AEROBIKA POSITIVE EXPIRATORY PRESSURE DEVICE

K123400 · Trudell Medical Intl. · BWF · May 16, 2013 · Anesthesiology

Device Facts

Record IDK123400
Device NameAEROBIKA POSITIVE EXPIRATORY PRESSURE DEVICE
ApplicantTrudell Medical Intl.
Product CodeBWF · Anesthesiology
Decision DateMay 16, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5690
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Aerobika* Oscillating Positive Expiratory Pressure device is intended for use as a Positive Expiratory Pressure (PEP) device. The Aerobika* Oscillating PEP device may also be used simultaneously with nebulized aerosol drug delivery. The device is intended to be used by patients capable of generating an exhalation flow of 10 lpm for 3 - 4 seconds.

Device Story

Aerobika is a hand-held, single-patient-use device for secretion clearance and lung expansion. It functions by creating vibrating positive expiratory pressure (PEP) during patient exhalation. The device is mechanically driven by the patient's own breath; no external power source is required. It features adjustable exhalation resistance and can be used simultaneously with a nebulizer for aerosol drug delivery. The device is intended for use by patients capable of maintaining an exhalation flow of at least 10 lpm for 3-4 seconds. It is non-sterile and designed to be disassembled for cleaning. By providing oscillating PEP, the device assists in mucociliary clearance, potentially benefiting patients with respiratory conditions requiring airway clearance.

Clinical Evidence

Bench testing only. Mechanical testing compared oscillation frequency and amplitude against the predicate, showing clear overlap and mean values within 15%. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, irritation). Cascade impactor testing confirmed comparable aerosol drug delivery performance to the predicate.

Technological Characteristics

Mechanically driven by patient exhalation; no external energy source. Materials meet ISO 10993-1:2009 for surface devices with mucosal contact (limited exposure). Adjustable exhalation resistance mechanism. Non-sterile, single-patient-use, hand-held form factor. Disassemblable for cleaning.

Indications for Use

Indicated for patients capable of generating an exhalation flow of 10 lpm for 3-4 seconds requiring secretion clearance and lung expansion therapy; may be used with nebulized aerosol drug delivery.

Regulatory Classification

Identification

An incentive spirometer is a device that indicates a patient's breathing volume or flow and that provides an incentive to the patient to improve his or her ventilation.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K123400 Premarket Notification 510(k) Aerobika* Oscillating Positive Expiratory Pressure (PEP) Device # Trudell Medical International MAY 1 6 2013 # Section 5 – 510(k) Summary Prepared: 10 May 2013 ### 510(k) Owner | Trudell Medical International | | |-------------------------------|-----------------------------------------------| | 725 Third Street | | | London, Ontario N5V 5G4 | | | CANADA | | | Contact Person | Darryl Fischer | | | Associate Director, Global Regulatory Affairs | | Phone | 1-519-455-7060 ext 2140 | | Fax | 1-519-455-6329 | | e-mail | dfischer@trudellmed.com | ## Device Name | Proprietary | Aerobika* Oscillating Positive Expiratory Pressure (PEP)<br>Device | |-----------------------|--------------------------------------------------------------------| | Common/Classification | Spirometer, Therapeutic (Incentive) | Product Code Classification Regulation 868.5690 Predicate Device 510(k) Number K002768 Trade/Model Name Acapella BWF Manufacturer DHD Healthcare Corp. ## Device Description The Aerobika* Oscillating Positive Expiratory Pressure (PEP) device is indicated as a single patient use, hand held secretion clearance and lung expansion device that creates vibrating positive expiratory pressure when a patient exhales through the device. The Aerobika* device may be used simultaneously with aerosol drug delivery from a nebulizer. {1}------------------------------------------------ # Section 5 – 510(k) Summary ## Intended Use The Aerobika* Oscillating Positive Expiratory Pressure device is intended for use as a Positive Expiratory Pressure (PEP) device. The Aerobika* Oscillating PEP device may also be used simultaneously with nebulized aerosol drug delivery. The device is intended to be used by patients capable of generating an exhalation flow of 10 lpm for 3 - 4 seconds. # Technological Characteristic Comparison to Predicate Device(s) | Common Device Characteristics | | |------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Aerobika* PEP Device | Acapella (predicate device - K002768) PEP Device | | • combines PEP therapy with oscillations upon exhalation | • combines PEP therapy with oscillations upon exhalation | | • may be used with a nebulizer to deliver aerosol drug | • may be used with a nebulizer to deliver aerosol drug | | • mechanically driven using only the patient's exhaled breath | • mechanically driven using only the patient's exhaled breath | | • not orientation dependent | • not orientation dependent | | • can be adjusted to increase or decrease the exhalation resistance | • can be adjusted to increase or decrease the exhalation resistance | | • the device has a removable mouthpiece | • the device has a removable mouthpiece | | • the device can remain in the mouth through the treatment | • the device can remain in the mouth through the treatment | | Relevant Differences in Device Characteristics | | | • The Aerobika* Oscillating PEP can be disassembled for cleaning | • Only the mouthpiece of the Acapella device can be disassembled for cleaning | | • The Aerobika* Oscillating PEP may be used for patients with an exhalation flow of 10 lpm and above | Two devices are available.<br>• The green or high flow device suitable for most patients. The expiratory flow of the patient must be 15 lpm or greater<br>• The blue or low flow device is for those patients who have low expired lung volumes / less than 15 lpm | {2}------------------------------------------------ Premarket Notification 510(k) Aerobika* Oscillating Positive Expiratory Pressure (PEP) Device Trudell Medical International # Section 5 - 510(k) Summary #### Non-Clinical Test Summary . Mechanical testing was conducted to characterize the operating parameters (amplitude and frequency) of the Aerobika* device to the predicate device. The test parameters evaluated are a) oscillation frequency and b) oscillation amplitude. It was observed that the ranges of data for each parameter are similar for both devices, showing clear overlap. The mean Frequency and Amplitude values for the Aerobika* device are also within 15% of the respective mean values of the predicate device. The slightly lower minimum frequency value observed for the Aerobika* device is considered clinically insignificant in terms of therapeutic effect. In addition to the fact that the predicate device has exhibited lower frequencies of 8.0 Hz in previous evaluations (Volsko et al, 2003), it is also known that other medical devices utilizing wave oscillation to aid mucocillary clearance are effective down to 3Hz (Vest device, Douglas and Bond-Kendall) and 6Hz (Quake device, Okeson and McGowen, 2007). The patient contacting components of the Aerobika* oscillating positive expiratory pressure device meets the requirements of ISO 10993-1:2009, Biological evaluation of medical devices, and are classified according to the within ISO 10993-1 as follows; - . Category: Surface device - . Contact: Mucosal membranes - Contact Duration: A - Limited Exposure (≤ 24hrs) - Evaluation Testing for Biological Effect - o Cytotoxicity - according to ISO 10993-5 - o Sensitization - according to ISO 10993-10 - Irritation or Intracutaneous Reactivity according to ISO 10993-10 o Section 15 of this submission contains the information regarding the biocompatibility profile of the device under review. Cascade impactor testing was also conducted comparing the Aerobika* oscillating positive expiratory pressure device to the Acapella using a nebulizer and three (3) different drugs. The results demonstrate that the Aerobiko* device performs comparably to the Acapella device, and does not raise any new safety or efficacy related issues. #### Clinical Performance Summary Not applicable, the determination of substantial equivalence is not based on Clinical Performance Data. {3}------------------------------------------------ Premarket Notification 510(k) Aerobika* Oscillating Positive Expiratory Pressure (PEP) Device Trudell Medical International # Section 5 – 510(k) Summary ### Conclusions from Testing The Aerobika* oscillating positive expiratory pressure device has been evaluated against the currently marketed (predicate) Acapella PEP device for the determination of substantial equivalency. The Aerobika* oscillating positive expiratory pressure device and the predicate device share common indications for use, operating characteristics and usage environments. The devices are both single patient use, non-sterile and are available by prescription. The Aerobica* PEP device raises no new issues regarding safety or efficacy. > Section 5 Page 4 of 4 {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings outstretched. ### DEPARTMENT OF HEALTH & HUMAN SERVICES . Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 # May 16, 2013 Mr. Darryl Fischer Associate Director, Global Regulatory Affairs Trudell Medical International 725 Third Street LONDON Ontario N5V 5G4 Re: K123400 Trade/Device Name: Aerobika* Oscillating Positive Expiratory Pressure (PEP) Device Regulation Number: 21 CFR 868.5690 Regulation Name: Incentive Spirometer Regulatory Class: II Product Code: BWF Dated: January 29, 2013 Received: April 17, 2013 Dear Mr. Fischer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Image /page/5/Picture/8 description: The image shows the text "Kwame O. Ulmer for -S FDA". The text is arranged in three lines. The first line contains the name "Kwame O. Ulmer" followed by the word "for". The second line contains the letters "FDA" in a stylized font. The third line contains the text "-S". Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use 510(k) Number: K123400 Device Name: Aerobika* Oscillating Positive Expiratory Pressure (PEP) Device Indications for Use: The Aerobika* Oscillating Positive Expiratory Pressure device is intended for use as a Positive Expiratory Pressure (PEP) device. The Aerobika* Oscillating PEP device may also be used simultaneously with nebulized aerosol drug delivery. The device is intended to be used by patients capable of generating an exhalation flow of 10 lpm for 3 – 4 seconds. Prescription Use: (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE) Paul H. Shin=S 2013.05.15(10:10:35)-04'00' 'Division Sign-Off) Juvision of Anesthesiology, General Hospital -mection Control, Dental Devices 510(k) Number: K123400 Page 1 of
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...