Flexicare Heated Wire Breathing Systems are intended for use to connect a patient's airway to a Ventilator and Humidification Chamber as part of a complete system to provide warmed and humidified inspired respiratory gases to ventilated patients and those receiving respiratory support. Intended for Adult patients within a hospital environment. Compatible with Fisher & Paykel MR850 Respiratory Gas Humidifier. Available in Adult (038-31-748U) size. Humidification Chambers are intended for use to hold water required to humidify the air being delivered to patients. Intended for any patient requiring active humidification within a hospital environment.
Device Story
Device consists of sterile, single-patient-use heated wire breathing circuits and autofill humidification chambers; used to connect patient airway to ventilator and external humidifier (Fisher & Paykel MR850). Heated wires within tubing lumen warm inspired gases to reduce condensation. Operated by clinicians in hospital settings. Output is warmed, humidified gas delivered to patient. Benefits include reduced condensation and maintenance of gas temperature (max 37°C).
Clinical Evidence
Bench testing only. Testing included visual/dimensional inspection, internal volume, heated wire resistance, flow resistance, humidifier compatibility, connector strength, leak testing, drop testing, tensile strength, tube compliance, humidification output, shelf life, EMC, electrical safety, and biocompatibility (ISO 10993). All tests passed.
Technological Characteristics
Materials: Low density polyethylene tubing, injection molded/extruded plastics. Design: Heated wire assemblies, 22mm conical connectors (ISO 5356-1:2004). Energy: Passive heating via external humidifier (MR850). Sterilization: Ethylene Oxide. Connectivity: Compatible with standard ventilators and MR850 humidifier. Standards: ISO 5367, ISO 8185, IEC 60601-1, ISO 10993.
Indications for Use
Indicated for adult patients in hospital environments requiring respiratory support or ventilation via a heated humidification system.
Regulatory Classification
Identification
A breathing system heater is a device that is intended to warm breathing gases before they enter a patient's airway. The device may include a temperature controller.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three faces in profile, stacked on top of each other. The faces are oriented to the right and are depicted in a simple, line-art style. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 15, 2015
Flexicare Medical Ltd. % Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, MN 55313
Re: K150900
Trade/Device Name: Flexicare Heated Wire Breathing Systems Regulation Number: 21 CFR 868.5270 Regulation Name: Breathing System Heater Regulatory Class: Class II Product Code: BZE, BTT Dated: April 30, 2015 Received: May 1, 2015
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mark Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name
Flexicare Heated Wire Breathing Systems
#### Indications for Use (Describe)
Flexicare Heated Wire Breathing Systems are intended for use to connect a patient's airway to a Ventilator and Humidification Chamber as part of a complete system to provide warmed and humidified inspired respiratory gases to ventilated patients and those receiving respiratory support. Intended for Adult patients within a hospital environment. Compatible with Fisher & Paykel MR850 Respiratory Gas Humidifier. Available in Adult (038-31-748U) size.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name
Flexicare Autofill Humidification Chamber
#### Indications for Use (Describe)
Humidification Chambers are intended for use to hold water required to humidify the air being delivered to patients. Intended for any patient requiring active humidification within a hospital environment.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Flexicare. The logo has the word "flexicare" in blue and red, with the "flexi" in blue and the "care" in red. To the left of the word is a stylized "f" in red and blue. Below the word is the phrase "TOTAL QUALITY - TOTAL CARE" in red.
# SECTION 4
510(k) Summary
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Flexicare. The word "flexicare" is written in a stylized font, with the "f" in red and the rest of the word in blue. Above the word "flexicare" is the phrase "TOTAL QUALITY - TOTAL CARE" in red. The logo is simple and modern, and it conveys a sense of quality and care.
Cynon Valley Business Park, Mountain Ash, Mid Glamorgan. CF45 4ER UK. Tel: +44 (0)1443 474647 Fax: +44 (0)1443 474222 Customer Services: +44 (0)1443 471580 Purchasing Fax: +44 (0)1443 476332 Email: enquiries@flexicare.com www.flexicare.com
## 510(k) Summary
## 510(k) Sponsor, Contact Person and Date Summary Prepared:
Flexicare Medical Limited Cynon Valley Business Park Mountain Ash, Mid. Glamorgan CF45 4ER. United Kingdom
Joel Biddle Compliance Engineer Telephone: 00 44 1443 474 647 Fax: 00 44 1443 474 222
Summary prepared on: May 2015
Device Name:
Trade Name: Flexicare Heated Wire Breathing Systems
Common/Usual Name: Heated Breathing System
Classification Name: Breathing System Heater: 21 CFR 868.5270
Product Codes: BZE (breathing system heater) and BTT (respiratory gas humidifier).
## Legally Marketed Equivalent Device:
Flexicare's Adult Heated Wire Breathing Systems are substantially equivalent to Fisher & Paykel's Adult RT380 Heated Breathing Circuit cleared under K122432.
Flexicare's Autofill Humidification Chamber is substantially equivalent to Fisher & Paykel's MR290 Autofill Humidification chamber cleared under K934140.
## Device Description:
Flexicare Heated Wire Breathing Systems are sterile single patient use devices which form part of a respiratory humidification system. In this system the inspiratory limb delivers heated humidified gas to the patient and expiratory limb carries the expired gas away from the patient.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for "flexicare". The logo has the word "flexicare" in blue and red letters. Below the word "flexicare" is the phrase "TOTAL QUALITY - TOTAL CARE" in red letters.
Flexicare Heated Wire Breathing Systems are supplied with an Auto-fill Humidification chamber. When in use a Humidfication Chamber holds a volume of water and is placed onto a heater unit and fills automatically from a suspended water source.
The air from the ventilator is passed through the Humidification Chamber, gaining heat and humidity. This air is then channeled through the Heated Wire Breathing System to the patient.
The inclusion of a heated wire within the lumen of both the System tubes the amount that the humidified air cools when travelling to/from the patient. This in turn reduces the amount of condensation and water build-up within the System.
Flexicare's Heated Wire Breathing Systems are comprised of disposable connectors, tubing and heated wire assemblies. The systems are intended for Adult patients.
Intended Use:
Flexicare Heated Wire Breathing Systems are intended for use to connect a patient's airway to a Ventilator and Humidification Chamber as part of a complete system to provide warmed and humidified inspired respiratory gases to ventilated patients and those receiving respiratory support. Intended for Adult patients within a hospital environment. Compatible with Fisher & Paykel MR850 Respiratory Gas Humidifier. Available in Adult (038-31-748U) size.
Humidification Chambers are intended for use to hold water reguired to humidify the air being delivered to patients. Intended for any patient requiring active humidification within a hospital environment.
Substantial Equivalence:
Flexicare's Heated Wire Breathing Systems have the same intended use as the predicate device.
Flexicare's Heated Wire Breathing Systems and the Predicate device are single patient use devices. Supplied in Adult sizes.
Neither manufacturer's devices are life supporting or life sustaining.
Patient Contact -
Skin Contact & Externally Communicating - Prolonged duration <30days (for max 7 days use)
Neither Flexicare's Heated Wire Breathing Systems nor the predicate device by Fisher & Paykel require software to operate/function.
Flexicare's Heated Wire Breathing Systems & the predicate device require connection to an electronically powered Fisher & Paykel MR850 Respiratory Gas Humidifier.
Flexicare's Heated Wire Breathing Systems are supplied sterile in individual EO permeable pouches. Fisher & Paykel's devices are supplied non-sterile in sealed Polybags.
{7}------------------------------------------------
Both devices are able to be used with industry standard devices such as ventilators and catheter mounts.
Both Flexicare's Heated Wire Breathing Systems & the predicate device are designed for the same intended use in the same intended conditions.
Both designs consist of components made from injection molded & extruded plastics. During comparison testing it was determined that there were no invasive components in either of the manufacturer's devices.
The gas delivering tubing of both the Flexicare and predicate device are manufactured from Low density Polyethylene. Both manufacturers' device's feature an inspiratory limb angled 22mm female connector with temperature port. Both manufacturers' device's also feature a Y-piece connector with temperature port at the patient end.
All connectors on both Flexicare and Predicate devices are conical connectors compliant with ISO 5356-1:2004.
Flexicare Heated Wire Breathing System tubes are blue and/or colorless with all connectors blue in color except for the angled machine-end connectors which are colorless. The Fisher and Paykel Heated Wire Breathing System tubes are blue and/or colorless with colorless connectors.
Any differences in color/shade between the Flexicare devices and the predicate devices is by manufacturer's choice/ branding, and is not related to sizing, intended use, gender of patient or performance of device.
Both manufacturer's devices are supplied with an automatically filling humidification chamber.
Both Flexicare Heated Wire Breathing Systems and Predicate device have 1,6m breathing tube length.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image is a logo for "flexicare". The logo is made up of the word "flexicare" in blue and red, with the "f" in "flexi" being stylized with two curved lines, one red and one blue. Below the word "flexicare" is the phrase "TOTAL QUALITY - TOTAL CARE" in smaller blue letters.
# Substantial equivalence comparison table - Adult Heated Wire Breathing Systems
Flexicare's Heated Wire Breathing Systems are substantially equivalent to RT380 Heated Wire Breathing System manufactured by Fisher & Paykel (510(k) K122432) The Table below shows the similarities and differences between the Flexicare Medical Adult Heated Wire Breathing Systems and Fisher & Paykel predicate device.
| Characteristic compared | Flexicare Adult Heated Wire Breathing Systems | Predicate Device F&P RT380 |
|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510K | K: Unknown | K:122432 |
| Intended use | Flexicare Heated Wire Breathing Systems are<br>intended for use to connect a patient's airway<br>to a Ventilator and Humidification Chamber as<br>part of a complete system to provide warmed<br>and humidified inspired respiratory gases to<br>ventilated patients and those receiving<br>respiratory support. Intended for Adult<br>patients within a hospital environment.<br>Compatible with Fisher & Paykel MR850<br>Respiratory Gas Humidifier. Available in Adult<br>(038-31-748U) size. | The RT380 and RT385 'Adult Evaqua 2' dual-<br>heated breathing circuits are intended as<br>conduits of breathing gas for ventilation of<br>adult patients, and to maintain the<br>temperature of humidified inspired gas.<br>Source: K122432 SE letter from FDA |
| Target population | Adult | Adult |
| Indications for use | Instruction leaflet | Instruction leaflet |
| Environment used | Hospital | Hospital |
| Product labelling | Heated Wire Breathing System | Heated Wire Breathing System |
| Volume (ml) | 926 | 1090 |
| Tube length (m) | Ins - 1.6m<br>Exp - 1.6m | Ins - 1.6m<br>Exp - 1.6m |
| Wire resistance (ohms) | Ins - 18.55<br>Exp - 25.9 | Ins - 17.5<br>Exp - 22.6 |
| Tube material and designated diameter | Ins - Corrugated 22MM<br>Exp - Smoothbore 22MM | Ins - Corrugated 22MM<br>Exp - Smoothbore 22MM |
| Rated Flow | 40L.min | 40L.min |
| Compliance (ml/Kpa) | 7.68ML/kPa | 8.93ML/kPa |
| Resistance to flow @ 30 lpm (mb) | 0.3 | 0.2 |
| Compatibility with the environment and other devices. | To be used with Flexicare Autofill<br>Humidification Chamber. Compatible with F &<br>P MR850 Respiratory Gas Humidifier. | To be used with F & P MR290 Humidification<br>Chamber. Compatible with F & P MR850<br>Respiratory Gas Humidifier. |
| Active<br>controller | No - Humidifier controlled | No - Humidifier controlled |
| Energy used<br>and or<br>delivered | Due to the wire resistance, heat is dispersed<br>into the breathing tube lumen. As a result, the<br>air passing through the tubing is warmed<br>whilst travelling to patient, reducing the water<br>condensation in the breathing System. The<br>raising of the gas temperature does not<br>exceed 37ºC | Due to the wire resistance, heat is dispersed<br>into the breathing tube lumen. As a result, the<br>air passing through the tubing is warmed<br>whilst travelling to patient, reducing the water<br>condensation in the breathing System. The<br>raising of the gas temperature does not<br>exceed 37°C |
| Sterility<br>Standards Met | Sterile – Ethylene Oxide Gas<br>ISO 5367<br>ISO 5356<br>ISO 8185<br>IEC 60601-1<br>ISO 10993<br>BS EN 556 | Non-Sterile<br>ISO 5367<br>ISO 5356<br>ISO 8185<br>IEC 60601-1<br>ISO 10993 |
| Biocompatibility | ISO 10993 compliant | ISO 10993 compliant |
| Electrical Safety | IEC 60601-1 compliant<br>IEC 60601-1-2 compliant | IEC 60601-1 compliant<br>IEC 60601-1-2 compliant |
| Non-clinical<br>Test Results | Verification tests were performed to establish the safety and efficacy of Flexicare's 038-31-<br>748U Heated Wire Breathing System. These Non-clinical tests included Visual<br>inspection/comparison, Dimensional inspection, Internal Volume, Heated wire resistance,<br>resistance to flow, humidifier start up/compatibility, conical connector compliance, connector<br>strength, leaking, drop testing, tensile strength, tube compliance, humidification output, shelf<br>life verification, Electromagnetic Capability, Electrical Safety and Biocompatibility. Testing<br>demonstrated that the relevant features, design and performance of each manufacturer's<br>device are substantially equivalent. | |
| Conclusion | Flexicare's 038-31-748U Heated Wire Breathing System is considered to be substantially<br>equivalent to the Fisher & Paykel RT380 Breathing Circuit. The comparison of features,<br>performance, materials and intended use demonstrate that Flexicare's 038-31-748U Heated<br>Wire Breathing System is as safe and effective as the predicate device for its intended purpose. | |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image is a logo for a company called "flexicare". The word "flexicare" is written in a stylized font, with the "f" in red and the rest of the word in blue. Below the word "flexicare" is the tagline "TOTAL QUALITY - TOTAL CARE", which is written in red. The logo is simple and modern, and it conveys a sense of quality and care.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image is a logo for "flexicare". The logo is made up of the word "flexicare" in blue and red, with the "f" in "flexi" being stylized with two curved lines, one red and one blue. Below the word "flexicare" is the phrase "TOTAL QUALITY - TOTAL CARE" in smaller blue letters.
## Substantial equivalence comparison table - Autofill Humidification Chamber
Flexicare's Autofill Humidification Chamber is substantially equivalent to MR290 Autofill Humidification Chamber manufactured by Fisher & Paykel (510(k) K934140) The Table below shows the similarities and differences between the Flexicare Medical Autofill Humidification chamber and Fisher & Paykel predicate device.
| Characteristic compared | Flexicare Autofill Humidification Chamber | Predicate Device<br>F&P MR290<br>K:934140 |
|-------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| 510K | | |
| Intended use | Humidification Chambers are intended for use to hold water required to humidify the air being delivered to patients. Intended for any patient requiring active humidification within a hospital environment. | Humidification chambers are intended for use to hold water required to humidify the air being delivered to patients. |
| Target population | Any patient using a heated humidifier | Any patient using a heated humidifier |
| Indications for use | Instruction leaflet | Instruction leaflet |
| Environment used | Hospital | Hospital |
| Product labelling | Autofill Humidification chamber | Autofill Humidification chamber |
| Fill Volume (ml) | Full = 292<br>Max = 114.9 | Full = 340<br>Max = 83.9 |
| Weight (g) | 124 | 109.5 |
| Leakage (ml/min) | No leakage | No leakage |
| Compliance at empty (ml/Kpa) | 5.5 | 5.5 |
| Resistance to flow @ 60 lpm (mb) | 0.1 | 0.3 |
| Moisture output (mg/l) at 10lpm | 46.6 | 53.9 |
| Compatibility with the environment and other devices. | To be used with Flexicare's Heated Wire Breathing Systems and F & P MR850 Respiratory Gas Humidifier. | To be used with F & P Heated Wire Breathing Systems and F & P MR850 Respiratory Gas Humidifier. |
| Energy used and or delivered | Electrical energy within heater unit is used to raise the temperature and humidity of the gas delivered to the patient | Electrical energy within heater unit is used to raise the temperature and humidity of the gas delivered to the patient |
| Sterility | Sterile - Ethylene Oxide Gas | Non-Sterile |
| Standards Met | ISO 5367<br>ISO 5356<br>ISO 8185<br>IEC 60601-1<br>ISO 10993<br>BS EN 556 | ISO 5367<br>ISO 5356<br>ISO 8185<br>IEC 60601-1<br>ISO 10993 |
| Biocompatibility | ISO 10993 compliant | ISO 10993 compliant |
| Electrical Safety | IEC 60601-1 compliant<br>IEC 60601-1-2 compliant | IEC 60601-1 compliant<br>IEC 60601-1-2 compliant |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image is a logo for "Flexicare". The word "Flexicare" is written in a stylized font, with the "F" in red and the rest of the letters in blue. Above the word "Flexicare" is the phrase "TOTAL QUALITY - TOTAL CARE" in red. The logo is simple and modern, and it conveys a sense of quality and care.
| Non-clinical Test<br>Results | Verification tests were performed to establish the safety and efficacy of Flexicare's Autofill<br>Humidification Chamber. These Non-clinical tests included Visual inspection/comparison,<br>Dimensional inspection, Internal Volume, resistance to flow, conical connector<br>compliance, connector strength, leaking, tensile strength, compliance, humidification<br>output, shelf life verification, Electromagnetic Capability, Electrical Safety and<br>Biocompatibility. Testing demonstrated that the relevant features, design and<br>performance of each manufacturer's device are substantially equivalent. |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion | Flexicare's Autofill Humidification Chamber is considered to be substantially equivalent to<br>the Fisher & Paykel MR290 Humidification Chamber. The comparison of features,<br>performance, materials and intended use demonstrate that Flexicare's Autofill<br>Humidification Chamber is as safe and effective as the predicate device for its intended<br>purpose. |
Summary of performance Testing: Flexicare's Heated Wire Breathing Systems have been evaluated in accordance with standards listed in table:
| Test | Standard / Pre-Determined<br>Acceptance Criteria | Outcome |
|------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|---------|
| Visual inspection | Pre-Determined Acceptance Criteria | Pass |
| Dimensional inspection | Pre-Determined Acceptance Criteria | Pass |
| System internal volume | Pre-Determined Acceptance Criteria | Pass |
| Wire resistance | Pre-Determined Acceptance Criteria | Pass |
| Tubing resistance to flow | Pre-Determined Acceptance Criteria | Pass |
| MR850 start-up test/ system compatibility check | Pre-Determined Acceptance Criteria | Pass |
| Means of connection | ISO 5367:2000 | Pass |
| Tubing resistance to flow | | |
| Increase in flow with bending | | |
| Leakage | | |
| Compliance | | |
| Packaging Pouch Integrity | | |
| | ASTM F1886-09<br>ASTM F88-09<br>ASTM F1929-12 | Pass |
| Conical Connector compliance | ISO 5356-1:2004 | Pass |
| Leak testing | | |
| Drop testing | | |
| Cytotoxicity, Irritation, Sensitization, Systemic<br>Toxicity, Genotoxicity, Implantation, Sub-Acute<br>Toxicity | | |
| | 10993-10:2010<br>10993-5:2009<br>10993-3:2014<br>10993-6:2009 | Pass |
| Testing to ISO 5367, temp probe tensile testing,<br>temp port leaking | ISO 8185:2007 | Pass |
| Electromagnetic capability & Electrical safety<br>testing | BS EN 60601-1:2006<br>BS EN 60601-1-2:2007<br>IEC 60601-1:2005 +CORR.1:2006<br>CORR. 2:2007 | Pass |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image is a logo for "flexicare". The logo has the word "flexicare" in blue, with the first letter "f" stylized in red. Below the word "flexicare" is the phrase "TOTAL QUALITY - TOTAL CARE" in red.
All Samples passed the performance testing when tested against methods and criteria from both pre-determined acceptance criteria methods and relevant FDA Recognized standards. The results of this testing show that Flexicare Heated Wire Breathing Systems pass all performance tests and performs at least as well as the marketed predicate devices.
Although very similar in design and function there are some differences, as described below, between the Flexicare Heated Wire Breathing Systems and the predicate devices from Fisher & Paykel.
The Flexicare Heated Wire Breathing System has a corrugated inspiratory limb and a smoothbore expiratory limb whilst the Fisher and Paykel device has corrugated limbs for both inspiratory and expiratory.
Another difference between Flexicare's device and its predicate device from Fisher & Paykel is the colour of tubing, connectors and accessories. However, these differences in colour are only due to individual company branding/marketing.
The overall conclusion from the comparison testing is that Flexicare's Heated Wire Breathing Systems are considered to be substantially equivalent to that of the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.