K133057 · Westmed, Inc. · SDT · May 23, 2014 · Anesthesiology
Device Facts
Record ID
K133057
Device Name
VIBRALUNG ACOUSTICAL PERCUSSOR
Applicant
Westmed, Inc.
Product Code
SDT · Anesthesiology
Decision Date
May 23, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5665
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Vibralung Acoustical Percussor is indicated as an airway secretion clearance device that creates vibrations in the airways and as a lung expansion device that applies Positive Expiratory Pressure (PEP) as a patient breathes through the device. It may be used to promote bronchial drainage, airway clearance and expectoration. The Vibralung may be used simultaneously with aerosol drug delivery.
Device Story
Vibralung Acoustical Percussor provides airway clearance therapy (ACT) and lung expansion via Positive Expiratory Pressure (PEP). Device consists of a Treatment Control Unit (TCU) frequency generator and a hand-held transducer (HHT) containing an electro-acoustical speaker. HHT connects to patient via Y-adapter and mouthpiece. TCU generates 5–1,200 Hz frequencies, inducing oscillatory sound waves in the airway gas column to promote mucokinesis. Device supports simultaneous nebulized aerosol drug delivery. Used in hospital or home settings by patients or clinicians. Output is acoustic vibration; healthcare providers monitor patient sputum clearance and respiratory status. Benefits include improved bronchial drainage and secretion expectoration.
Clinical Evidence
Two clinical studies performed. Study 1 (n=10, CF patients): compared treatment with vs. without sound; results showed greater improvement in Lung Clearance Index (LCI) with sound, confirming safety. Study 2 (n=11): compared Vibralung to Vest predicate over 5 days; results showed equivalent performance in sputum clearance (wet/pellet/dry weight).
Technological Characteristics
Electro-acoustical transducer (audio speaker) driven by frequency generator. Frequency range 5–1,200 Hz. Y-adapter waveguide for mouthpiece interface. Variable resistor valve for PEP (up to 20 cmH2O). Materials utilized in prior Westmed devices. Electrical safety per IEC 60601-1, IEC 60601-1-11, IEC 62133; EMC per EN 5501.
Indications for Use
Indicated for patients with cystic fibrosis, COPD, asthma, lung diseases with secretory problems, or neuromuscular disease affecting cough ability. Intended for patients capable of reading or following verbal instructions. Contraindicated in patients with untreated air leaks, tension pneumothorax, bronchopleural fistula, recent hemoptysis, or pulmonary hemorrhage.
Regulatory Classification
Identification
A powered percussor is a device that is intended to transmit vibration through a patient's chest wall to aid in freeing mucus deposits in the lung in order to improve bronchial drainage and that may be powered by electricity or compressed gas.
{0}------------------------------------------------
| Page 1 of 12 | | | | |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
| | MAY 23 2014 | | | |
| Date Prepared | 23-May-14 | | | |
| Company | Westmed, Inc.<br>5580 South Nogales Highway | | | |
| | Tucson, AZ 85706 | | | |
| | Tel - 520-294-7987<br>Fax - 520-294-2780 | | | |
| Official Contact: | John McKinnon - CEO | | | |
| Proprietary or Trade Name: | Vibralung Acoustical Percussor | | | |
| Common/Usual Name: | Electric powered percussor | | | |
| Classification / CFR: | BYI / CFR 868.5665 / Class 2 | | | |
| Device: | Vibralung Acoustical Percussor | | | |
| Predicate Devices: | Dymedso - Frequencer - K063645 and K103176 (Primary)<br>DHD Healthcare (Smiths) - Acapella - K991561 and K002768<br>Medical Acoustics - Lung Flute - K091557 | | | |
Device Description: The proposed Vibralung Acoustical Percussor provides airway clearance therapy (ACT) and promotes mucokinesis by inducing vibration in airways by vibrating the column of gas in the airways at a variety of different frequencies. It is an acoustic device that induces oscillatory sound waves by means of an electro-acoustical transducer (audio speaker), referred to as the "Hand-held Transducer" or HHT, which is interfaced to the patient's airway through a mouthpiece. The transducer unit is connected to a frequency generator, referred to as the "Treatment Control Unit" or TCU, which is set to produce frequencies between 5 and 1,200 Hz in different ranges that are user-selectable.
The HHT directs sound waves into the patient's airways through a waveguide and a mouthpiece.
In addition, the Vibralung can provide Positive Expiratory Pressure (PEP) and may simultaneously provide nebulized aerosol drug delivery.
There are two different Y-adapters available that act as "waveguides" to direct sound waves from the audio speaker in the HHT to the mouthpiece that is held in the patient's mouth in the same manner as with other breathing treatment mouthpieces.
The purpose of the Y-adapters is two-fold:
- · To direct the sound waves into the airway opening, and
- · To provide separate pathways for inhalation and exhalation gas flow with minimal mixing and rebreathing of exhaled gas.
{1}------------------------------------------------
# 510(k) Summary Page 2 of 12 23-May-14
The Standard Y-adapter allows ambient breathing without aerosol therapy. It contains an ambient air intake port with a valve for one-way inspiratory flow while preventing exhalation through the port. The exhalation is directed through one branch of the Y-adapter and out to the room through a variable expiratory resistor that also provides Positive Expiratory Pressure (PEP).
The other Y-adapter (Aerosol Y-adapter) allows the device to be coupled to the airway opening (mouth) while interfaced with Westmed's Circulaire aerosol drug delivery system to enable aerosolized medication to be delivered concomitantly with ACT. In this configuration, inhalation and exhalation pathways plus PEP are incorporated into the Circulaire device.
#### Indications for Use:
The Vibralung Acoustical Percussor is indicated as an airway secretion clearance device that creates vibrations in the airways and as a lung expansion device that applies Positive Expiratory Pressure (PEP) as a patient breathes through the device. It may be used to promote bronchial drainage, airway clearance and expectoration. The Vibralung may be used simultaneously with aerosol drug delivery.
Patient Population - Cystic Fibrosis, COPD, asthma and lung diseases with secretory problems, patients with neuromuscular disease affecting the ability to effectively cough. Anyone who is able to read and / or follow verbal instructions
Environment of Use - Hospital and Home
Contraindications: ACT or use of the Vibralung Acoustical Percussor, especially with Positive Expiratory Pressure, may be contraindicated in patients who have untreated air leaks, tension pneumothorax, bronchopleural fistula, recent hemoptysis, or pulmonary hemorrhage because it may exacerbate those conditions. Prescribers should weigh benefits against risks in patients with these conditions.
Adverse Reactions: If the patient complains of dry or sore throat or mouth brought about by treatment with the Vibralung, consider adding nebulization with normal saline if it is not already being done. If the patient complains of sore mouth, jaw or teeth brought about by using the Vibralung device, the healthcare practitioners should assess the patient complains of dizziness or light-headedness, assess the patient for possible hyperventilation while using the device. If the patient appears to be hyperventilating, pause the treatment and coach the patient to alter their breathing pattern appropriately. Any other adverse reactions should be fully assessed before continuing therapy with the Vibralung Acoustical Percussor.
# Level of Patient Contact and Duration:
In accordance with G95-1 and ISO 10993-1 the Vibralung components are primarily considered as:
{2}------------------------------------------------
# 510(k) Summary Page 3 of 12 23-May-14
Components in the nebulized aerosol gas pathway
- External Communicating (indirect contact but in the gas pathway) .
- Tissue Contacting .
- . Permanent Duration of Use for those components in the nebulizer aeroso! path
Components in the gas pathway but NOT in the nebulized aerosol gas pathway
- External Communicating (indirect contact but in the gas pathway) ●
- Tissue Contacting .
- Limited Duration of Use (less than 24 hr of accumulated exposure) .
Components in Direct Contact patient contact
- Surface Contacting ●
- . Mucosal membrane
- . Limited duration of use (less than 24h)
All the materials in the Vibralung have been utilized in Westmed devices which have substantially equivalent levels of patient contact and duration of use.
# Predicate Device Comparison:
# Table 1: Predicates by Indications for Use and Airway Clearance technology
| Predicate | Airway<br>Clearance | PEP | Aerosol<br>delivery | Contact<br>with airway | Applied<br>During | Technology |
|---------------------------------------------|---------------------|-----|---------------------|------------------------|----------------------------|------------------------------------------------------|
| Frequencer<br>K063645<br>K100749<br>K103176 | X | - | - | External<br>Chest wall | Inhalation /<br>exhalation | Acoustical<br>power head<br>applied to<br>chest wall |
| Acapella<br>K991561<br>K002768 | X | X | X | Direct<br>Mouthpiece | Exhalation | Interrupted<br>airflow at a<br>frequency |
| Lung Flute<br>K091557 | X | X | - | Direct<br>Mouthpiece | Exhalation | Interrupted<br>airflow with<br>reed for<br>frequency |
{3}------------------------------------------------
# 510(k) Summary
# Table 2 - Substantial Equivalence Table - Frequencer
| Criteria | Vibralung | Dymedso Frequencer<br>K063645, K100749,<br>K103176 |
|---------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA Classification | BYI - Percussor, power-electric<br>868.5665 | BYI - Percussor, power-electric<br>868.5665 |
| Indications for Use | The Vibralung Acoustical Percussor is<br>indicated as an airway secretion clearance<br>device that creates vibrations in the<br>airways and as a lung expansion device<br>that applies Positive Expiratory Pressure<br>(PEP) as a patient breathes through the<br>device. It may be used to promote<br>bronchial drainage, airway clearance and<br>expectoration. The Vibralung may be<br>used simultaneously with acrosol drug<br>delivery. | The Frequencer provides airway<br>clearance therapy and promotes<br>bronchial drainage by inducing vibration<br>in the chest walls. This device is<br>intended to be a component of postural<br>drainage therapy by providing a<br>convenient method of external thorax<br>manipulation. |
| Other indications | Applies Positive Expiratory Pressure<br>(PEP)<br>Simultaneously used with nebulizer | No |
| Environments of use | Hospital and home settings | Hospital and home settings |
| Patient Population | Cystic Fibrosis, COPD, asthma and lung<br>diseases with secretory problems, patients<br>with neuromuscular disease affecting the<br>ability to effectively cough. Anyone who<br>is able to read and / or follow verbal<br>instructions. | Patients having respiratory ailments<br>which involves defective mucociliary<br>clearance, as is typical in patients<br>suffering from Cystic Fibrosis, chronic<br>bronchitis, bronchiectasis, ciliary<br>dyskinesia syndrome, asthma, muscular<br>dystrophy, neuromuscular degenerative<br>disorders, post-operative atelectasis and<br>thoracic wall defects. |
| Prescription/OTC | Prescription use | Prescription use |
| Principle of Operation | Acoustical generator uses sound waves to<br>vibrate the gas column in the airway at<br>various frequencies | Acoustical generator applied externally<br>to the chest that transmits sound through<br>the chest wall, vibrating the airways. |
| Technology employed | Loudspeaker controlled to apply a range<br>of frequencies | Loudspeaker controlled to apply a range<br>of frequencies |
| How it interfaces with<br>the patient | The patient uses a mouthpiece and the<br>acoustical sound is directed to the airway | In contact with the chest wall, but the<br>vibrations of the loudspeaker create a<br>vibration and frequency through the<br>chest wall to the airways |
| Patient Effort | Effort independent | Effort independent |
| Patient Breathing<br>Maneuver | Normal inspiratory & expiatory pattern | Normal inspiratory & expiatory pattern |
| Chest wall Frequency | N/A | 20 - 100 Hz |
| Airway Acoustical<br>Frequency | 5 - 1,200 Hz generated; up to 7,000 Hz<br>actual resulting from harmonics and<br>resonance | 20 - 65 Hz generated and applied to<br>external chest wall; up to 9,000 Hz<br>actual resulting from harmonics and<br>resonance |
| Criteria | Vibralung | DHD - Acapella<br>K991561<br>K002768 |
| FDA Classification | BYI - Percussor, power-electric<br>868.5665 | BWF - Incentive Spirometer<br>868.5690 |
| Indications for Use . | The Vibralung Acoustical Percussor is<br>indicated as an airway secretion clearance<br>device that creates vibrations in the<br>airways and as a lung expansion device<br>that applies Positive Expiratory Pressure<br>(PEP) as a patient breathes through the<br>device. It may be used to promote<br>bronchial drainage, airway clearance and<br>expectoration. The Vibralung may be<br>used simultaneously with aerosol drug<br>delivery. | Indicated for use as a PEP device.<br>Improves clearance of secretions, reduce<br>the need of postural drainage, facilitates<br>opening of airways, prevents or reverses<br>atelectasis.<br>It may be used simultaneously with<br>nebulized aerosol drug delivery. |
| Criteria | Vibralung | DHD - Acapella<br>K991561<br>K002768 |
| Environments of use | Hospital and home settings | Hospital and home settings |
| Patient Population | Cystic Fibrosis, COPD, asthma and lung<br>diseases with secretory problems, patients<br>with neuromuscular disease affecting the<br>ability to effectively cough. Anyone who<br>is able to read and / or follow verbal<br>instructions. | Cystic Fibrosis, COPD, asthma and lung<br>diseases with secretory problems<br>Patients with atelectasis. |
| Prescription/OTC | Prescription use | Prescription use |
| Principle of Operation<br>Secretion Clearance | Acoustical generator uses sound waves to<br>vibrate the "column" of gas in the<br>airways at various frequencies | Creates airflow interruption during<br>exhalation which vibrates the "column"<br>of gas in the airways at various<br>frequencies |
| Technology employed<br>for<br>Secretion Clearance | Loudspeaker controlled to apply a range<br>of frequencies | Unstable counterweighted lever and<br>magnet mechanism action during<br>exhalation, to interrupt the exhaled<br>airflow as it is directed through the<br>device. This deliberately unstable<br>mechanism repeatedly interrupts<br>airflow, affecting both frequency and<br>amplitude. Backpressure (PEP) is<br>simultaneously created by directing<br>airflow through an adjustable small<br>diameter restriction. |
| Technology employed<br>for Positive<br>Expiratory Pressure<br>(PEP) | Variable resistor valve is placed in the<br>expiratory part of the wye; generates PEP<br>up to 4 cmH2O during resting breathing<br>or 10 – 20 cmH2O during PEP maneuver. | Airflow is diverted through a restriction<br>in the device creating a backpressure<br>which is PEP up to 60 cmH2O at 30<br>L/min |
| Principle of Operation<br>PEP | Creating resistance to exhalation creates<br>backpressure known as PEP which opens<br>the airways | Creating resistance to exhalation creates<br>backpressure known as PEP which<br>opens the airways |
| Technology employed<br>for in-line nebulizer | Y-adapter that allows for a nebulizer to<br>be placed in-line | Connector that allows for a nebulizer to<br>be placed in-line |
| Principle of Operation<br>Nebulizer | Standard off-the-shelf nebulizer that is<br>fitted in-line and has a reservoir, spacer<br>like device. Circulaire II (K955047 /<br>K926055) | Standard off-the-shelf nebulizer that is<br>fitted in-line |
| How it interfaces with<br>the patient? | The patient uses a mouthpiece | The patient uses a mouthpiece |
| Patient Effort | Effort independent | Effort dependent |
| Patient Breathing<br>Maneuver | Normal inspiratory & expiratory pattern | Deliberate exhalation maneuver |
| Airway Acoustical<br>Frequency | 5 – 1,200 Hz generated; up to 7,000 Hz<br>actual resulting from harmonics and<br>resonance | ~11 Hz up to >10,000 Hz actual<br>resulting from harmonics and resonance |
{4}------------------------------------------------
# 510(k) Summary Page 5 of 12 23-May-14
The Vibralung is viewed as substantially equivalent to the predicate, Dymedso The Frequencer, K063645, K100749, and K103176, because:
#### Indications
The airway clearance and bronchial drainage indications for use are substantially equivalent. Discussion - The indications for use are substantially equivalent for the proposed device and the predicate, Dymedso Frequencer, K063645, K100749, and K103176.
# Technology
The Vibralung technology of creating a frequency for vibrating the airways is substantially equivalent to the predicate Dymedso Frequencer, K063645, K100749, and K103176. They both use a loudspeaker that can generate acoustical sound at various frequencies that vibrate the gas in the airways.
Discussion - The technology is substantially equivalent for the proposed device and the predicate, Dymedso Frequencer, K063645, K100749, and K103176.
#### Environment of Use
The environments of use, hospital and home are substantially equivalent to the predicate. Discussion - The environments of use are substantially equivalent to the predicate, Dymedso The Frequencer, K063645, K100749, and K103176.
# Patient Population
The patient population of those with Cystic Fibrosis, COPD, asthma and lung diseases with secretory problems is substantially equivalent to the predicate.
Discussion - The patient population for which airway clearance is indicated, is substantially equivalent to the predicate.
# Table 3 - Substantial Equivalence Table - Acapella
{5}------------------------------------------------
# 510(k) Summary
、
{6}------------------------------------------------
The Vibralung is viewed as substantially equivalent to the predicate DHD- Acapella - K991561 and K002768 because:
Indications: There are 3 indications for use for comparing the Vibralung to the DHD - Acapella.
- . The airway clearance and bronchial drainage indications for use are substantially equivalent.
- Indicated for use as a PEP device which facilitates opening of airways is substantially . equivalent.
- . It may be used simultaneously with aerosol drug delivery.
Discussion - The indications for use are substantially equivalent for the proposed device and the predicate, Dymedso Frequencer, K063645, K100749, and K103176.
# Technology
Secretion Clearance: The Vibralung technology of creating a frequency for vibrating the airways is different than the predicate DHD - Acapella, K991561 and K002768. Vibralung uses a loudspeaker to generate sound waves that create acoustical vibrations in the airways. The predicate, DHD-Acapella, uses an unstable counterweighted balance mechanism that varies the frequency and amplitude of a pressure wave and creates vibration in the airway during exhalation. Nevertheless, the physiological effects are the same regardless of the mechanism that produces the vibrations. Discussion - The technology for creating the vibrations in the airway is done differently but the results are the same. The physiological results are substantially equivalent to the predicate DHD -Acapella - K991561 and K002768.
Positive Expiratory Pressure (PEP): Both the Vibralung and predicate. DHD Acapella, K991561 and K002768, create expiratory resistance / pressure during exhalation. The Vibralung does it with an adjustable resistor valve in the exhalation side of the Y-adapter and the predicate has a diverting air flow path that creates backpressure (resistance) during exhalation, Discussion - Both devices create PEP during exhalation.
Simultaneous nebulized aerosol drug delivery: Both the Vibralung and predicate DHD -
Acapella, K991561 and K002768 can accommodate an in-line nebulizer should the clinician wish to provide aerosol treatment while a patient is using the devices.
Discussion - The use of an in-line nebulizer is substantially equivalent - both devices allow for an in-line nebulizer to be attached.
Environment of Use: The environments of use, hospital and home, are substantially equivalent to the predicate.
Discussion - The environments of use are substantially equivalent to the predicate DHD - Acapella. K991561 and K002768.
Patient Population: The patient population of those with Cystic Fibrosis. COPD, asthma and lung diseases with secretory problems is substantially equivalent to the predicate.
Discussion - The patient population is substantially equivalent to the predicate DHD - Acapella, K991561 and K002768.
{7}------------------------------------------------
# Page 8 of 12
23-May-14
# Table 4 - Substantial Equivalence Table - Lung Flute
| Criteria | Vibralung | Medical Acoustics<br>Lung Flute<br>K091557 |
|---------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| FDA Classification | BYI - Percussor, powered-electric<br>868.5665 | BWF - Incentive Spirometer<br>868.5690 |
| Indications for Use | The Vibralung Acoustical Percussor is<br>indicated as an airway secretion clearance<br>device that creates vibrations in the<br>airways and as a lung expansion device<br>that applies Positive Expiratory Pressure<br>(PEP) as a patient breathes through the<br>device. It may be used to promote<br>bronchial drainage, airway clearance and<br>expectoration. The Vibralung may be<br>used simultaneously with aerosol drug<br>delivery. | Mentioned in 510(k) summary –<br>facilitates mucus clearing by vibrating<br>the airway.<br>The Lung Flute is indicated for PEP<br>therapy. |
| Environments of use | Hospital and home settings | Hospital and home settings |
| Patient Population | Cystic Fibrosis, COPD, asthma and lung<br>diseases with secretory problems,<br>patients with neuromuscular disease<br>affecting the ability to effectively cough.<br>Anyone who is able to read and / or<br>follow verbal instructions. | Not specific but mentions COPD<br>patients |
| Prescription/OTC | Prescription use | Prescription use |
| Principle of Operation<br>Secretion Clearance | Acoustical generator uses sound waves to<br>vibrate the "column" of gas in the<br>airways at various frequencies | Uses acoustical impedance to produce<br>sound waves that vibrate the "column"<br>of gas in the airways at various<br>frequencies |
| Technology employed<br>for<br>Secretion Clearance | Loudspeaker controlled to apply a range<br>of frequencies | Reed which vibrates during exhalation<br>to produce sound frequencies |
| Technology employed<br>for Positive<br>Expiratory Pressure<br>(PEP) | Variable resistor valve is placed in the<br>expiratory part of the wye; generates PEP<br>up to $4 cmH_2O$ during resting breathing<br>or 10 – 20 $cmH_2O$ during PEP maneuver. | Airflow encounters restriction through<br>the device creating a back pressure<br>which is PEP 2.5 $cm H_2O$ . |
| Principle of Operation<br>PEP | Creating resistance to exhalation creates<br>backpressure known as PEP which opens<br>the airways | Creating resistance to exhalation<br>creates backpressure known as PEP<br>which opens the airways |
| How it interfaces with<br>the patient? | The patient uses a mouthpiece | The patient uses a mouthpiece |
| Patient Effort | Effort independent | Effort dependent |
| Patient Breathing<br>Maneuver | Normal inspiratory & expiratory<br>breathing pattern | Deliberate exhalation maneuver |
| Airway Acoustical<br>Frequency | 5 – 1,200 Hz generated; up to 7,000 Hz<br>actual resulting from harmonics and | ~11 Hz up to >10,000 Hz actual<br>resulting from harmonics and<br>resonance |
{8}------------------------------------------------
#### Page 9 of 12 23-May-14
The Vibralung is viewed as substantially equivalent to the predicate Medical Acoustics ~ Lung Flute - K091557 because:
Indications: There are 2 indications for use for comparing the Vibralung to the Medical Acoustics -Lung Flute.
- The airway clearance and bronchial drainage indications for use are substantially equivalent. .
- . Indicated for use as a PEP device that facilitates opening of airways is substantially equivalent.
Discussion - The indications for use are substantially equivalent the proposed Vibralung and to the predicate Medical Acoustics - Lung Flute - K091557.
# Technology
Secretion Clearance: The Vibralung technology for creating a frequency for vibrating the airways is different than the predicate Medical Acoustics - Lung Flute - K091557. Vibralung uses a loudspeaker to generate sound waves that create acoustical vibrations in the airways, while the predicate. Medical Acoustics - Lung Flute uses a vibrating reed that provides a sound frequency that creates vibration in the airway during exhalation.
Discussion - The technology for creating the vibrations in the airway is done differently but the results are the same. The physiological results are substantially equivalent to the predicate Medical Acoustics - Lung Flute - K091557.
Positive Expiratory Pressure (PEP): Both the Vibralung and predicate Medical Acoustics - Lung Flute - K091557 create expiratory resistance / pressure during exhalation. The Vibralung does it with an adjustable resistor valve in the exhalation side of the Y-adapter and the predicate has a diverting air flow path that creates some backpressure (resistance) during exhalation. Discussion - Both devices created PEP during exhalation. Both devices produce about a substantially equivalent level of expiratory resistance (PEP), 4 cm H2O for Vibralung and 2.5 cm H2O for the Lung Flute.
Environment of Use: The environments of use - hospital and home are substantially equivalent to the predicate.
Discussion - The environments of use are substantially equivalent to the predicate Medical Acoustics - Lung Flute - K091557.
Patient Population: The patient population is not specifically listed in the predicates for Use or 510(k) Summary. However, there are references to patients who require secretion clearance as well as PEP treatment, thus the patients would, by implication, include those with Cystic Fibrosis, COPD, asthma and lung diseases with secretory problems as substantially equivalent to the predicate.
Discussion - The patient population is substantially equivalent to the predicate Medical Acoustics -Lung Flute - K091557.
{9}------------------------------------------------
# Page 10 of 12 23-May-14
# Non-clinical Testing Summary -
We performed a number of non-clinical tests to demonstrate that the Vibralung Acoustical Percussor to be substantially equivalent and will perform within specification.
- . Verification and Validation Testing
- Inspiratory Resistance .
- o Inspiratory Resistance
- o Airway Pressure and Sound Pressure Level
- . Mechanical Dead Space
- Aerosol Performance During Treatment ●
- Particle Characterization with and without aerosolization o
- Particle Characterization of the Acapella Predicate with and without aerosolization o
- Acoustic Aerosol Performance o
- . Comparative Acoustical Testing vs. Predicates
- . Treatment Control Unit
- . Human Factors Study
- . Accelerated Aging
- . Cleaning
- Electrical Safety, EMC, EMI .
- O IEC 60601
- . O EN 5501
- lEC 60601-1 236815
- o IEC 60601-1-11 236815-1
- o IEC 62133
### Comparison of Predicate Devices:
We have performed comparative testing for the Vibralung and the Predicates. The spectral analysis patterns for the Vibralung are different than the predicate devices for a number of reasons, but they also contain the same frequencies.
- The Vibralung lung device generates sound frequencies in the range of 5 to 1,200 Hz. ● However, when these frequencies are reproduced by the loudspeaker, mounted in the HHT and conducted to the patient through a waveguide Y-adapter, the sound experiences the normal phenomena of harmonics and resonance with the following ranges:
| Low Mode | 5 Hz to ~1,700 Hz |
|-------------------|-------------------|
| Medium Mode | 5 Hz to 3,000 Hz |
| High Mode | 5 Hz to 3,000 Hz |
| Random Noise Mode | 5 Hz to >7,000 Hz |
{10}------------------------------------------------
## Page 11 of 12 23-May-14
- Likewise, the predicate devices generate sound frequencies as a consequence of their . operation. These frequencies likewise experience harmonics and resonance that yields resulting frequencies closely substantially equivalent to the Vibralung:
| Blue Acapella | <11 Hz to >10,000 Hz |
|----------------|--------------------------------------|
| Green Acapella | <11 Hz to >10,000 Hz |
| Lung Flute | <11 Hz to >10,000 Hz |
| The Frequencer | 30 Hz to ~9,000 Hz (@ 65 Hz setting) |
- Like the predicate devices, the Vibralung has a number of frequency components. But . unlike the predicates, many of these are deliberately created.
- . Like the predicate devices, the Vibralung is also subject to resonance whereby additional frequencies are created due to internal resonance and harmonics.
- . Like the predicate devices, the Vibralung demonstrates low frequency components. Unlike the predicate devices, they are not as defined and pronounced.
Summary: These studies demonstrate that all the predicate devices demonstrate acoustic frequencies that are inclusive of, and even higher than, those frequencies generated by the Vibralung Acoustical Percussor.
Both the predicate devices and the Vibralung generate resonance and harmonic frequencies. This is a normally occurring phenomenon any time sound is directed through tubes.
The Vibralung and the predicate devices are substantially equivalent in terms of the acoustic frequencies and amplitudes they employ for airway clearance therapy.
# Clinical Testing:
Two (2) studies were performed with the Vibralung Acoustical Percussor.
Study 1 included 10 subjects to determine if Vibralung can be a positive treatment option to improve mucous discharge in subjects with cystic fibrosis (CF); and to assess if the treatment was safe for the patient. The subjects with mild to moderate CF were subjected to two treatments: (1) with sound, and (2) without sound, each for a period of 20 minutes on two different days. On each visit we measured pulmonary function, lung clearance index (LCI), symptoms and peripheral oxygen saturation (SaO2) at baseline and 1 and 4 hours post-treatment, and total sputum production was collected over 4 hours. Although the results indicated no change from baseline to 4 hours post treatment in FVC, FEV1, and SaO2, there was a greater improvement to LCI with the device use with sound than without sound. The results demonstrated that the Vibralung device is safe for patient use.
Study 2 was designed to assess the effects of the Vibralung on sputum clearance when compared to the Vest predicate for five days. There were 11 subjects. Similar results in sputum wet weight, pellet weight, or dry weight, when comparing the 2 devices. The results demonstrated that the Vibralung is equivalent in performance for sputum clearance.
{11}------------------------------------------------
# Page 12 of 12 23-May-14
# Substantial Equivalence Conclusion:
The sponsor has demonstrated through performance testing, design and features, non-clinical, and clinical testing that the proposed Vibralung Acoustic Percussor device does not raise any new safety concerns and is substantially equivalent to the identified predicates.
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 27, 2014
Westmed, Inc. C/O Mr. Paul Dryden President. Promedic, Inc. 5580 S. Nogales Hwy. Tucson, AZ 85706 USA
Re: K133057
Trade/Device Name: Vibralung Acoustic Processor Regulation Number: 21 CFR 868.5665 Regulation Name: Electric powered percussor Regulatory Class: Class II Product Code: BYI Dated: April 24th, 2014 Received: April 25th, 2014
Dear Mr. Dryden,
This letter corrects our substantially equivalent letter of May 23, 2014.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{13}------------------------------------------------
Page 2 - Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purobit-Sheth, M.D. Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith. M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{14}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known) K133057
Device Name Vibralung Acoustical Percussor
Indications for Use (Describe)
The Vibralung Acoustical Percussor is indicated as an airway secretion clearance device that creates vibrations in the airways and as a lung expansion device that applies Positive Expiratory Pressure (PEP) as a patient breather through the device. It may be used to promote bronchial drainage, airway clearance and expectoration. The Vibralung may be used simultaneously with aerosol drug delivery.
Patient Population - Cystic Fibrosis, COPD, asthma and lung diseases with secretory problems, with neuromuscular disease affecting the ability to effectively cough. Anyone who is able to read and / or follow verbal instructions
Environment of Use - Hospital and Home
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
[ ] Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
A CONSULTION CONTROLLER FOR EDALUSE ONLY CONTROLLER CONTROLLER CONTRACTOR CONSULTION CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR CONTRACTOR C Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/14/Picture/14 description: The image shows the logo for the Food and Drug Administration (FDA). The logo is in black and white and features the letters "FDA" in a stylized font. The letters are bold and have a slightly rounded appearance. The logo is simple and recognizable, and it is often used on products that have been approved by the FDA.
Anya C. Harry -S 2014.05.23 03:23:19 -04'00'
FORM FDA 3881 (9/13)
**PSC Publishing Services (301) 443-4760 EP**
{15}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.