The Juno VPAP ST-A is indicated to provide noninvasive ventilation for patients weighing more than 30lbs (13 kg) with respiratory insufficiency or obstructive sleep apnoea (OSA). The iVAPS mode is indicated for patients weighing more than 66lbs (30 kg). The Juno VPAP ST-A is intended for home and hospital use. The humidifier is intended for single patient use in the home environment and re-use in a hospital/institutional environment.
Device Story
Juno VPAP ST-A is a positive pressure ventilator for noninvasive ventilation. It uses a microprocessor-controlled blower to deliver CPAP (4-20 cmH2O) or bilevel pressures (3-30 cmH2O) via patient tubing and mask. Designed for home and hospital use by patients under clinician supervision. Features include integrated humidifier, color LCD, simple controls, and wireless data transfer. Therapy modes include CPAP, S, ST, T, PAC, and iVAPS. Device maintains 'air splint' for OSA or provides pressure support for respiratory insufficiency. Integrated alarm module monitors therapy. Clinicians use device output to adjust therapy settings; remote changes possible but discouraged in hospital settings without coordination. Benefits include effective treatment of respiratory conditions through controlled airway pressure.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including pressure, flow, pressure support, trigger/cycling, and apnea/hypopnea detection protocols compared against the predicate device. Compliance with IEC 60601-1, 60601-1-2, 60601-1-8, and 60601-1-11 standards confirmed.
Technological Characteristics
Positive pressure ventilator; brush-less low voltage DC motor; pressure range 3-30 cmH2O. Microchip STM32F405ZG microcontroller with ARM32-bit Cortex-M4 CPU. Integrated heated humidifier (HumidAir). Connectivity via SD card and integrated wireless module. Non-sterile. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-11.
Indications for Use
Indicated for patients >30lbs (13kg) with respiratory insufficiency or obstructive sleep apnoea (OSA). iVAPS mode indicated for patients >66lbs (30kg).
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 13, 2016
ResMed Ltd Larissa D'Andrea Director, Government & Regulatory Affairs 9001 Spectrum Center Boulevard San Diego, California 92123
Re: K153061
Trade/Device Name: Juno VPAP ST-A Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: MNS Dated: March 9, 2016 Received: March 11, 2016
Dear Ms. D'Andrea:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153061
Device Name Juno VPAP ST-A
Indications for Use (Describe)
The Juno VPAP ST-A is indicated to provide noninvasive ventilation for patients weighing more than 30lbs (13 kg) with respiratory insufficiency or obstructive sleep apnoea (OSA).
The iVAPS mode is indicated for patients weighing more than 66lbs (30 kg).
The Juno VPAP ST-A is intended for home and hospital use.
The humidifier is intended for single patient use in the home environment and re-use in a hospital/institutional environment.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the ResMed logo. The logo features a curved line made up of a series of dots that transition in color from blue to red. Below the line of dots is the company name, "ResMed," in a bold, sans-serif font.
Juno VPAP ST-A Traditional 510(k)
# 510(k) SUMMARY [As required by 21 CFR 807.92(c)]
- April 12th, 2016 1. Date prepared
## 2. Applicant information
| Company Name/<br>Owner | ResMed Ltd<br>1 Elizabeth Macarthur Drive<br>Bella Vista NSW 2153 Australia |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Name | Mr. Jean-Nicolas Boudaud<br>Regulatory Affairs Manager<br>+61 2 88841000 (phone)<br>+61 2 88842004 (fax)<br>Jean.boudaud@resmed.com.au |
| Correspondent Details/<br>Official Contact | Ms. Larissa D'Andrea<br>Director, Government & Regulatory Affairs<br>9001 Spectrum Center Blvd<br>San Diego CA 92123 USA<br>(858) 836 6837 (phone)<br>(858) 836 5519 (fax)<br>Larissa.D'Andrea@resmed.com |
## 3. Device details and substantial equivalence claim [807.92(a)(3)]
| Trade/Device Names | Juno VPAP ST-A |
|--------------------|-----------------------------------------------|
| Device Common Name | Ventilator, continuous, non-life supporting |
| Regulation Number | 21 CFR §868.5895 |
| Regulation Name | Anesthesiology devices, Continuous Ventilator |
| Regulatory Class | Class II |
| Product Code | MNS |
| Predicate Device | ResMed S9 VPAP ST-A (K113288) |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for ResMed. The logo consists of a curved line of dots that transition in color from blue to red. Below the line of dots is the word "ResMed" in a bold, sans-serif font.
## 4. Device description
The Juno VPAP ST-A System is a positive pressure device which design is based on the previously cleared ResMed S9 WANDA VPAP ST (K140159) platform. The device is used in a wider breathing circuit which typically comprises patient tubing and a mask (patient interface) to deliver the prescribed positive pressure treatment to patients.
Some of the key features of the platform include: in-line power supply; fully integrated humidifier; heater controller; colour LCD and simple controls for ease of use. An integrated wireless module for data transfer/connectivity is also included (note: Remote changes in a hospital setting may not be appropriate, as these setting changes may not be communicated to all hospital personnel treating the patient. Hospital staff should liaise with the patient's reqular care provider such that the desired therapy outcome is achieved).
Juno VPAP ST-A contains a Micro-processor controlled blower system that generates positive airway pressure (CPAP) between 4-20 cmHyO as required to maintain an "air splint" for effective treatment of OSA and (Bilevel) pressures between 3-30 cmH2O for the treatment respiratory insufficiency.
The device also incorporates an Alarm function as a fully integrated module.
The therapy modes available in the Juno VPAP ST-A come from the predicate ResMed S9 VPAP ST-A (K113288). These include CPAP, Spontaneous, Spontaneous/Timed, Timed, PAC and iVAPS.
Juno VPAP ST-A is intended to be used under the conditions and purposes indicated in the labelling provided with the product.
It is a prescription device, supplied non-sterile.
#### 5. Indications For Use
The Juno VPAP ST-A is indicated to provide noninvasive ventilation for patients weighing more than 30lbs (13 kg) with respiratory insufficiency or obstructive sleep apnoea (OSA).
The iVAPS mode is indicated for patients weighing more than 66lbs (30 kg).
The Juno VPAP ST-A is intended for home and hospital use.
The humidifier is intended for single patient use in the home environment and re-use in a hospital/institutional environment.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the ResMed logo. The logo consists of a series of dots that transition in color from blue to red, forming a curved line. Below the dots, the word "ResMed" is written in a bold, sans-serif font.
# 6. Comparison with previously cleared devices
| Characteristics<br>Intended Use | Juno VPAP ST-A<br>(new device) | S9 VPAP ST-A (K113288)<br>Predicate device | S9 WANDA VPAP ST<br>(K140159)<br>Reference device only | Comments |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | The Juno VPAP ST-A is<br>indicated to provide<br>noninvasive ventilation for<br>patients weighing more<br>than 30lbs (13 kg) with<br>respiratory insufficiency or<br>obstructive sleep apnoea<br>(OSA).<br>The iVAPS mode is<br>indicated for patients<br>weighing more than 66lbs<br>(30 kg).<br>The Juno VPAP ST-A is<br>intended for home and<br>hospital use.<br>The Humidifier is intended<br>for single patient use in<br>the home environment<br>and re-use in a<br>hospital/institutional<br>environment | The S9 VPAP ST-A is<br>indicated to provide<br>noninvasive ventilation for<br>patients weighing more<br>than 30lbs (13 kg) or more<br>than 66lbs (30 kg) in<br>iVAPS mode with<br>respiratory insufficiency or<br>obstructive sleep apnoea<br>(OSA). The S9 VPAP ST-<br>A is intended for home<br>and hospital use. | The S9 WANDA VPAP<br>ST is indicated for the<br>treatment of Obstructive<br>Sleep Apnea (OSA) in<br>patients weighing more<br>than 66 lb (30 kg).<br>It is intended for use in the<br>hospital and home. | The new device has the same<br>intended use as the predicate S9<br>VPAP ST-A (K113288).<br>Equivalent |
| Environment of use | Hospital/Home | Hospital/Home | Hospital/Home | Equivalent |
| Prescription status | Prescription only | Prescription only | Prescription only | Equivalent |
| Therapies | | | | |
| Modes available | • CPAP<br>• S, ST, T<br>• PAC<br>• iVAPS | • CPAP<br>• S, ST, T<br>• PAC<br>• iVAPS | • CPAP (w/o EPR mode)<br>• S, ST, T<br>• VAuto | The new device includes the same<br>therapy modes as the predicate S9<br>VPAP ST-A (K113288).<br>Equivalent |
| Pressure ranges | • 4-20 cmH2O (CPAP)<br>• 3-30 cmH2O (bi-level) | • 4-20 cmH2O (CPAP)<br>• 3-30 cmH2O (bi-level) | • 4-20 cm H2O (CPAP)<br>• 4-25 cmH2O (VAuto)<br>EPR + 3 cmH2O<br>• 3-25 cmH2O (bi-level) | The new device operates in the same<br>pressure ranges as the predicate S9<br>VPAP ST-A (K113288).<br>Equivalent |
| Ramp settings | • User selected as "Off"<br>to 45 minutes in 5<br>minute increments<br>• Max Ramp time set at<br>clinician's discretion | • User selected as "Off"<br>to 45 minutes in 5<br>minute increments<br>• Max Ramp time set at<br>clinician's discretion | • User selected as "Off"<br>to 45 minutes in 5<br>minute increments<br>• Max Ramp time set at<br>clinician's discretion | Equivalent |
| Features | | | | |
| Alarms module | Yes | Yes | No | Both the new device and the<br>predicate S9 VPAP ST-A (K113288)<br>include an Alarm module.<br>Equivalent |
| Humidifier | Yes - HumidAir | Yes - H5i | Yes - HumidAir | The new device includes the last<br>generation humidifier that was first<br>used in S9 WANDA VPAP ST<br>(K140159). The humidifier function<br>and performance is equivalent to that<br>of the predicate S9 VPAP ST-A<br>(K113288).<br>All humidifiers are intended for single<br>patient use in the home environment<br>and re-use in a hospital/institutional<br>environment<br>Equivalent |
| | | | | |
| Motor type | Brush-less low voltage DC | Brush-less low voltage DC | Brush-less low voltage DC | Equivalent |
| Operating system | Microchip STM32F405ZG<br>micro-controller with<br>ARM32-bit Cortex™-M4<br>CPU | Micrium uC/OS-II<br>Software/digital | Microchip STM32F405ZG<br>micro-controller with<br>ARM32-bit Cortex™-M4<br>CPU | The new device includes an updated<br>micro and CPU compared to the<br>predicate. Those components are<br>coming from S9 WANDA VPAP ST<br>(K140159).<br>Equivalent |
| Supplemental oxygen | Labeled for use with<br>Supplemental Oxygen | Labeled for use with<br>Supplemental Oxygen | Labeled for use with<br>Supplemental Oxygen | Equivalent |
| Data transfer medium | • SD Card<br>• Fully integrated<br>wireless module | • SD Card<br>• Optional wireless<br>module | • SD Card<br>• Fully integrated<br>wireless module | The new device and the predicate<br>both provide connectivity to ResMed's<br>ECO platform (K132371) however<br>access to ECO is provided via an<br>optional module on the predicate<br>whilst the new device includes the<br>built-in module previously cleared with<br>S9 WANDA VPAP ST (K140159).<br>Equivalent |
| Hardware platform | • Gen 10 | • Gen 9 | • Gen 10 | The new device reuses the S9<br>WANDA VPAP ST (K140159)<br>platform so that numerous hardware<br>components are common with this<br>previously cleared platform.<br>Equivalent |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for ResMed. The logo consists of a curved line made up of a series of dots that transition in color from blue to red. Below the line of dots is the word "ResMed" in a bold, italicized font.
Juno VPAP ST-A
Traditional 510(k)
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for ResMed. The logo consists of a curved line of dots that transition in color from blue to red. Below the dots is the word "ResMed" in a bold, sans-serif font.
Juno VPAP ST-A
Traditional 510(k)
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the ResMed logo. The logo consists of a curved line made up of a series of dots that transition in color from blue to red. Below the line of dots is the word "ResMed" in a bold, sans-serif font.
#### 7. Non Clinical data submitted
Bench test data is presented to demonstrate that the new device meets its published specifications and back-to-back testing shows substantial equivalence to the predicate device S9 VPAP ST-A (K113288).
All bench tests confirmed that the product met the predetermined acceptance criteria. this included Pressure, Flow, Pressure Support, Trigger and Cycling, Hypopnea and Apnea tests against the predicate devices using common protocols for both devices.
The Juno VPAP ST-A has also been tested to the relevant FDA consensus standards and other applicable requirements passing all test protocols. The Juno VPAP ST-A with and without the integrated heated humidifier (HumidAir) was designed and tested according to:
- IEC 60601-1:2005+AMD1:2012, Medical electrical equipment - Part 1: General requirements for safety Medical electrical equipment - General requirements for basic safety and essential performance
- IEC 60601-1-2:2014, Medical electrical equipment Part 1-2: General requirements for ● basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- . IEC 60601-1-8:2006, Medical electrical equipment -- Part 1-8: General requirements for basic safety and essential performance -- Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- IEC 60601-1-11:2010, Medical electrical equipment Part 1-11: General requirements for ● basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
The device complies with the applicable requirements referenced in the FDA quidance documents:
- FDA Draft Reviewer Guidance for Ventilators (July 1995)
- . FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005)
## 8. Clinical data
Clinical data for the Juno VPAP ST-A is not required as the predicate devices have been subjected to clinical trial requirements or validated patient simulation models have been used during the bench testing phases
#### 9. Substantial Equivalence Conclusion
This submission demonstrates that the new Juno VPAP ST-A is substantially equivalent to the predicate S9 VPAP ST-A (K113288).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.