Revitalair 430F

K171899 · Oxavita Srl · CBF · Nov 22, 2019 · Anesthesiology

Device Facts

Record IDK171899
Device NameRevitalair 430F
ApplicantOxavita Srl
Product CodeCBF · Anesthesiology
Decision DateNov 22, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5470
Device ClassClass 2
AttributesTherapeutic

Intended Use

The intended use of the Portable Hyperbaric Chamber Revitalair® 430F is to treat acute mountain sickness under the prescription of a health professional. The medical device is designed for use at physician offices and health institutions.

Device Story

Portable hyperbaric chamber for low-pressure therapy (max 1.4 ATA); encloses patient within pressure boundary. Chamber constructed of polyester-based plastic fabric with aluminum rings; includes 10 transparent windows for patient monitoring. Operated via electric oil-less compressor; manual valves control inflation/pressure. Safety relief valves activate above 1.3 ATA. Used in physician offices and health institutions; operated by healthcare professionals or patients (internal/external controls). Provides hyperbaric environment to treat acute mountain sickness; enables visual monitoring of patient comfort. Benefits patient by providing controlled pressure environment for symptom relief.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical safety (IEC 60601-1), electromagnetic disturbance (IEC 60601-1-2), biocompatibility (ISO 10993, ISO 18562), and pressure testing (ASME PVHO-1).

Technological Characteristics

Portable hyperbaric chamber; polyester-based plastic fabric with aluminum rings; 900mm x 1850mm. Operating pressure max 1.4 ATA. Electric oil-less compressor. Relief valves (metal/plastic) set at 1.3 ATA. 10 transparent windows. Complies with ASME PVHO 1-2012. Biocompatibility per ISO 10993/18562. Electrical safety per IEC 60601-1; EMD per IEC 60601-1-2.

Indications for Use

Indicated for treatment of acute mountain sickness in patients with medical conditions related to altitude sickness. Contraindicated for patients with colds or flu symptoms, recent alcohol consumption, blocked ear canals, blocked sinuses, otic barotraumas, excessive CO2 exposure, pulmonary hyper-expansion, or decompression sickness.

Regulatory Classification

Identification

A hyperbaric chamber is a device that is intended to increase the environmental oxygen pressure to promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure. This device does not include topical oxygen chambers for extremities (§ 878.5650).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Oxavita S.R.L. % Diane Sudduth Sr. RA/QA Consultant Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, Texas 78746 Re: K171899 Trade/Device Name: Revitalair 430F Regulation Number: 21 CFR 868.5470 Regulation Name: Hyperbaric Chamber Regulatory Class: Class II Product Code: CBF Dated: April 10, 2019 Received: April 11, 2019 Dear Diane Sudduth: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K171899 Device Name Revitalair® 430F Indications for Use (Describe) The intended use of the Portable Hyperbaric Chamber Revitalair® 430F is to treat acute mountain sickness under the prescription of a health professional. The medical device is designed for use at physician offices and health institutions. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X | Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary # Revitalair® 430F # K171899 #### 1. Submission Sponsor Oxavita S.R.L. Tamborini 5813 Buenos Aires, Argentina Contact: Eduardo Northing Title: Exports and Logistics #### 2. Submission Correspondent Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746 USA Office Phone: (512) 327.9997 Contact: Dr. Diane Sudduth Title: Senior Consultant, RA #### 3. Date Prepared 11/22/19 #### 4. Device Identification | Trade/Proprietary Name: | Revitalair® 430F | |-------------------------|-----------------------------| | Common/Usual Name: | Portable Hyperbaric Chamber | | Classification Name: | Chamber, Hyperbaric | | Regulation Number: | 21 CFR 868.5470 | | Product Code: | CBF | {4}------------------------------------------------ Device Class: Class II Classification Panel: Anesthesiology Devices ## 5. Legally Marketed Predicate Device(s) K072757, The Dive HD250 and The Grand Dive HD400, SUMMIT TO SEA ### 6. Indication for Use Statement The intended use of the Portable Hyperbaric Chamber Revitalair® 430F is to treat acute mountain sickness under the prescription of a health professional. The medical device is designed for use at physician offices and health institutions. ### 7. Device Description The Portable Hyperbaric Chamber Revitalair® 430F is a hyperbaric chamber for low pressures (operating at pressures of no greater than 1.4 atmospheres absolute (ATA)). The operational design pressure of a hyperbaric chamber that encloses a human within its pressure boundary falls within the scope of the American Society of Mechanical Engineers Pressure Vessels Human Occupancy 1 (ASME PVHO 1-2012). Revitalair® 430F consists of 2 parts, the cabin or chamber and the compression system or compressor's cabinet. The chamber, weighing 62 pounds, is constructed of an airtight polyester-based plastic fabric joined by aluminum rings, forming a cylinder 900 mm in diameter and 1850 mm in length (Figure 1). The chamber is inflated with atmospheric air through an electric compressor. Fittings allow the chamber to be connected to compressed air by means of manually controlled valves. The safety or relief valves are operated at pressures above 1.3 ATM which ensures safe operations. The compressor has an additional safety valve for any obstruction of the supply hose from the compressor to the chamber. The Revitalair® 430F has 10 transparent windows to let in light, 360º viewing and enable easy verification of the patient's comfort from the outside. The Revitalair® 430F can be operated from the interior as well as from the exterior. After folding it up, the Revitalair® 430F is placed in its transportation box or in the optional carrying case. ### 8. Substantial Equivalence Discussion The following table compares the Portable Hyperbaric Chamber Revitalair® 430F (abbreviated as Revitalair® 430F) to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance testing. {5}------------------------------------------------ | Manufacturer | SUMMIT TO SEA | SUMMIT TO SEA | Oxavita S.R.L. | Significant<br>Difference | |-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------| | Trade Name | The Grand Dive<br>HD400 | The Dive HD250 | Revitalair® 430F | | | 510(k) | K072757 | K072757 | | | | Product Code | CBF | CBF | CBF | Same | | Regulation<br>Number | 868.5470 | 868.5470 | 868.5470 | Same | | Regulation Name | Hyperbaric<br>Chamber | Hyperbaric<br>Chamber | Hyperbaric Chamber | Same | | Intended Use | Acute Mountain<br>Sickness AMS and<br>its associated mild<br>symptoms | Acute Mountain<br>Sickness AMS and<br>its associated mild<br>symptoms | Acute Mountain<br>Sickness AMS and its<br>associated mild<br>symptoms | Same | | Prescription -<br>Rx Only | Yes | Yes | Yes | Same | | Target Population | People with medical<br>conditions by<br>altitude sickness | People with<br>medical conditions<br>by altitude sickness | People with medical<br>conditions by<br>altitude sickness | Same | | Places of Use | Homes, gyms,<br>offices, outdoor,<br>hospitals and clinics | Homes, gyms,<br>offices, outdoor,<br>hospitals and clinics | Hospitals and clinics | Similar | | Weight | 32 lb | 15 lb | 62 lb | Different | | Measures<br>Diameter<br>×<br>Length | 39.37in x 106.29in<br>(1000mm x<br>2700mm) | 26.6in x 82.7in<br>(675mm x<br>2100mm) | 35.4in x 74.8in<br>(900mm x 1900mm) | Similar | | Windows | 4 | 2 | 10 | Different | | Transport Belts | 4 | 4 | 2 | Different | | Relief Valves | Yes | Yes | Yes, Metal | Same | | Manufacturer | SUMMIT TO SEA | SUMMIT TO SEA | Oxavita S.R.L. | Significant | | Trade Name | The Grand Dive<br>HD400 | The Dive HD250 | Revitalair® 430F | Difference | | Drain Valves | Yes, integrated<br>emptying and filling | Yes, integrated<br>emptying and filling | Yes, 2 units only for<br>emptying | Same | | Operating<br>Pressure | 4psi / 1.3 ATA | 4psi / 1.3 ATA | 4psi / 1.3 ATA | Same | | Inflation Method | Gast compressor<br>0523 | Gast compressor<br>0523 | Oil-less compressor | Same | | Locking System | Simple air tight<br>closure | Simple air tight<br>closure | System of metal<br>rings | Different | | Materials | 33-39 oz uretane<br>coated nylon | 33-39 oz uretane<br>coated nylon | Polyester hatched<br>1100 18x18 Vinyl<br>with phthalate free | Similar | | Relief Valves | Plastic & Metal | Plastic & Metal | 2 by Plastic & Metal | Similar | | Compressor | Gast oil-less | Gast oil-less | Oil-less | Similar | | Air Filter in the<br>Compressor | Yes | Yes | Yes | Same | | Pressure<br>Measurement | Yes | Yes | Yes | Same | | Air Filter Chamber | Yes | Yes | Yes | Same | | Skills of the<br>Person to Leave<br>the Chamber | Yes | Yes | Yes | Same | | Operating<br>Temperature °C | -50°F to 125°F | -50°F to 125°F | 59 °F / 113 °F | Similar | | Manufacturer | SUMMIT TO SEA | SUMMIT TO SEA | Oxavita S.R.L. | Significant<br>Difference | | Trade Name | The Grand Dive<br>HD400 | The Dive HD250 | Revitalair® 430F | Difference | | Contraindications | Patients with:<br>Colds, flu<br>symptoms,<br>- Recent Alcohol<br>consumption,<br>- Blocked ear<br>canals,<br>- Blocked Sinuses,<br>- Otic barotraumas,<br>- Excessive CO2<br>Exposure,<br>- Pulmonary hyper<br>expansion<br>Decompression<br>sickness | Patients with:<br>Colds, or flu<br>symptoms,<br>- Recent Alcohol<br>consumption<br>- Blocked Ear<br>Canals,<br>- Blocked Sinuses,<br>- Otic barotraumas,<br>- Excessive C02<br> | Patients with: Colds,<br>or flu symptoms,<br>- Recent Alcohol<br>consumption<br>- Blocked Ear Canals,<br>- Blocked Sinuses,<br>- Otic barotraumas,<br>- Excessive C02<br>Exposure,<br>- Pulmonary hyper<br>expansion<br>Decompression<br>Sickness | Same | | Labeling | Comparable | Comparable | Comparable | Same | # Table 5A – Comparison of Characteristics {6}------------------------------------------------ {7}------------------------------------------------ # 9. Non-Clinical Performance Data As part of demonstrating safety and effectiveness of the Revitalair® 430F Hyperbaric Chamber and in showing substantial equivalence to the predicate device that are subject to this 510(k) submission, Oxavita completed an number of non-clinical performance tests. Testing was performed as determined by the hazard analysis. The Revitalair® 430F Hyperbaric Chamber meets all the requirements for overall design, biocompatibility, and electrical safety results confirming that the design output meets the design inputs and specifications for the device. The Revitalair® 430F Hyperbaric Chamber passed all the testing in accordance with internal requirements, and the standards shown below to support substantial equivalence of the subject device: - . Electrical safety testing per IEC 60601-1 - Electromagnetic Disturbance (EMD) testing per IEC 60601-1-2 - Biocompatibility testing - Cytotoxicity, Sensitization, Gas emission VOC, Particulates, Inorganic gases (Ozone, CO, CO2) per ISO 10993 and ISO 18562 {8}------------------------------------------------ In addition to the guidance and standards testing, the following life testing was performed: - Device Life Report which address's the device's useful life based on reliability analysis of the individual components - . Pressure Testing based on ASME PVHO-1-2007, sections: 2-7.2 al 2-7.7; 4-8.1.1; 4-8.1.2; 4-8.1.4 which address's the device's response to excessive pressures. #### 10. Clinical Performance Data There was no human clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with proven safety and efficacy for the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device. #### 11. Statement of Substantial Equivalence The Portable Hyperbaric Chamber Revitalair® 430F, as designed and manufactured, is determined to be substantially equivalent to the cited predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...