K962871 · Reimers Systems, Inc. · CBF · Oct 23, 1996 · Anesthesiology
Device Facts
Record ID
K962871
Device Name
ODS1 - HOOD DRIVER
Applicant
Reimers Systems, Inc.
Product Code
CBF · Anesthesiology
Decision Date
Oct 23, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5470
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Model ODS1 - Hood Driver is intended for use with monoplace hyperbaric chambers. It is normally used to permit the patient to breath oxygen while the chamber is pressurized on air. However, it can be used to supply the patient with any physiologically appropriate gas which may or may not be the same as the gas used to pressurize the chamber.
Device Story
ODS1 Hood Driver; hyperbaric chamber breathing station. Supplies steady flow of breathing gas (oxygen or air) to patient via 'Sea-Long' hood or free-flow mask. Controls exhaust flow to chamber exterior. Used in monoplace hyperbaric chambers; operated by clinician (inside or outside chamber). Protects patient from suction injury via flow regulation. Benefits patient by enabling controlled gas delivery during hyperbaric therapy.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Materials: Copper, brass, stainless steel; 100% oxygen compatible softgoods. Connectivity: Standard 22 mm respiratory fittings. Flow control: 20 - 200 scfh. Mechanical gas flow regulation system.
Indications for Use
Indicated for patients undergoing hyperbaric oxygen therapy for conditions listed by the Undersea Hyperbaric & Medical Association.
Regulatory Classification
Identification
A hyperbaric chamber is a device that is intended to increase the environmental oxygen pressure to promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure. This device does not include topical oxygen chambers for extremities (§ 878.5650).
K072427 — SIGMA SERIES HYPERBARIC CHAMBER · Perry Baromedical Corp. · Jun 11, 2008
K011565 — INFANT OXYGEN TREATMENT HOOD · Gulf Coast Hyperbarics, Inc. · Aug 16, 2001
K040145 — ATA HCM MONOPLACE HYPERBARIC SYSTEM · Ata Hyperbaric Chamber Manufacturing · May 4, 2005
K052713 — MODEL 3300E/ER MONOPLACE HYPERBARIC OXYGEN CHAMBER · Sechrist Industries, Inc. · Oct 28, 2005
K022214 — MSI I2010 DUAL PLACE HYPERBARIC CHAMBER · Mechidyne Systems, Inc. · Feb 26, 2003
Submission Summary (Full Text)
{0}
OCT 23 1996
K962871
# REIMERS SYSTEMS, INC.
7668 FULLERTON ROAD
SPRINGFIELD, VIRGINIA 22153
S. D. REIMERS
PRESIDENT
# 510(k) SUMMARY - ODS1 HOOD DRIVER
TELEPHONE
(703) 866-5490
TELECOPIER
(703) 866-9066
DATE: July 18, 1996
SUBMITTER: Stephen D. Reimers, PE
President
CONTACT: Carlyle H. Smith, PE
Staff Engineer
# DEVICE INFORMATION:
Trade name: ODS1 - Hood Driver
Common name: Hyperbaric Chamber Breathing Station
Classification name: Chambers, Hyperbaric 73CBF
# DESCRIPTION:
The Model ODS1 - Hood Driver is intended for use with monoplace hyperbaric chambers. It is normally used to permit the patient to breath oxygen while the chamber is pressurized on air. However, it can be used to supply the patient with any physiologically appropriate gas which may or may not be the same as the gas used to pressurize the chamber.
The ODS1 circuit is designed to supply a steady flow of breathing gas to the breathing apparatus used in the chamber. The breathing apparatus can be either a "Sea-Long" hood or a free-flow mask arrangement.
# INDICATIONS:
Hyperbaric chambers are used for treating the indications as currently listed by the Undersea Hyperbaric & Medical Association for hyperbaric oxygen therapy.
# COMPARISON TO EXISTING EQUIPMENT:
## Similarities:
The ODS1 is substantially equivalent to the breathing stations found in larger hyperbaric chamber systems, such as the Lifeforce Hyperbaric Chamber System (K951125) built by Reimers Engineering. They are similar in the following ways:
1. Supplies a steady flow of breathing gas to the breathing apparatus.
2. Controls a steady exhaust flow from the breathing apparatus to the chamber exterior.
Page 1 of 2
{1}
3. Can be used with both hoods and constant flow masks.
4. Designed around standard 22 mm respiratory fittings.
5. Patient is adequately protected from suction injury caused by excessive exhaust flow or inadequate supply flow.
6. Requires minimal flow adjustments after initial setup.
7. Same range of supply flow control: ODS1: 20 - 200 scfh
Lifeforce: 21 - 212 scfh
8. Works with oxygen or air.
9. Same wetted materials (Copper, Brass, Stainless-Steel)
10. 100% oxygen compatible softgoods
## Differences
The significant differences between the ODS1 - Hood Driver and the Lifeforce breathing stations are as follows:
1. The Lifeforce breathing stations are located inside the chamber. The ODS1 controls can be placed inside or outside of the chamber. With the controls outside of the chamber, an outside attendant sets the controls instead of an inside attendant.
2. The Lifeforce breathing stations contain an additional supply flowmeter and throttle valve for a separate nebulizer flow. The ODS1 has only one supply flowmeter and throttle valve. The ODS1 can still be used with certain types of nebulizers if desired. However the additional dedicated nebulizer circuit has been found unnecessary for most monoplace chamber applications.
## DISCUSSION:
Neither of these differences affect the safety or effectiveness of the ODS1 which is inherent in the circuit design. The differences are only related to the packaging of the circuit.
## CONCLUSION:
The ODS1 - Hood Driver is substantially equivalent to existing, legally marketed devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.