PURITAN-BENNETT HELIOS PORTABLE LIQUID OXYGEN SYSTEM
Applicant
Puritan Bennett Corp.
Product Code
BYJ · Anesthesiology
Decision Date
Oct 22, 1999
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5655
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Puritan-Bennett Helios Portable Liquid Oxygen System is intended to provide supplemental oxygen to oxygen therapy patients who may have difficulty extracting oxygen from the air that they breathe. It is intended to be used as an ambulatory source of oxygen both inside and outside the patient's home. It is not intended as life sustaining or life supporting device. The device has no contraindications.
Device Story
Portable liquid oxygen system; vacuum-insulated cryogenic container; stores liquid oxygen under pressure. Patient-operated flow control valve releases liquid into vaporizing coil; ambient heat converts liquid to gas; gas delivered via nasal cannula at 12 metered flow settings. Includes pneumatic conserving device (4:1 ratio) to extend duration. Used by patients for ambulatory oxygen therapy; filled by mating to larger stationary reservoir. Purely mechanical; no electrical/electronic components. Benefits patient by providing portable, long-lasting supplemental oxygen source.
Clinical Evidence
Bench testing only. Performance testing included mechanical/climatic and device performance evaluations. No clinical data presented.
Technological Characteristics
Vacuum-insulated cryogenic container; plumbing with pressure relief valves; heat exchanger tubes; pneumatic conserving device. Purely mechanical; no energy source. 12 flow settings. 4:1 conservation ratio. Operable in horizontal or vertical orientation.
Indications for Use
Indicated for oxygen therapy patients requiring supplemental oxygen; ambulatory use inside and outside the home. No contraindications. Not for life support.
Regulatory Classification
Identification
A portable liquid oxygen unit is a portable, thermally insulated container of liquid oxygen that is intended to supplement gases to be inhaled by a patient, is sometimes accompanied by tubing and an oxygen mask. An empty portable liquid oxygen unit is a device, while the oxygen contained therein is a drug.
Predicate Devices
Puritan-Bennett Companion 550 Liquid Oxygen (Lox) System (K933930)
Related Devices
K041122 — PRECISION LIQUID OXYGEN SYSTEM · Precision Medical, Inc. · Oct 4, 2004
K103324 — PRECISION LIQUID OXYGEN SYSTEM · Precision Medical, Inc. · Apr 21, 2011
K072723 — RESPIRONICS GOLOX · Respironics, Inc. · Mar 20, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
OCT 22 1999
Image /page/0/Picture/2 description: The image shows the word "MALLINCKRODT" in a bold, serif font. The "M" is white and set against a black square. The rest of the letters are black and set against a white background.
## 510(k) Summary
## Puritan-Bennett Helios Portable Liquid Oxygen System
#### Submitter Information
| Submitter: | Puritan-Bennett Corporation<br>(a subsidiary of Mallinckrodt Inc.)<br>3 Missouri Research Park Drive<br>St. Charles, MO 63304. |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Tony Keaveney<br>Regulatory Affairs Manager |
| Tel: | (314) 498 3365 |
| Fax: | (314) 939 5535 |
| Preparation Date: | 24 September 1999 |
#### Device Name
| Proprietary Name: | Helios Portable Liquid Oxygen System |
|----------------------|-------------------------------------------------------------|
| Common Name: | Portable Liquid Oxygen Unit |
| Classification Name: | Portable Liquid Oxygen Unit (73 BYJ) as per 21 CFR 868.5655 |
#### Predicate Device Equivalence
Puritan-Bennett is claiming substantial equivalence to the Puritan-Bennett Companion 550 Liquid Oxygen (Lox) System, K933930.
{1}------------------------------------------------
#### Device Description
The Helios portable liquid oxygen system consists of a vacuum insulated cryogenic container and plumbing that includes pressure relief valves, heat exchanger tubes and a pneumatic conserving device. The vacuum insulated container allows oxygen to be stored in its liquid state under pressure where the system pressure is controlled by the relief valve. When the patient opens the flow control valve, the liquid oxygen is withdrawn from the container into a vaporizing coil. Once in the vaporizing coil, ambient heat is absorbed by the liquid oxygen, converting it into a gas and warming it to near room temperature. The gas finally passes through the flow control valve and on to the patient at the prescribed metered rate. The device is intended to be used in conjunction with a larger liquid oxygen reservoir where it is filled by mating it to the fill connection on the top of the larger system. The Helios device is a purely mechanical device and contains no electrical or electronic components.
#### Intended Use
The Helios portable lox system is intended to provide supplemental oxygen to oxygen therapy patients who may have difficulty extracting oxygen from the air that they breathe. The patient would normally receive the oxygen via a nasal cannula. The device delivers 100% oxygen at 12 different flow settings. It is intended to be used as an ambulatory source of oxygen both inside and outside the patient's home. It is not intended as life sustaining or life supporting device. The device has no contraindications.
#### Comparison of Technological Characteristics
The intended use of the modified device remains unchanged as does its risk classification. The modified device will continue to be indicated for use for the same patients and will continue to be a prescription only device. Both devices are intended only as sources of supplemental oxygen and are not intended to be life supporting devices. Neither device has any contraindications.
Both the modified and unmodified device include a vacuum insulated cryogenic container and plumbing that includes pressure relief valves, heat exchanger tubes and a pneumatic conserving device. The cryogenic container in the Helios device is smaller than that in the Companion device and the design of the associated valve and heat exchanger assemblies have been optimized. The enclosure has also been changed to accommodate the change in physical dimensions of the modified device. The conservation ratio of the pneumatic conserver has been changed from 2:1 to 4:1 to allow the device to last longer between filling. The Helios device can be used in the horizontal position whereas the Companion device may only be used in the vertical position.
{2}------------------------------------------------
#### Summary of Performance Testing
The Helios liquid oxygen system successfully passed tests in the following areas;
- Mechanical/Climatic ●
- Device Performance .
#### Conclusions
In summary, Puritan-Bennett has demonstrated that the Helios portable liquid oxygen system is safe and effective. The combined testing and analysis of results provides assurance that the device meets its specifications and is safe and effective for its intended use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three wings, symbolizing health, human services, and well-being. The eagle is positioned to the right of the department's name, which is arranged in a circular pattern around the logo.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 22 1999
Mr. Tony Keaveney Puritan-Bennett Corporation 3 Missouri Research Park Drive St. Charles, MO 63304
Re: K993220 Helios Portable Liquid Oxygen System Requlatory Class: II (two) Product Code: 73 BYJ September 24, 1999 Dated: September 27, 1999 Received:
Dear Mr. Keaveney:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Mr. Tony Keaveney
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Joanne H. Weateober, fec
Wolf Sapirstein, M.D. Acting Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Statement of Indications for Use
| Device name: | Helios Portable Liquid Oxygen System |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use: | The Puritan-Bennett Helios Portable Liquid Oxygen System is intended to provide supplemental oxygen to oxygen therapy patients who may have difficulty extracting oxygen from the air that they breathe. It is intended to be used as an ambulatory source of oxygen both inside and outside the patient's home. It is not intended as life sustaining or life supporting device. The device has no contraindications. |
Prescription Use:
Yes
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use:
\
or
Over-The-Counter Use:
Premarket Notification 510(k) Number:
KA93220
Tothu.extshos
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number .
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.