The RT380/RT385 'Adult Evaqua 2' dual-heated breathing circuits are intended to deliver humidified breathing gases for administration to an adult patient. The RT385 differs only from the RT380 in that the RT385 is supplied with a ventilator pressure line and connector port. Gases available for medical use do not contain sufficient moisture and may damage or irritate the respiratory tract, or desiccate secretions of patients whose supraglottic airways have been bypassed. Thus humidified gases via a heated breathing circuit may be indicated for patients requiring mechanical ventilation, positive pressure breathing assistance, or general medical gases. These gases may be delivered by facemask or through bypassing the upper airways via an endotracheal tube.
Device Story
RT380/RT385 'Adult Evaqua 2' are dual-heated breathing circuits for adult patients; function as conduits for humidified breathing gases; integrate into respiratory humidification systems. Circuits feature inspiratory and expiratory limbs with embedded heater wires to minimize condensate formation. Expiratory limb utilizes water vapor permeable 'Evaqua 2' technology; foaming technique creates tiny air bubbles in tube walls for thermal insulation and structural robustness. RT385 includes ventilator pressure line and connector port. Used in clinical settings for patients on mechanical ventilation or positive pressure support. Healthcare providers connect circuits to humidification systems to prevent respiratory tract damage or secretion desiccation. Benefits include consistent gas temperature and reduced condensate accumulation. Verified for 14 days continuous use.
Clinical Evidence
No clinical data. Safety and efficacy established via bench testing including condensate performance, humidity performance, pneumatic testing, single fault conditions, patient leakage current, enthalpy, connector strength, duration of use (14 days), shelf life (5 years), and biocompatibility.
Technological Characteristics
Dual-heated breathing circuit; corrugated polymer tubing; heater wires in inspiratory and expiratory limbs; water vapor permeable expiratory limb; foaming technique for thermal insulation and wall robustness; 14-day duration of use; 5-year shelf life.
Indications for Use
Indicated for adult patients requiring mechanical ventilation, positive pressure breathing assistance, or general medical gases via facemask or endotracheal tube to deliver humidified breathing gases and maintain inspired gas temperature.
Regulatory Classification
Identification
A breathing system heater is a device that is intended to warm breathing gases before they enter a patient's airway. The device may include a temperature controller.
Predicate Devices
RT265 and RT266 Dual Heated Infant Breathing Circuit (K103767)
Submission Summary (Full Text)
{0}------------------------------------------------
### 5 510(k) Summary
# DEC 0 6 2012
Website: www fohcare.com
| Contact person | Robert Petry |
|---------------------|---------------------------------------------------------------------------------------------------------|
| Date prepared | 06 August 2012 |
| Trade name | RT380 and RT385 'Adult Evaqua 2' Dual Heated Breathing Circuits <sup>1</sup> |
| Common name | Heated Breathing Circuit |
| Classification name | Breathing System Heater<br>Class II (21 CFR § 868.5270), product code BZE |
| Predicate device | K103767 Fisher & Paykel Healthcare<br>RT265 and RT266 Dual Heated Infant Breathing Circuit <sup>2</sup> |
1 Referred to throughout this document as RT380/RT385
2 Referred to throughout this document as RT265/RT266
#### 5.1 Description
The RT380/RT385 'Adult Evaqua 2' dual heated breathing circuits are classified as 'Breathing System Heater' according to 21 CFR §868.5270.
The RT380/RT385 breathing circuits form part of a respiratory humidification system in which the inspiratory limb delivers humidified gas to the patient and the expiratory limb carries the Heater wires in the inspiratory and expiratory limbs expired gas away from the patient. minimise the formation of condensate.
#### 5.2 Intended use
The RT380/RT385 'Adult Evaqua 2' dual-heated breathing circuits are intended to deliver humidified breathing gases for administration to an adult patient. The RT385 differs only from the RT380 in that the RT385 is supplied with a ventilator pressure line and connector port. Gases available for medical use do not contain sufficient moisture and may damage or irritate the respiratory tract, or desiccate secretions of patients whose supraglottic airways have been bypassed. Thus humidified gases via a heated breathing circuit may be indicated for patients requiring mechanical ventilation, positive pressure breathing assistance, or general medical gases. These gases may be delivered by facemask or through bypassing the upper airways via an endotracheal tube.
{1}------------------------------------------------
#### 5.3 Indications for use
The RT380 and RT385 'Adult Evaqua 2' dual-heated breathing circuits are intended as conduits of breathing gas for ventilation of adult patients, and to maintain the temperature of humidified inspired gas.
#### Technological characteristics comparison 5.4
The new adult breathing circuits (RT380/RT385) have been developed based on the cleared RT265/RT266 infant breathing circuit device. The RT380/RT385 are dual-heated adult breathing circuits that, like the predicate device (RT265/RT266), utilize corrugated polymer tubing, with a heater wire in both the inspiratory and expiratory limb.
The intended use of the RT380/RT385 is the same as that of the predicate device, differing only in the population indicated.
The RT380 and RT385 have a water vapor permeable expiratory limb. Permeability allows water vapor to pass into the environment, minimizing the amount of condensate that might form within the expiratory limb and also minimizing the amount of humidity delivered to a ventilator.
The inspiratory and expiratory limbs of the RT380/RT385 are made using a foaming technique that creates tiny air bubbles in the tube wall. In the case of the inspiratory limb the air bubbles improve the limb thermal insulation properties and thus reduce condensate formation. In the case of the expiratory limb, the air bubbles allow a thick tube wall for robustness against physical damage without compromising the water vapor permeability.
The RT380 and RT385 use the same expiratory limb technology and construction ('Evaqua 2') as the RT265/RT266 expiratory limb, but in an adult sized circuit.
Duration of use for the RT380/RT385 has been verified and labeled for 14 days continuous use, compared with 7 days for the predicate device.
Both the RT380/RT385 and the RT265/RT266 predicate device have been verified and labeled for a shelf life of 5 years.
#### Non-clinical tests 5.5
Verification tests were performed to establish the safety and efficacy of the RT380 and RT385. These included condensate performance, humidity performance, pneumatic testing, single fault conditions, patient leakage current, enthalpy, connector strength, duration of use testing, shelf life verification and biocompatibility. Testing was performed to the same standards as used for the predicate device.
#### 5.6 Conclusion
The RT380 and RT385 are considered to be substantially equivalent to the RT265/RT266 breathing circuit. The comparison of features, performance, materials and intended use demonstrate that the RT380 and RT385 'Adult Evaqua 2' dual-heated breathing circuits are safe and effective for their intended purpose.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. In the center of the logo is an abstract symbol that resembles an eagle or bird in flight.
## Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
# December 6, 2012
Mr. Robert Petry Regulatory Affairs Specialist Fisher & Paykel Healthcare Limited 15 Maurice Paykel Place, East Tamaki Auckland, New Zealand 2013
Re: K122432 ·
Trade/Device Name: RT380 and RT385 'Adult Evaqua 2' Dual Heated Breathing Circuits Regulation Number: 21 CFR 868.5270 Regulation Name: Breathing System Heater Regulatory Class: II Product Code: BZE Dated: November 5, 2012 Received: November 6, 2012
Dear Mr. Petry:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Petry
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Kwame O. Ulmer
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use Statement 4
510(k) Number
K122 432
Device Name
Fisher & Paykel Healthcare RT380 and RT385 'Adult Evaqua 2' Dual Heated Breathing Circuits
The RT380 and RT385 'Adult Evaqua 2' dual-heated breathing circuits are intended as conduits of breathing gas for ventilation of adult patients, and to maintain the temperature of humidified inspired gas.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
Please do not write below this line - continue on another page if needed
Concurrence of CDRH, Office of Device Evaluation (ODE)
Albert E.
Moyal
Digitally signed by Albert E. Moyal DN: <=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, cn=Albert E. Moyal, 0.9.2342.19200300.100.1.1=13000593331 Date: 2012,12.06 08:29:26 -05'00"
(Division Sign-Off) Division of Anesthestology, General Hospital Infection Control, Dental Devices
510(k) Number:_K122432
4-1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.