ACS FRT4 IMMUNOASSAY
Device Facts
| Record ID | K961510 |
|---|---|
| Device Name | ACS FRT4 IMMUNOASSAY |
| Applicant | Ciba Corning Diagnostics Corp. |
| Product Code | JJX · Clinical Chemistry |
| Decision Date | Jun 17, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1660 |
| Device Class | Class 1 |
Intended Use
For the quantitative determination of free thyroxine (FT4) in serum using the Ciba Corning Automated Chemiluminescence Systems.
Device Story
Competitive chemiluminescence immunoassay; measures free thyroxine (FT4) in serum samples. FT4 competes with acridinium ester-labeled T4 for polyclonal rabbit anti-T4 antibody coupled to paramagnetic particles. Indirect relationship between FT4 concentration and relative light units (RLUs). Used in clinical laboratory settings with Ciba Corning Automated Chemiluminescence Systems. Output provides quantitative FT4 levels to assist clinicians in evaluating thyroid status.
Clinical Evidence
Bench testing only. Correlation study (n=74) against RIA reference method showed r=0.97, equation: ACS FrT4 = 1.06(RIA) - 0.33. Sensitivity: 0.1 ng/dl (range up to 11.0 ng/dl). Precision: Total % CV 2.81-4.65%.
Technological Characteristics
Competitive chemiluminescence immunoassay. Solid phase: paramagnetic particles with covalently coupled polyclonal rabbit anti-T4 antibody. Tracer: acridinium ester-labeled T4. Analyte: free thyroxine (FT4). Platform: Ciba Corning Automated Chemiluminescence Systems.
Indications for Use
Indicated for quantitative measurement of free thyroxine (FT4) in human serum to assess thyroid function.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
- Amerlex-MAB FT4 (510(k) number not known)
Related Devices
- K173122 — Abbott Architect Free T4 · Abbott Laboratories · Oct 27, 2017
- K121951 — LIAISON FT4 LIAISON CONTROL THYROID 1 LIAISON CONTROL THYROID 2 LIAISON CONTROL THYROID 3 · DiaSorin, Inc. · Apr 4, 2013
- K080167 — ACS 180 AND ADVIA CENTAUR FT4 IMMUNOASSAYS · Siemens Healthcare Diagnostics · Aug 8, 2008