = 0.979x + 0.058$ , $n = 91$ , $r = 0.997$ .
In another study, 73 matched $\mathrm{K}_2\mathrm{EDTA}$ plasma samples were compared with serum with T4 values ranging from $0.17 - 11.4~\mathrm{ng / dL}$ . Samples include natural samples and spiked samples in order to cover the hard to find sample range. Mean $\%$ bias was found to be $-2.25\%$ . Linear regression of the data yielded the following values:
$\mathrm{y} = 0.967\mathrm{x} + 0.007$ $\mathrm{n = 73}$ $\mathrm{r = 0.998}$
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
{7}
b. Clinical specificity: Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable): Not applicable.
4. Clinical cut-off: Not applicable.
5. Expected values/Reference range: The expected results for the ACS:180® FrT4 assay were previously established. Data was obtained on serum samples from 388 apparently healthy individuals. Based on this population, the following reference ranges were established:
| Clinical Condition | FT4 Range (ng/dL) | FT4 Range (pmol/L) |
| --- | --- | --- |
| Euthyroid | 0.89 – 1.76 | 11.5 – 22.7 |
| Hypothyroid | less than 0.89 | less than 11.5 |
| Hyperthyroid | greater than 1.76 | greater than 22.7 |
The sponsor claims that these expected values are transferrable to the ADVA Centaur FT4 assay by analyzing 283 samples in the range of 0.14 to 11.1 ng/dL in the method comparison study above and obtaining a linear regression equation of $\mathrm{Y} = 0.979\mathrm{X} + 0.058$ , $n = 91$ , $r = 0.997$ with a mean bias of $4.33\%$ .
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.