K063645 · Dymedso, Inc. · BYI · Mar 15, 2007 · Anesthesiology
Device Facts
Record ID
K063645
Device Name
FREQUENCER, MODEL 1001
Applicant
Dymedso, Inc.
Product Code
BYI · Anesthesiology
Decision Date
Mar 15, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5665
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is intended to be a component of postural drainage therapy by providing a convenient method of external thorax manipulation. It is indicated for patients having respiratory ailments which involve defective mucociliary clearance, as is typical in patients suffering from cystic fibrosis as well as chronic bronchitis, bronchiectasis, ciliary dyskinesia syndromes, asthma, muscular dystrophy, neuromuscular degenerative disorders, post-operative atelectasis and thoracic wall defects.
Device Story
Frequencer™ provides airway clearance therapy via external chest wall vibration. Device uses electro-acoustical transducer (Power Head) placed on patient's chest to induce oscillatory sound waves (20-100 Hz). Frequency generator drives transducer; operator selects frequency by adjusting until patient feels sympathetic resonance. Device loosens mucus deposits to promote bronchial drainage. Used as component of postural drainage therapy. Provides gentler, less painful alternative to traditional manual 'clapping' (CCPT). Benefits patients unable to tolerate manual percussion. Clinical trial on 20 adult cystic fibrosis patients demonstrated comparable sputum expulsion to CCPT.
Clinical Evidence
Clinical trial on 20 adult patients with cystic fibrosis compared Frequencer™ to conventional chest physiotherapy (CCPT). Primary endpoint was quantity of expelled sputum per session. Results: Frequencer™ (9.24 ± 1.62 g) vs. CCPT (9.27 ± 1.62 g). Study indicates comparable efficacy.
Technological Characteristics
Electro-acoustical transducer (Power Head) driven by frequency generator. Operating frequency range: 20-100 Hz. Applied force: 0.4 - 3 N. Energy source: electric. Device is external, non-invasive.
Indications for Use
Indicated for patients with respiratory ailments involving defective mucociliary clearance, including cystic fibrosis, chronic bronchitis, bronchiectasis, ciliary dyskinesia syndromes, asthma, muscular dystrophy, neuromuscular degenerative disorders, post-operative atelectasis, and thoracic wall defects.
Regulatory Classification
Identification
A powered percussor is a device that is intended to transmit vibration through a patient's chest wall to aid in freeing mucus deposits in the lung in order to improve bronchial drainage and that may be powered by electricity or compressed gas.
{0}------------------------------------------------
Dymedso Frequencer™ Powered Percussor
K063645
#### II. 510(k) Summary
This summary of safety and effectiveness is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR Part 807.92.
| Date of Submission: | November 22, 2006 |
|-------------------------------|--------------------------------------------------------------------|
| Submitter: | DYMEDSO, Inc. |
| Address: | 901A, Boul. Du Curé Boivin<br>Boisbriand, Québec J7G 2S8<br>CANADA |
| Establishment Registration #: | 3004365906 |
Yvon Robert, President Contact: Phone: (450) 437-9601 Fax: (450) 437-2063
Proprietary Device Name: Frequencer™ Common Name of Device: Powered Percussor Percussor, Powered-Electric Classification Name: Device Class: で BYI Product Code:
## Predicate Devices:
K965192-Thairapy® Vest System, K972859-Flutter® D Mucus Clearance Device, K982889-Medpulse™ Respiratory Vest System, Model 1000, K012928-AbiVest™ Airway Clearance System, K972859-Flutter® D Mucus Clearance Device K031876-Electro Flo Percussor, Model 5000
## Description:
The Frequencer™ provides airway clearance therapy and promotes bronchial drainage by inducing vibration in the chest walls. It induces oscillatory sound waves in the chest by means of an electro-acoustical transducer (hereafter referred to as the "Power Head") which is placed externally on the patient's chest. The Power Head is connected to a frequency generator which is capable of producing frequencies between 20 and 100 Hz. The Power Head induces sound waves in the patient's chest which are effective in loosening mucus deposits.
#### Intended Use:
This device is intended to be a component of postural drainage therapy by providing a convenient method of external thorax manipulation. It is indicated for patients having respiratory ailments which involve defective mucociliary clearance, as is typical in patients suffering from cystic fibrosis as well as chronic bronchitis, bronchiectasis, ciliary dyskinesia syndromes, asthma,
{1}------------------------------------------------
muscular dystrophy, neuromuscular degenerative disorders, post-operative atelectasis and thoracic wall defects.
# Comparison of Technological Characteristics:
The Frequencer™ is similar to the predicate devices in terms of indications for use, contra-indications, and the intent to create vibrations in the chest wall via external stimulation. The operating principle of the proposed device differs from those of the predicate devices. Rather than mechanical percussion (Electro-Flo Percussor Model 5000), airflow-induced air pulses which are transferred through the air in the patient's throat (Flutter® D Mucus Clearance Device), or pneumatically created pressure oscillations (AbiVest™ Airway Clearance System, Medpulse™ Respiratory Vest System, Model 1000, Thairapy® Vest System), the Frequencer uses sound wave pulses from a transducer to stimulate vibration in the chest wall.
# Performance Testing
Conventional chest physiotherapy (CCPT) typically involves forces externally applied to the patient's chest of 58.10 +/- 15.32 N (13.06 +/- 3.44 lb). The Frequencer™ uses much smaller forces, namely 0.4 - 3 N (0.090 - 0.674 Ib). Technological data for pneumatic vests are typically expressed in units of pressure and therefore do not allow a basis for direct comparison. Frequencies for CCPT are typically 6.6 Hz, pneumatic vests (AbiVest™ Airway Clearance System, Medpulse™ Respiratory Vest System, Model 1000, Thairapy® Vest System) and the mechanical percussor (Electro-Flo Percussor Model 5000) are capable of delivering roughly 5-25 Hz. The Frequencer™ can deliver a range of frequencies from 20-100 Hz. The actual operating frequency is selected by varying until the patient feels a sympathetic resonance in the chest.
# Clinical Data:
The efficacy of the Frequencer™ as compared to CCPT was assessed in a clinical trial on 20 adult patients suffering from cystic fibrosis (CF). The results show that use of the Frequencer™ results in comparable quantities of expelled sputum per session as using CCPT (9.24 ± 1.62 g for the Frequencer™ vs. 9.27 ± 1.62 g for CCPT). This is despite the fact that patients treated with the Frequencer were sitting, while those receiving CCPT were lying down, thus benefiting from postural drainage.
# Conclusions:
The proposed Frequencer™ has the identical indications and intended use as well as comparable technological characteristics as the predicate devices. Only the operating principle differs, being electro-acoustic rather than pneumatic or electro-mechanical. Safety and efficacy have been demonstrated in both laboratory and clinical settings. Thus, the proposed Frequencer™ is substantially equivalent to the predicate devices.
4
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three wavy lines representing the department's mission to protect the health of all Americans and provide essential human services. The logo is surrounded by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# MAR 1 5 2007
Dymedso, Incorporated C/O Ms. Jean Bigoney Springfield Metallurgical Services, Incorporated 127 Main Street P.O. Box 826 Springfield, Vermont 05156-0826
Re: K063645
Trade/Device Name: The Frequencer Regulation Number: 21 CFR 868.5665 Regulation Name: Powered Percussor Regulatory Class: II Product Code: BYI Dated: February 21, 2007 Received: February 23, 2007
Dear Ms. Bigoney:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
# Page 2 - Ms. Bigoney
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Clues
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE
510(k) Number: K063645 Device Name: The Frequencer Indications for Use:
The Frequencer™ provides airway clearance therapy and promotes bronchial drainage by inducing vibration in the chest walls. This device is intended to be a component of postural drainage therapy by providing a convenient method of external thorax manipulation.
It is indicated for patients having respiratory ailments which involve defective mucociliary clearance, as is typical in patients suffering from cystic fibrosis as well as chronic bronchitis, bronchiectasis, ciliary dyskinesia syndromes, asthma. muscular dystrophy, neuromuscular degenerative disorders, post-operative atelectasis and thoracic wall defects. Indications for this form of therapy are described in the Clinical Practice Guidelines for Postural Drainage Therapy of the American Association for Respiratory Care (AARC) published in 1991. This particular device provides a gentler, less painful form of therapy than the traditional "clapping" method of postural drainage therapy, allowing it to be used on patients who cannot be treated by clapping.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-theCounter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Morf Thler
Page 1 of 1
ion Control, and Devices
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.