K110398 · Respironics, Inc. · BZE · Jun 1, 2011 · Anesthesiology
Device Facts
Record ID
K110398
Device Name
RESPIRONICS DISPOSABLE HEATED WIRE CIRCUITS
Applicant
Respironics, Inc.
Product Code
BZE · Anesthesiology
Decision Date
Jun 1, 2011
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 868.5270
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The disposable heated wire circuit is a heated wire breathing circuit intended to provide warmed and/or humidified breathing gases before they enter a patient's airway. The disposable heated wire circuit is indicated for use by a single adult or pediatric patient in the home, hospital and/or institutional settings. It may be used for both invasive and non-invasive ventilation.
Device Story
Disposable heated wire breathing circuit; 15 or 22 mm corrugated plastic tubing with integrated heater wire harness. Input: breathing gases from compatible heated humidifier (e.g., Fisher & Paykel HC500). Operation: voltage applied to heater wires; wire resistance dissipates heat through tube wall into gas lumen; maintains gas temperature at or above dew point to prevent condensation/pooling. Output: warmed/humidified gas delivered to patient airway. Used in home, hospital, nursing, extended care, and sleep labs. Operated by patients, caregivers, nurses, respiratory therapists, or physicians. Enhances patient comfort and prevents airway irritation/desiccation by optimizing gas temperature and humidity. Compatible with standard ventilator interfaces and supplemental oxygen.
Clinical Evidence
No clinical data. Bench testing only. Testing included mechanical, electrical, and temperature accuracy under environmental conditions, resistance to flow, compliance, compressible volume, and wire resistance. Biocompatibility evaluated per ISO 10993-1.
Technological Characteristics
Corrugated plastic tubing (15/22 mm); integrated heater wire harness (electrically operated). Materials ISO 10993 compliant. Interfaces with standard male connectors. Non-sterile, single-patient-use. Complies with ISO 5367 (breathing tubes), ISO 5356-1 (connectors), ISO 8185 (humidification systems), and IEC 60601-1 (electrical safety).
Indications for Use
Indicated for single adult or pediatric (>= 5kg) patients requiring invasive or non-invasive ventilation in home, hospital, or institutional settings to deliver warmed/humidified breathing gases.
Regulatory Classification
Identification
A breathing system heater is a device that is intended to warm breathing gases before they enter a patient's airway. The device may include a temperature controller.
Predicate Devices
Plastiflex Healthcare Hybernite Rainout Control System (K100104)
Intersurgical Heated Wire Breathing System (K092129)
{0}------------------------------------------------
# KID348
# Respironics Disposable Heated Wire Circuits
JUN - 1 2011 Premarket Notification -- Abbreviated 510(k)
# Section 5.0 510(k) Summary
17
# Administrative Information and Device Identification
:
| Name and address of the submitter, contract manufacturer and sponsor of the 510(k) submission: | Submitter: |
|------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Respironics, Inc. |
| | 1740 Golden Mile Highway |
| | Monroeville, PA 15146 |
| | Fax: 724-387-7490 |
| | Contract Manufacturer: |
| | WILAmed GMBH |
| | Aurachhoehe 5 7 |
| | Kammerstein, Bayern, 91126, Germany |
| | Fax: 049-9178-996778 |
| | United States Agent: |
| | Regulatory Insight, Inc. |
| | Kevin Walls |
| | 5401 S. Cottonwood CT |
| | Greenwood Village, CO 80121 USA |
| | Office: 720-962-5412 |
| | Fax: 720-962-5413 |
| | Email: kevin@reginsight.com |
| FDA registration number of the manufacturer of the new device: | Submitter: |
| | 2518422 (Establishment Registration Number) |
| | Contract Manufacturer: |
| | 9615827 (Establishment Registration Number) |
| | 10033146 (Owner/Operator Number) |
| Official contact person for all correspondence: | Joseph E. Olsavsky |
| | Sr. Regulatory Affairs Manager |
| | Philips Respironics |
| | 1740 Golden Mile Highway |
| | Monroeville, PA 15146 |
| | Office: 724-387-7562 |
| | Fax: 724-387-7490 |
| | Email: joseph.olsavsky@philips.com |
| Date Prepared: | February 10, 2011 |
| Device Name: | Respironics Disposable Heated Wire Circuits |
| Trade or Proprietary name of new | Disposable Pediatric, Heated Wire Active Circuit USA |
| device: | Disposable Adult, Heated Wire Active Circuit USA<br>Disposable Pediatric, Heated Wire Passive Circuit USA<br>Disposable Adult, Heated Wire Passive Circuit USA |
| Common or usual name of the<br>device: | Heated Breathing Tube |
| Philips/Respironics model number: | 1076043, 1076044, 1076045, 1076046 |
| Classification of new device: | Class II |
| Classification of the predicate<br>device: | Class II |
| Classification Panel: | Anesthesiology |
| Panel Code: | BZE - Breathing system, heater |
| CFR Regulation Number: | 21 CFR 868.5270<br>a) Identification. A breathing system heater is a device<br>that is intended to warm breathing gases before they<br>enter a patient's airway. The device may include a<br>temperature controller.<br>(b) Classification. Class II (performance standards). |
| Predicate Device Name(s) and<br>510(k) numbers: | 1. Plastiflex Healthcare Hybernite Rainout Control<br>System - K100104 date of concurrence 04/14/2010)<br>2. Intersurgical Heated Wire Breathing System -<br>K092129 (date of concurrence 05/18/2010)<br>3. Fisher & Paykel Respiratory Humidifier - K983112<br>(date of concurrence 11/10/1996) |
| Reason for submission: | New Device |
{1}------------------------------------------------
# Description of Device:
The Respironics heated wire breathing circuit is a disposable device compromised of 15 or 22 mm corrugated plastic tubing, and 22 mm plastic tube connectors, and an electrical heater wire harness subassembly. The Respironics disposable heated wire circuit consists of a single limb single lumen smooth interior tube (15 or 22 mm diameter) containing 2 heater wires that are located in the tubing construction of which the tube is formed having a supporting structure; the tube is spiral and the wire has a single loop form. After the gas is warmed and humidified in the water chamber it is delivered through the breathing circuit to the patient. Heating of the breathing tube is provided and controlled by a compatible heated humidifier. This disposable heated wire circuit is designed to be used with the HC500 Fisher and Paykel humidifier. The
{2}------------------------------------------------
heating wires are physically separated from the lumen of the tubing. As such, there is no direct contact between the heating wires and the air flow. When a voltage is applied, a current flows through the heating wires. Due to the wire resistance, heat is dissipated through the wall of the tube construction into the air flow in the lumen of the tubing.
There are four types of disposable breathing circuits:
- l . Disposable Pediatric, Heated Wire Active Circuit
- 2. Disposable Adult, Heated Wire Active Circuit
- 3. Disposable Pediatric, Heated Wire Passive Circuit
- 4. Disposable Adult, Heated Wire Passive Circuit
The purpose of the heated wire breathing circuits is to maintain or raise the gas temperature to or above the dew point thus reducing or eliminating water condensation and/or pooling of water in the breathing circuit.
The Respironics disposable heated wire breathing circuit has standard cuffs on both the machine-end cuff and mask-end cuff. As such, the disposable heated wire breathing circuit can be connected to heated humidifiers and flow generators that have standard male outlet connectors. The Respironics disposable heated wire breathing circuit is intended for incorporation into ventilator devices and is intended to act as a conduit for the breathing gasses delivered from the humidifier to the patient. It can also be used in conjunction with supplemental Oxygen.
The environment of use for the disposable heated wire breathing circuit will be for in the home, institutional, nursing, extended care, and clinical sleep lab settings.
Operators of the disposable heated wire breathing circuit are expected to be: Patients, Lay caregivers (includes family members of patients and aides), Nurses, Respiratory therapists, Physicians and Home care providers.
The disposable heated wire breathing circuit is intended to be used with ventilators that provide both pressure support and volume modes of therapy.
Other accessories such as water traps, etc. can be added in to the overall assembly creating different product variations.
# Comparison of Device Technological Characteristics to Predicate Devices:
The Respironics Disposable Heated Wire Circuits have the following similarities to those predicate devices listed in this submission which previously received 510(k) concurrence; the Respironics Disposable Heated Wire Circuits:
- Has the same intended use, .
- Uses the same operating principle, .
{3}------------------------------------------------
- Incorporates the same basic heated wire breathing circuit design elements for use with . ventilator devices including physical interfaces; and performance characteristics;
- . Incorporates similar materials & is ISO 10993 compliant;
- ls manufactured utilizing similar manufacturing processes; and .
- Complies with similar electrical, mechanical, chemical and performance standards .
According to FDA's Draft Reviewer Guidance for Premarket Notification Submissions -Anesthesiology and Respiratory Devices Branch; Division of Cardiovascular, Respiratory, and Neurological Devices (November 1993), the following characteristics for the submitted device and consistent with the predicate devices are identified below:
- The Respironics disposable heated wire breathing circuit is not an implantable device. ●
- The Respironics disposable heated wire breathing circuit is intended for life support or . life sustaining applications.
- The Respironics disposable heated wire breathing circuit is not sold as sterile. ●
- The Respironics disposable heated wire breathing circuit is a single-patient-use device. .
- The Respironics disposable heated wire breathing circuit must be prescribed by a . physician.
- The Respironics disposable heated wire breathing circuit does not contain a drug or . biological as a component.
- The Respironics disposable heated wire breathing circuit is not a kit. .
- The Respironics disposable heated wire breathing circuit is not software driven. .
- The Respironics disposable heated wire breathing circuit is electrically operated_ .
The intended use of the Respironics heated wire breathing circuit is comparable to the referenced predicate dcvices. The comparison of the data shows similar values for the key performance characteristics. The Respironics disposable heated wire circuit shows similar values for compliance, volume, resistance to flow, wire resistance and tube length and connectivity.
The reason for the Abbreviated 510(k) promarket notification submission for Respironics Disposable Heated Wire circuit is that this is a new device.
The new device as designed and manufactured does not raise any new issues of safety and effectiveness.
{4}------------------------------------------------
emarket Notification -- Abbreviated 510(k
he following table compares the Respironics Disposable Heated Wire Circuits with the legally marketed predicate dev
Table 12-1 Comparison of Respirants Disposable Heated Circuits with Platified Croup NV Hybernite Rainout Cyntrol System
| Feature/Function | New device: Respironics<br>Disposable Heated Wire<br>Circuits | Predicate Device: Plastiflex<br>Healthcare Hybernite<br>Rainout Control System<br>(K100104) | Predicate Device:<br>Intersurgical Heated Wire<br>Breathing System<br>(K092129) | Fisher & Paykel<br>Respiratory Humidifier<br>(K983112) | Similarity/<br>Difference | Impact on<br>safety and<br>effectiveness |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| Common or usual<br>name of the device | Heated Breathing Tube,<br>Breathing system heater | Heated Breathing Tube,<br>Breathing system heater | Heated Breathing Tube,<br>Breathing system heater | Respiratory humidifier with<br>accessories (including<br>heated wire circuits) | Similar.<br>F&P<br>clearance<br>covers a<br>system<br>(humidifier<br>+ heated<br>wire<br>circuit) | No impact on<br>safety and<br>effectiveness |
| Classification | Class II | Class II | Class II | Class II | Same | No impact on<br>safety and<br>effectiveness |
| Classification<br>Panel | Anesthesiology | Anesthesiology | Anesthesiology | Anesthesiology | Same | No impact on<br>safety and<br>effectiveness |
| Product Code: | BZE- heater, breathing<br>system w/wo controller | BZE- heater, breathing<br>system w/wo controller | BZE- heater, breathing<br>system w/wo controller | BTT - humidifier,<br>respiratory gas | Same | No impact on<br>safety and<br>effectiveness |
| CFR Regulation<br>Number | 21 CFR 868.5270<br>a) Identification. A<br>breathing system<br>heater is a device<br>that is intended to<br>warm breathing<br>gases before they<br>enter a patient's | 21 CFR 868.5270<br>a) Identification. A<br>breathing system<br>heater is a device that<br>is intended to warm<br>breathing gases<br>before they enter a<br>patient's airway. The | 21 CFR 868.5270<br>a) Identification. A<br>breathing system<br>heater is a device<br>that is intended to<br>warm breathing<br>gases before they<br>enter a patient's | 21 CFR 868.5270<br>a) Identification. A<br>breathing system<br>heater is a device<br>that is intended to<br>warm breathing<br>gases before they<br>enter a patient's | Same | No impact on<br>safety and<br>effectiveness |
| 10(k) numbers: | | | | | | |
| Intended Use | airway. The device<br>may include a<br>temperature<br>controller. | device may include a<br>temperature<br>controller. | airway. The device<br>may include a<br>temperature<br>controller. | airway. The device<br>may include a<br>temperature<br>controller. | | |
| | (b) Classification. Class II<br>(performance standards). | (b) Classification. Class II<br>(performance standards). | (b) Classification. Class II<br>(performance standards). | (b) Classification. Class II<br>(performance standards). | | |
| | Not yet assigned<br>The disposable heated wire<br>circuit is a heated wire<br>breathing circuit intended to<br>provide warmed and/or<br>humidified breathing gases<br>before they enter a patient's<br>airway. The disposable<br>heated wire circuit is<br>indicated for use by a single<br>adult or pediatric patient in<br>the home, hospital and/or | K100104<br>The Hybernite Rainout<br>Control System is a heated<br>breathing circuit intended to<br>provide warmed and/or<br>humidified breathing gases<br>before entering the patient<br>airway. The Hybernite device<br>is intended for incorporation<br>into CPAP (continuous<br>positive airway pressure)<br>devices and is intended to act | K092129<br>Breathing system heaters<br>are defined as a device that<br>is intended to warm<br>breathing gases before they<br>enter the patient's airway. | 21 CFR 868.5450<br>Respiratory gas humidifier.<br><br>(a)Identification. A<br>respiratory gas humidifier is<br>a device that is intended to<br>add moisture to, and<br>sometimes to warm, the<br>breathing gases for<br>administration to a patient.<br>Cascade, gas, heated, and<br>prefilled humidifiers are<br>included in this generic type<br>of device. | K983112<br>The Fisher & Paykel<br>Healthcare MR 850<br>Humidifier is a Respiratory<br>Gas Humidifier as per 73<br>BTT, 21 CFR 868.5450. It<br>is intended to add moisture<br>to and warm breathing gases<br>for administration to a<br>patient. The MR850 is<br>intended to be used to warm<br>and add humidity to gases | N/A<br>No impact on<br>safety and<br>effectiveness |
| | | | | (b)Classification. Class II<br>(performance standards. | | N/A<br>Similar |
{5}------------------------------------------------
Premarket Notification -- Abbreviated 510(k)
Tab 5 Page 6 of 14
{6}------------------------------------------------
| Respironics Disposable Heated Wire Circuits | Premarket Notification -- Abbreviated 510(k) | |
|------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| be used for both invasive and<br>non-invasive ventilation. | gasses delivered form the<br>humidifier to the patient.<br>After the gas is warmed and<br>humidified in the humidifier, it | requiring mechanical<br>ventilation, positive<br>pressure breathing<br>assistance or general |
.
| be used for both invasive and<br>non-invasive ventilation. | gasses delivered form the<br>humidifier to the patient.<br>After the gas is warmed and<br>humidified in the humidifier, it<br>is delivered through the heated<br>tubing to the patient. The<br>purpose of the Hybernite<br>Rainout Control System is to<br>maintain or raise the gas<br>temperature to or above the<br>dew point (of the air exiting<br>the humidifier) reducing or<br>eliminating water<br>condensation and/or pooling<br>of water in the breathing<br>circuit, and problems<br>associated with such. The<br>Hybernite is indicated for use<br>in the home or sleep-lab<br>setting by a single adult<br>patient. It can also be used in<br>conjunction with supplemental<br>Oxygen. Hybernite is<br>indicated for non-invasive<br>ventilation. | requiring mechanical<br>ventilation, positive<br>pressure breathing<br>assistance or general<br>medical gases. Gases<br>available for medical use do<br>not contain sufficient<br>moisture and may damage<br>or irritate the respiratory<br>tract, or desiccate secretions<br>of patients whose<br>supraglottic airways have<br>been bypassed. Heat is used<br>to increase the water output<br>of the humidifier. | Heated breathing tubes are<br>also utilized in order to<br>increase operating<br>efficiency and reduce<br>excessive water and heat<br>loss. | | | | | heating wires | heating wires | heating wires | | | safety and<br>effectiveness |
|------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|---------|---------------------------------------------|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|---------------------------------------------|
| Anatomical Sites | Invasive & Non-invasive | Non-invasive | Not Specified<br>(intended use includes any<br>patient using a heated<br>humidifier which would be<br>inclusive of Invasive<br>patients) | Invasive & Non-invasive | Same | No impact on<br>safety and<br>effectiveness | energy used and or<br>delivered | current flows through the<br>heating wires. Due to the wire<br>resistance, heat is dissipated<br>through the wall of the tube<br>construction into the air flow<br>in the lumen of the tubing. As<br>a result, the air passing<br>through the tubing is warmed<br>to or above the dew point (of<br>the air existing the humidifier)<br>reducing or eliminating water<br>condensation and/or pooling<br>of water in the breathing<br>circuit. | Due to the wire resistance,<br>heat is dissipated through the<br>wall of the tube construction<br>into the air flow in the lumen<br>of the tubing. As a result, the<br>air passing through the tubing<br>is warmed to or above the dew<br>point (of the air existing the<br>humidifier) reducing or<br>eliminating water<br>condensation and/or pooling<br>of water in the breathing<br>circuit. | Rising of the delivered gas<br>temperature from 37 to 40C<br>increases its enthalpy. | Due to the wire resistance,<br>heat is dissipated through<br>the wall of the tube<br>construction into the air<br>flow in the lumen of the<br>tubing. As a result, the air<br>passing through the tubing<br>is warmed to or above the<br>dew point (of the air<br>existing the humidifier)<br>reducing or eliminating<br>water condensation and/or<br>pooling of water in the<br>breathing circuit. | Same | No impact on<br>safety and<br>effectiveness |
| Target Patient<br>Population | Pediatric (>= 5kg) to Adult | Adult | Any patient using a heated<br>humidifier | Any patient using a heated<br>humidifier | Same | No impact on<br>safety and<br>effectiveness | Reusable | The raising of the gas<br>temperature does not exceed<br>41C. | The raising of the gas<br>temperature does not exceed<br>40C. | | The raising of the gas<br>temperature does not exceed<br>41C. | Similar | No impact on<br>safety and<br>effectiveness |
| Environment of<br>Use | Home, Institution/Hospital<br>Setting, Extended Care and<br>Clinical Sleep settings | Home or Sleep Lab Setting | Hospital Setting | Home, Institution/Hospital<br>Setting, Extended Care and<br>Clinical Sleep settings | Same | No impact on<br>safety and<br>effectiveness | | Single Patient Use -<br>Reusable | Single Patient Use - Reusable<br>Cleaning Regime: Mild soap<br>and water after use. | Not Specified | Single Patient Use --<br>Reusable | Similar | No impact on<br>safety and<br>effectiveness |
| Sterility | Non-Sterile | Non-Sterile | Not Specified | Non-Sterile | Similar | No impact on<br>safety and<br>effectiveness | | | | | | | |
{7}------------------------------------------------
Premarket Notification -- Abbreviated 510(k)
{8}------------------------------------------------
Premarket Notification -- Abbreviated 510(k)
| Compatible with<br>multiple<br>humidifiers<br>(standard<br>connectors) | For use with F&P 500 | Universal | For use with Intersurgical<br>model humidifiers | For use with F&P model<br>humidifiers | Similar | No impact on<br>safety and<br>effectiveness |
|------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|---------------------------------------------|
| Breathing Gases<br>Specified | Air & Supplemental Oxygen | Air & Supplemental Oxygen | No t Specified | Air & Supplemental<br>Oxygen | Similar | No impact on<br>safety and<br>effectiveness |
| Standard<br>Breathing Circuit<br>Polymeric<br>materials | Yes | Yes | Yes | Yes | Similar | No impact on<br>safety and<br>effectiveness |
| Tube Diameter | 15 and 22 mm configurations | 15 and 22 mm configurations | 22mm | 15 and 22 mm<br>configurations | Similar | No impact on<br>safety and<br>effectiveness |
| Tube Length | 1.83 m | ~1.5 m | ~1.5 m | 1.5 m | Same | No impact on<br>safety and<br>effectiveness |
| Power Source | Incorporated (humidifier<br>controlled) | Separate | Incorporated (humidifier<br>controlled) | Incorporated (humidifier<br>controlled) | Similar | No impact on<br>safety and<br>effectiveness |
| Heating Wire | Encased | Encased | Encased | Encased | Same | No impact on<br>safety and<br>effectiveness |
| Wire Resistance<br>(ohms) | 30 +/- 5% | Not specified | 14.7 Ins & 11.9 Exp | Not specified | Similar | No impact on<br>safety and<br>effectiveness |
| Active Controller | No – Humidifier Controlled | No – Power source controlled | No - Humidifier Controlled | No – Humidifier Controlled | Similar | No impact on<br>safety and<br>effectiveness |
| Standards of<br>Conformity/Perfor<br>mance | ISO 5367 – Breathing tubes<br>intended for use with<br>anesthetic apparatus and<br>ventilators<br>ISO 5356-1- Anesthetic and<br>respiratory equipment –<br>Conical connectors: Part 1: | ISO 5367 – Breathing tubes<br>intended for use with<br>anesthetic apparatus and<br>ventilators<br>ISO 5356-1 - Anesthetic and<br>respiratory equipment –<br>Conical connectors: Part 1: | ISO 5367 – Breathing tubes<br>intended for use with<br>anesthetic apparatus and<br>ventilators<br>ISO 5356-1 – Anesthetic<br>and respiratory equipment –<br>Conical connectors: Part 1: | ISO 5367 – Breathing tubes<br>intended for use with<br>anesthetic apparatus and<br>ventilators<br>ISO 5356-1 – Anesthetic<br>and respiratory equipment –<br>Conical connectors: Part 1: | Similar | No impact on<br>safety and<br>effectiveness |
Tab 5 Page 9 of 14
.
{9}------------------------------------------------
ت
remarket Notification -- Abbreviated 510(k
| ISO 8185 – Respiratory<br>humidification systems –<br>requirements (as applicable<br>to breathing tubes). | IEC 60601-1: Medical<br>electrical equipment – Part 1:<br>General requirements for<br>basic safety and essential<br>performance | IEC 60601-1-2 Medical<br>Electrical Equipment Part 1-<br>2: Collateral Standard:<br>Electromagnetic<br>Compatibility –<br>Requirements and Tests | AAMI/ANSI/ISO 10993-1<br>Biological Evaluation of<br>Medical Devices – Part 1:<br>Evaluation and Testing | ISO 14971 Medical devices -<br>Application of risk<br>management to medical<br>devices. | ISO 5367 compliant | | ISO 10993 | | No impact on<br>safety and<br>effectiveness |
|-----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|--------------------|-----------------------------|---------------------|------------------|---------------------------------------------|
| ISO 8185 – Respiratory<br>humidification systems –<br>requirements (as applicable to<br>breathing tubes). | IEC 60601-1: Medical<br>electrical equipment – Part 1:<br>General requirements for<br>basic safety and essential<br>performance | IEC 60601-1-2 Medical<br>Electrical Equipment Part 1-<br>2: Collateral Standard:<br>Electromagnetic<br>Compatibility –<br>Requirements and Tests | AAMI/ANSI/ISO 10993-1<br>Biological Evaluation of<br>Medical Devices - Part 1:<br>Evaluation and Testing | | ISO 5367 compliant | | ISO 10993 | | Same |
| ISO 8185 – Respiratory<br>humidification systems –<br>requirements (as applicable<br>to breathing tubes). | IEC 60601-1: Medical<br>electrical equipment – Part<br>1: General requirements for<br>basic safety and essential<br>performance | IEC 60601-1-2 Medical<br>Electrical Equipment Part 1-<br>2: Collateral Standard:<br>Electromagnetic<br>Compatibility –<br>Requirements and Tests | AAMI/ANSI/ISO 10993-1<br>Biological Evaluation of<br>Medical Devices - Part 1:<br>Evaluation and Testing | | ISO 5367 compliant | | ISO 10993 | | No impact on<br>safety and<br>effectiveness |
| ISO 8185 - Respiratory<br>humidification systems -<br>requirements (as applicable<br>to breathing tubes). | IEC 60601-1: Medical<br>electrical equipment - Part<br>1: General requirements for<br>basic safety and essential<br>performance | IEC 60601-1-2 Medical<br>Electrical Equipment Part 1-<br>2: Collateral Standard:<br>Electromagnetic<br>Compatibility -<br>Requirements and Tests | AAMI/ANSI/ISO 10993-1<br>Biological Evaluation of<br>Medical Devices - Part 1:<br>Evaluation and Testing | ISO 14971 Medical devices<br>- Application of risk<br>management to medical<br>devices. | ISO 5367 compliant | | ISO 10993 compliant | | Same |
| Compliance (mlpa) | | | | | | Resistance to Flow<br>(mb)² | Tube Volume (ml) | Biocompatibility | |
{10}------------------------------------------------
| 011180<br>1<br>) |
|--------------------------------------------------------------------------------------|
| t<br>1<br>1<br>t<br>1 |
| ۲۵<br>1<br>1<br>1 |
| 4<br>1<br>tet<br>A<br>3<br>1<br>l<br>t<br>)<br>1<br>1<br>T<br>2<br>1 |
| 1<br>0<br>1<br>1<br>1<br>C<br>3<br>ﻟ<br>t<br>1<br>(<br>1<br>Circ<br>2<br>1<br>1<br>1 |
| 1 |
| 1<br>1<br>C<br>1<br>1<br>+7 ><br>1<br>1 |
| 1100<br>2<br>1<br>1<br>o<br>C<br>3<br>t |
| 1<br>œ<br>t |
arket Notification -- Abbrev
| | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | The first and the count of the first to the country of the first and the first and the many of the first and the may be and | | | | . |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|----------------------|----------------------|------|---------------|
| Electrical Safe | IEC 60601-1 compliant | 91-1 compliant<br>IEC 6060 | EC 60601-1 compliant | EC 60601-1 compliant | Same | No impact or |
| | | | | | | safety and |
| | | | | | | effectiveness |
| | . | Annual And Annual Commend Commend of Children Annual Property And<br>1 | | | | |
ﮯ ﮯ
# of Comparison of Device Technological Characteristics to Predicate Dev
onics Disposable Heated Circuits are substantially equivalent to the predicate devices listed in this Summary and
s not raise any new issues of safety and effectiveness.
ddressed by the Reviewer's Checklist are unchanged from the predicate devices identified in this submitta
.0 – Substantial Equivalence Discussion.
# erformance Testing Summ
ce has been tested to appropriate ISO and IEC standards and other applicate requirements passing all test protocol
ics disposable heated wire circuit was designed and tested
Draft Reviewer Guidance for Premarket Notification Submissions - Anesthesiology and Respiratory Devices
of Cardiovascular,Respiratory, and Neurological Devices (November 199
Oraft Reviewer Guidance for Ventilators July 1995
as suggested by TDA's Noveller Denided "Reviewed with Premarket Nothical on Pomission Ancatesional Respiracy: Division Cariorscular, Ind Newsled I bevices…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.