K122438 · Hamilton Medical AG · CBK · Nov 20, 2013 · Anesthesiology
Device Facts
Record ID
K122438
Device Name
HAMILTON-MR1
Applicant
Hamilton Medical AG
Product Code
CBK · Anesthesiology
Decision Date
Nov 20, 2013
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The HAMILTON-MR1 ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics. Intended areas of use: • In the MRI department • In the intensive care ward or in the recovery room • During transfer of ventilated patients within the hospital The HAMILTON-MR1 ventilator is classified as MR Conditional with the use of 1.5 T and 3.0 Tesla static magnetic field scanners. The HAMILTON-MR1 ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.
Device Story
HAMILTON-MR1 is an MR-conditional, turbine-powered ventilator providing invasive and non-invasive ventilation (NIV) and Adaptive Support Ventilation (ASV). It functions as a continuous ventilator for adult and pediatric patients in ICU, recovery, and MRI environments. The device features an integrated gaussmeter that monitors magnetic field strength, providing audible and visual alarms (green/yellow/red LEDs) to ensure safe positioning at the 50mT (500 gauss) line. It is designed with non-ferrous materials and a 'dead man's' braking trolley to prevent attraction to MRI magnets. Clinicians operate the device via touchscreen or single-turn wheel. The system includes a high-performance turbine (up to 210 L/min) and dual batteries for 6-hour operation. It provides continuous ventilation during patient transport to/from the MRI suite. The device is substantially equivalent to the HAMILTON-C1, with modifications for MR compatibility, including shielding, specialized alarms, and reduced ferromagnetic components.
Clinical Evidence
Bench testing only. No clinical data presented. Testing included verification of ventilation performance, alarm systems, and MR compatibility per ASTM F2213-06, F2052-06, F2119-07, and F2503-08. Gaussmeter accuracy validated against calibrated third-party equipment (±10% accuracy). Performance confirmed at 1.5T and 3.0T using Philips Achieva scanners.
Technological Characteristics
MR-conditional ventilator; turbine-powered; non-ferrous trolley with dead-man's brake. Integrated gaussmeter (magnetic sensor) for proximity monitoring. Connectivity: USB port (not tested for wireless). Power: AC 100-240V or dual Li-ion batteries. Software: Version 1.2.0. Standards: IEC 60601-1, 60601-1-2, 60601-1-8, 62304, 62366, ISO 5356-1, 5367, 14971, 21647, ASTM F2213-06, F2052-06, F2119-07, F2503-08, MIL-STD-461F.
Indications for Use
Indicated for positive pressure ventilatory support in adult and pediatric patients. Intended for use in MRI departments, intensive care wards, recovery rooms, and during hospital transport. MR Conditional for 1.5T and 3.0T scanners. Restricted to use by qualified personnel under physician direction.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Hamilton Medical. The logo consists of the word "HAMILTON" in bold, sans-serif font, with a stylized "H" symbol to the right. Below the word "HAMILTON" is the word "MEDICAL" in a similar bold, sans-serif font. The logo is simple and modern, and the use of bold fonts gives it a strong and professional look.
ST ILTON "T TRI
# 510(k) SUMMARY
Image /page/0/Picture/3 description: The image is a black square with a white icon in the center. The icon depicts a radar-like symbol with concentric circles emanating from a central point, with the number "1m" above the circles. There are also left and right arrowheads on either side of the circles, suggesting a range or distance measurement.
MRIcompliant Compatibility up to 1 n
Image /page/0/Picture/5 description: The image is of an app icon. The icon is a black square with rounded corners. Inside the square is a flower-like design with six petals. The petals are different shades of white, gray, and black.
NoCompA
Image /page/0/Picture/7 description: The image shows a gray square with rounded corners. Inside the square is a gray triangle with the letters "MR" in bold, yellow font. The letters are stacked on top of each other, with "M" on top and "R" on the bottom.
Image /page/0/Picture/8 description: The image is a square icon with rounded corners. The icon is gray and features a simple line drawing of a person lying on a gurney. The gurney has wheels and a handle, suggesting it is a medical stretcher.
Image /page/0/Picture/9 description: The image is a square sign with rounded corners. The sign is gray with the word "HOSPITAL" at the top. Below the word is a large letter "H" in black.
| SUBMITTER: | HAMILTON MEDICAL AG<br>Via Crusch 8<br>Bonaduz, 7402<br>SWITZERLAND |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON: | Ralph Aguila<br>Regulatory Affairs / Quality Engineer<br>Phone: +41 81 660 6845<br>Fax: +41 81 660 6020<br>e-mail: raaguila@hamilton-medical.ch |
| ESTABLISHMENT<br>REGISTRATION<br>NUMBER: | 3001421318 |
| PREPARATION DATE: | November 8th, 2013 |
| TRADE NAME: | HAMILTON-MR1 |
| CLASSIFICATION NAME: | CLASS II Ventilator, Continuous |
| REGULATION NUMBER: | 21 CFR 868.5895 |
| PRODUCT CODE: | CBK |
| PREDICATE DEVICE:<br>(PRIMARY) | HAMILTON-C1<br>510(k) Number: K120574 |
| PREDICATE DEVICE:<br>(SECONDARY) | UNI-VENT Eagle II MR ventilator<br>510(k) Number: K111473<br><br>AESTIVA/5 MRI anesthesia system<br>510(k) Number: K050055 |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Hamilton Medical. The logo is in black and white and features the words "HAMILTON" stacked on top of "MEDICAL". The "O" in Hamilton is replaced with a symbol that looks like a stylized "H" inside a circle.
#### DEVICE DESCRIPTION
Image /page/1/Picture/3 description: The image shows a black square button with a white icon on it. The icon depicts a person standing within a circular radar-like display, with the number "1m" at the top. There are left and right arrowheads on either side of the circle.
Image /page/1/Picture/5 description: The image is of a logo. The logo is a black square with rounded corners. Inside the square is a flower-like design with six petals. The petals are different shades of gray, ranging from light gray to dark gray.
Image /page/1/Picture/7 description: The image is a gray square with rounded corners. Inside the square is a gray triangle with the letters "MR" in bold, dark gray font. The letters are centered within the triangle.
Image /page/1/Picture/8 description: The image is a gray square containing a black icon. The icon depicts a person lying on a gurney or hospital bed, which is supported by a wheeled frame. The person is shown in profile, lying flat on their back. The gurney has an X-shaped support structure with wheels at each of the four corners.
Image /page/1/Picture/17 description: The image is a square icon for a hospital. The icon is gray with rounded corners. The letter "H" is in the center of the icon in a dark gray color. The word "HOSPITAL" is written in a smaller font above the "H".
The HAMILTON-MR1 is an MR-Conditional ventilator which increases the availability of appropriate modes of therapy for ventilated hospital patients requiring MR imaging. It covers a full range of clinical requirements such as: invasive ventilation, automated ventilation with Adaptive Support Ventilation (ASV), and non-invasive Ventilation (NIV). It can be used at the 500 gauss line, in the presence of either 1.5T or 3T magnets.
In the MRI environment, where strong magnetic fields pose a danger for both the patient and the operator, safety is the highest priority. With the effectively shielded, MRI-compatible HAMILTON-MR1 ventilator, ventilation performance and MR image quality is guaranteed throughout the procedure. The integrated gaussmeter is programmed to alarm when the clinician is placing the HAMILTON-MR1 too close to the MRI magnet, which helps the clinician to properly position the HAMILTON-MR1 at the 50mT (500 gauss) line or less.
The HAMILTON-MR1 can be used in ICU special care areas, cardiac surgery recovery rooms. step-down or sub-acute care units, and when transporting patients to the MRIdepartment. In these cases, the HAMILTON-MR1 guarantees uncompromised continuous ventilation care from the ICU to the MRI and back with its 6-hour long batteries. Alternatively, the HAMILTON-MR1 can be used as an MRI-proprietary ventilator, waiting in the MRI department for the patient. In the MR environment -- when the clinician is unable to stay close to the patient for routine adjustments -- the ASV mode adapts to the patient's lung condition.
Positioning a medical device too close to the MRI can have fatal consequences and cause serious injury to the patient or clinician. In addition, significant financial losses can occur if an MRI shutdown is required. The HAMILTON-MR1's integrated gaussmeter continuously monitors the magnetic field and gives the clinician both an audible and a visual signal if the HAMILTON-MR1 is getting too close to the MRI magnet. For increased MRI safety and ease of use in the MR environment, the integrated gaussmeter continues monitoring -- even when the ventilator is not in use. Close proximity of the ventilator to the MRI machine is crucial. The HAMILTON-MR1 is a ventilator able to be used at a magnetic field strength of 500 gauss, without creating any MR image artifacts.
Both the HAMILTON-MR1's software and ventilation modes are identical to the HAMILTON-C1. One can operate the HAMILTON-MR1 with the touch-screen or with a single-turn wheel. Hard keys give direct access to the most important functions. The two devices are identical with the exception of modifications related to the MR environment (e.g. integrated gaussmeter, special alarm indicators, MR-specific labeling, the reduction in ferromagnetic materials, the removal of the options board, and increased safety measures for battery removal or replacement). Due to these modifications, the HAMILTON-MR1 is able to withstand the challenging conditions found in the MR environment. With the large alarm lamp, a clinician can immediately identify an alarming HAMILTON-MR1 ventilator because the alarm lamp is located at the top of the device, even if the clinician is a long distance away or when several different devices are operating simultaneously in the same room. The high-performance turbine can deliver up to 210 L/min flow, which is potentially helpful when using NIV modes of ventilation. The HAMILTON-MR1 includes a trolley made out of nonferrous materials, which will not be attracted to the powerful electromagnetic forces emanating from the MRI's magnet. The trolley also has a "fail-safe" braking system (i.e., Dead man's brake).
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Hamilton Medical. The logo is in black and white and features the words "HAMILTON" and "MEDICAL" stacked on top of each other. To the right of the word "HAMILTON" is a symbol that appears to be a stylized letter "H" inside of a circle. The logo is simple and modern, and it is likely used to represent the company's brand.
## INTENDED USE
Image /page/2/Picture/3 description: The image is a square icon with rounded corners, set against a dark background. The icon features a stylized representation of a surveillance camera or sensor, with concentric circles emanating from it, suggesting a range or field of view. Above the camera, the text '1m' indicates a distance, possibly the range of the sensor. Arrows pointing left and right are positioned on either side of the camera, implying a scanning or detection capability in both directions.
# #
-
Image /page/2/Picture/6 description: The image is a square icon with rounded corners. The background of the icon is dark gray. In the center of the icon is a stylized flower or star shape composed of six petals or points. The petals are colored in a gradient from white to dark gray, creating a sense of depth and dimension.
NoCompAir
Independency
with turbine.
Image /page/2/Picture/8 description: The image shows a gray square with rounded corners. Inside the square is a triangle, also gray, with the letters "MR" in black inside the triangle. The letters are in a sans-serif font and are slightly overlapping.
Image /page/2/Picture/9 description: The image shows a symbol that is commonly used to represent medical assistance or emergency services. The symbol features a person lying on a gurney or stretcher, which is a wheeled platform used to transport patients. The gurney is depicted with its wheels visible, and there is a person standing next to it, possibly a medical professional or caregiver. The symbol is often used in hospitals, ambulances, and other healthcare settings to indicate the availability of medical assistance or emergency services.
Image /page/2/Picture/10 description: The image shows a square sign for a hospital. The sign is gray with rounded corners. The letter "H" is in the center of the sign, and the word "HOSPITAL" is at the top.
The HAMILTON-MR1 ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics.
Intended areas of use:
- · In the MRI department
- · In the intensive care ward or in the recovery room
- · During transfer of ventilated patients within the hospital
The HAMILTON-MR1 ventilator is classified as MR Conditional with the use of 1.5T and 3.0 Tesla static magnetic field scanners. The HAMILTON-MR1 ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.
## DISCUSSION OF THE NON-CLINICAL TESTS
| IEC 60601-1 | FDA Draft Reviewer Guidance for Ventilators (1995) |
|----------------|----------------------------------------------------------|
| IEC 60601-1-2 | General Requirements for Safety |
| IEC 60601-1-4 | Electromagnetic Compatibility |
| IEC 60601-1-8 | Programmable electrical medical systems |
| IEC 60601-2-12 | Alarm Systems |
| IEC 62304 | Critical Care Ventilators |
| IEC 62366 | Software life-cycle processes |
| ISO 5356-1 | Application of usability engineering to medical devices. |
| ISO 5367 | Conical connectors: Part 1: Cones and sockets |
| AAMI/ANSI HE75 | Breathing tubes intended for use with ventilators |
| ISO 14971 | Human factors engineering. Design of medical devices |
| ISO 21647 | Application of risk management to medical devices |
| | Basic safety and performance of respiratory gas monitors |
Other internationally recognized standards which the HAMILTON-MR1 meets or exceeds:
| ASTM F2213-06 | Standard Test Method for Measurement of Magnetically<br>Induced Torque on Medical Devices in the MR Environment |
|---------------|-----------------------------------------------------------------------------------------------------------------|
| ASTM F2052-06 | Measurement of Magnetically Induced Displacement Force<br>on Medical Devices in the MR Environment |
| ASTM F2119-07 | Evaluation of MR Image Artifacts from Passive Implants |
| ASTM F2503-08 | Marking medical devices for safety in the MR environment |
| IEC 62133 | Battery safety standard. Non-spillable. |
| IEC 60601-1-6 | Usability engineering process |
| ASTM F1100-90 | Waveform performance and volume comparison testing |
| MIL-STD-461F | Immunity to conducted electromagnetic energy |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Hamilton Medical. The logo is in black and white and features the words "HAMILTON" in a bold, sans-serif font, with the letter "O" replaced by a stylized "H" symbol. Below "HAMILTON" is the word "MEDICAL" in a similar bold font, creating a stacked logo design.
# AT ILTON "T TRI
# COMPARISON WITH THE PRIMARY PREDICATE DEVICE
| | | PARAMETERS | HAMILTON-MR1<br>(Proposed device) | HAMILTON-C1<br>Predicate device: K120574 | COMMENTS |
|--|--------------------------------------------------|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Image: MRicompliant Compatibility for up to 1 m. | Intended Use | The HAMILTON-MR1<br>ventilator is intended to<br>provide positive pressure<br>ventilatory support to adults<br>and pediatrics.<br>Intended areas of use:<br>• In the MRI department<br>• In the intensive care ward<br>or in the recovery room<br>• During transfer of<br>ventilated patients within<br>the hospital<br>The HAMILTON-MR1<br>ventilator is classified as<br>MR Conditional with the<br>use of 1.5 T and 3.0 Tesla<br>static magnetic field<br>scanners.<br>The HAMILTON-MR1<br>ventilator is a restricted<br>medical device intended for<br>use by qualified, trained<br>personnel under the<br>direction of a physician and<br>within the limits of its stated<br>technical specifications. | The HAMILTON-C1<br>ventilator is intended to<br>provide positive pressure<br>ventilatory support to<br>adults and pediatrics.<br>Intended areas of use:<br>• In the intensive care<br>ward or in the recovery<br>room<br>• During transfer of<br>ventilated patients within<br>the hospital<br>The HAMILTON-C1<br>ventilator is a medical<br>device intended for use by<br>qualified, trained<br>personnel under the<br>direction of a physician<br>and within the limits of its<br>stated technical<br>specifications. | Substantially equivalent.<br>The differences in the<br>intended use of the<br>HAMILTON-MR1 are<br>highlighted in <b>bold</b> .<br>The only major difference<br>between the HAMILTON-<br>C1 and the HAMILTON-<br>MR1 has to do with the<br>additional protections<br>which the HAMILTON-<br>MR1 has, in order to<br>withstand the high<br>electromagnetic fields<br>present near an MRI<br>scanner.<br>The HAMILTON-MR1 has<br>passed rigorous testing to<br>make sure that it will<br>function properly in the<br>MRI environment. Testing<br>included: magnetic field<br>interactions, heating,<br>induced electrical fields,<br>and artifacts. |
| | Image: NoCompAir Independency with turbine. | Intended patient<br>population | Patients include adults and<br>pediatrics. | Patients include adults and<br>pediatrics. | Equivalent |
| | Image: MR | Maximum<br>inspiratory flow | 210 L/min | 210 L/min | Equivalent |
| | Image: HOSPITAL H | MRI-room<br>compatible | Yes | No | Not equivalent |
| | | Temperature<br>Range | 5 to 40° C (operating),<br>-20 to 60° C (storage) | 5 to 40° C (operating),<br>-20 to 60° C (storage) | Equivalent |
| | | Software<br>version | Version 1.2.0 | Version 1.1.2 | Substantially equivalent |
| | | Number of<br>batteries | 2 | 1 | Substantially equivalent |
| | | Weight | 6.8 kg (15 lb), or<br>21 kg (46 lb) w/ trolley | 4.9 kg (10.8 lb) | Substantially equivalent |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Hamilton Medical. The text "HAMILTON" is in bold, black letters, with a stylized "G" symbol next to it. Below "HAMILTON" is the word "MEDICAL", also in bold, black letters.
# Host JILTO. ". 151
| PARAMETERS | HAMILTON-MR1<br>(Proposed device) | HAMILTON-C1<br>Predicate device: K120574 | COMMENTS |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Operation<br>environmental<br>requirements | •5 to 40 °C (41 to 104 °F)<br>•10 to 95%, non-condensing<br>•1100 to 700 hPa | •5 to 40 °C (41 to 104 °F)<br>•10 to 95%, non-condensing<br>•1100 to 600 hPa | Substantially<br>Equivalent |
| Input Power | AC: 100 to 240 V, 50/60 Hz | AC: 100 to 240 V, 50/60 Hz | Equivalent |
| Power<br>Consumption | 50 VA typical | 50 VA typical | Equivalent |
| Battery | Two batteries, Li-ion, sealed,<br>maintenance-free | One battery, Li-ion, sealed,<br>maintenance-free | Substantially<br>Equivalent |
| Battery operating<br>time (typical) | 5.5 hours | 2 hours | Substantially<br>Equivalent |
| Safety features | • Apnea backup ventilation<br>• Automatic self-tests<br>• Alarms (operator-<br>adjustable / non-adjustable)<br>• Alarm backup buzzer<br>• External flow sensor failure<br>mode<br>• Safety mode in case of<br>technical failures<br>• Air inlet HEPA filter<br>monitoring<br>• Monitored fan<br>• Event log | • Apnea backup ventilation<br>• Automatic self-tests<br>• Alarms (operator-<br>adjustable / non-adjustable)<br>• Alarm backup buzzer<br>• External flow sensor<br>failure mode<br>• Safety mode in case of<br>technical failures<br>• Air inlet HEPA filter<br>monitoring<br>• Monitored fan<br>• Event log | Equivalent |
| Emergency air<br>intake | In case of a power supply,<br>technical, or pneumatics<br>failure the ambient valve<br>allows spontaneous<br>breathing. | In case of a power supply,<br>technical, or pneumatics<br>failure the ambient valve<br>allows spontaneous<br>breathing. | Equivalent |
| Maximum working<br>pressure limit | 60 cmH20 | 60 cmH20 | Equivalent |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Hamilton Medical. The logo is in black and white and features the words "HAMILTON" stacked on top of "MEDICAL". To the right of the word "HAMILTON" is a stylized "H" inside of a circle.
Image /page/5/Picture/1 description: The image shows the word "HAMILTON.MR1" in a bold, sans-serif font. The letters are black, and the background is white. The letters are evenly spaced and easy to read. The word appears to be a brand name or product name.
1m
**MRIcompliant**
Compatibility
for up to 1 m.
Image /page/5/Picture/4 description: The image is a square icon with rounded corners. The background of the icon is dark gray. In the center of the icon is a six-petal flower design. The petals are colored in shades of gray, ranging from light gray to dark gray, creating a gradient effect.
NoCompAir
independency
with turbine.
Image /page/5/Picture/6 description: The image shows a gray square with a triangle inside. The triangle is also gray and has the letters "MR" in bold, yellow font inside. The letters are centered within the triangle.
Image /page/5/Picture/7 description: The image is a gray square with rounded corners. Inside the square is a black icon of a person lying on a gurney. To the right of the gurney is a person standing with a cane.
Image /page/5/Picture/8 description: The image is a gray square with rounded corners. Inside the square is a large, dark gray letter "H". Above the "H" is the word "HOSPITAL" in smaller, dark gray letters. The image likely represents a hospital sign or symbol.
## MAGNETIC FIELD THRESHOLDS
| Alarm/action | Magnetic<br>field range | Accuracy |
|---------------------------------------------|-------------------------|-----------|
| Green light - acceptable | $\leq$ 50mT | $\pm$ 10% |
| Yellow light - too close, alarm sounds | 50 - 80mT | |
| Red light - too close, alarm sounds | > 80mT | |
| Red X light - technical fault, alarm sounds | -- | |
The intended-use statement for the HAMILTON-MR1 ventilator is comparable to the primary predicate device. The only difference is that the HAMILTON-MR1 can also be used in the MR-environment. However, the technological characteristics (i.e., design and energy source) and the performance specifications of the HAMILTON-MR1 ventilator are equivalent to those of the HAMILTON-C1 predicate device. The HAMILTON-MR1 has not been tested for the inclusion of wireless connections into its USB-port, e.g. plug-in "dongles", as
The AESTIVA/5 MRI anesthesia system (K050055) is referred to as a predicate device since it also has an integrated gaussmeter, which notifies the user when the device is in excess of its labeled operating parameters. This is comparable to the integrated gaussmeter found on the HAMILTON-MR1, which notifies the user when the device is in excess of its labeled operating parameters (i.e. 50mT / 500 gauss). Non-clinical performance testing has shown that the HAMILTON-MR1's integrated gaussmeter has an accuracy level of ± 10%.
described in the FDA guidance on RF Wireless Technology in Medical Devices.
#### Note:
These values are based on a comparison with a commercially available, calibrated thirdparty gaussmeter. The HAMILTON-MR1's integrated gaussmeter's magnetic sensors are located at the center of the HAMILTON-MR1's enclosure; it consequently measures the gauss levels at the center of device. Furthermore, performance bench-testing has shown that gauss levels of 29mT in the center of the HAMILTON-MR1, correspond to ≤ 50mT on the outside of the HAMILTON-MR1's enclosure. Nevertheless, the HAMILTON-MR1 can function in accordance to its "essential performance criteria" at qauss levels of more than 100mT (1,000 gauss), although this is not recommended by HAMILTON MEDICAL. These performance tests were conducted using both the Philips Achieva 3.0T and 1.5T MRI scanners. For the dynamic field testing, HAMILTON MEDICAL used the compatibility protocols for third-party equipment validation recommended by Philips (e.g. MaxGrad. MaxB1+SAR, MaxGrad+RF).
In non-clinical testing, the device was found to be MRI safe to operate at (or less than) a fringe magnetic of 50mT. This testing was based on both ASTM F2213-06 (Magnetically Induced Torque) and ASTM F2052-06 (Magnetically Induced Displacement Force), which showed that the HAMILTON-MR1 with its brake activated, will only be pulled towards the MRI magnet at gauss levels of ≥ 445mT (4,450 gauss) with a magnetically induced force of 64.2 newtons. This test was conducted using a Philips Achieva 3.0 MRI scanner.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Hamilton Medical. The logo is in black and white and features the words "HAMILTON" and "MEDICAL" stacked on top of each other. To the right of the word "HAMILTON" is a stylized "H" inside of a circle.
### INTEGRATED GAUSSMETER
Image /page/6/Picture/3 description: The image shows a black button with a white icon. The icon depicts a person standing within a circular range, labeled "1m" at the top. There are left and right arrows on either side of the circle, suggesting a range or distance adjustment feature. The button appears to be part of a device or interface related to distance measurement or proximity detection.
**MRIcompliant**
Compatibility
for up to 1 m.
Image /page/6/Picture/5 description: The image is a square icon with a dark background. In the center of the icon is a stylized flower or star shape composed of six petals or leaves. The petals are arranged in a circular pattern, radiating outward from the center. The petals are colored in varying shades, creating a sense of depth and visual interest.
NoCompAir
Independency
with turbine
Image /page/6/Picture/7 description: The image is a gray square button with rounded corners. Inside the square is a triangle, and inside the triangle are the letters "MR" in bold, sans-serif font. The letters are also gray, but slightly lighter than the background of the triangle.
Image /page/6/Picture/8 description: The image describes the gauss levels and LED indicators for the HAMILTON-MR1 ventilator in relation to MRI scanner proximity. A green LED indicates safe distance at 290 gauss (29mT) or less. A yellow LED indicates a range of 300 gauss (30mT) to 690 gauss (69mT), triggering a warning alarm to move the ventilator away from the MRI scanner. A red LED indicates at least 700 gauss (70mT), causing continuous flashing and an audible alarm, requiring servicing by a HAMILTON-MEDICAL specialist.
The HAMILTON-MR1 includes similar LED indicators to that of the AESTIVA/5 (K050055). The purpose for the three LED indicators on the integrated gaussmeter is to provide a visual representation of a range of gauss levels. For example, in the HAMILTON-MR1, the green LED shows that the integrated gaussmeter at the center of the ventilator is measuring gauss
### COMPARISON WITH OTHER MR-COMPATIBLE VENTILATORS
| | Proposed<br>device:<br>HAMILTON-MR1 | Maquet<br>Servo-i<br>(K063404) | CareFusion<br>LTV 1200 MR<br>(K083688) | GE Versamed<br>iVent 201 MRI<br>(K073694) | Impact<br>Eagle II MR<br>(K111473) |
|-----------------------------------------------------|-------------------------------------|--------------------------------|----------------------------------------|-------------------------------------------|------------------------------------|
| Maximum<br>recommended<br>magnetic<br>induction | 50mT<br>(500 gauss) | ∼20mT<br>(∼200 gauss) | ∼10mT<br>(∼100 gauss) | ∼10mT<br>(∼100 gauss) | ∼20mT<br>(∼200 gauss) |
| Nearest<br>distance to<br>the MRI's bore<br>opening | ∼1 m<br>(3.25 ft) | ∼2.25 m<br>(7.5 ft) | ∼2.75 m<br>(9 ft) | ∼2.75 m<br>(9 ft) | ∼2 m<br>(6.5 ft) |
| Minimal Tidal<br>Volume (VT) | ≥ 20 mL | ≥ 100 mL | ≥ 50 mL | ≥ 100 mL | ≥ 50 mL |
| Integrated<br>Gaussmeter | Yes | No | No | No | No |
#### CONCLUSION
The intended use of the HAMILTON-MR1 is substantially equivalent to the predicate devices. The HAMILTON-MR1's software has passed through verification/validation tests. A complete revision level history, hazard analysis, and a traceability analysis linking requirements to validation were done. The HAMILTON-MR1 includes a risk management report, system-level validation, verification testing according to the applicable standards, and testing in an MR environment. The conclusions drawn from the non-clinical tests demonstrate that the HAMILTON-MR1 is substantially equivalent to the legally marketed devices cited above.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three tail feathers, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. The logo is black and white.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
· November 20, 2013
Hamilton Medical AG Mr. Ralph Aguila Regulatory Affairs/Quality Engineer Via Crusch 8 BONADUZ. GRISONS 7402 SWITZERLAND
Rc: KI22438
Trade/Device Name: HAMILTON-MR1 Regulation Number: 21 CFR 868.5895 Regulation Name: Ventilator, Continuous Regulatory Class: II Product Code: CBK Dated: November 8, 2013 Received: November 12, 2013
Dear Mr. Aguila:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it nay be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{8}------------------------------------------------
#### Page 2 - Mr. Aguila
or any Federal statuies and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
Intp://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.hum. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Image /page/8/Picture/8 description: The image contains a combination of text and decorative elements. The text includes "Tejashri Purohit-Sheth, M.D.", "Clinical Deputy Director", and "DAGRID". There is also a decorative flourish on the left side of the image, and the word "FOR" is visible in the bottom right corner.
Erin Keith, M.S. Acting Division Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
### INDICATIONS FOR USE STATEMENT
### K122438
510(k) Number:
Device Name: HAMILTON-MR1
Indication for Use:
The HAMILTON-MR1 ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics.
Intended areas of use:
- · In the MRI department
- · In the intensive care ward or in the recovery room
- · During transfer of ventilated patients within the hospital
The HAMILTON-MR1 ventilator is classified as MR Conditional with the use of 1.5 T and 3.0 Tesla static magnetic field scanners.
The HAMILTON-MR1 ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anya C. Harı
Digitally signed by Anya C. Hany -S DN: colls, DaU.S. Government, ouatHHS, ou=FDA Poorta, consinva C Harry -S. 0.9.2342.19200300.100.1.3 =0011315590 Date: 2013.11.19 15:21:16 -05 00
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.