To provide mild hyperbaria for the treatment of Acute Mountain Sickness (AMS) and its associated mild symptoms. Environment of Use: Home, Physicians office, Outdoor environments, Hospital, Sub-acute Institutions, Emergency Services. Prescriptive Use: Caution, Federal law restricts this device to sale by or on the order of a physician.
Device Story
Flexi-Lite is a portable, mild hyperbaric chamber (pressures < 5 psi) for treating Acute Mountain Sickness (AMS). Device uses atmospheric air supplied by GAST 0523 oil-less compressor to pressurize chamber. Construction features inner bag (pressure containment) and outer bag (structural support). Equipped with metal relief valves, air addition/depressurization valves, pressure gauges, and air sampling port. Operated in home, clinic, hospital, or field settings. Occupant can self-extricate. Healthcare providers prescribe device to provide mild hyperbaric environment, potentially alleviating AMS symptoms.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on descriptive comparisons of design, materials, and intended use to legally marketed predicate devices.
Technological Characteristics
Mild hyperbaric chamber; operating pressure 2-4 psi. Materials: inner bag 880 denier urethane-coated nylon; outer bag 2x2 basket weave urethane-coated nylon. Components: brass/stainless steel relief valves, GAST 0523 oil-less compressor, air filtration, pressure gauges. Dimensions: 8 ft x 30 in. Portable, non-networked, mechanical operation.
Indications for Use
Indicated for treatment of mild symptoms of Acute Mountain Sickness (AMS). Contraindications: colds/flu, recent alcohol consumption, blocked ear canals, otic barotrauma, excessive CO2 exposure, pulmonary hyperexpansion, decompression sickness.
Regulatory Classification
Identification
A hyperbaric chamber is a device that is intended to increase the environmental oxygen pressure to promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure. This device does not include topical oxygen chambers for extremities (§ 878.5650).
{0}------------------------------------------------
K 051759
NOV 1 7 2005
Performance Hyperbarics
2599 A Olinda Rd
Makawao, HI 96768
# Non-Confidential Summary of Safety and Effectiveness, 510(k) Submission
June 15, 2005
| Performance Hyperbarics | Tel - 808-573-8166 |
|----------------------------|-------------------------------------------------|
| 2599 A Olinda Rd | Fax- 808-573-8976 |
| Makawao, HI 96768 | email- info@hyperbaric-oxygen-chamber.com |
| Official Contact: | Spencer Feldman |
| Proprietary or Trade Name: | Flexi-Lite Hyperbaric Chamber |
| Common Used Name: | Mild Hyperbaric Chamber |
| Classification Name: | Hyperbaric Chamber |
| Regulatory Class: | Class II (two) |
| Product Code: | 73 CBF |
| Device: | Flexible Hyperbaric Chamber with Air Compressor |
| Predicate Devices: | HTI - Gamow Bag -K874752A |
| | HTI - Mild Hyperbaric Chamber -K001409 |
### Device Description
The Flexi-Lite flexible hyperbaric chamber is a mild hyperbaric chamber for pressures less than 5 psi. This lightweight and portable chamber utilizes atmospheric Air as supplied by a GAST model 0523 "oiless, breathable air compressor" to pressurize the chapplice und provide a suitable environment for the occupant. The Flexi-Lite construction utilizes and bag design with the inner bag containing the pressure and an exterior hag to provide structural support. All components are attached to the inner bag utilizing oulkhead connections. It is outfitted with two externally mounted metal relief valves (set out nedit two metal Air addition valves (one external & one internal), two metal depressurization valves (one external & one internal), two pressure gauges (one external & one internal, and one metal Air sampling port. In total it weighs 39 lbs (compressor adds another 54 lbs).
{1}------------------------------------------------
## Intended Use
:
| Indicated Use: | To provide mild hyperbaria for the treatment of Acute Mountain<br>Sickness (AMS) and its associated mild symptoms |
|---------------------|-------------------------------------------------------------------------------------------------------------------|
| Environment of Use: | Home, Physicians office, Outdoor environments, Hospital, Sub-<br>acute Institutions, Emergency Services. |
| Prescriptive Use: | Caution, Federal law restricts this device to sale by or on the order<br>of a physician. |
.
- ﺟﺒ
.
{2}------------------------------------------------
## Comparison To Predicate Devices
| Attribute | Gamow<br>K874752A | HTI Model-2<br>K001409 | HTI Model-3<br>K001409 | Flexi-Lite<br>K051759 |
|-----------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| Intended Use | Acute Mountain<br>Sickness (AMS)<br>& associated<br>symptoms | Acute Mountain<br>Sickness (AMS)<br>& associated<br>symptoms | Acute Mountain<br>Sickness (AMS)<br>& associated<br>symptoms | Acute Mountain<br>Sickness (AMS)<br>& associated<br>symptoms |
| Prescription Use | Yes | Yes | Yes | Yes |
| Intended<br>Population | Persons with high<br>altitude mountain<br>sickness | Persons with high<br>altitude mountain<br>sickness | Persons with high<br>altitude mountain<br>sickness | Persons with high<br>altitude mountain<br>sickness |
| Intended<br>Environment of<br>Use | Home, Physician<br>Office, Outdoor,<br>Hospital, Sub-<br>acute facility,<br>EMS | Home, Physician<br>Office, Outdoor,<br>Hospital, Sub-<br>acute facility,<br>EMS | Home, Physician<br>Office, Outdoor,<br>Hospital, Sub-<br>acute facility,<br>EMS | Home, Physician<br>Office, Outdoor,<br>Hospital, Sub-<br>acute facility,<br>EMS |
| Weight (lbs) | 6.5 | 11 | 17 | 39 |
| Size<br>Length x<br>inflated dia. | 7 ft x 21 in | 7 ft x 21 in | 7 ft x 21 in | 8 ft x 30 in |
| Windows | 2 | 3 | 3 | 4 |
| Straps for<br>Transport | 2 | none | none | none |
| Relief Valves | Yes-Plastic | Yes-Metal | Yes-Metal | Yes, Qty-2-Metal |
| Dump Valve | yes | yes, 2-way | yes, 2-way | yes-Separate |
| Operating<br>Pressure (psi) | 2-4 | 2-4 | 2-4 | 2-4 |
| Methods of<br>Inflation | Foot pump /<br>Compressor | Compressor | Compressor | GAST Model<br>0523 Compressor |
{3}------------------------------------------------
| Zipper | Pressure seal | Double zipper w/<br>seal-flap, 2-way | Double zipper w/<br>seal-flap, 2-way | Single zipper |
|---------------------------------|-----------------------------------------|--------------------------------------------------------|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Materials | | | | |
| Attribute | Gamow<br>K874752A | HTI Model-2<br>K001409 | HTI Model-3<br>K001409 | Flexi-Lite<br>K051759 |
| Chamber | 420 denier-<br>urethane coated<br>nylon | 420 denier and<br>33-39 oz<br>urethane coated<br>nylon | 33-39 oz urethane<br>coated nylon | inner bag-880<br>denier-urethane<br>coated nylon<br>outer bag- 2 x 2<br>basket weave, 2-<br>side-urethane<br>coated nylon |
| Relief valves | Plastic | Stainless Steel | Stainless Steel | Brass & Stainless |
| Compressor | GAST oil-less | GAST oil-less | GAST oil-less | GAST 0523 oil-<br>less |
| Air Filtration on<br>compressor | yes | yes | yes | yes |
| Pressure Gauge | yes | yes | yes | yes |
| Air Filtration on<br>chamber | ? | ? | ? | Yes |
۔۔۔ ﺟ
:
{4}------------------------------------------------
| Ability for<br>person inside to<br>extricate<br>themselves if<br>needed | no | yes | yes | yes |
|-------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Attribute | Gamow<br>K874752A | HTI Model-2<br>K001409 | HTI Model-3<br>K001409 | Flexi-Lite<br>K051759 |
| Operating<br>Temperatures | -50F to + 120F | -50F to + 120F | -50F to + 120F | -50F to + 120F |
| Contraindicatio<br>ns | Patients with:<br>-Colds or flu<br>symptoms<br>-Recent alcohol<br>consumption<br>-Blocked ear<br>canals for any<br>reason<br>-Otic barotrauma<br>-Excessive CO2<br>exposure<br>-Pulmonary<br>hyperexpansion<br>-Decompression<br>Sickness | Patients with:<br>-Colds or flu<br>symptoms<br>-Recent alcohol<br>consumption<br>-Blocked ear<br>canals for any<br>reason<br>-Otic barotrauma<br>-Excessive CO2<br>exposure<br>-Pulmonary<br>hyperexpansion<br>-Decompression<br>Sickness | Patients with:<br>-Colds or flu<br>symptoms<br>-Recent alcohol<br>consumption<br>-Blocked ear<br>canals for any<br>reason<br>-Otic barotrauma<br>-Excessive CO2<br>exposure<br>-Pulmonary<br>hyperexpansion<br>-Decompression<br>Sickness | Patients with:<br>-Colds or flu<br>symptoms<br>-Recent alcohol<br>consumption<br>-Blocked ear<br>canals for any<br>reason<br>-Otic barotrauma<br>-Excessive CO2<br>exposure<br>-Pulmonary<br>hyperexpansion<br>-Decompression<br>Sickness |
| Labeling | Comparable | Comparable | Comparable | Comparable |
# Differences Between Other Legally Marketed Devices:
. .
There are no significant differences between the FlexiLite and the predicates - Gamow Bag K874752A -HTI Models 2 or 3 K001409 ـ - ﺣﺘﻰ
.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the edge. In the center of the seal is a stylized eagle emblem. The eagle is depicted with three curved lines representing its wings and body.
Public Health Service
NOV 1 7 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Spencer Feldman President Performance Hyperbarics 2599 A Olinda Road Makawao, Hawaii 96768
Re: K051759
Trade/Device Name: Flexi-Lite Regulation Number: 868.5470 Regulation Name: Hyperbaric Chamber Regulatory Class: II Product Code: CBF Dated: October 21, 2005 Received: October 21, 2005
Dear Mr. Feldman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Feldman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Susan Gunno
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications For Use
#### 510(k) Number (if known): K051759
Device Name: Flexi-Lite
The Flexi-Lite Hyperbaric chamber is a rugged & portable Indications For Use: hyperbaric chamber intended to be used in treating mild symptoms consistent with Acute Mountain Sickness (AMS) as prescribed by or under the direction of a physician. Caution: Federal law restricts this device to sale by or on the order of a physician.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use (21 CFR 807 Subpart C )
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ciara Sullivan
്ഥാ ടിğn-Off)
Sion of Anesthesiology, General Hospital, Control, Dental Devices
ා වලදා) Number:
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.