MEDPULSE RESPIRATORY VEST SYSTEM, MODEL 2000EZ

K040367 · Electromed, Inc. · BYI · Apr 7, 2004 · Anesthesiology

Device Facts

Record IDK040367
Device NameMEDPULSE RESPIRATORY VEST SYSTEM, MODEL 2000EZ
ApplicantElectromed, Inc.
Product CodeBYI · Anesthesiology
Decision DateApr 7, 2004
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 868.5665
Device ClassClass 2
AttributesTherapeutic

Intended Use

The intended use of the Electromed MedPulse® Respiratory Vest System Model 2000ez is to deliver high-frequency chest wall oscillation to promote airway clearance and improve bronchial drainage. The MedPulse® System is indicated when external chest manipulation is the physician's treatment of choice to enhance mucus transport.

Device Story

MedPulse® Respiratory Vest System Model 2000ez delivers high-frequency chest wall oscillation (HFCWO) to promote airway clearance and bronchial drainage. System comprises an Air-Pulse Generator, inflatable vest, and connector hose. Generator uses dual-acting pneumatic diaphragm technology to produce oscillating pressurized air-pulses, causing the vest to rapidly inflate/deflate against the patient's chest wall. Operated by patients or caregivers in clinical or home settings as prescribed by a physician. Device features a brushless motor and digital frequency control, replacing previous analog components to improve reliability and enable a timer feature that stores previous user settings. Output performance remains identical to the predicate. Benefits include improved mucus transport and bronchial secretion mobilization.

Clinical Evidence

Bench testing only. Functional performance and electrical safety testing compared the Model 2000ez to the predicate Model 2000 (K982889), demonstrating no change in output performance or safety.

Technological Characteristics

Powered percussor (Class II). Dual-acting pneumatic diaphragm technology for air-pulse generation. Components: Air-Pulse Generator, inflatable vest, connector hose. Features brushless motor and digital frequency control. Connectivity: None. Sterilization: Not specified. Software: Embedded digital control for frequency and timer functionality.

Indications for Use

Indicated for patients requiring airway clearance and bronchial drainage where external chest manipulation is the physician's treatment of choice to enhance mucus transport.

Regulatory Classification

Identification

A powered percussor is a device that is intended to transmit vibration through a patient's chest wall to aid in freeing mucus deposits in the lung in order to improve bronchial drainage and that may be powered by electricity or compressed gas.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ APR - 7 2004 Image /page/0/Picture/1 description: The image shows a series of alphanumeric characters, likely a code or identifier. The characters are "K0403657" and are oriented diagonally in the image. The text appears to be handwritten or printed with a slightly uneven quality. # ELECTROMED, INC. 502 Sixth Avenue N.W. New Prague, MN 56071 Phone: 952-758-9299 · Fax: 952-758-1941 February 7, 2004 # SUMMARY OF SAFETY AND EFFECTIVENESS This summary of safety and effectiveness is being submitted in accordance with the requirements of SMDA 1990. | Submitter: | Chet Sievert<br>Regulatory Affairs<br>Electromed, Inc.<br>502 Sixth Avenue N.W.<br>New Prague, MN 56071<br>Phone: 952.758.9299<br>Fax: 952.758.1941 | |-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------| | Contact person: | Chet Sievert | | Name of Device: | MedPulse® Respiratory Vest System<br>Model 2000ez | | Classification: | Powered Percussor, Class II | | Predicate Device: | MedPulse® Respiratory Vest System<br>Model 2000<br>510(K) number: K982889 | ## Description of Device: The proposed MedPulse® Respiratory Vest System Model 2000ez is designed to deliver high-frequency chest wall oscillation to promote airway clearance and improve bronchial drainage. The MedPulse® System is indicated when external chest manipulation is the physician's treatment of choice to enhance mucus transport. The primary components of the MedPulse® System include an Air-Pulse Generator, an Inflatable Vest and Vest/Generator Connector Hose. The Air-Pulse Generator produces oscillating pressurized air-pulses that are delivered to the Inflatable Vest by the Vest/Generator Hose. The air-pulses produced by the Generator cause the Vest to rapidly inflate and deflate against the external chest wall of a patient to promote airway clearance by creating high-frequency chest wall oscillation (HFCWO) resulting in mobilization of bronchial secretions. {1}------------------------------------------------ #### Intended Use: The intended use of the Electromed MedPulse® Respiratory Vest System Model 2000ez is to deliver high-frequency chest wall oscillation to promote airway clearance and improve bronchial drainage. The MedPulse® System is indicated when external chest manipulation is the physician's treatment of choice to enhance mucus transport. #### Comparison of Technological Characteristics: The proposed MedPulse® Respiratory Vest System Model 2000ez has the same technological characteristics and principles of operation as Electromed's previously cleared Medpulse® Respiratory Vest System Model 2000 (K982889). The proposed MedPulse® System will continue to use the same method of generating pressurized air-pulses as the Predicate, i.e., by dual acting pneumatic diaphragm technology. The user of the Proposed System will continue to be able to adjust air pulse frequency and pressure amplitude as the Predicate System. The Inflatable Vest and Vest/Generator Connector Hose components of the Modified MedPulse® System will remain unchanged from the Predicate System. The reason for this submission is due to two component changes to the Air-Pulse Generator's motor. The first proposed change is to switch from a servo brush motor to a brushless motor. The motor's output performance will remain the same. A change to a brushless technology is intended to make the motor more reliable and have a longer life span. The second proposed change is to change the motor's frequency control from an analog component (i.e., potentiometer) to a digital control component. The change from analog to digital frequency control is considered to make the Air-Pulse Generator's motor more robust and reliable. The motor's digital frequency control also presents the opportunity to use the same component to control the Timer feature on the Control Panel that regulates the duration of treatment. This change allows the System to "remember" the user's previous settings for ease and more reliable use. ## Performance Testing: The proposed MedPulse® Respiratory Vest System Model 2000ez was tested and compared to the predicate Medpulse® Respiratory Vest System Model 2000 (K982889). Functional performance and electrical safety tests were performed that demonstrated no change in output performance or safety issues to result from the changes proposed. #### Substantial Equivalence: The proposed MedPulse® Respiratory Vest System Model 2000ez has the identical indications and intended use, technological characteristics and principles of operation as the Predicate device. Functional performance and electrical safety testing have demonstrated no changes in safety or effectiveness. Thus, the proposed MedPulse® Respiratory Vest System Model 2000ez is substantially equivalent to the proposed Medpulse® Respiratory Vest System Model 2000 (K982889). {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with three wavy lines instead of snakes. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR - 7 2004 Mr. Chet Sievert Regulatory Consultant Electromed, Incorporated 502 Sixth Avenue N.W. New Prague, MN 56071 Re: K040367 Trade Name: Medpulse Respiratory Vest System, Model 2000ez Regulation Number: 21 CFR 868.5665 Regulation Name: Powered Percussor Regulatory Class: II Product Code: BYI Dated: March 16, 2004 Received: March 19, 2004 Dear Mr. Sievert: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, TDAravay, publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements af the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Mr. Chet Sievert forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. for Sincerely yours, stt Wuttersheue Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number: K040367 Device Name: MEDPULSE RESPIRATORY VEST SYSTEM, MODEL 2000ez Indications For Use: The Electromed MedPulse® Respiratory Vest System Model 2000ez is designed to deliver high-frequency chest wall oscillation to promote airway clearance and improve bronchial drainage. The MedPulse® System is indicated when external chest manipulation is the physician's treatment of choice to enhance mucus transport. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) JtW.h sthesiology, General Hospital, 510(k) Number: K040367 Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...