K132794 · Electromed, Inc. · BYI · Dec 19, 2013 · Anesthesiology
Device Facts
Record ID
K132794
Device Name
SMARTVEST AIRWAY CLEARANCE SYSTEM
Applicant
Electromed, Inc.
Product Code
BYI · Anesthesiology
Decision Date
Dec 19, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5665
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SmartVest® Airway Clearance System, Model SQL is designed to deliver high frequency chest wall oscillation to promote airway clearance and improve bronchial drainage. The SmartVest System is indicated when external chest manipulation is the physician's treatment of choice to enhance mucus transport.
Device Story
SmartVest SQL is an electrically powered percussor for high-frequency chest wall oscillation (HFCWO). System components include an air pulse generator, inflatable vest, and connecting air hose. The generator produces pressurized air pulses (5-20 Hz) delivered to the vest, which imparts oscillatory pressure to the patient's external chest wall. This promotes airflow within the lungs to loosen mucus for expulsion. Used in clinical or home settings under physician prescription. The modified generator features a smaller, lighter, and quieter design with digital pressure control, child-lock, and enhanced ramping/pause features. Healthcare providers use the device to manage patients with impaired respiratory function; the output assists in mucus clearance, potentially reducing respiratory complications.
Clinical Evidence
Bench testing only. Performance testing compared air pulse pressure, frequency, and treatment time between the modified and predicate devices. Design verification and validation, including usability/human factors testing, confirmed that design outputs met requirements. Safety testing included IEC 60601-1, 60601-1-2, 60601-1-6, 60601-1-11, and UL standards.
Technological Characteristics
Electrically powered percussor; HFCWO delivery via air pulse generator and inflatable vest. Connectivity: Standalone. Standards: IEC 60601-1 (3rd ed), 60601-1-2 (3rd ed), 60601-1-6 (3rd ed), 60601-1-11 (3rd ed), UL, and ASTM D4169-09 DC13 for shipping. Features digital pressure control, child-lock, and enhanced ramping/pause functions.
Indications for Use
Indicated for patients requiring airway clearance and bronchial drainage via external chest manipulation to enhance mucus transport, as prescribed by a physician.
Regulatory Classification
Identification
A powered percussor is a device that is intended to transmit vibration through a patient's chest wall to aid in freeing mucus deposits in the lung in order to improve bronchial drainage and that may be powered by electricity or compressed gas.
Predicate Devices
SmartVest Airway Clearance System, Model TL (K053248)
Submission Summary (Full Text)
{0}------------------------------------------------
DEC 1 9 2013
# ELECTROMED, INC.
Creating superior care through innovation®
# SPECIAL 510(k) SUMMARY
This Special 510(k) Summary is submitted in accordance with 21 CFR 807.92.
Date prepared: September 4, 2013
## SUBMITTER
Electromed, Inc. 502 Sixth Ave. NW New Praque, MN 56071 Telephone: 952-758-9299
# CONTACT INFORMATION
Chet Sievert Director of Regulatory and Clinical Affairs Telephone: 952-758-0384 direct 651-246-8621 cell Fax: 888-877-3371 E-mail: csievert@electromed.com
DEVICE INFORMATION
Trade Name: SmartVest® Airway Clearance System, Model SQL Common Name: Percussor, Powered-Electric
CLASSIFICATION
Regulation Number: 21 CFR 868.5665, Powered percusson Product Code: BYI Class: Il
Panel: Anesthesiology
�
#### PREDICATE DEVICE
SmartVest Airway Clearance System, Model TL: K053248
#### INDICATIONS FOR USE
The Indications for Use of the proposed modified device are identical to the legally marketed predicate device:
"The SmartVest Airway Clearance System, Model SQL is designed to deliver high-frequency chest wall oscillation to promote airway clearance and improve bronchial drainage. The SmartVest® System is indicated when external chest manipulation is the physician's treatment of choice to enhance mucus transport."
> Corporate Headquarters 500 Sixth Ave. N.W. New Prague, MN 56071 Phone: 952-758-9299 Fax: 952-758-1941 www.electromed.com
{1}------------------------------------------------
## DEVICE DESCRIPTION
The modified Electromed SmartVest® Airway Clearance System, Model SQL is an electrically powered percussor device designed to deliver high frequency chest wall oscillation (HFCWO) to aid in freeing mucus deposits in the lungs in order to improve bronchial drainage under the order of a physician's prescription. The primary components of both the proposed modified and predicate HFCWO Airway Clearance Systems consist of an Air Pulse Generator, an Inflatable Vest, and an Air Hose that connects the Generator to the Inflatable Vest. The Air Pulse Generator produces small volumes of pressurized air pulses that are rapidly delivered to the Inflatable Vest via the Air Hose at a selected oscillatory frequency between 5-20 times per minute (Hz). The Inflatable Vest imparts the oscillatory air pulses as pressure forces to the patient's external chest wall. These pressurized air pulses promote transient increases in airflow within the lungs that loosens mucus sufficiently to facilitate expulsion by the patient when normal respiratory function is not capable.
ﻧﮯ
### DESCRIPTION OF THE CHANGES TO THE PREDICATE
Modifications were made to the Air Pulse Generator only. The modified generator was designed primarily to be smaller, quieter and lighter compared to the predicate. In addition, the following features were added to the predicate generator and two existing features were enhanced:
- Child-lock option on treatment settings
- Digital control of pressure settings ●
- . On/Off switch
- Enhanced ramping feature .
- . Enhanced pause feature
- More "user friendly" graphic user interface .
## PERFORMANCE, FUNCTIONAL and SAFETY TESTING
Testing included:
- Comparison output performance testing; proposed modified vs. predicate (air . pulse pressure, air pulse frequency and treatment time)
- Design verification testing (including software) .
- Design validation testing (usability/human factors) .
- IEC 60601-1 3th edition, 60601-1-2 30 edition, 60601-1-6 30 edition, 60601-1-. 11 300 edition and UL testing
- ASTM D4169-09 DC13. Standard Practice for Performance Testing of Shipping . Containers and Systems
## COMPLETION OF DESIGN CONTROL ACTIVITIES
Managing and documenting the design process of making modifications to the predicate device was performed under Electromed's Quality System which is compliant with 21 CFR 820. All modification and design activities were conducted in conformance to 21 CFR 820.30 Design Control including documentation of all design and development activities including risk analysis, project planning, design
{2}------------------------------------------------
requirements, design specifications, design transfer, design changes, verification and validation procedures and formal design reviews. These design control activities and testing ensured that the design output specifications met the design input requirements as well as meeting the device's intended use and the needs of the user population.
#### SUBTANTIAL EQUIVALENCE
A formal risk analysis of the proposed modified generator design revealed no new risks compared to the predicate generator design. All material changes were to a type of material that has been used in other legally marketed devices. Design verification testing, including software testing, demonstrated that the design output specifications met the design input requirements. Comparison performance testing of the proposed modified generator and the predicate generator including measurements of the air pulse pressure, air pulse frequency and treatment time demonstrated equivalence between the two devices. Usability/human factors testing demonstrated enhanced user operation with less potential for errors compared to the predicate. The modified design passed electromagnetic compatibility testing, electrical safety testing and UL testing.
The modified design does NOT change the device's operating principle or mechanism of action. Compared to the predicate, the proposed modified device has the identical indications for use, the same patient population, equivalent output performance and the same fundamental scientific technology which in total does not raise additional questions of safety or effectiveness thus meeting the definition of substantial equivalence.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an eagle or bird-like figure.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002
December 19, 2013
Electromed, Incorporated Mr. Chet Sievert Director of Regulatory and Clinical Affairs 502 Sixth Avenue NW NEW PRAGUE MN 56071
Re: K132794
Trade/Device Name: SmartVent Airway Clearance System, Model SQL Regulation Number: 21 CFR 868.5665 Regulation Name: Powered Percussor Regulatory Class: Class II Product Code: BYI Dated: November 15, 2013 Received: November 20, 2013
Dear Mr. Sievert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subiect to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Sievert
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/4/Picture/8 description: The image contains a logo or emblem with the text "Teashri Purohit Sheth, M.D." and "Clinical Deputy Director." The logo also includes the acronym "DAGRID." The text and logo are surrounded by decorative elements, giving it a formal or official appearance. The word "FOR" is present in the lower right corner.
Erin I. Keith M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Special 510k SmartVest® Airway Clearance System, Model SQL
Section 5. Indications for Use Statement
# Indications for Use Statement
510(k) Number: K132794
Device Name: SMARTVEST® AIRWAY CLEARANCE SYSTEM, MODEL SQL
Indications for Use:
The SmartVest® Airway Clearance System, Model SQL is designed to deliver high frequency chest wall oscillation to promote airway clearance and improve bronchial drainage. The SmartVest System is indicated when external chest manipulation is the physician's treatment of choice to enhance mucus transport.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
1993年09月 - 198 ﺔ ﺍﻟﻘﻠﻴﻤﻴﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟ 1 :
Anya C. Harry -S 2013.12.18 17:11:59 -05'00'
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.