The INOmax DS delivery system delivers INOmax® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators. The INOmax DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system. The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source. The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup. The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.
Device Story
Device delivers nitric oxide (NO) gas into patient breathing circuits; tracks ventilator waveforms via injector module to provide synchronized, proportional dosing. Dual-channel design: first channel manages delivery (CPU, flow controller, injector); second channel monitors gas concentrations (NO, NO2, O2) and manages alarms. Includes backup delivery mode (fixed 250 mL/min NO flow). Used in NICU and transport; operated by clinicians. Provides continuous monitoring of inspired gases; battery-powered for 6 hours. Output informs clinician of gas concentrations and delivery status; enables precise titration of NO therapy to benefit neonates with respiratory conditions. This submission adds compatibility with Vapotherm Precision Flow system.
Clinical Evidence
No clinical studies were required to support substantial equivalence. Evidence consists of bench testing validating compatibility with the Vapotherm Precision Flow system, confirming O2 dilution, NO delivery accuracy, and NO2 generation levels remain within specifications.
Technological Characteristics
Dual-channel architecture (delivery CPU/flow controller vs. monitoring CPU/gas cells). Components include delivery unit, blender, stand/cart, and NO gas tanks. Battery-powered (6 hours). Connectivity includes ventilator waveform tracking. Complies with FDA guidance for Nitric Oxide Delivery Apparatus (January 24, 2000).
Indications for Use
Indicated for neonates requiring nitric oxide inhalation therapy. Used in NICU and transport settings.
Regulatory Classification
Identification
The nitric oxide administration apparatus is a device used to add nitric oxide to gases that are to be breathed by a patient. The nitric oxide administration apparatus is to be used in conjunction with a ventilator or other breathing gas administration system.
Special Controls
The special control for this device is FDA's “Guidance Document for Premarket Notification Submissions for Nitric Oxide Administration Apparatus, Nitric Oxide Analyzer, and Nitrogen Dioxide Analyzer.”
The special control developed by the agency is a guidance document, entitled Guidance Document for Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer. This guidance document identifies the risks associated with these types of devices and contains information that will help manufacturers address those risks. This document is available on FDA's website at http://www.fda.gov/cdrh/ode/1157.pdf.
{0}------------------------------------------------
Appendix A – K110344 Section 5; Summary Additional Information Response.
::
·
. :
. .
::
# 510(k) Summary
JUN 2 4 2011
| In accordance with 21 CFR 807.92 the following summary of information is provided: | |
|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Date:</b> | December 16, 2010 |
| <b>Submitter:</b> | INO Therapeutics/Ikaria |
| | 2902 Dairy Drive |
| | Madison, Wisconsin 53718 |
| <b>Primary Contact Person:</b> | Larry Lepley |
| | Associate Director, Regulatory Affairs |
| | INO Therapeutics/Ikaria |
| | T: 608-226-3415 |
| | F: 608-226-3402 |
| <b>Secondary Contact Person:</b> | David Trueblood |
| | Director, Regulatory Affairs |
| | INO Therapeutics/Ikaria |
| | T: 608-395-3910 |
| | F: 608-226-3402 |
| <b>Device: Trade Name:</b> | INOmax DS (Delivery System) |
| <b>Common/Usual Name:</b> | Nitric Oxide Administration Apparatus (primary) |
| | Nitric Oxide Administration Apparatus, Back-up System |
| | Nitric Oxide Analyzer |
| | Nitrogen Dioxide Analyzer |
| <b>Classification Names:</b> | Apparatus, Nitric Oxide Delivery, or Apparatus, Nitric Oxide<br>Backup Delivery, Class II - 21 CFR 868.5165 |
| <b>Product Code:</b> | MRN (Primary), MRQ, MRP |
| <b>Predicate Device(s):</b> | INOmax DSIR K061901, K070867, K071516, K080484,<br>K081691, K090958, K092545, K093922 |
| <b>Device Description:</b> | The INOmax DSIR uses a "dual-channel" design to ensure the<br>safe delivery of INOmax. The first channel has the delivery CPU<br>the flow controller and the injector module to ensure the accurate<br>delivery of NO. The second channel is the monitoring system,<br>which includes a separate monitor CPU, the gas cells (NO, NO2,<br>and O2 cells) and the user interface including the display and<br>alarms. The dual-channel approach to delivery and monitoring<br>permits INOmax delivery independent of monitoring but also<br>allows the monitoring system to shutdown INOmax delivery if it<br>detects a fault in the delivery system such that the NO<br>concentration could become greater than 100 ppm. |
{1}------------------------------------------------
Appendix A - K110344 Section 5; Summary Additional Information Response.
Intended Use: The INOmax DS delivery system delivers INOmax® (nitric oxide of inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators. The INOmax DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system. The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source. The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup. The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates. Technology: All revisions of INOmax DSIR utilize component technology to deliver Nitric Oxide gas to the patient. The components consist of the Delivery System unit, the blender, a stand/cart and the NO gas tanks. Determination of Summary of Non-Clinical Tests: Substantial Equivalence: The testing concluded four requirements necessary for the operation of the INOmax DSrg and Vapotherm Precision Flow system to be compatible: O2 dilution The measured O2 was within ± 4% v/v of the expected value (see Pass/Fail Table on page 8; does not include tolerance of the Precision Flow). INOmax DSm delivery accuracy The measured values were + 15% of setting or 1.3 ppm whichever is greater which is within specification of the INOmax DSIR (see Pass/Fail Table on page 8). NO2 generation The maximum generated NO2 was at 80 ppm and 100% oxygen, as would be expected. NO2 levels did not exceed 1.3 ppm on any of the flow settings (per INOmax label < 3 ppm on 100% FiO2 and any dose setting). The maximum generated NO2 at 40 ppm
{2}------------------------------------------------
Appendix A - K110344 Section 5; Summary Additional Information Response,
and 60% oxygen was 0.4 ppm (worse case per FDA guideline of < 1 ppm on 60% FiO2 at 40 ppm dose setting).
#### Backup Delivery
The delivered NO during backup delivery was within 5% (see Pass/Fail table, NO Measured section) of the INOmax DSIR specification at flow rates above 5 L/min. Measured NO2 levels did not exceed 0.2 ppm on any settings tested and 100% oxygen (per INOmax label < 3 ppm on 100% FiO2 and any dose setting). The measured O2 was within + 3% v/v of the expected value (see Pass/Fail table 100% Oxygen section of test report).
The INOmax DS revision complies to "Guidance Document for Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer" (January 24, 2000) FDA-1157
The results of testing validated that the additional interface to the Vapotherm Precision Flow system did not alter the safety or effectiveness of the INOmax DS operation.
### Summary of Clinical Tests:
The subject of this premarket submission, INOmax DSJR interfaced to Vapotherm Precision Flow did not require clinical studies to support substantial equivalence.
Comparison of the Similarities and Differences with the Predicate Device.
Other than the item listed below, all functional/technological characteristics remain the same between the reviewed INOmax DSIR and the predicate INOmax DSJR.
#### Similarities:
No functional or technological change was made to INOmax DSJR for this 510(k) submission. The INOmax DS18 submitted for this review retains the same as the predicate device(s).
#### Differences:
Labeling change; listing and connection criteria for Vapotherm Precision Flow as an accepted interface for INOmax DSIR to be added in the INOmax DSIR Instructions for Use (IFU).
#### Conclusion:
INO Therapeutics/Ikaria considers the INOmax DSB to be as safe, as effective, and performance is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, with a wavy line underneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Larry Lepley Associate Director, Regulatory Affairs INO Therapeutics 2902 Dairy Drive Madison, Wisconsin 53718
JUN 2 4 2011
Re: K110344
Trade/Device Name: INOmax DS Regulation Number: 21 CFR 868.5165 Regulation Name: Nitric Oxide Administration Apparatus Regulatory Class: II Product Code: MRN Dated: June 2, 2011 Received: June 3, 2011
Dear Mr. Lepley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Lepley
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Anthony D. Watson
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## 510(k) Number (if known):
INOmax DS Device Name:
Indications for Use:
The INOmax DS delivery system delivers INOmax® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.
The INOmax DS provides continuous integrated monitoring of inspired 02, NO2 and NO, and a comprehensive alarm system.
The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.
The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup.
The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.
Prescription Use_X____________________________________________________________________________________________________________________________________________________________ AND/OR Use (Part 21 CFR 801 Subpart D)
Over-The-Counter
(Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schulther
(Division Sign-Uff) (Division Sign-On)
Division of Anesthesiology, General Hospital
Division of Anesthesion of Boxicos ാസ്ട്രിംസ് 6.
Infection Control, Dental Devices
510(k) Number: K110344
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.