The INOmax DS delivery system delivers INOmax® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators. The INOmax DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system. The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source. The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 ml/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup. The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.
Device Story
INOmax DS is a nitric oxide (NO) delivery system for neonates in NICU or transport settings. It inputs NO gas from high-pressure cylinders and ventilator flow data. A dual-channel architecture separates delivery (CPU, flow controller, injector module) from monitoring (CPU, electrochemical gas cells for NO, NO2, O2). The system tracks ventilator waveforms to deliver a synchronized, proportional dose of NO into the inspiratory limb of the breathing circuit. It provides continuous monitoring and alarms. If a fault is detected, the system shuts down and provides a fixed-flow pneumatic backup (20 ppm at 10 L/min flow). Clinicians set the desired NO concentration via the user interface; the device maintains this constant dose regardless of ventilator flow patterns. Benefits include precise, synchronized NO delivery and integrated safety monitoring. The device includes a 6-hour battery backup.
Clinical Evidence
Bench testing only. Validation included neonatal ventilator mode testing and software version 1.3.4 verification. No clinical data presented.
Technological Characteristics
Dual-channel delivery/monitoring system. Electrochemical gas sensors for NO, NO2, O2. Proportional solenoid valve for gas delivery. Aluminum gas cylinders. Battery backup (6 hours). Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-1-8, CGA V-1, ASTM F-1462-93, and ASTM F-1054-87.
Indications for Use
Indicated for neonates requiring nitric oxide (NO) inhalation therapy. Used in NICU or during transport. Contraindications are governed by INOmax drug labeling.
Regulatory Classification
Identification
The nitric oxide administration apparatus is a device used to add nitric oxide to gases that are to be breathed by a patient. The nitric oxide administration apparatus is to be used in conjunction with a ventilator or other breathing gas administration system.
Special Controls
The special control for this device is FDA's “Guidance Document for Premarket Notification Submissions for Nitric Oxide Administration Apparatus, Nitric Oxide Analyzer, and Nitrogen Dioxide Analyzer.”
The special control developed by the agency is a guidance document, entitled Guidance Document for Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer. This guidance document identifies the risks associated with these types of devices and contains information that will help manufacturers address those risks. This document is available on FDA's website at http://www.fda.gov/cdrh/ode/1157.pdf.
{0}------------------------------------------------
K081691
Page 1 of 6
AUG 2 8 2008
Image /page/0/Picture/2 description: The image shows the logo for INC Therapeutics. The logo is in black and white and features the letters "INC" in a bold, sans-serif font. The "C" in "INC" is larger than the other letters and curves around the word "Therapeutics", which is written in a smaller, sans-serif font to the right of "INC". A line is underneath the word "Therapeutics".
2820 Walton Commons West Madison, WI 53718 USA Tel. 608-395-3900 / Fax. 608-395-3905
- June 12, 2008 Date Prepared:
510(k) Summary of Safety and Effectiveness Information for Subject: the INOmax DS. This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 1992 and 21CFR 807.92.
Proprietary Name: INOmax DS
- Common Name(s): Nitric Oxide Administration Apparatus (primary) Nitric Oxide Administration Apparatus, Back-up System Nitric Oxide Analyzer Nitrogen Dioxide Analyzer
- Class II 21CFR868.5165, MRN (primary) Classification(s): Class II - 21CFR868.5165, MRO Class II - 21CFR868.2380, MRP Class II - 21CFR868.2385, MRQ
Anesthesiology Panel:
1NO Therapeutics' INOmax™ DS (Delivery System) Predicate Device(s): 510(k) Number K061901 (primary) INO Therapeutics' INOmax DS (Delivery System) 510(k) Number K070867 INO Therapeutics' INOmax DS (Delivery System) 510(k) Number K071516 Datex-Ohmeda INOvent Delivery System (now GE Healthcare) 510(k) Number K974562
- David M. Trueblood, Director, Regulatory Affairs Contact Person:
- Patient Population: The targeted patient population is neonates. This is derived from the INOmax® drug package labeling.
{1}------------------------------------------------
K081691
Page 2 of 6
- The primary targeted clinical setting is the Neonatal Intensive Clinical Setting: Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates as derived from the INOmax drug package labeling.
- The technology utilized in the INOmax DS (Delivery System) Technology: is very similar to the Datex-Ohmeda INOvent Delivery System (now GE Healthcare), with many of the subsystems and components being unchanged.
| Specification | INOvent Delivery System | INOmax DS (Delivery System) |
|--------------------------|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Similar to INOmax DS | Except battery backup is 6 hours<br>and provides an integrated backup<br>system for delivering a fixed flow of<br>NO into a constant flow of breathing<br>circuit gas to produce a constant<br>concentration of NO. |
| Physical Dimensions | Height 215mm<br>Width 350mm<br>Depth 430mm<br>Weight 21Kg | Height 220mm<br>Width 350mm<br>Depth 160mm<br>Weight 5.5Kg |
| Ventilator Compatibility | Similar to INOmax DS | Except 2-60 L/min for inspiratory<br>flow for neonatal and 4-120 L/min<br>for adult. |
| NO Delivery | Similar to INOmax DS | Except NO set resolution is 0.1/1/5<br>ppm depend on range, accuracy is<br>+/- 20% indicated or 2 ppm<br>whichever greater, and NO inlet<br>pressure is 1.7 to 2.3 Bar (25 to 33<br>psig). |
| NO Cylinder INOmax | Same as INOmax DS | Same |
| Gas Monitoring | Similar to INOmax DS | Except NO range resolution is 0 to<br>10 ppm +/- (20% of reading +0.5<br>ppm), calibration span is daily zero<br>with span at pre-use test if needed,<br>and NO accuracy is +/- 20% of<br>reading +0.5 ppm. |
| Injector Module | Same as INOmax DS | Same |
| NO Delivery Shut Off | Same as INOmax DS | Same |
{2}------------------------------------------------
K08116091
Page 308 6
| Specification | INOvent Delivery System | INOmax DS (Delivery System) |
|---------------------------------|-----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Calibration Gas Cylinders | NO Cal gas 40-80 ppm<br>+/- 4%<br>NO2 Cal gas 10-15 ppm +/- 10% | INOcal calibration gas, INO<br>Therapeutics LLC<br>NO Cal gas 45 ppm +/- 4%<br>NO2 Cal gas 10 ppm +/- 10% |
| Electrical Specifications | Similar to INOmax DS | Except no nurse call and battery<br>back-up is 6 hours. |
| Environmental<br>Specifications | Similar to INOmax DS | Except ambient operation and<br>storage pressure are 57 to 110 kPa<br>(430 to 825 mmHg) and |
| No Back-up Delivery | 20 ppm @ 15 L/min. | 0.25 L/min when delivered into 10<br>L/min provides 20 ppm, plus<br>INOblender may also be used. |
The materials selected were based upon INO Therapeutics' Material(s) 1NOmax DS (Delivery System) and Datex-Ohmeda INOvent Delivery System (now GE Healthcare).
Device Description: The INOmax DS (Delivery System) provides a constant dose of INOmax (nitric oxide) therapy gas into the inspiratory limb of the ventilator circuit. The INOmax DS (Delivery System) uses a "dual-channel" design to ensure the safe delivery of INOmax.
> The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2 and O2 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOmax delivery independent of monitoring but also allows the monitoring system to shutdown INOmax delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm. INOmax drug is stored as a gas mixture of NO/N2 in an aluminum cylinder at a nominal pressure of 2200 psig.
> The cylinder is attached to a high pressure regulator which incorporates a pressure gauge that indicates cylinder pressure when the cylinder valve is open. The cylinder requlator is attached via tubing to the INOmax DS (Delivery System) using one of the two NO/NO2 quick connect inlets on the back of the machine.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows handwritten text that appears to be a combination of alphanumeric characters. The top line reads 'KOS816691', while the bottom line reads 'Page 4 of 6'. The text is written in a casual, handwritten style, and the overall image has a simple, unadorned appearance.
The INOmax enters the back of the INOmax DS (Delivery System), passes through a filter, then a safety shutoff valve, which is open under normal operations.
An injector module is placed in the ventilator gas flow between the ventilator inspiratory outlet and the humidifier. Based on the ventilator flow, the INOmax cylinder concentration and set INOmax dose, the proportional solenoid delivers INOmax into the ventilator circuit via the injector module. This allows the INOmax DS (Delivery Svstem) to deliver a constant dose of INOmax regardless of the ventilator flow pattern or breath rate.
A flow sensor inside the INOmax DS (Delivery System) also monitors the NO flow out of the machine. A check valve is included prior to the INOmax DS (Delivery System) drug outlet to prevent pressure effects from the ventilator breathing circuit interfering with the NO flow sensor reading.
The INOmax DS (Delivery System) gas monitoring system provides monitored values for inspired NO, NO2, and O2 The sample gas is withdrawn from the breathing circuit and goes through a water trap to remove excess water, a zero valve, a sample pump and finally a sample flow sensor to the gas monitoring cells. The zero valve allows the gas cells to be zeroed (during low calibration) without having to disconnect the sample line from the breathing circuit. The pump and sample flow sensor ensure a constant sample gas flow rate is maintained to the monitoring cells.
The gas monitoring cells are electrochemical; they are specific to each gas and provide an electronic signal, which is proportional to the concentration of gas present.
If the delivery system does go into shutdown, the INOmax DS (Delivery System) has an integrated backup function which provides a fixed flow of INOmax (0.25L/min) into the injector module using a pneumatic on/off switch and a restrictor built into the delivery side of the system. This fixed flow of INOmax will provide 20 ppm of NO when the continuous ventilator gas flow is 10 L/min. The backup is only for short term use until a replacement delivery system can be obtained. An alarm will warn the user if the backup system is turned on while the main delivery system is in use for INOmax delivery.
{4}------------------------------------------------
K081691
Page 5 of 6
Indications for Use: The INOmax DS delivery system delivers INOmax® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.
> The INOmax DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system.
The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.
The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 ml/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup.
The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.
The testing indicated the INOmax DS (Delivery System) met Testing: its design input specifications, design output specifications and risk analysis requirements. Testing completed included:
- Validation of neonatal ventilator mode 1.
- Validation of software version 1.3.0 and intermediate 2. versions, leading to the release of software version 1.3.4 with no anomalies.
{5}------------------------------------------------
K081691
Page 6 of 6
The INOmax DS (Delivery System) was designed to comply Standard(s): with the applicable portions of the following product standards
1. FDA Guidance Document for Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer (Special Controls).
2. IEC 60601-1, plus amendments: Medical Electrical Equipment
- 3. IEC 60601-1-2: EMC/EMI
- 4. IEC 60601-1-4: Programmable Systems
- 5. IEC 60601-1-8: Alarms
- 6. CGA V-1: Medical Cylinder Connections
- 7. ASTM F-1462-93: Oxygen Analyzers
- 8. ASTM F-1054-87: Conical Fittings
The INOmax DS (Delivery System) device described in this Conclusions: Notice has the same intended use and indications, technological characteristics, and principles of operations and software applications as the previously cleared predicate device (INOmax DS (Delivery System): K061901). The only differences are that a newly validated neonatal ventilator mode has been qualified with the INOmax DS (Delivery System) and minor software changes were made in order to accommodate very low, non-adjustable bias flow rates, to respond to user feedback and to improve gas cell ambient pressure (altitude) compensation and the detection of incorrect calibration gas. These differences do not present any new issues of safety or effectiveness. This device is substantially equivalent to the predicate devices.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The bird is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 8 2008
Mr. David M. Trueblood Director, Regulatory Affairs INO Therapeutics 2820 Walton Commons West Madison, Wisconsin 53718
Re: K081691
Trade/Device Name: INOmax DS (Delivery System) Regulation Number: 868.5165 Regulation Name: Nitric Oxide Administration Apparatus Regulatory Class: II Product Code: MRN, MRO, MRP, MRP, MRQ Dated: July 31, 2008 Received: August 1, 2008
Dear Mr. Trueblood:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
CWR
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Indications for Use
510(k) Number (if known): K0811691
Device Name: INOmax DS (Delivery System)
Indications for Use:
The INOmax DS delivery system delivers INOmax® (nitric oxide of inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.
The INOmax DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system.
The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.
The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup.
The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.
× Prescription Use (Part 21 CFR 801 Subpart D)
and/or
Over-the-Counter-Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
h
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
510(k) Number:
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.